← Product Code [NHA](/submissions/DE/subpart-d%E2%80%94prosthetic-devices/NHA) · K132749

# NOBELPROCERA OVERDENTURE BAR (K132749)

_Nobel Biocare USA, LLC · NHA · Jul 24, 2014 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/NHA/K132749

## Device Facts

- **Applicant:** Nobel Biocare USA, LLC
- **Product Code:** [NHA](/submissions/DE/subpart-d%E2%80%94prosthetic-devices/NHA.md)
- **Decision Date:** Jul 24, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3630
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

The NobelProcera Overdenture Bar is indicated for use as an overdenture bar that attaches to implants or abutments in the treatment of partially or totally edentulous jaws for the purpose of restoring chewing function.

## Device Story

NobelProcera Overdenture Bar is a patient-specific, computer-aided design/manufacturing (CAD/CAM) precision-milled superstructure; provides retention/support for removable or fixed dentures. Input: master cast impression scanned at manufacturing facility. Process: CAD software designs superstructure; CNC milling system machines titanium alloy from solid piece. Output: custom-milled bar delivered to dental laboratory for integration with denture. Used in dental clinics/laboratories; operated by dental professionals. Healthcare providers use the bar to provide prosthetic support for patients; benefits include restored chewing function and stable dental restoration.

## Clinical Evidence

No clinical trials performed. Evidence consists of a retrospective study of devices placed globally since 2010 and non-clinical bench testing, including fatigue testing per FDA guidance and MR conditional testing.

## Technological Characteristics

Material: Titanium alloy. Design: Patient-specific, CAD/CAM machined. Form factor: Fixed (Dolder, Hader, Round) or Free Form (Milled, Montreal, Paris, Wrap-around) shapes. Connectivity: None. Sterilization: Non-sterile. Software: CAD/CAM design and manufacturing system.

## Regulatory Identification

An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.

## Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)

## Predicate Devices

- Procera Implant Bridge Overdenture ([K090069](/device/K090069.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# 1111 2 4 2014 432749

#### 1.4 510(k) Summary

| Submitted by:                          | Phuong Nguyen Son<br>Senior Regulatory Affairs Manager<br>Nobel Biocare USA LLC<br>22715 Savi Ranch Parkway<br>Yorba Linda, CA 92887 |
|----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|
| Submitted for:                         | Nobel Biocare AB<br>Västra Hamngatan 1<br>Goteborg, SE 411 17<br>Sweden                                                              |
| Telephone:                             | (714) 282-4800 x7830                                                                                                                 |
| Facsimile:                             | (714) 998-9348                                                                                                                       |
| Date of Submission:                    | August 30, 2013                                                                                                                      |
| Classification Name:<br>Product Code:  | Endosseous Dental Implant Abutment (21 CFR 872.3630)<br>NHA                                                                          |
| Trade or Proprietary<br>or Model Name: | NobelProcera Overdenture Bar                                                                                                         |
| Legally Marketed Devices:              | Procera® Implant Bridge Overdenture (K090069)                                                                                        |

#### Device Description:

NobelProcera Overdenture Bars are computer aided design (CAM) precision milled, individually designed Superstructures manufactured for each patient. The NobelProcera Overdenture Bar provides retention and support for a removable or fixed denture made of standard laboratory dental materials such as resin composite or porcelain veneer.

The Superstructures can be either Fixed Shaped Bars such as Dolder, Hader, Round, or Free Form Shaped Bars such as Free Form Milled, Montreal, Paris, Wrap-around; or a combination of shapes. The Overdenture Bar shapes and design parameters remain unchanged from the predicate device, Procera Implant Bridge Overdenture (K090069). Additionally, the Superstructure can also be designed per specific prescription order by qualified health care professionals.

The Superstructure is a multiple unit restoration for fixed or removeable dental prosthetic framework milled from one solid piece of material. The NobelProcera Overdenture Bar accommodates commercially available denture attachments for removable dental prosthesis, or retention elements for fixed dental prosthesis.

The many abutment and implant interfaces that can be used in the Superstructure are precision milled and are an integral part of the Overdenture Bar.

The NobelProcera Overdenture Bar is packaged as non-sterile, and delivered to a dental laboratory for completion. Once received at the laboratory, the NobelProcera Overdenture Bar is matched to a denture for final placement.

Nobel Biocare Traditional 510(k) Notification NobelProcera Overdenture Bar August 2013

{1}------------------------------------------------

#### Indications for Use:

The NobelProcera Overdenture Bar is indicated for use as an overdenture bar that attaches to implants or abutments in the treatment of partially or totally edentulous jaws for the purpose of restoring chewing function.

| Manufacturer                  | Name                                   | Sizes                                           |
|-------------------------------|----------------------------------------|-------------------------------------------------|
| 3i® (BioMet)                  | (Certain) Internal                     | 3.4, 3.75/4.0, 5.0, 6.0                         |
|                               | Conical Abutment                       | 3.4, 4.1/5.0, 6.0                               |
|                               | External (hex)                         | 3.4, 4.1, 5.0, 6.0                              |
|                               | IOL® Abutment                          | -                                               |
|                               | Low Profile abutments                  | -                                               |
|                               | Standard abutment                      | -                                               |
| AB Dental Devices             | Narrow Platform                        | -                                               |
| Alpha Bio                     | Internal Hex                           | 4.5                                             |
|                               | TCT System                             |                                                 |
| Anthogyr                      | Anthofit® HE System Conical Abutment   | 4.1                                             |
|                               | Anthofit® Ol System                    | 3.5                                             |
|                               | Anthofit® OI System Conical Abutment   | 3.5                                             |
|                               | Axiom REG/PX Platform                  | 4.0                                             |
|                               | Axiom REG/PX Straight Conical Abutment | -                                               |
| BEGO Semados®                 | S/RI-Line                              | 3.25/3.75/4.1/4.5, 5.5                          |
| BioHorizons®                  | Abutment for Screw                     | 3.5/4.0, 5.0                                    |
|                               | Conical internal connection            | 3.0, 3.5, 4.5, 5.7                              |
|                               | External implant system                | 3.5, 4.0, 5.0, 6.0                              |
|                               | Single Stage Restoration               | 3.5, 4.5, 5.7                                   |
| Biotech Dental<br>Implants    | B.I.S. Conic                           | 3.5/3.9, 4.0/4.4, 5/5.4                         |
|                               | B.I.S. Conic Base Analog               | 4.4                                             |
| BTI®                          | Externa® Platform (abutment level)     | 4.1                                             |
|                               | Externa® Platform (implant level)      | 4.1                                             |
|                               | Interna® Platform                      | 4.1                                             |
|                               | Tiny® Platform                         | 3.5                                             |
| Camlog®                       | Camlog Bar Abutment                    | 3.3, 3.8, 6.0                                   |
|                               | Camlog Implant System                  | 3.3, 3.8, 4.3, 5.0, 6.0                         |
|                               | CONELOG Implant System                 | 3.8                                             |
| Dentsply (Ankylos®)           | Balance Base Abutment                  | -                                               |
|                               | Balance Base Abutment Narrow           | -                                               |
|                               | Balance Base C                         | -                                               |
|                               | Regular C/X Abutment                   | -                                               |
| Dentsply (Astra<br>Tech®)     | 20° UniAbutment                        | 3.5/4.0                                         |
|                               | 45° UniAbutment                        | 3.5/4.0                                         |
|                               | Astra ST                               | 3.5                                             |
|                               | Astra Tech                             | Yellow, Aqua, Lilac                             |
| Dentsply (XIVE®)              | FRIADENT®                              | 3.0, 3.4, 3.8, 4.5, 5.5                         |
|                               | FRIADENT® MP                           | 3.8, 4.5, 5.5                                   |
|                               | TG Platform                            | 3.4-4.5                                         |
| Hiossen                       | ET III SA                              | mini, regular                                   |
| HI-TEC implants               | Universal Unit System                  | -                                               |
| Implant Direct                | Attachment International               | 3.25, 3.3                                       |
|                               | Legacy™                                | 3.0, 3.5, 4.5, 5.7                              |
|                               | RePlant®/RePlus®                       | 3.5, 4.3, 5.0                                   |
|                               | ScrewIndirect                          | -                                               |
|                               | ScrewIndirect                          | 5.0                                             |
| Manufacturer                  | Name                                   | Sizes                                           |
| Implant Direct<br>(Sybron)    | ScrewPlant®/ScrewPlus®                 | 3.7, 4.7, 5.7                                   |
|                               | Endopore® Internal Hex                 | 4.1                                             |
|                               | Entegra™ External Hex                  | 3.5, 4.1, 5.0                                   |
|                               | PITT-EASY® Paracentric Line            | 3.75, 4.0, 4.9                                  |
|                               | SybronPRO™ Series Hex                  | 3.5, 4.1                                        |
| LifeCore Dental<br>(Keystone) | Genesis System                         | 3.8, 4.5, 5.5                                   |
|                               | PrimaConnex® System                    | 3.5, 4.1, 5.0/7.0/8.0                           |
|                               | PrimaConnex® System Multi-unit         | 3.5/4.1, WD                                     |
|                               | Renova® Implant System                 | SDI 3.75, RDI 4.5                               |
|                               | Restore® Implant System                | SD 3.3, RD 3.75/4.0, WD                         |
| MIS®                          | Stage-1 system RDS                     | -                                               |
|                               | Internal Hex                           | -                                               |
|                               | Multi-Unit system                      | -                                               |
|                               | Standard Platform                      | -                                               |
|                               | Wide Interial Hex                      | -                                               |
| Neoss                         | Access Abutment                        | -                                               |
|                               | Neoss Implant System                   | 3.25                                            |
| Nobel Biocare                 | Brånemark System (External Hex)        | NP, RP, WP                                      |
|                               | Hex                                    | 2.0                                             |
|                               | Internal Hexagon                       | -                                               |
|                               | Multi-Unit Abutment                    | NP/RP, WP                                       |
|                               | Nobel Novum                            | -                                               |
|                               | Nobel Replace (Internal Trilobe)       | NP, RP, WP, 6.0                                 |
|                               | NobelActive (Conical Connection)       | NP, RP, WP                                      |
|                               | Replace Hex Abutment                   | 3.5, 4.3, 3.25HL,<br>3.8/4.5 HL, 5.0 HL, 6.0 HL |
| Osstem                        | TS & GS System                         | Mini 4.0, standard 4.5                          |
|                               | US System Esthetic Low                 | -                                               |
|                               | US System Regular platform             | -                                               |
| SIC Implants                  | Sic Platform                           | 3.3, 4.2                                        |
| Simpler Implants              | Implant Analog                         | 3.25, 4.0                                       |
| Southern Implants             | External Hex                           | 4.0, 5.0                                        |
| Straumann®                    | Bone Level                             | 3.3, 4.1/4.8                                    |
|                               | Multi-base                             | NC 3.5, RC 4.5, RC 6.5                          |
|                               | Octagon                                | 3.5, 4.8, 6.5                                   |
|                               | Ortho System                           | 4.2                                             |
|                               | SynOcta®                               | RN, WN                                          |
| Thommen Medical               | SPI®RETAIN DIRECT System               | -                                               |
|                               | SPI®VARIOmulti System                  | 4.0, 4.5, 5.0                                   |
|                               | VARIOeco system                        | 3.5, 4.0, 4.5, 5.0, 6.0                         |
| Titan implant                 | Internal Hex platform                  | 4.5, 5.7                                        |
| Zimmer® Dental                | AdVent                                 | 4.5                                             |
|                               | Calcitek                               | 3.25, 4.0                                       |
|                               | Screw-Vent                             | 3.5, 4.5, 5.7                                   |
|                               | Specta-Cone                            | -                                               |
|                               | Spline® Implant System                 | 3.25, 3.75/4.0, 5.0                             |
|                               | SwissPlus® Implant System              | 3.8, 4.8                                        |

The NobelProcera Overdenture Bar is compatible with the below implants and abutments:

ﺮ ﺍﻟﻤﺮﺍﺟﻊ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ Nobel Biocare Traditional 510(k) Notification NobelProcera Overdenture Bar

August 2013

{2}------------------------------------------------

Nobel Biocare Traditional 510(k) Notification
NobelProcera Overdenture Bar
August 2013

:

・

{3}------------------------------------------------

#### Summary of testing to demonstrate Substantial Equivalence

A retrospective study of NobelProcera Overdenture Bar placed in the market globally since 2010 was performed to establish the safety and effectiveness of the additional implant and abutment interfaces. The NobelProcera Overdenture Bar data set of compatible implant and abutment interfaces were generated through measurements of original manufacturer's bar interfaces. Additionally, non-clinical test data was used to demonstrate substantial equivalence. Non-clinical testing consisted of performance of fatigue testing in accordance with the FDA guidance Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments.

MR conditional testing was conducted according to FDA Guidance Document "Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment".

{4}------------------------------------------------

| CHARACTERISTIC                        | CANDIDATE                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | PREDICATE                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name                            | NobelProcera Overdenture Bar                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Procera Implant Bridge Overdenture (K090069)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Anatomical Site                       | Oral Cavity                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Oral Cavity                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Design/Construction                   | Patient specific / machined                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Patient specific / machined                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Device Material                       | Titanium alloy                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Titanium alloy                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Design Method<br>Bar Order            | Individually designed for each patient by order of<br>prescription                                                                                                                                                                                                                                                                                                                                                                                                             | Individually designed for each patient by order of<br>prescription                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Bar Shape                             | Fixed shaped bars<br>Dolder, Hader, Round                                                                                                                                                                                                                                                                                                                                                                                                                                      | Fixed shaped bars<br>Dolder, Hader, Round                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Platform Compatibility                | Free form shaped bars<br>Milled, Montreal, Paris, Victoria, Wrap-around                                                                                                                                                                                                                                                                                                                                                                                                        | Free form shaped bars<br>Milled, Montreal, Paris, Victoria, Wrap-around                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|                                       | See 'Indications for use statement'                                                                                                                                                                                                                                                                                                                                                                                                                                            | Branemark System NP, RP, WP<br>NobelReplace NP, RP, WP, 6.0<br>NobelActive NP, RP<br>Multi-unit NP/RP, WP<br>Astra Tech Yellow, Aqua, Lilac<br>Straumman Octagon 3.5,4.8, 6.5<br>Straumman Bone Level 3.3, 4.1/4.8<br>BioMet 3i Internal 3.4, 3.75/4.0, 5.0, 6.0<br>BioMet 3i External 3.4, 4.1, 5.0, 6.0<br>Zimmer Screw-Vent 3.5, 4.5, 5.7<br>Zimmer AdVent 4.5                                                                                                                                                                                                                                                                                                                                                                         |
| RACTERISTIC                           | CANDIDATE                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | PREDICATE                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Manufacturing                         | Nobel Procera Overdenture Bar<br><br>A master cast is received with the patient's impression.<br>The Nobel Biocare's manufacturing facility scans the<br>master cast to create the manufacturing files. Then the<br>design files are compiled for CAM using a CNC milling<br>system. The machined Overdenture bar is inspected for<br>fit against the master cast.<br>The overdenture bar is sent to the laboratory for<br>verification of fit and integration with a denture. | Procera Implant Bridge Overdenture<br>(K090069).<br><br>Procera Implant bridges can be designed either using<br>traditional wax-up technique or by using Nobel's Procera<br>Software. When the wax-up technique is used, either the<br>master cast or a scan of the master cast can be sent to<br>Nobel for production. If the Procera Software is used, the<br>design information is transmitted to Nobel for production.<br><br>If necessary Nobel Biocare's manufacturing facility scans<br>the master cast to create the manufacturing files. Then<br>the design files are compiled for CAM using a CNC milling<br>system. The machined overdenture bar is sent to the<br>laboratory for verification and integration with a denture. |
| Denture Bar/<br>Abutment<br>Interface | Nobel Biocare and additional manufacturers' implant<br>and abutment interfaces                                                                                                                                                                                                                                                                                                                                                                                                 | Nobel Biocare and other manufacturers' implant interfaces                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Packaging                             | Plastic bag and cardboard box                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Plastic bag and cardboard box                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Sterility                             | Non-sterile                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Non-sterile                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Intended Use                          | Removable or fixed superstructure                                                                                                                                                                                                                                                                                                                                                                                                                                              | Removable or fixed superstructure                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Indications for Use                   | Compatible platforms added to indications for use.                                                                                                                                                                                                                                                                                                                                                                                                                             | The Procera Implant Bridge Overdenture is indicated for<br>use as an overdenture bar that attaches to implants or<br>abutments in the treatment of partially or totally<br>edentulous jaws for the purpose of restoring chewing<br>function.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| MR Safety Information                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | No MR Safety information                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |

Comparison of Technological Characteristics

Nobel Biocare Traditional 510(k) Notification
NobelProcera Overdenture Bar
August 2013

{5}------------------------------------------------

Conclusion provided in this submission demonstrates that the device is substantially equivalent to the predicate device.

Nobel Biocare Traditional 510(K) Notification
NobelProcera Overdenture Bar
August 2013

{6}------------------------------------------------

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 24, 2014

Nobel Biocare AB C/O Ms. Phuong Nguyen Son Senior Regulatory Affairs Manager Nobel Biocare USA, Limited Liability Company 22715 Savi Ranch Parkway Yorba Linda. CA 92887

Re: K132749

Trade/Device Name: NobelProcera Overdenture Bar Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: June 26, 2014 Received: June 26, 2014

Dear Ms. Son:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{7}------------------------------------------------

### Page 2 - Ms. Son

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note

the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Image /page/7/Picture/7 description: The image shows the text "Mary Sumner - FDA". The text is in a bold, sans-serif font. The letters are black and the background is white. The letters are slightly distorted, as if they were printed on a textured surface. The text is arranged in a single line, with the words "Mary Sumner" on the left and "FDA" on the right. There is a hyphen between the words "Sumner" and "FDA".

Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

## Indications for Use

510(k) Number (if known): K132749

#### Device Name: NobelProcera Overdenture Bar

Indications For Use:

The NobelProcera Overdenture Bar is indicated for use as an overdenture bar that attaches to implants or abutments in the treatment of partially edentulous jaws for the purpose of restoring chewing function.

The NobelProcera Overdenture Bar is compatible with the below implants and abutments:

| Manufacturer               | Name                                   | Sizes                   |
|----------------------------|----------------------------------------|-------------------------|
| 3I® (BioMet)               | (Certain) Internal                     | 3.4, 3.75/4.0, 5.0, 6.0 |
|                            | Conical Abutment                       | 3.4, 4.1/5.0, 6.0       |
|                            | External (hex)                         | 3.4, 4.1, 5.0, 6.0      |
| AB Dental Devices          | Narrow Platform                        | -                       |
| Alpha Bio                  | Internal Hex                           | 4.5                     |
|                            | TCT System                             | -                       |
| Anthogyr                   | Anthofit® HE System Conical Abutment   | 4.1                     |
|                            | Anthofit® OI System                    | 3.5                     |
|                            | Anthofit® OI System Conical Abutment   | 3.5                     |
|                            | Axiom REG/PX Platform                  | 4.0                     |
|                            | Axiom REG/PX Straight Conical Abutment | -                       |
| BEGO Semados®              | S/RI-Line                              | 3.25/3.75/4.1/4.5, 5.5  |
| BioHorizons®               | Abutment for Screw                     | 3.5/4.0, 5.0            |
|                            | Conical internal connection            | 3.0, 3.5, 4.5, 5.7      |
|                            | External implant system                | 3.5, 4.0, 5.0, 6.0      |
|                            | Single Stage Restoration               | 3.5, 4.5, 5.7           |
| Biotech Dental<br>Implants | B.I.S. Conic                           | 3.5/3.9, 4.0/4.4, 5/5.4 |
|                            | B.I.S. Conic Base Analog               | 4.4                     |

AND/OR Over-The-Counter Use Prescription Use X (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 3

{9}------------------------------------------------

| Manufacturer                  | Name                               | Sizes                   |
|-------------------------------|------------------------------------|-------------------------|
| BTI®                          | Externa® Platform (abutment level) | 4.1                     |
|                               | Externa® Platform (implant level)  | 4.1                     |
|                               | Interna® Platform                  | 4.1                     |
|                               | Tiny® Platform                     | 3.5                     |
| Camlog®                       | Camlog Bar Abutment                | 3.3, 3.8, 6.0           |
|                               | Camlog Implant System              | 3.3, 3.8, 4.3, 5.0, 6.0 |
|                               | CONELOG Implant System             | 3.8                     |
| Dentsply (Ankylos®)           | Balance Base Abutment              | -                       |
|                               | Balance Base Abutment Narrow       | -                       |
|                               | Balance Base C                     | -                       |
|                               | Regular C/X Abutment               | -                       |
| Dentsply (Astra Tech®)        | 20° UnlAbutment                    | 3.5/4.0                 |
|                               | 45° UniAbutment                    | 3.5/4.0                 |
|                               | Astra ST                           | 3.5                     |
|                               | Astra Tech                         | Yellow, Aqua, Lilac     |
| Dentsply (XIVE®)              | FRIADENT®                          | 3.0, 3.4, 3.8, 4.5, 5.5 |
|                               | FRIADENT® MP                       | 3.8, 4.5, 5.5           |
|                               | TG Platform                        | 3.4-4.5                 |
| Hiossen                       | ET III SA                          | mini, regular           |
| HI-TEC Implants               | Universal Unit System              | -                       |
| Implant Direct                | Attachment International           | 3.25, 3.3               |
|                               | Legacy™                            | 3.0, 3.5, 4.5, 5.7      |
|                               | RePlant®/RePlus®                   | 3.5, 4.3, 5.0           |
|                               | Screwindirect                      | -                       |
|                               | Screwindirect                      | 5.0                     |
|                               | ScrewPlant®/ScrewPlus®             | 3.7, 4.7, 5.7           |
| Implant Direct<br>(Sybron)    | Endopore® Internal Hex             | 4.1                     |
|                               | Entegra™ External Hex              | 3.5, 4.1, 5.0           |
|                               | PITT-EASY® Paracentric Line        | 3.75, 4.0, 4.9          |
|                               | SybronPRO™ Series Hex              | 3.5, 4.1                |
| LifeCore Dental<br>(Keystone) | Genesis System                     | 3.8, 4.5, 5.5           |
|                               | PrimaConnex® System                | 3.5, 4.1, 5.0/7.0/8.0   |
|                               | PrimaConnex® System Multi-unit     | 3.5/4.1, WD             |
|                               | Renova® Implant System             | SDI 3.75, RDI 4.5       |
|                               | Restore® Implant System            | SD 3.3, RD 3.75/4.0, WD |
|                               | Stage-1 system RDS                 | -                       |
| MIS®                          | Internal Hex                       | -                       |
|                               | Multi-Unit system                  | -                       |
|                               | Standard Platform                  | -                       |
|                               | Wide Interial Hex                  | -                       |

Prescription Use _____________________________________________________________________________________________________________________________________________________________

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 2 of 3

{10}------------------------------------------------

| Manufacturer      | Name                             | Sizes                                           |
|-------------------|----------------------------------|-------------------------------------------------|
| Neoss             | Access Abutment                  | -                                               |
|                   | Neoss Implant System             | 3.25                                            |
| Nobel Blocare     | Brånemark System (External Hex)  | NP, RP, WP                                      |
|                   | Hex                              | 2.0                                             |
|                   | Internal Hexagon                 | -                                               |
|                   | Multi-Unit Abutment              | NP/RP, WP                                       |
|                   | Nobel Novum                      | -                                               |
|                   | Nobel Replace (Internal Trilobe) | NP, RP, WP, 6.0                                 |
|                   | NobelActive (Conical Connection) | NP, RP, WP                                      |
|                   | Replace Hex Abutment             | 3.5, 4.3, 3.25HL, 3.8/4.5<br>HL, 5.0 HL, 6.0 HL |
| Osslem            | TS & GS System                   | Mini 4.0, standard 4.5                          |
|                   | US System Esthetic Low           | -                                               |
|                   | US System Regular platform       | -                                               |
| SIC Implants      | Sic Platform                     | 3.3, 4.2                                        |
| Simpler Implants  | Implant Analog                   | 3.25, 4.0                                       |
| Southern Implants | External Hex                     | 4.0, 5.0                                        |
| Straumann®        | Bone Level                       | 3.3, 4.1/4.8                                    |
|                   | Multi-base                       | NC 3.5, RC 4.5, RC 6.5                          |
|                   | Octagon                          | 3.5, 4.8, 6.5                                   |
|                   | Ortho System                     | 4.2                                             |
|                   | SynOcta®                         | RN, WN                                          |
| Thommen Medical   | SPI®RETAIN DIRECT System         | -                                               |
|                   | SPI®VARIOmulti System            | 4.0, 4.5, 5.0                                   |
|                   | VARIOeco system                  | 3.5, 4.0, 4.5, 5.0, 6.0                         |
| Titan Implant     | Internal Hex platform            | 4.5, 5.7                                        |
| Zimmer® Dental    | AdVent                           | 4.5                                             |
|                   | Calcitek                         | 3.25, 4.0                                       |
|                   | Screw-Vent                       | 3.5, 4.5, 5.7                                   |
|                   | Specta-Cone                      | -                                               |
|                   | Spline® Implant System           | 3.25, 3.75/4.0, 5.0                             |
|                   | SwissPlus® Implant System        | 3.8, 4.8                                        |
|                   | Tapered abutment                 | 4.5                                             |

Prescription Use _____________________________________________________________________________________________________________________________________________________________

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NÈEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Sheena A. Green
2014.07.24 14:04'00'

Page 3 of 3

---

**Source:** [https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/NHA/K132749](https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/NHA/K132749)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
