K132067 · Hiossen, Inc. · NHA · Dec 20, 2013 · Dental
Device Facts
Record ID
K132067
Device Name
MULTI ANGLED ABUTMENT SYSTEM
Applicant
Hiossen, Inc.
Product Code
NHA · Dental
Decision Date
Dec 20, 2013
Decision
SESE
Submission Type
Special
Regulation
21 CFR 872.3630
Device Class
Class 2
Indications for Use
Multi Angled Abutment system is intended for use with a dental implant to provide support for prosthetic restorations such as bridges, or overdentures.
Device Story
Multi Angled Abutment system is a titanium alloy dental abutment used to support prosthetic restorations (bridges, overdentures) on endosseous dental implants. The system includes the abutment, an abutment screw, and a combined carrier tool to facilitate placement. It is used by dental professionals in a clinical setting. The device is functionally identical to the predicate (K123755), with the addition of a carrier to improve handling convenience during the surgical/prosthetic procedure. The abutment connects to the HTIII SA Fixture System.
Clinical Evidence
No clinical studies were submitted. Substantial equivalence is based on design and intended use similarity to the predicate device.
Technological Characteristics
Material: Titanium alloy. Design: Multi-angled (17°, 30°) abutment with 4.8mm platform. Includes combined carrier tool for placement. No software or electronic components.
Indications for Use
Indicated for patients requiring prosthetic restoration support, such as bridges or overdentures, via dental implants.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
Predicate Devices
Multi Angled Abutment, OSSTEM Implant Co., Ltd. (K123755)
K151375 — Inclusive Titanium Abutments compatible with: Hiossen HG Implant System · Prismatik Dentalcraft, Inc. · Oct 7, 2015
K161103 — US SA Implant System · Osstem Implant Co., Ltd. · Feb 2, 2017
K210903 — IS Multi Unit Abutment System · Neobiotech Co., Ltd. · Jul 23, 2021
K093643 — MULTI- UNIT ABUTMENTS FOR STRAUMANN AND ASTRA TECH IMPLANT SYSTEMS · Nobel Biocare AB · Jun 24, 2010
Submission Summary (Full Text)
{0}------------------------------------------------
KNB2067
DEC 2 0 2013
Image /page/0/Picture/2 description: The image shows the logo and contact information for OSSTEM Implant Co., Ltd. The address is #507-8 Geoje3-Dong Yeonje-Gu Busan, 611-804 Republic of Korea. The telephone number is +82 51 850-2500, the fax number is +82 51 850-4341, and the website is www.osstem.com.
## 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
> OSSTEM Implant Co., Ltd. #507-8 Geoje3-Dong Yeonje-Gu
Mr. Hee Kwon Son
+82 51 850 2575
HIOSSEN Inc.
Patrick Lim 888 678 0001
Busan, 611-804, Republic of Korea
85 Ben Fairless Dr. Fairless Hills, PA 19030
Date: June 27, 2013
1. Company and Correspondent making the submission:
- Submitter`s Name :
- Address :
- Contact :
- Phone:
- Correspondent`s Name:
- Address:
- Contact:
- Phone:
- 2. Device :
Trade or (Proprietary) Name : Common or usual name : Classification Name :
Multi Angled Abutment system Dental Abutment Abutment, implant, dental, endosseous 21CFR872.3630 Class II NHA
- 3. Predicate Device :
Multi Angled Abutment, OSSTEM Implant Co., Ltd., K123755
- 4. Description :
The Multi Angled Abutment is device made of titanium alloy intended for use as an aid in prosthetic restoration.
The Multi Angled Abutment is used with Esthetic -low Cylinders (only Non-Hex) in the US System, K62030 and connected to HTIII SA Fixture in the HTIII SA Fixture System, K101096
- I) The Multi Angled Abutment system consists of Abutment combined carrier that is tool to carry Multi angled abutment. and Abutment Screw.
QS-Q1-505-3(Rev.0)
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the text "OSSTEM Implant Co., Ltd." The text is in bold, black font and is centered on a white background. The text appears to be the name of a company.
#507-8 Geoje3-Dong Yeonje-Gu Busan, 611-804 Republic of Korea Tel: +82 51 850-2500 Fax: +82 51 850-4341 www.osstem.com
- 2) The Multi Angled Abutment system is exactly same with Multi Angled Abutment (K123755) except combined carrier
| | Multi Angled Abutment system | Multi Angled Abutment |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Osstem Implant Co., Ltd | Osstem Implant Co., Ltd |
| 510(k) Number | New | K123755 |
| Design | Image: Multi Angled Abutment system design | Image: Multi Angled Abutment design |
| Intended use | Multi Angled Abutment is intended for<br>use with a dental implant to provide<br>support for prosthetic restorations such<br>as bridges, or overdentures. | Multi Angled Abutment is intended for<br>use with a dental implant to provide<br>support for prosthetic restorations such<br>as bridges, or overdentures. |
| Abutment<br>Angle(°) | 17, 30 | 17, 30 |
| platform(Ø) | 4.8 | 4.8 |
| Connection | The Multi Angled Abutment system is exactly same with Multi Angled<br>Abutment(K123755) except combined carrier.<br>Carrier is just to carry the Multi Angled Abutment to implanted Fixture | |
### - Substantial Equivalence Matrix
5. Indication for use :
The Multi Angled Abutment system is intended for use with a dental implant to provide support for prosthetic restorations such as bridges, or overdentures.
ਾਂ
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for OSSTEM Implant. The logo is in black and white and features the word "OSSTEM" in a stylized font. Below the word "OSSTEM" is the word "IMPLANT" in a smaller, sans-serif font. The logo is simple and modern, and it is likely used to represent a company that manufactures or sells dental implants.
OSSTEM Implant Co., Ltd.
#507-8 Geoje3-Dong Yeonje-Gu Busan, 611-804 Republic of Korea Tel: +82 51 850-2500 Fax: +82 51 850-4341 www.osstem.com
### 6. Review :
The Multi Angled Abutment system is merely improved for movement convenience of abutment by combination of carrier therefore, The Multi Angled Abutment is nothing changed from predicated Multi Angled Abutment (K123755) except combined carrier
#### 7. Summary of nonclinical testing
The risks associated with the modification were evaluated via a DCAS table and it was determined that no performance testing is needed. However, minor modifications were made to the labeling.
- 8. Summary of clinical testing No clinical studies are submitted
#### 9. Conclusions
In accordance with the Federal Food, Drug and Cosmetic Act. 21 CFR Part 807, and based on the information provided in this premarket notification Osstem Implant Co., Ltd. concludes that the Multi Angled Abutment system is substantially equivalent to the predicate devices as described herein.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird-like figure.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 20, 2013
OSSTEM Implant Company, Limited C/O Mr. Patrick Lim HIOSSEN Incorporated 85 Ben Fairless Drive Fairless Hills, PA 19030
Re: K132067
Trade/Device Name: Multi Angled Abutment System Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: October 31, 2013 Received: November 25, 2013
Dear Mr. Lim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
ff your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Lim
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
mer -
for
Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
DSSTEM Implant Co., Lt #507-8 Geoje3-Dong Yeonje-Gu Busan, Tel: +82 51 850-2500 Fax: +82 51 850-4341 www.osstem.com
# Indications for Use
510(k) Number K 132067
... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ..
Device Name : Multi Angled Abutment system
Indication for use : Multi Angled Abutment system is intended for use with a dental implant to provide support for prosthetic restorations such as bridges, or overdentures.
Prescription Use_ X (Per 21CFR801 Subpart D) OR Over-The-Counter Use (Per 21CFR807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mary Mary S. Runner-5
Susan Run DDS, MA 2013-12-20 08:45:50-05-00
QS-Q1-505-2(Rev.0)
Letter(8.5 X I I in)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.