ET PROSTHETIC SYSTEM

K130662 · Osstem Implant Co., Ltd. · NHA · Jan 10, 2014 · Dental

Device Facts

Record IDK130662
Device NameET PROSTHETIC SYSTEM
ApplicantOsstem Implant Co., Ltd.
Product CodeNHA · Dental
Decision DateJan 10, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3630
Device ClassClass 2

Indications for Use

ET Prosthetic System is intended for use with a dental implant to provide support for prosthetic restorations such as crowns, bridges, or overdentures.

Device Story

ET Prosthetic System consists of titanium/titanium alloy abutments and screws used to support dental prosthetic restorations (crowns, bridges, overdentures). Device is compatible with HIOSSEN HTIII SA Fixture. Components include rigid/transfer abutments, FreeForm ST abutments, multi-abutments, and multi-angled abutments. Used by dental professionals in clinical settings to connect prosthetic restorations to endosseous dental implants. Provides mechanical support for dental rehabilitation. Benefits include restoration of oral function and aesthetics for patients with missing teeth.

Clinical Evidence

No clinical studies were submitted. Evidence is based on bench testing, including fatigue testing conducted in accordance with FDA guidance and ISO 14801, demonstrating compliance and similarity to predicate devices.

Technological Characteristics

Materials: Titanium Alloy Ti 6Al 4V (ASTM F136). Surface: Partially TiN coated and uncoated. Design: Rigid/transfer, FreeForm ST, multi, and multi-angled abutments. Energy source: None (mechanical). Connectivity: None. Sterilization: Not specified. Software: None.

Indications for Use

Indicated for patients requiring prosthetic restoration (crowns, bridges, or overdentures) supported by dental implants. No specific age or gender restrictions provided.

Regulatory Classification

Identification

An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.

Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ JAN 1 0 2014 ### OSSTEM Implant Co., Ltd #507-8 Geoje3-Dong Yeonje-Gu Busan, 611-804 Republic of Korea Tel: +82 51 850-2500 Fax: +82 51 850-4341 www.osstem.com # 510(k) Summary This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92. Date: February 26, 2013 1. Company and Correspondent making the submission: - Submitter's Name : - Address : - Contact : - Phone: - Correspondent's Name: - Address: - Contact: - Phone: 2. Device : Trade or (Proprietary) Name : Common or usual name : Classification Name : ET Prosthetic System Dental Abutment Endosseous dental implant abutment 21CFR872.3630 Class II NHA OSSTEM Implant Co., Ltd. Mr. Hee Kwon Son +82 51 850 2575 HIOSSEN Inc. Patrick Lim 888 678 000 I #507-8 Geoje3-Dong Yeonje-Gu Busan, 611-804, Republic of Korea 85 Ben Fairless Dr. Fairless Hills, PA 19030 - 3. Predicate Device: Prosthetic System, K110308. Osstem Implant Co., Ltd ET/SS Implant System, K120847, Osstem Implant Co., Ltd - 4. Description: - 1) The ET Prosthetic System is device made of titanium and titanium alloy intended for use as an aid in prosthetic restoration. It consists of Abutments and Abutment Screws. It's surfaces are partially TiN coated and uncoated. - 2) The ET Prosthetic System is similar to other commercially available products based on the intended use, the technology used, the claims, the material composition employed and performance characteristics. QS-Q1-505-3(Rev.0) Letter(8.5 X 11in) {1}------------------------------------------------ ## OSSTEM® OSSTEM Implant Co., Ltd. #507-8 Geoje3-Dong Yeonje-Gu Busan, 611-804 Republic of Korea Tel: +82 51 850-2500 Fax: +82 51 850-4341 www.osstem.com - 3) The ET Prosthetic System is substantially equivalent in design, function and intended use to the Prosthetic System(K110308) and ET/SS Implant System (K120847)of Osstem Implant Co., Ltd. - 4) These proposed abutments are compatible with the HIOSSEN Implant fixture only. HTIII SA Fixture (K101096) #### - Substantial Equivalence Matrix | | ET Prosthetic System | ET/SS Implant System | |---------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------| | Manufacturer | Osstem Implant Co., Ltd | Osstem Implant Co., Ltd | | 510(k) Number | New | K120847 | | Design | Image: Rigid / Transfer Abutment | Image: Rigid / Transfer Abutment | | | Rigid / Transfer Abutment | Rigid / Transfer Abutment | | Design | Image: FreeForm ST Abutment | Image: FreeForm ST Abutment | | | FreeForm ST Abutment | FreeForm ST Abutment | | Intended use | ET Prosthetic System is intended<br>for use with a dental implant to<br>provide support for prosthetic<br>restorations such as crowns, bridges,<br>or overdentures. | Intended for use as an aid in<br>prosthetic rehabilitation. | | Material | Titanium Alloy<br>Ti 6Al 4V (ASTM F136) | Titanium Alloy<br>Ti 6Al 4V (ASTM F136) | | SE | - Rigid / Transfer Abutment:<br>No changes in function and intended use. Fully the same except G/H<br>Shape.<br>- FreeForm ST Abutment:<br>No changes in function and intended use Fully the same.<br>Art# of diameter 5.0 mm type us added . | | QS-QI-505-3(Rev.0) Letter(8.5 X 11in) {2}------------------------------------------------ OSTEM® # THEM Implant Co., I 507-8 Geoje3-Dong Yeonie-Gu Busan. 611-804 Republic of K Tel: +82 51 850-2500 Fax: +82 51 850-4341 www.osstem.com | Part Name | Multi Abutment | Convertible<br>Abutment | Multi Angled<br>Abutment | |---------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------|-------------------------------------| | 510K | New | K120847 | K110308 | | Material | Titanium Alloy | Titanium Alloy | Titanium Alloy | | Design | Image: Multi Abutment Design | Image: Convertible Abutment Design | Image: Multi Angled Abutment Design | | Type | One body | One body | Two piece<br>(Abutment + Screw) | | Material | Titanium Alloy | Titanium Alloy | Titanium Alloy | | Compatible<br>Fixture | HTIII SA Fixture<br>(K101096). | HTIII SA Fixture<br>(K101096). | HTIII SA Fixture<br>(K101096). | | Upper<br>Connection<br>Compatible<br>Device | US Esthetic low<br>Cylinders | Convertible Cylinders | US Esthetic low<br>Cylinders | | SE | No changes in function and intended use from the predicate device, convertible abutment_K120847.<br>Upper connection part is changed to use cylinder that is used with the<br>predicate device, multi angled abutment_K110308. | | | #### 5. Indication for use : ET Prosthetic System is intended for use with a dental implant to provide support for prosthetic restorations such as crowns, bridges, or overdentures. #### 6. Review : ET Prosthetic System has same material and indication for use and similar design and technological characteristics as the predicate device. Safety tests including biocompatibility have been considered to ensure the devices comply with the applicable International and US regulations. - 7. Summary of nonclinical testing The Fatigue testing was conducted according to the "Guidance for industry and FDA staff Class II Special Controls Guidance Document Root-form Endosseous Dental Implants and Endosseous Dental Abutment" and ISO 14801 Dentistry - Fatigue test for endosseous dental implants with the worst case scenario. The results are in compliance with it and were similar to previously cleared predicate devices." QS-QI-505-3(Rev.0) Letter(8.5 X 11in) {3}------------------------------------------------ OSSTEM Implant Co., Ltd. OSSTEM® #507-8 Geoje3-Dong Yeonje-Gu Busan, 611-804 Republic of Korea Tel: +82 51 850-2500 Fax: +82 51 850-4341 www.osstem.com - 8. Summary of clinical testing No clinical studies are submitted - 9. Conclusions In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, and based · on the information provided in this premarket notification Osstem Implant Co., Ltd. concludes that the ET Prosthetic System is substantially equivalent to the predicate devices as described herein QS-QI-505-3(Rev.0) Letter(8.5 X 11in) - 4/4 - {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three legs and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the eagle. The text is in all caps and appears to be in a sans-serif font. The logo is black and white. #### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 January 10, 2014 OSSTEM Implant Company, Limited C/O Mr. Patrick Lim Manager HiOSSEN Incorporated 85 Ben Fairless Drive Fairless Hills, PA 19030 Re: K130662 Trade/Device Name: ET Prosthetic System Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: November 29, 2013 Received: December 4, 2013 Dear Mr. Lim: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {5}------------------------------------------------ Page 2 - Mr. Lim Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address Sincerely yours, http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. FOR Erin I. Keith. M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ ## OSSTEM Implant Co., #507-8 Geoje3-Dong Yeonje-Gu Busan, 611-804 Republic of Korea Tel: +82 51 850-2500 Fax: +82 51 850-4341 www.osstem.com ### Indications for Use 510(k) Number K _130662 Device Name : ET Prosthetic System Indication for use : ET Prosthetic System is intended for use with a dental implant to provide support for prosthetic restorations such as crowns, bridges, or overdentures. Prescription Use X (Per 21CFR801 Subpart D) OR Over-The-Counter Use (Per 21CFR807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Marys. Runner -S 2074.01.09 18.462-05'00' Susan R QS-QI-505-2(Rev.0) Letter(8.5 X 1 I I in) - | -
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...