The ISUS Implant Suprastructures are indicated for attachment to dental implants or abutments in the treatment of partially or totally edentulous jaws for the purpose of restoring chewing function. The ISUS Implant Suprastructures are intended for attachment to a minimum of two (2) implants. ISUS Implant Suprastructures are indicated for compatibility with the following implant and abutment systems: Implants: Nobel Biocare Replace Select: NP (3.5mm), RP (4.3mm), WP (5.0mm), and Replace Select 6.0mm; Nobel Biocare Active Internal: NP (3.5mm), RP (4.3mm, 5.0mm); Zimmer Screw Vent: D3.5, D4.5, D5.7; Straumann: NN (3.5mm), RN (4.8mm), WN (6.0mm); Straumann Bone Level: NC (3.3mm), RC (4.1mm, 4.8mm); 3I Internal Connection: D3.4, D4.1, D5, D6; Friadent XiVE S: D3, D3.4, D3.8, D4.5, D5.5. Abutments: ASTRA TECH 20° and 45° UniAbutment; ASTRA TECH UniAbutment EV: 3.6; ANKYLOS Balance Base Abutment D5.5 and Narrow Abutment D4.2; Nobel Biocare Multi-Unit Abutment RP: 4.0 mm; Zimmer Tapered Abutment: 4.5mm; Straumann RN (4.8mm), WN (6.5 mm); Straumann Bone Level: Multi-Base Abutment D3.5, D4.5; Straumann Bone Level Angled Abutment: 4.0 mm; 3I Low Profile Abutment; Friadent XiVE MP D3.8, D4.5, D5.5; Friadent XiVE TG D3.8, D4.5, D5.5.
Device Story
ISUS Implant Suprastructures are customized metallic dental restorative devices (bars, bridges, hybrids) designed to restore chewing function in edentulous patients. Dental professionals provide patient dentition models; DENTSPLY converts these to digital representations using CAD/CAM technology to manufacture patient-specific suprastructures. Devices are screw-retained to dental implants or abutments. Bars support removable prostheses; bridges allow veneering for fixed prostheses; hybrids serve as substructures for bonded dentures. Used in dental clinics; operated by dental professionals. Output is a custom-fit metallic structure that integrates with existing implant systems to provide a stable foundation for dental restorations, benefiting patients by restoring oral function.
Clinical Evidence
No clinical data. Bench testing only. Dynamic fatigue testing performed per ISO 14801:2007 in simulated implant-supported constructs. Material properties verified per ISO 22674:2006 and ISO 9693:1999 for chemical composition, strength, and corrosion resistance.
Technological Characteristics
Customized metallic restorative devices; unalloyed CP titanium or cobalt-chromium alloy. CAD/CAM manufacturing. Screw-retained connection. Standards: ISO 14801 (fatigue), ISO 22674 (materials), ISO 9693 (metal-ceramic systems), ISO 17665-1/2 (sterilization).
Indications for Use
Indicated for partially or totally edentulous patients requiring restoration of chewing function via screw-retained dental prostheses attached to a minimum of two dental implants or abutments.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{1}------------------------------------------------
K122424
# SECTION 5. 510(k) SUMMARY
for
### ISUS Implant Suprastructures
5 2013 JUL
#### 5.1 Submitter Information:
DENTSPLY International Susquehanna Commerce Center 221 West Philadelphia Street York, PA 17405
| Contact Person: | Helen Lewis |
|-------------------|--------------|
| Telephone Number: | 717-849-4229 |
| Fax Number: | 717-849-4343 |
June 20, 2013 Date Prepared:
#### 5.2 Device Name:
●
.
ISUS Implant Suprastructures
Classification Name: Endosseous dental implant abutment.
I I
- CFR Number: 21CFR 872.3630 .
- Device Class: ●
Proprietary Name:
- Product Code: NHA .
#### 5.3 Predicate Device:
| CAM StructSURE® Overdenture Bar | - K101582. |
|-------------------------------------|------------|
| Procera® Implant Bridge Overdenture | - K090069 |
| FRIALIT-2® EstheticBase Abutments | - K013438 |
#### 5.4 Description of Device:
The ISUS Implant Suprastructures are metallic dental restorative devices which are intended for attachment by screw retention to dental implants to aid in the treatment of partially and totally edentulous patients for the purpose of restoring chewing function. The ISUS system consists of the Suprastructure devices themselves as well as retention screw accessories which facilitate their attachment to the dental implants.
The design of the ISUS Implant Suprastructures is derived from models of patient dentition which are produced by the dental professional using standard dental techniques. The dental models are converted to digital representations by DENTSPLY to facilitate DENTSPLY's Computer Aided Design (CAD) and Computer Aided Manufacturing (CAM) of the customized, patient-specific ISUS Implant Suprastructure.
{2}------------------------------------------------
The ISUS Implant Suprastructures consist of three (3) device types:
- Bars Intended to act as a supporting structure to facilitate attachment of . removable dental prostheses.
4122424
- . Bridges - Intended for direct veneering using dental ceramics or resin composites resulting in a fixed, screw-retained prosthesis.
- Hybrids Intended as a metallic substructure to which finished dentures . are bonded to form a fixed prosthesis.
The ISUS Implant Suprastructures are manufactured in versions composed of titanium and cobalt-chromium alloy and feature customized designs derived from individual patient dental impressions.
#### 5.5 Indications for Use:
The ISUS Implant Suprastructures are indicated for attachment to dental implants or abutments in the treatment of partially or totally edentulous jaws for the purpose of restoring chewing function. The ISUS Implant Suprastructures are intended for attachment to a minimum of two (2) implants.
ISUS Implant Suprastructures are indicated for compatibility with the following implant and abutment systems:
Implants:
- . Nobel Biocare Replace Select: NP (3.5mm), RP (4.3mm), WP (5.0mm), and Replace Select 6.0mm
- . Nobel Biocare Active Internal: NP (3.5mm), RP (4.3mm, 5.0mm)
- Zimmer Screw Vent: D3.5, D4.5, D5.7 .
- Straumann: NN (3.5mm), RN (4.8mm), WN (6.0mm) .
- Straumann Bone Level: NC (3.3mm), RC (4.1mm, 4.8mm) .
- 31 Internal Connection: D3.4, D4.1, D5, D6 .
- Friadent XiVE S: D3, D3.4, D3.8, D4.5, D5.5 .
Abutments:
- ASTRA TECH 20° and 45° UniAbutment .
- ASTRA TECH UniAbutment EV: 3.6 .
- ANKYLOS Balance Base Abutment D5.5 and Narrow Abutment D4.2 .
- `Nobel Biocare Multi -Unit Abutment RP: 4.0 mm .
- Zimmer Tapered Abutment: 4.5mm .
- Straumann RN(4.8mm), WN (6.5 mm) .
- Straumann Bone Level: Multi-Base Abutment D3.5, D4.5 .
- . Straumann Bone Level Angled Abutment:4.0 mm
- 3I Low Profile Abutment .
- Friadent XiVE MP D3.8. D4.5. D5.5 .
- . Friadent XiVE TG D3.8, D4.5, D5.5
{3}------------------------------------------------
#### ર .6 Description of Safety and Substantial Equivalence:
# . Technological Characteristics.
The ISUS Implant Suprastructures are customized metallic restorative devices. The CAD design of the devices is derived by DENTSPLY from individual patient dental impressions produced by dental professionals. The fabrication of the devices is performed by DENTSPLY using automated CAM manufacturing methods. The ISUS Implant Suprastructures feature the following technical characteristics:
- Screw retained with connection to dental implants or abutments. .
- Unalloyed, commercially-pure (CP) titanium or cobalt-chromium . alloy construction.
- . Fabricated from homogenous, single-block raw material (CP titanium or cobalt-chromium alloy).
# Non-Clinical Performance Data.
Dynamic fatigue testing was conducted to evaluate the mechanical integrity of the ISUS device constructs. Representative worst-case test samples were constructed and subjected to cyclic loading in a simulated, implant-supported construct. Dynamic testing was conducted according to the method described in ISO 14801:2007 (Dentistry - Implants - Dynamic fatigue test for endosseous dental implants). The results of the dynamic testing were compared to the results of equivalent testing which was conducted on legally marketed predicate dental abutments as well as to published in vitro data regarding maximum bite and chewing forces in edentulous patients. The results of the dynamic fatigue testing showed that the lowest fatigue strength of the ISUS Implant Suprastructures supports substantial equivalence.
Additionally, the materials which comprise the the ISUS Implant Suprastructures have been tested according to ISO 22674:2006 (Dentistry -- Metallic materials for fixed and removable restorations and appliances) and ISO 9693:1999 (Metal- ceramic dental restorative systems) to verify that the requirements for: chemical composition. proof strength, elongation after fracture, corrosion resistance, linear thermal expansion, and metal-ceramic bond strength are met. The results of the testing showed that the materials of which the ISUS Implant Suprastructures are composed conform to the requirements for these characteristics as specified in the relevant standards.
Recommended sterilization parameters which have been validated according to ISO 17665-1: 2006 (Sterilization of health care products -- Moist heat -- Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices) and ISO 17665-2: 2009 (Sterilization of health care products - Moist heat - Part 2: Guidance on the application of ISO 17665-1) are provided in accordance with ANSI/AAMI ST79:2010 (Comprehensive guide to steam sterilization and sterility assurance in health care facilities).
{4}------------------------------------------------
# KIZZ424
### Clinical Performance Data.
Not applicable. No human clinical study was performed to support substantial equivalence.
# Conclusion as to Substantial Equivalence
The proposed ISUS. Implant Suprastructures are substantially equivalent to the predicate Biomet 3i CAM StrucSURE Overdenture Bar (K101582) and Procera®. Implant Bridge Overdenture (K090069). The proposed devices have the same intended use and indications for use, are composed of the same or similar materials, and are characterized by the same fundamental product technology as the predicate devices. Testing has been conducted on the materials which compose the ISUS Implant Suprastructures and the results show conformity to the relevant standards for the properties which were studied. Additionally, in vitro test data has been to confirm the mechanical integrity of of the included ISUS Implant Suprastructures through dynamic fatigue testing in representative worst case constructs. The fatigue strength results for the ISUS Implant Suprastructures support substantial equivalence.
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three bars representing the agency's commitment to health, services, and people. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
# July 5, 2013
Ms. Helen Lewis Director, Corporate Regulatory Affairs DENTSPLY International, Incorporated Susquehanna Commerce Center 221 West Philadelphia Street YORK PA 17405
Re: K122424
Trade/Device Name: ISUS Implant Suprastructures Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: March 26, 2013 Received: March 27, 2013
### Dear Ms. Lewis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Lester 2013.07.05 13:02:20
Acting for Kwame Ulmer, M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
### Enclosure
{7}------------------------------------------------
## SECTION 4. INDICATIONS FOR USE STATEMENT
510(k) Number (if known): K122424
Device Name:
ISUS Implant Suprastructures
# Indications for Use:
The ISUS Implant Suprastructures are indicated for attachment to dental implants or abutments in the treatment of partially or totally edentulous jaws for the purpose of restoring chewing function. The ISUS Implant Suprastructures are intended for attachment to a minimum of two (2) implants.
ISUS Implant Suprastructures are indicated for compatibility with the following implant and abutment systems:
### Implants:
- . Nobel Biocare Replace Select: NP (3.5mm), RP (4.3mm), WP (5.0mm) and Replace Select 6.0mm
- Nobel Biocare Active Internal: NP (3.5mm), RP (4.3mm, 5.0mm) .
- Zimmer Screw Vent: D3.5, D4.5, D5.7
- Straumann: NN (3.5mm), RN (4.8mm), WN (6.0mm) �
- . Straumann Bone Level: NC (3.3mm), RC (4.1mm, 4.8mm)
- . 31 Internal Connection: D3.4. D4.1, D5, D6
- Friadent XIVE S: D3, D3.4, D3.8, D4.5, D5.5 .
#### Abutments:
- ASTRA TECH 20° and 45° UniAbutment: .
- � ASTRA TECH UniAbutment EV: 3.6
- ANKYLOS Balance Base Abutment D5.5 and Narrow Abutment D4.2 �
- Nobel Biocare Multi -Unit Abutment RP: 4.0 mm .
- . Zimmer Tapered Abutment: 4.5 mm
- Straumann RN (4.8 mm), WN (6.5 mm) �
- Straumann Bone Level: Multi-Base Abutment D3.5, D4.5 ●
- Straumann Bone Level Angled Abutment: 4.0 mm .
- ◆ 3I Low Profile Abutment
- Friadent XiVE MP D3.8, D4.5, D5.5 .
- Friadent XIVE TG D3.8, D4.5, D5.5 ◆
{8}------------------------------------------------
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Sheena ក្រុមកមួយ
2013.07 07 07 07 07 07 07 07 07 07 07 07 07 07 00 13.07 00 13.07 00 13.07 00 13.07 00 13.07 00 13.07 00 13.07 00 13.07 00 13.07 00 13.07 00 13.07 00 13.07 00 -04'00' y
for M. Susan Runner, DDS, ખેત
Page 101 of 333 . : :
(Division Sign-Off) (Division Sign-On)
Division of Anesthesiology, General Hospital Division of Anasulosio.org/w/infection Controllers
**510(k) Number:** k122424
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.