← Product Code [NHA](/submissions/DE/subpart-d%E2%80%94prosthetic-devices/NHA) · K112259

# NOBELACTIVE 3.0 GOLDADAPT ABUTMENTS (K112259)

_Nobel Biocare AB · NHA · Oct 28, 2011 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/NHA/K112259

## Device Facts

- **Applicant:** Nobel Biocare AB
- **Product Code:** [NHA](/submissions/DE/subpart-d%E2%80%94prosthetic-devices/NHA.md)
- **Decision Date:** Oct 28, 2011
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 872.3630
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

The NobelActive 3.0mm implant is indicated for use in the treatment of missing maxillary lateral incisors or the mandibular central and lateral incisors to support prosthetic devices, such as artificial teeth, in order to restore chewing function in partially edentulous patients. The NobelActive 3.0 implants may be put into immediate function provided that stability requirements detailed in the manual are satisfied.

## Device Story

NobelActive 3.0 GoldAdapt Abutment is a gold alloy dental abutment designed for use with the NobelActive 3.0 implant system. It serves as an interface between the endosseous implant and a prosthetic device (artificial tooth) to restore chewing function. Used in clinical dental settings by trained professionals. The device features an internal hex connection and is placed in the oral cavity. It functions by providing structural support for dental restorations. Clinical benefit includes restoration of esthetics and masticatory function in patients with specific tooth loss. Safety is supported by fatigue testing demonstrating strength to withstand anticipated oral forces.

## Clinical Evidence

No clinical testing was performed. Safety and effectiveness were supported by non-clinical bench testing, specifically fatigue testing conducted in accordance with the FDA guidance 'Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments'.

## Technological Characteristics

Material: Gold alloy. Connection: Internal hex. Angulation: 0 degrees. Intended for use with NobelActive 3.0 implant system. Fatigue testing performed per FDA Class II Special Controls guidance.

## Regulatory Identification

An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.

## Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)

## Predicate Devices

- NobelActive GoldAdapt Abutment ([K083100](/device/K083100.md))
- NobelActive 3.0 ([K102436](/device/K102436.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

510(k) Summary of Safety and Effectiveness 1.4

Kl b2259

| Submitted by:                          | Herbert Crane<br>Director, Global Regulatory Affairs                       | OCT 28 2011 |
|----------------------------------------|----------------------------------------------------------------------------|-------------|
| Address:                               | Nobel Biocare USA LLC<br>22715 Savi Ranch Parkway<br>Yorba Linda, CA 92887 |             |
| Telephone:                             | (714) 282-5074                                                             |             |
| Facsimile:                             | (714) 998-9348                                                             |             |
| Date of Submission:                    | August 5, 2011                                                             |             |
| Classification Name:                   | Endosseous Dental Implant Abutment (21 CFR 872.3630)                       |             |
| Product Code:                          | NHA                                                                        |             |
| Trade or Proprietary<br>or Model Name: | NobelActive 3.0 GoldAdapt Abutment                                         |             |
| Legally Marketed Devices:              | Nobel Biocare -- NobelActive 3.0 (K102436)                                 |             |

Device Description:

The NobelActive 3.0 GoldAdapt Abutment is an abutment intended to be used with the Nobel Biocare NobelActive 3.0 implant system. The abutments are made of a gold alloy.

Nobel Biocare - NobelActive GoldAdapt Abutment (K083100)

#### Indications for Use:

The NobelActive 3.0mm implant is indicated for use in the treatment of missing maxillary lateral incisors or the mandibular central and lateral incisors to support prosthetic devices, such as artificial teeth, in order to restore chewing function in partially edentulous patients. The NobelActive 3.0 implants may be put into immediate function provided that stability requirements detailed in the manual are satisfied.

### Non-Clinical Testing:

Non-clinical test data was used to support the decision of substantial equivalence. Non-clinical testing consisted of performance of fatigue testing in accordance with the FDA guidance Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments. The testing indicates that the device is strong enough to withstand the anticipated forces.

#### Clinical Testing;

No clinical testing was performed. Non-clinical test data was used to support the decision of safety and effectiveness.

## Conclusions:

The testing indicates that the abutments are substantially equivalent to the identified predicates.

Nobel Biocare Special 510(k) Notification NobelActive 3.0 GoldAdapt Abutment August 2011

{1}------------------------------------------------

5
6
6
11
↓

510(k) Summary of Safety and Effectiveness 1.4

-

Substantial Equivalence Comparison to Predicate Devices

| CHARACTERISTIC                 | PREDICATE<br>NobelActive GoldAdapt Abutment<br>(K083100)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | PREDICATE<br>NobelActive 3.0<br>(K102436)                                                                                                                                                                                                                                                                                                                                                                                                                         | CANDIDATE<br>NobelActive 3.0 GoldAdapt<br>Abutment                                                                                                                                                                                                                                                                                                                                                                                                                   |
|--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name                     | NobelActive                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | NobelActive                                                                                                                                                                                                                                                                                                                                                                                                                                                       | NobelActive                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Anatomical Site                | Oral Cavity                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Oral Cavity                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Oral Cavity                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Abutment Material              | Gold Alloy                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Titanium Vanadium Alloy                                                                                                                                                                                                                                                                                                                                                                                                                                           | Gold Alloy                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Abutment angulation            | 0 deg                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | 0 deg                                                                                                                                                                                                                                                                                                                                                                                                                                                             | 0 deg                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Implant/Abutment<br>Connection | Internal Hex                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Internal Hex                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Internal Hex                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Indications for Use            | Nobel Biocare's NobelActive implants<br>are endosseous implant intended to<br>be surgically placed in the bone of the<br>upper or lower jaw arches to provide<br>support for prosthetic devices, such<br>as an artificial tooth, in order to<br>restore patient esthetics and chewing<br>function. Nobel Biocare's NobelActive<br>implants are indicated for single or<br>multiple unit restorations in splinted or<br>non-splinted applications. Nobel<br>Biocare's NobelActive implants may<br>be placed immediately and put into<br>immediate function provided that initial<br>stability requirements detailed in the<br>manual are satisfied. | The NobelActive 3.0mm implant is<br>indicated for use in the treatment of<br>missing maxillary lateral incisors or the<br>mandibular central and lateral incisors<br>to support prosthetic devices, such as<br>artificial teeth, in order to restore<br>chewing function in partially edentulous<br>patients. The NobelActive 3.0 implants<br>may be put into immediate function<br>provided that stability requirements<br>detailed in the manual are satisfied. | The NobelActive 3.0mm implant is<br>indicated for use in the treatment of<br>missing maxillary lateral incisors or<br>the mandibular central and lateral<br>incisors to support prosthetic<br>devices, such as artificial teeth, in<br>order to restore chewing function in<br>partially edentulous patients. The<br>NobelActive 3.0 implants may be put<br>into immediate function provided<br>that stability requirements detailed in<br>the manual are satisfied. |

Nobel Biocare Special 510(k) Notification
NobelActive 3.0 GoldAdapt Abutment
August 2011

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

OCT 2 8 2011

Nobel Biocare AB C/O Mr. Herbert Crane Director, Global Regulatory Affairs Nobel Biocare USA LLC 22715 Savi Ranch Parkwav Yorba Linda, California 92887

Re: K112259

Trade/Device Name: NobelActvie 3.0 GoldAdapt Abutment Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: September 27, 2011 Received: September 28, 2011

Dear Mr. Crane:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

lf your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

# Page 2- Mr. Crane

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm 115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Wh for

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the word "Kiiasa" written in cursive. The letters are connected and slanted. The word is written in black ink on a white background.

510(k) Number (if known):

Device Name: NobelActive 3.0 GoldAdapt Abutment

## Indications For Use:

The NobelActive 3.0mm implant is indicated for use in the treatment of missing maxillary lateral incisors or the mandibular central and lateral incisors to support prosthetic devices, such as artificial teeth, in order to restore chewing function in partially edentulous patients. The NobelActive 3.0 implants may be put into immediate function provided that stability requirements detailed in the manual are satisfied.

Prescription Use × (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Susan Rumer

(Division Sign-Off) (Division Sign-On)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices

Page 1 of 1

510(k) Number

---

**Source:** [https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/NHA/K112259](https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/NHA/K112259)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
