INCLUSIVE TITANIUM ABUTMENTS FOR ASTRA TECH OSSEOSPEED IMPLANTS
K100993 · Prismatik Dentalcraft, Inc. · NHA · Dec 28, 2010 · Dental
Device Facts
Record ID
K100993
Device Name
INCLUSIVE TITANIUM ABUTMENTS FOR ASTRA TECH OSSEOSPEED IMPLANTS
Applicant
Prismatik Dentalcraft, Inc.
Product Code
NHA · Dental
Decision Date
Dec 28, 2010
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 872.3630
Device Class
Class 2
Indications for Use
The Inclusive® Titanium Abutments for Astra OsseoSpeed™ Implants are premanufactured prosthetic components directly connected to endosseous dental implants and are intended for use as an aid in prosthetic rehabilitation. They are compatible with the Astra Tech OsseoSpeed™ 3.0, 3.5, 4.0, 4.5, 5.0 implants.
Device Story
Inclusive® Titanium Abutments are premanufactured prosthetic components for endosseous dental implants; used to support cemented restorations. Device consists of titanium (Ti-6Al-4V ELI) abutment and screw; connects to implant via internal hexagon interface. Placed by dental professionals in clinical settings. Abutment provides structural support for prosthetic rehabilitation; screw secures abutment to implant. Benefits include restoration of dental function and aesthetics for edentulous or partially edentulous patients.
Clinical Evidence
No clinical data. Safety and effectiveness supported by non-clinical bench testing, specifically fatigue testing of worst-case platform samples in accordance with FDA guidance for Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments.
Technological Characteristics
Material: Titanium grade Ti-6Al-4V ELI (ASTM F-136). Design: Endosseous dental implant abutment with internal hexagon connection. Dimensions: Compatible with 3.0, 3.5, 4.0, 4.5, 5.0mm platform diameters. Energy source: None (mechanical).
Indications for Use
Indicated for patients requiring prosthetic rehabilitation via cemented restorations supported by Astra Tech OsseoSpeed™ 3.0, 3.5, 4.0, 4.5, or 5.0 endosseous dental implants.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
Atlantis™ Abutment for Astra Tech OsseoSpeed 3.0 Implant System
Submission Summary (Full Text)
{1}------------------------------------------------
K100993
DEC 2 8 2010
# 510(k) Summary
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA and 21 CFR 807.92.
### A. SUBMITTER INFORMATION
| Company Name: | Prismatik Dentalcraft, Inc. |
|------------------------|------------------------------------------------------|
| Company Address: | 4141 Mac Arthur Boulevard<br>Newport Beach, CA 92660 |
| Company Phone: | (949) 440-2683 |
| Company FAX: | (949) 440-2787 |
| Contact Person: | Kathleen Dragovich, (949) 399-1940 |
| Date Summary Prepared: | December 20, 2010 |
### B. DEVICE IDENTIFICATION
| Trade/Proprietary Name: | Inclusive <sup>®</sup> Titanium Abutments for- Astra OsseoSpeed <sup>™</sup> |
|-------------------------|------------------------------------------------------------------------------|
| | Implants |
| 21 CFR Reference: | 872.3630 |
| 21 CFR Common Name: | Endosseous Dental Implant Abutment |
| Classification: | Class II |
| Panel: | Dental NHA |
### C. IDENTIFICATION OF PREDICATE DEVICE
Astra Tech AB OsseoSpeed™ Narrow, K081666 (10-07-08) Trade/Proprietary Name: Astra Tech Implant - Dental System, K990304 (07-15-99) Astra Tech Implant – Dental System, K931767 (02-08-94)
### D. DEVICE DESCRIPTION
Inclusive® Titanium Abutments for- Astra OsseoSpeed™ Implants are endosseous implant abutments which are placed in to the dental implant to provide support for a prosthetic restoration. The subject abutments are indicated for cemented restorations. These abutments are made of titanium grade Ti-6Al-4V ELI (meets ASTM Standard F-136). The abutment is placed over the implant shoulder and is mounted into the implant with a screw. These abutments are compatible with the Astra OsseoSpeed™ Implants.
### INDICATIONS FOR USE ن
The Inclusive® Titanium Abutments for Astra OsseoSpeed™ Implants are premanufactured prosthetic components directly connected to endosseous dental implants and are intended for use as an aid in prosthetic rehabilitation. They are compatible with the Astra Tech OsseoSpeed™ 3.0, 3.5, 4.0, 4.5, 5.0 implants.
{2}------------------------------------------------
# F. SUBSTANTIAL EQUIVALENCE
The Inclusive® Abutments for Astra OsseoSpeed™ Implants are substantially equivalent to the Atlantis™ Abutment for Astra Tech OsseoSpeed 3.0 Implant System, and the Astra Tech Implants Dental System Abutments. These abutments are substantially equivalent in intended use, indications for use, material, design and performance.
| Element of<br>Comparison | Prismatik's<br>Inclusive® Abutments for<br>Astra OsseoSpeed™<br>Implants | Astra Tech AB<br>OsseoSpeed™<br>Narrow, K081666 | Astra Tech Implant –<br>Dental System,<br>K990304 |
|-------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|
| Material | Ti-6Al-4V ELI | Ti-6Al-4V ELI | CP Titanium |
| Platform<br>Diameters<br>(mm) | 3.0, 3.5, 4.0, 4.5, 5.0mm | 3.0, 3.5, 4.0, 4.5, 5.0mm | 3.0, 3.5, 4.0, 4.5,<br>5.0mm |
| Indications | The Inclusive® Titanium<br>Abutments for Astra<br>OsseoSpeed™ Implants are<br>premanufactured prosthetic<br>components directly<br>connected to endosseous<br>dental implants and are<br>intended for use as an aid in<br>prosthetic rehabilitation.<br>They are compatible with the<br>Astra Tech OsseoSpeed™ 3.0,<br>3.5, 4.0, 4.5, 5.0 implants. | The Atlantis Abutment is<br>intended for use with an<br>endosseous implant to<br>support a prosthetic device<br>in a partially or completely<br>edentulous patient. It is<br>intended for use to support<br>single and multiple tooth<br>prosthesis, in the maxillary<br>lateral incisors and<br>mandibular lateral and<br>central incisors. The<br>prosthesis can be cement<br>retained to the abutment.<br>The abutment screw is<br>intended to secure the<br>abutment to the<br>endosseous implant. The<br>device is compatible with<br>the Astra Tech OsseoSpeed<br>3.0mm Implant | For use in selected fully<br>edentulous and partially<br>edentulous arches. |
| Design | implant/Abutment assembly<br>with abutment screw.<br>Abutment connection to<br>implant is an internal<br>hexagon. | Implant/Abutment<br>assembly with abutment<br>screw. Abutment<br>connection to implant is an<br>internal hexagon. | Implant/Abutment<br>assembly with abutment<br>screw. Abutment<br>connection to implant is<br>an internal hexagon. |
| Performance | Fatigue testing conducted by<br>Straumann® on stock Astra<br>Tech TiDesign™ abutments<br>and OsseoSpeed™ implants<br>showed mean fatigue<br>strength at 30° was around<br>110N and 125N for the 3.0<br>and 2.5mm implant<br>diameters, respectively. | Fatigue testing conducted<br>at Engineering Materials<br>Laboratory on angled<br>Inclusive® abutments and<br>Astra Tech OsseoSpeed™<br>implants showed mean<br>fatigue strength at 20" was<br>166.8N and 244.7N for the<br>3.0 and 3.5mm implant<br>diameters, respectively.<br>The higher fatigue strength<br>results can be attributed to | |
{3}------------------------------------------------
### G. PERFORMANCE DATA
Non-clinical test data was used to support the decision of safety and effectiveness. Clinical testing was not necessary. Non-clinical testing consisted of analysis of platforms to identify worst-case test samples and performance of fatigue testing of the worst-case samples in accordance with the FDA guidance Class II Special controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments. The testing indicates that the device is safe and effective for its intended use and performs as well or better than the predicate device.
## H. COMPARISON OF TECHNOLOGICAL DIFFERENCES
There are no known technological differences between the Inclusive® Titanium Abutments for- Astra OsseoSpeed™ Implants and those of the predicate devices.
4
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an emblem featuring a stylized depiction of a human figure embracing a bird, symbolizing care and protection.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Kathleen Dragovich Regulatory Affairs Specialist Prismatik Dentalcraft, Incorporated 4141 Mac Arthur Boulevard Newport Beach, California 92660
DEC 2 8 2000
Re: K100993
Trade/Device Name: Inclusive® Titanium Abutments for Astra OsseoSpeed™ Implants Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: December 20, 2010 Received: December 22, 2010
Dear Ms. Dragovich:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
# Page 2 - Ms. Dragovich
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Anthony Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Prismatik Dentalcraft, Inc. The logo consists of a black triangle on the left and the company name on the right. The company name is written in a sans-serif font, with "PRISMATIK" on the top line and "DENTALCRAFT. INC" on the bottom line.
Prismatik Dentalcraft, Inc. 4141 Mac Arthur Boulevard Newport Beach, CA 92660
### INDICATIONS FOR USE STATEMENT
DEC 2 8 2010
510(K) Number: K100993
Device Name: Inclusive® Titanium Abutments for Astra OsseoSpeed™ Implants
Indications for Use:
The Inclusive® Titanium Abutments for Astra OsseoSpeed™ Implants are premanufactured prosthetic components directly connected to endosseous dental implants and are intended for use as an aid in prosthetic rehabilitation. They are compatible with the Astra Tech OsseoSpeed™ 3.0, 3.5, 4.0, 4.5, 5.0 implants.
Prescription Use _ X (Part 21 CFR 801 Subpart D) AND/OR
Over-the-Counter Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
R.S. Betz, DDS for Dr. Susan Renner
Division of Anesthesiology, General Hospital
Infection Control and Dental Devices
510(k) Number: K100993
б
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.