K100356 · Pou YU Biotechnology Co., Ltd. · NHA · Oct 5, 2010 · Dental
Device Facts
Record ID
K100356
Device Name
TDS PRECISION DENTAL BAR
Applicant
Pou YU Biotechnology Co., Ltd.
Product Code
NHA · Dental
Decision Date
Oct 5, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
TDS Precision Dental Bars are indicated for use with dental implants as a prosthetic framework to support and/or retain removable or fixed dental prostheses in the treatment of partially or totally edentulous jaws to restore chewing function. The abutment screws are intended to secure the bar to the endosseous implants.
Device Story
TDS Precision Dental Bars are custom-fabricated prosthetic frameworks; machined from titanium, zirconia, or zirconia-titanium; secured to endosseous dental implants via Ti-6Al-4V ELI alloy screws. Device serves as support/retention structure for fixed or removable dental prostheses. Used in clinical dental settings; operated by dental professionals. Compatibility spans multiple major implant systems (e.g., NobelBiocare, Straumann, Zimmer). Output is a physical framework restoring patient chewing function.
Clinical Evidence
Bench testing only. Mechanical fatigue testing performed per ISO 14801 on worst-case construction (angled abutment) to verify strength. Compatibility testing conducted to evaluate dimensions, tolerances, and rotation parameters for fit with designated implant systems.
Technological Characteristics
Materials: Machined titanium, zirconia, and zirconia-titanium. Abutment screws: Ti-6Al-4V ELI alloy. Design: Custom-fabricated prosthetic framework. Compatibility: Multi-brand dental implant systems. Testing: ISO 14801 fatigue testing. No software or electronic components.
Indications for Use
Indicated for partially or totally edentulous patients requiring prosthetic frameworks to support/retain fixed or removable dental prostheses. Contraindicated for highly angled abutments (30°) on implants <4 mm diameter in the posterior region due to limited implant strength.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
K10035p
# 510(k) Summary Pou Yu Biotechnology Co., Ltd TDS Precision Dental Bar
- ﺗﺸﺘﻪ
OCT - 5 2010
#### APPLICANT'S NAME AND ADDRESS
| Applicant's Name: | |
|-------------------|--|
| Address: | |
| Telephone: |
|-----------------|
| Fax: |
| Contact person: |
| Date Prepared: |
Pou Yu Biotechnology Co., Ltd. No. 6 Fugong Rd. Fusing Township Changhua County 506, Taiwan +886-(0)4-768-5660 x5122 +886-(0)4-768-9032 Daniel Tsao February 2, 2010
#### NAME OF THE DEVICE
| Trade/Proprietary Name: | TDS Precision Dental Bar |
|-------------------------|------------------------------------|
| Common Name: | Dental implant bar |
| Classification Name: | Endosseous dental implant abutment |
| | Class II, 21 CFR 872.3630 |
| Product Code: | NHA |
#### LEGALLY MARKETED DEVICES
BIOMET 3i , Incorporated. - CAM StructSURE® Precision Milled Bars - K080864 Cagenix, Incorporated. - Accuframe TM - K081157 Nobel Biocare USA LLC. - NobelProcera Implant Bridge Zirconia - K091907
### DESCRIPTION OF THE DEVICE
TDS Precision Dental Bars are prosthetic frameworks and made of machined titanium, zirconia-titanium, and zirconia materials. They are connected with dental implants utilizing the TDS screw which is made of Ti-6A1-4V ELI (Extra Low Interstitial) alloy. TDS Precision Dental Bars are
{1}------------------------------------------------
fabricated individually following instructions to each patient and serve to support and/or retain fixed dental prosthesis or removable dental prosthesis. They are compatible with dental implants manufactured by NobelBiocare, Camlog, BIOMET 3i, Zimmer, Lifecore, Astra, BioHorizons, Straumann, Dentsply Friadent, and Osstem.
## INTENDED USE OF THE DEVICE
TDS Precision Dental Bars are indicated for use with dental implants as a prosthetic framework to support and/or retain removable or fixed dental prostheses in the treatment of partially or totally edentulous jaws to restore chewing function. The abutment screws are intended to secure the bar to the endosseous implants.
| Implant brand | Model |
|-------------------|----------------------------------------------------------------------------|
| NobelBiocare | Branemark - NP, RP, WP; Replace - NP, RP, WP, 6.0 mm; NobelActive - NP, RP |
| Camlog | 3.8, 4.3, 5.0, 6.0 mm |
| BIOMET 3i | External - 3.4, 4.1, 5.0, 6.0 mm; Certain - 3.4, 4.1, 5.0, 6.0 mm |
| Zimmer | Tapered Screw-vent - 3.5, 4.5, 5.7 mm |
| Lifecore | External Hex - SD, RD, WD; Internal Hex - SDI 3.75, RDI 4.5/4.75 mm |
| Astra | OsseoSpeed - 3.5/4.0, 4.5/5.0 mm |
| BioHorizons | External - 3.5, 4.0, 5.0, 6.0 mm; Internal - 3.5, 4.5, 5.7 mm |
| Straumann | NN, RN, WN; Bone Level - NC 3.3, RC 4.1, RC 4.8 mm |
| Dentsply Friadent | XiVE - 3.0 mm; Frialit(XiVE) - 3.4, 3.8, 4.5, 5.5 mm; Frialit - 6.5 mm |
| Osstem | GS - Mini, Standard ; SS - Regular, Wide ; US - Mini, Regular, Wide |
TDS Titanium Precision Dental Bars, made of machined titanium material, are compatible with following implant systems:
TDS Zirconia Precision Dental Bars, made of machined zirconia material, are compatible with following implant systems:
| Implant brand | Model |
|---------------|----------------------------------|
| NobelBiocare | Branemark - NP, RP, WP |
| BIOMET 3i | External - 3.4, 4.1, 5.0, 6.0 mm |
| Lifecore | External Hex - SD, RD, WD |
| BioHorizons | External - 3.5, 4.0, 5.0, 6.0 mm |
| Straumann | NN |
| Osstem | US - Mini, Regular, Wide |
{2}------------------------------------------------
TDS Hybrid Precision Dental Bars, made of machined zirconia-titanium materials, are compatible with following implant systems:
| Implant brand | Model |
|-------------------|------------------------------------------------------------------------|
| NobelBiocare | Replace - NP, RP, WP, 6.0 mm; NobelActive - NP, RP |
| Camlog | 3.8, 4.3, 5.0, 6.0 mm |
| BIOMET 3i | Certain - 3.4, 4.1, 5.0, 6.0 mm |
| Zimmer | Tapered Screw-vent - 3.5, 4.5, 5.7 mm |
| Lifecore | Internal Hex - SDI 3.75, RDI 4.5/4.75 mm |
| Astra | OsseoSpeed - 3.5/4.0, 4.5/5.0 mm |
| BioHorizons | Internal - 3.5, 4.5, 5.7 mm |
| Straumann | RN, WN; Bone Level - NC 3.3, RC 4.1, RC 4.8 mm |
| Dentsply Friadent | XiVE - 3.0 mm; Frialit(XiVE) - 3.4, 3.8, 4.5, 5.5 mm; Frialit - 6.5 mm |
| Osstem | GS - Mini, Standard ; SS - Regular, Wide |
Note: Highly angled abutments (i.e. 30°) on implants with diameters less than 4 mm are not intended for the posterior region of the mouth due to limited strength of the implant.
#### TECHNOLOGICAL CHARACTERISTICS
The TDS Precision Dental Bar has the following similarities to the predicate devices which have been determined by FDA:
- · has the same intended use,
- · uses the same operating principle,
- · incorporates the same basic design,
- · incorporates the same materials, and
- · is produced using the same processes.
#### NON-CLINICAL TESTING DATA
Mechanical testing, according to ISO 14801 Dentistry - Fatigue test for endosseous dental implants, was conducted on a worst-case scenario with construction of an angle abutment to ensure that the strength of the abutment for TDS Precision Dental Bar is appropriate for its intended use. Compatibility testing was conducted on the abutments and corresponding dental implants with designated screws, the dimensions, tolerances and rotation parameters were evaluated in determining
{3}------------------------------------------------
appropriate fit.
These testing results show that TDS Precision Dental Bar made of titanium, zirconia-titanium and zirconia materials for their respective dental implant systems have sufficient mechanical strength for their intended clinical application and are compatible with the implant systems for which they are indicated for use.
### BASIS FOR SUBSTANTIAL EQUIVALENCE
Pou Yu Biotechnology Co. Ltd demonstrated that, for the purposes of FDA's regulation of medical devices, TDS Precision Dental Bar is substantially equivalent in intended use, materials, design principles and produced processes to predicate devices, each of which has been determined by FDA to be substantially equivalent to pre-amendment devices.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows a logo for the Department of Health & Human Services USA. The logo features a stylized eagle or bird symbol on the right side. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the bird symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Daniel Tsao Manager Pou Yu Biotechnology Company, Limited No. 6 Fugong Road Fusing Township, Changhua City CHINA (Taiwan) 506
OCT - 5 2010
Re: K100356
Trade/Device Name: TDS Precision Dental Bar Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: September 3, 2010 Received: September 9, 2010
Dear Mr. Tsao:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Tsao
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
h for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
#### Enclosure
{6}------------------------------------------------
## Indications for Use
510(k) Number (if known): 1100356
Device Name: TDS Precision Dental Bar
Indications for Use:
TDS Precision Dental Bars are indicated for use with dental implants as a prosthetic framework to support and/or retain removable or fixed dental prostheses in the treatment of partially or totally edentulous jaws to restore chewing function. The abutment screws are intended to secure the bar to the endosseous implants.
TDS Titanium Precision Dental Bars, made of machined titanium material, are compatible with following implant systems:
| Implant brand | Model |
|-------------------|----------------------------------------------------------------------------|
| NobelBiocare | Branemark - NP, RP, WP; Replace - NP, RP, WP, 6.0 mm; NobelActive - NP, RP |
| Camlog | 3.8, 4.3, 5.0, 6.0 mm |
| BIOMET 3i | External - 3.4, 4.1, 5.0, 6.0 mm; Certain - 3.4, 4.1, 5.0, 6.0 mm |
| Zimmer | Tapered Screw-vent - 3.5, 4.5, 5.7 mm |
| Lifecore | External Hex - SD, RD, WD; Internal Hex - SDI 3.75, RDI 4.5/4.75 mm |
| Astra | OsseoSpeed - 3.5/4.0, 4.5/5.0 mm |
| BioHorizons | External - 3.5, 4.0, 5.0, 6.0 mm; Internal - 3.5, 4.5, 5.7 mm |
| Straumann | NN, RN, WN; Bone Level - NC 3.3, RC 4.1, RC 4.8 mm |
| Dentsply Friadent | XiVE - 3.0 mm; Frialit(XiVE) - 3.4, 3.8, 4.5, 5.5 mm; Frialit - 6.5 mm |
| Osstem | GS - Mini, Standard ; SS - Regular, Wide ; US - Mini, Regular, Wide |
TDS Zirconia Precision Dental Bars, made of machined zirconia material, are compatible with following implant systems:
| Implant brand | Model |
|---------------|----------------------------------|
| NobelBiocare | Branemark - NP, RP, WP |
| BIOMET 3i | External - 3.4, 4.1, 5.0, 6.0 mm |
| Lifecore | External Hex - SD, RD, WD |
| BioHorizons | External - 3.5, 4.0, 5.0, 6.0 mm |
| Straumann | NN |
| Osstem | US - Mini, Regular, Wide |
OCT - 5 2010
{7}------------------------------------------------
TDS Hybrid Precision Dental Bars, made of machined zirconia-titanium materials, are compatible with following implant systems:
| Implant brand | Model |
|-------------------|------------------------------------------------------------------------|
| NobelBiocare | Replace - NP, RP, WP, 6.0 mm; NobelActive - NP, RP |
| Camlog | 3.8, 4.3, 5.0, 6.0 mm |
| BIOMET 3i | Certain - 3.4, 4.1, 5.0, 6.0 mm |
| Zimmer | Tapered Screw-vent - 3.5, 4.5, 5.7 mm |
| Lifecore | Internal Hex - SDI 3.75, RDI 4.5/4.75 mm |
| Astra | OsseoSpeed - 3.5/4.0, 4.5/5.0 mm |
| BioHorizons | Internal - 3.5, 4.5, 5.7 mm |
| Straumann | RN, WN; Bone Level - NC 3.3, RC 4.1, RC 4.8 mm |
| Dentsply Friadent | XiVE - 3.0 mm; Frialit(XiVE) - 3.4, 3.8, 4.5, 5.5 mm; Frialit - 6.5 mm |
| Osstem | GS - Mini, Standard ; SS - Regular, Wide |
Note: Highly angled abutments (i.e. 30°) on implants with diameters less than 4 mm are not intended for the posterior region of the mouth due to limited strength of the implant.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evatuation (ODE)
(Division Sign-Off)
Division of Anesthesiology, General Hospital
Infection Control, Dental Devices
Page *2*of*2*
510(k) Number: K100354
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.