K073032 · Lifecore Biomedical, Inc. · NHA · Nov 16, 2007 · Dental
Device Facts
Record ID
K073032
Device Name
PRIMACONNEX SD ESTHETIC CONTOUR ZI ABUTMENTS
Applicant
Lifecore Biomedical, Inc.
Product Code
NHA · Dental
Decision Date
Nov 16, 2007
Decision
SESE
Submission Type
Special
Regulation
21 CFR 872.3630
Device Class
Class 2
Indications for Use
PrimaConnex SD Esthetic Contour Zi Abutment is intended for use in conjunction with the PrimaConnex Internal Connection Implant System in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit cement retained restorations.
Device Story
Ceramic abutment component for dental implant systems; used in conjunction with PrimaConnex Internal Connection Implant System. Provides structural support for single or multiple-unit cement-retained restorations in edentulous mandibles and maxillae. Device functions as interface between endosseous implant and dental crown. Operated by dental professionals in clinical settings. Benefits patient by restoring dental function and aesthetics through stable prosthetic support.
Clinical Evidence
No clinical data. Substantial equivalence is based on design, material, and intended use comparisons to legally marketed predicate devices.
Technological Characteristics
Ceramic dental abutment; designed for internal connection to dental implants. Materials and design are identical to previously cleared predicate devices. Non-active, mechanical device.
Indications for Use
Indicated for patients with partially or fully edentulous mandibles and maxillae requiring single or multiple-unit cement-retained dental restorations using the PrimaConnex Internal Connection Implant System.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
Predicate Devices
PrimaConnex RD and WD Esthetic Contour Zi Abutment (K072572)
Esthetic Contour Straight and Angled Abutment for the PrimaConnex Internal Connection Implant System (K051614)
{0}------------------------------------------------
Special 510(k) ifecore Biomedical, Inc. PrimaConnex® SD Esthetic Contour Zi Abutment KXXXXXX
73032
# Section 8: 510(k) Summary (per 21 CFR 807.92)
### 1. Submitter's Name and Contact Person
| Lifecore Biomedical, Inc. | Judith Medlock-Hayes |
|---------------------------|-------------------------------------|
| 3515 Lyman Blvd | Regulatory Affairs Specialist |
| Chaska, MN 55318 | Ph: 952-368-6364; Fax: 952-368-4278 |
### 2. General Information
| Trade Name | PrimaConnex® SD Esthetic Contour Zi<br>Abutment |
|-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name | Ceramic Abutment |
| Classification Name | Endosseous Implant Abutment |
| Identification of Predicate Devices | PrimaConnex RD and WD Esthetic<br>Contour Zi Abutment, Lifecore<br>Biomedical (K072572)<br>Esthetic Contour Straight and Angled<br>Abutment for the PrimaConnex Internal<br>Connection Implant System, Lifecore<br>Biomedical (K051614)<br>PrimaConnex Ceramic Abutment,<br>Lifecore Biomedical (K062876) |
#### 3. Device Description
PrimaConnex SD Esthetic Contour 7: Abutment is intended for use in conjunction with the PrimaConnex Internal Connection Implant System in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit cement retained restorations.
#### 4. Intended Use
PrimaConnex SD Esthetic Contour Zi Abutment is intended for use in conjunction with the PrimaConnex Internal Connection Implant System in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit cement retained restorations.
{1}------------------------------------------------
Special 510(k) ifecore Biomedical, Inc.
#### 5, Substantial Equivalence Comparison
The summary of how the PrimaConnex® SD Esthetic Contour Zi Abutment is substantially equivalent to the PrimaConnex RD and WD Esthetic Contour Zi Abutment (K072572) is provided below:
- Both have the same intended use
- Both are manufactured from the same material .
- Both incorporate the same design .
- Both are packaged using the same materials and processes. .
The summary of how the PrimaConnex® SD Esthetic Contour Zi Abutment is substantially equivalent to the PrimaConnex Esthetic Contour Abutment (K051614) is provided below:
- Both have the same intended use
- Both incorporate the same design .
- Both are packaged using the same materials and processes. .
The summary of how the PrimaConnex® SD Esthetic Contour Zi Abutment is substantially equivalent to the PrimaConnex Ceramic Abutment (K062876) is provided below:
- Both have the same intended use
- . Both are manufactured from the same material
- Both incorporate the same design .
- Both are packaged and using the same materials and processes. .
In summary, it is the belief of Lifecore Biomedical, Inc. that the PrimaConnex SD Esthetic Contour Zi Abutment described in this submission is substantially equivalent to the predicate devices.
Confidential
{2}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 1 6 2007
Ms. Judith Medlock-Hayes Regulatory Affairs Specialist Lifecore Biomedical, Incorporated 3515 Lyman Boulevard Chaska, Minnesota 55318
Re: K073032
Trade/Device Name: PrimaConnex SD Esthetic Contour ZiAbutment Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: October 25, 2007 Received: October 29, 2007
Dear Ms. Medlock-Hayes:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
#### Page 2 - Ms. Clement
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Nil R.P. Ogden
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
cial 510(k) core Biomedical, Inc. PrimaConnex® SD Esthetic Contour Zi Abutments
KXXXXXX
## tachment B: Indications for Use Statement
0(k) Number (if known):
evice Name: PrimaConnex SD Esthetic Contour Zi Abutment
ndications for Use:
imaConnex SD Esthetic Contour Zi Abutment is intended for use in conjunction with he PrimaConnex Internal Connection Implant System in partially or fully edentulous andibles and maxillae, in support of single or multiple-unit cement retained estorations.
rescription Use X AND/OR Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 807 Subpart C)
LEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE F NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Susan Russoe
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
Page 1 of 1
510(k) Number. K673032
Confidential