ZIMMER CONTOUR RESTORATIVE SYSTEM

K061043 · Zimmer Dental, Inc. · NHA · Jun 27, 2006 · Dental

Device Facts

Record IDK061043
Device NameZIMMER CONTOUR RESTORATIVE SYSTEM
ApplicantZimmer Dental, Inc.
Product CodeNHA · Dental
Decision DateJun 27, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3630
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Zimmer® Contour Healing Collar is used to shape the gingival tissue during healing to allow for a suitable emergence profile of the prosthesis. The Zimmer® Contour Healing Cap is for use with a Hex-Lock™ Contour Abutment or a Zimmer® One-Piece Implant to prevent irritation of soft tissue due to rubbing against the restorative area of the abutment or implant, and to prevent material from lodging in any undercuts or openings. The Zimmer® Contour Provisional Coping is used for fabricating a cement-retained provisional restoration for a Hex-Lock™ Contour Abutment or a Zimmer® One-Piece Implant. Use of the provisional cap is not to exceed 28 days.

Device Story

Zimmer Contour Restorative System comprises healing collars, healing caps, and provisional copings; used with Zimmer Dental abutment-implant combinations or Zimmer One-Piece Implants. System components facilitate gingival tissue shaping, protect soft tissue from irritation, prevent debris lodging in implant undercuts, and support fabrication of cement-retained provisional restorations. Devices are manufactured from titanium or acrylic. Used in dental clinical settings by dental professionals during the restorative phase of implant treatment. Output is a physical component that aids in prosthesis preparation and tissue management; benefits include improved emergence profiles and patient comfort during the healing period.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Components available in titanium or acrylic. Designed for use with Hex-Lock™ Contour Abutments or Zimmer® One-Piece Implants. Mechanical function for tissue shaping and provisional restoration support.

Indications for Use

Indicated for patients requiring gingival tissue shaping, soft tissue protection, or provisional restoration fabrication in conjunction with endosseous dental implants. Healing collars shape gingival tissue; healing caps prevent soft tissue irritation and debris accumulation; provisional copings facilitate cement-retained provisional restorations (limited to 28 days).

Regulatory Classification

Identification

An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.

Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo for Zimmer Dental. The logo consists of a circle with a large, bold letter "Z" inside. Below the circle, the word "zimmer" is written in a stylized font. Underneath "zimmer", the word "dental" is written in a smaller, simpler font. Zimmer Dental 1900 Aston Avenue Carlsbad, CA 92008 760.929.4300 (ph) 510k No .: Page No .: A5-1 JUN 2 7 2006 ## 510(k) SUMMARY (21CFR807.92(a)) - 1. Submitter's Information: | Name: | Zimmer Dental Inc. | |----------------|--------------------| | Address: | 1900 Aston Ave. | | | Carlsbad, CA 92008 | | Phone: | 760-929-4300 | | Contact: | Erin L. McVey | | Date Prepared: | April 12, 2006 | 2. Zimmer® Contour Restorative System Device Name: > Device Classification Name: Endosseous Dental Implant Abutment - 3. Predicate Device(s): - . Zimmer Dental Advent® Implant Extender/ Healing Collar - . 3i EP Two Piece Healing Abutment - Lifecore Quick-Abutment Temporary/ Healing Cap . - Zimmer Dental Pure Form Ceramic Coping . - 4. Device Description: The Zimmer Contour Restorative System is used to aid in the preparation of dental prostheses, as well as, to form the gingival tissue around an endosseous implant site prior to prosthesis fixation. The system is used with Zimmer Dental abutment-endosseous dental implant combinations, as well as, the Zimmer® One-Piece Implant. Devices are available in titanium or acrylic. - ട്. Intended Use: The Zimmer® Contour Healing Collar is used to shape the gingival tissue during healing to allow for a suitable emergence profile of the prosthesis. The Zimmer® Contour Healing Cap is for use with a Hex-Lock™ Contour Abutment or a Zimmer® One-Piece Implant to prevent irritation of soft tissue due to rubbing against the restorative area of the abutment or implant, and to prevent material from lodging in any undercuts or openings. The Zimmer® Contour Provisional Coping is used for fabricating a cement-retained provisional restoration for a Hex-Lock™ Contour Abutment or a Zimmer® One-Piece Implant. Use of the provisional cap is not to exceed 28 days. - Device Comparison: 6. Zimmer Dental Inc. believes the Zimmer® Contour Healing Collar. Healing Cap and Provisional Coping to be substantially equivalent to their respective predicates. They are equivalent in intended use, design, and materials. {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the United States Department of Health and Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" around the perimeter. Inside the circle is an abstract image of an eagle, represented by three curved lines. The eagle is facing to the right. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUN 27 2006 Mr. Eric L. McVev Regulatory Affairs Specialist Zimmer Dental, Incorporated 1900 Aston Avenue Carlsbad, California 92008-7308 Re: K061043 Trade/Device Name: Zimmer® Contour Restorative System Regulation Number: 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: June 12, 2006 Reccived: June 15, 2006 Dear Mr. McVey: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class H (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register. {2}------------------------------------------------ Page 2 -- Mr. McVey Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation cntitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours. Clare Chiu Lin. Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indications for Use 510(k) Number (if known): Device Name: Zimmer® Contour Healing Collar Zimmer® Contour Healing Cap Zimmer® Contour Provisional Coping Indications For Use: The Zimmer® Contour Healing Collar is used to shape the gingival tissue during healing to allow for a suitable emergence profile of the prosthesis. The Zimmer® Contour Healing Cap is for use with a Hex-Lock™ Contour Abutment or a Zimmer® One-Piece Implant to prevent irritation of soft tissue due a more against the restorative area of the abutment or implant, and to prevent materials from lodging in any undercuts or openings. The Zimmer® Contour Provisional Coping is used for fabricating a cement one neg provisional restoration for a Hex-Lock™ Contour Abutment or a Zimment of College Implant. Use of the provisional coping is not to exceed 28 days. × Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Union ... .. (21 CFR 801 Subpart C : (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER FOR A NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Sunil Kumar 13 hasiology, General Hospital, n Continel, Dental Devic K061043 Page 1 of _ : 1 Page: Section A4-1
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...