K241186 · Shenzhen Dazhou Medical Technology Co., Ltd. · LYC · Feb 21, 2025 · Dental
Device Facts
Record ID
K241186
Device Name
Synthetic Bone Graft Particulate
Applicant
Shenzhen Dazhou Medical Technology Co., Ltd.
Product Code
LYC · Dental
Decision Date
Feb 21, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3930
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Synthetic Bone Graft Particulate is intended to be used to augment the alveolar bone in tooth extraction procedures. (i.e., use in extraction sockets only)
Device Story
Synthetic Bone Graft Particulate is a synthetic, absorbable, osteoconductive bone graft substitute composed of 45S5 bioactive glass. Supplied as particulates (0.5 mm to 1 mm size), the device is mixed with sterile saline at the point of care to create a wet sandy paste. Clinicians (dentists/oral surgeons) pack the paste into tooth extraction sockets. The material acts as a non-structural, biocompatible scaffold; it is progressively resorbed and replaced by the patient's natural bone tissue via osteoconduction during the healing process. The device is intended for use in dental clinics.
Clinical Evidence
Non-clinical evidence includes biocompatibility (ISO 10993 series), chemical/crystallinity testing (ASTM F1538-03), and bioactivity (ISO 23317). A comparative pre-clinical animal study in a canine dental defect model evaluated in vivo absorbability and bone regeneration at 4, 12, and 26 weeks. Assessments included micro-CT scans and histological evaluations (H&E and Masson-Goldner staining). Results demonstrated the device is substantially equivalent to the predicate in bone healing and resorption characteristics.
Indicated for augmentation of alveolar bone in tooth extraction sockets.
Regulatory Classification
Identification
Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Special Controls
*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
Predicate Devices
PerioGlas - Bioglass Bone Graft Particulate (K040278)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The symbol on the left is a stylized representation of a human figure, and the text on the right is written in a combination of blue and white. The logo is simple and modern, and it is easily recognizable.
Shenzhen Dazhou Medical Technology Co., Ltd. February 21, 2025 % Mandy Li Regulatory Affair Cosmos Biomed Consulting Co., Ltd. Room 1201, No.1, 188 Alley, Shuangliu Road Changning District Shanghai City, CHINA
Re: K241186
Trade/Device Name: Synthetic Bone Graft Particulate Regulation Number: 21 CFR 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: Class II Product Code: LYC Dated: April 26, 2024 Received: January 21, 2025
#### Dear Mandy Li:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
Page
3
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Sherrill Lathrop Blitzer
for Andrew Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K241186
Device Name Synthetic Bone Graft Particulate
Indications for Use (Describe)
Synthetic Bone Graft Particulate is intended to be used to augment the alveolar bone in tooth extraction procedures. (i.e., use in extraction sockets only)
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X | Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) SUMMARY
#### 1. Submission Information
| Submitter: | Shenzhen Dazhou Medical Technology Co., Ltd.<br>Suite B605, Building No.1, YESUN INTELLIGENT COMMUNITY II,<br>1301-76 Guanguang Rd., Longhua District, Shenzhen, Guangdong,<br>518110, P.R. China |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter contact: | Dr. Jingzhou Yang<br>Tel: +86-755-21077756<br>E-mail: yangjingzhou@dazhoumedical.com |
| Date of preparation: | February 21, 2025 |
## 2. Device Name and Classification
| Product Name: | Synthetic Bone Graft Particulate |
|-----------------------|-----------------------------------|
| Classification Name: | Bone Grafting Material, Synthetic |
| Common or Usual Name: | Bone grafting material |
| Regulation Number: | 21 CFR 872.3930 |
| Product Code: | LYC |
### 3. Predicate Device(s)
| Product Name: | PerioGlas - Bioglass Bone Graft Particulate<br>(K040278) |
|-----------------------|----------------------------------------------------------|
| Common or Usual Name: | Bone grafting material |
| Regulation Number: | 21 CFR 872.3930 |
| Product Code: | LYC |
#### 4. Device Description
Synthetic Bone Graft Particulate is a synthetic absorbable osteoconductive bone graft substitute manufactured from 45S5 bioactive glass. The device is in particulate form with a size range of 0.5 mm to 1 mm.
The device is intended for augmenting alveolar bone in tooth extraction procedures. At time of use, the device is mixed with sterile saline to form a wet sandy paste that is applied to the defect. Synthetic Bone Graft Particulate is progressively resorbed and replaced by new bone tissue during the healing process.
It is supplied sterile, packaged in a rubber stopper-sealed glass bottle within a sterile barrier
{5}------------------------------------------------
package (Tyvek-sealed PETG box). The device packages are protected by carboard box.
## 5. Indications for Use
Synthetic Bone Graft Particulate is intended to be used to augment the alveolar bone in tooth extraction procedures. (i.e., use in extraction sockets only)
| Substantial<br>Equivalence<br>Comparison | Predicate Device [K040278]<br>PerioGlas -Bioglass Synthetic Bone<br>Graft Particulate | Proposed Device<br>Synthetic Bone Graft Particulate |
|------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification<br>& Product<br>Code | Class II, LYC | Same as predicate |
| Indications for<br>Use | PerioGlas – Bioglass Bone Graft<br>Particulate is indicated to be packed<br>into bony voids or gaps to fill and/or<br>augment oral, dental intraosseous, and<br>craniofacial defects. These defects may<br>include: periodontal/infrabony defects;<br>alveolar ridge augmentation<br>(sinusotomy, osteotomy, cystectomy);<br>dental extraction sites (ridge<br>maintenance, implant<br>preparation/placement); sinus lifts;<br>cystic defects; craniofacial<br>augmentation. PerioGlas may be used<br>alone in a manner comparable to<br>autogenous bone graft chips or<br>allograft bone particulate<br>(demineralized freeze dried bone), or it<br>may be mixed with either as a bone<br>graft extender. | Synthetic Bone Graft Particulate is<br>intended to be used to augment the<br>alveolar bone in tooth extraction<br>procedures. (i.e., use in extraction<br>sockets only) |
| Application | Gently packed into defect sites as non-<br>structural scaffold for body's natural | Same as predicate |
| Substantial<br>Equivalence<br>Comparison | Predicate Device [K040278]<br>PerioGlas -Bioglass Synthetic Bone<br>Graft Particulate | Proposed Device<br>Synthetic Bone Graft Particulate |
| | healing and bone regeneration<br>processes | |
| Material | Bioactive glass | Same as predicate |
| Morphology | Irregular dense particulate | Same as predicate |
| Particle Size | 90-710 microns | 500-1000 microns |
| Resorption<br>Rate | Majority absorbed by six months | Same as predicate |
| Mechanical | Particulate material; not intended for<br>use in load-bearing defects without<br>proper internal or external fixation | Particulate material; requires an<br>adequate healing period prior to<br>load-bearing |
| Performance | Bone infiltration occurs throughout the<br>grafts site via osteoconduction,<br>resulting in increased graft site<br>mechanical stiffness and strength | Bone infiltration occurs throughout<br>the grafts site |
| Bone<br>Remodeling | New bone grows into the graft area via<br>osteoconduction. The material is<br>slowly absorbed and replaced by the<br>host bone. | Same as predicate |
# 6. Comparison to the Predicate Device
{6}------------------------------------------------
Synthetic Bone Graft Particulate can be used in the same intended way as PerioGlas, without bone extender materials, and it thus substantially equivalent to PerioGlas for the same indication for use.
The technological characteristics of Synthetic Bone Graft Particulate are similar to the predicate in terms of the device material. Both devices are calcium phospho-silicate inorganic materials. The particle size of Synthetic Bone Graft Particulate is similar to that of PerioGlas. Synthetic Bone Graft Particulate and the predicate are designed to be gently packed into defect sites, functioning as a non-structural scaffold for the body's natural healing and bone regeneration process. The device acts as a synthetic, inorganic, biocompatible and osteoconductive scaffold into which new bone will grow.
### 7. Performance Data
{7}------------------------------------------------
### Non-clinical tests were performed on the proposed device.
Sterilization complies with ISO 11137. Product shelf-life testing was evaluated to ensure the labeled shelf life. Assessment of biocompatibility was completed per ISO 10993-1. The evaluation included cytotoxicity per ISO 10993-5, sensitization per ISO 10993-10, irritation per ISO 10993-23, pyrogenicity per USP<151> and ISO 10993-11, acute systemic toxicity per ISO 10993-11, subchronic systemic toxicity per ISO 10993-11 and ISO 10993-6, implantation per ISO 10993-6, and genotoxicity per ISO 10993-3. Chemical composition and crystallinity testing were completed per ASTM F1538-03. Morphology testing was completed per ISO 9276-6. Particle size distribution testing was completed per USP<786>. pH value testing was completed per USP<791>. Bioactivity testing was completed per ISO 23317. Water solubility testing was completed per OECD Test No. 105. The FT-IR method was utilized to analyze the additive residue. Bacterial endotoxin testing was performed using the LAL test method, following the USP <85> Bacterial Endotoxins Test, and met the limits (<20 EU/device).
#### Animal Study
Pre-clinical study was conducted in a canine dental defect model to evaluate the in vivo absorbability and bone healing/regeneration ability. The study was comparative between the subject and predicate devices. General observations, hematology tests, gross pathological examinations, micro-CT scans, and histological evaluations (H&E Staining and Masson-Goldner Staining) were performed at 4, 12, and 26 weeks post-operation.
#### 8. Conclusion
The test results indicate that Synthetic Bone Graft Particulate is substantial equivalence to the predicate device PerioGlas- Bioglass Bone Graft Particulate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.