Bond Apatite™ is a synthetic osteoconductive, bioresorbable bone grafting material composed of hydroxyapatite and biphasic calcium sulfate in granulated powder form. intended to fill. augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Device Story
Bond Apatite is a synthetic, osteoconductive, bioresorbable bone graft material; supplied as granulated powder (38.6% hydroxyapatite, 61.4% biphasic calcium sulfate). Used in dental/oral/maxillofacial surgery to fill or reconstruct bony defects. Clinician mixes dry powder with sterile saline inside provided disposable syringe applicator; delivers putty directly to treatment site. Material acts as scaffold for bone regeneration; resorbs over time. Benefits patient by facilitating bone repair in periodontal or maxillofacial defects.
Clinical Evidence
Bench testing only. Evidence includes material characterization (FTIR, XRD, ICP-OES), physical characterization (gravimetric, sieve, SEM), and performance testing (setting time, reaction temperature, compressive properties, pH, dissolution, and simulated usage handling).
Technological Characteristics
Synthetic bone graft; hydroxyapatite (ISO 13779-1/3) and biphasic calcium sulfate (ASTM F2224). Supplied as dry powder in 1cc disposable syringe applicator for mixing with saline. Osteoconductive, bioresorbable. No electronic or software components.
Indications for Use
Indicated for filling, augmenting, or reconstructing periodontal or bony defects of the oral and maxillofacial region in patients requiring bone grafting.
Regulatory Classification
Identification
Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Special Controls
*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
HAPSET Hydroxylapatite Bone Graft Plaster (K910432)
Submission Summary (Full Text)
{0}------------------------------------------------
510(k) Summary K121177
DEC 0 5 2013
# 510(k) Summary Augma Biomaterials, Ltd. Bond Apatite™ K121177
December 5, 2013
# ADMINISTRATIVE INFORMATION
Augma Biomaterials. Ltd. Manufacturer Name Usishkin 8 Netanya. Israel 42273 Telephone: +972-(0)77-559-1945 +972-(0)4-627-5337 Fax:
Official Contact
Dr. Amos Yahav, DMD
Representative/Consultant
Kevin A. Thomas, PhD Linda K. Schulz, BSDH. RDH PaxMed International. LLC 11234 El Camino Real. Suite 200 San Diego. CA 92130 Telephone: +1 (858) 792-1235 Fax: +1 (858) 792-1236 kthomas@paxmed.com Email: lschulz@paxmed.com
# DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name | Bond Apatite™ |
|----------------------------|------------------------------------|
| Classification Name | Bone Grafting Material, Synthetic. |
| Classification Regulations | 21 CFR 872.3930, Class II |
| Product Code | LYC |
| Classification Panel | Dental Products Panel |
| Reviewing Branch | Dental Devices Branch |
#### INTENDED USE
Bond Apatite™ is a synthetic osteoconductive, bioresorbable bone grafting material composed of hydroxyapatite and biphasic calcium sulfate in granulated powder form. intended to fill. augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
{1}------------------------------------------------
## 510(k) Summary K121177
## DEVICE DESCRIPTION
Bond Apatite is composed of hydroxyapatite and biphasic calcium sulfate in granulated powder form. The hydroxyapatite component is sintered hydroxyapatite granules that conform to ISO 13779-1 Implants for surgery - Hydroxyapatite - Part 1: Ceramic Hydroxyapatite and ISO 13779-3 Implants for surgery – Hydroxyapatite – Part 3: Chemical analysis and characterization of crystallinity and phase purity. The calcium sulfate material is Bond Bone™, a mixture of surgical grade calcium sulfate dihydrate and calcium sulfate hemihydrate, and cleared under K083858. The calcium suffate component conforms to the chemical requirements of ASTM F2224 Standard Specification for High Purity Calcium Sulfate Hemihydrate or Dihydrate for Surgical Implants. Bone Apatite is provided sterile, in a single unit size of 1 cc in a disposable applicator that is used for mixing the dry powder with sterile saline and for delivery to the treatment site.
### EQUIVALENCE TO MARKETED DEVICE
Augma Biomaterials, Ltd., demonstrated that, for the purposes of FDA's regulation of medical devices Bond Apatite is substantially equivalent in indications and design principles to the following legally marketed predicate devices:
Augma Biomaterials, Ltd., Bond Bone™, K083858;
BONESUPPORT AB, CERAMENT™BONE VOID FILLER (A0210-12), K090871; and
Lifecore Biomedical. Inc., HAPSET Hydroxylapatite Bone Graft Plaster, K910432.
The intended use, materials, design, and functional characteristics of Bond Apatite and the predicate devices are substantially the same. The subject device and the predicate device Bond Bone are indicated for use in dental bone regeneration procedures. The intended use of the subject device and both predicate devices is to fill bony defects. The calcium sulfate material in the subject device Bond Apatite is the predicate device. Bond Bone, cleared under K083858. By weight, Bone Apatite is approximately 38.6% hydroxyapatite and 61.4% biphasic calcium sulfate. The predicate devices CERAMENT™|BONE VOID FILLER and HAPSET Hydroxylapatite Bone Graft Plaster consist of powdered hydroxyapatite and calcium sulfate hemihvdrate.The subject device and the predicate devices are supplied as a dry powder to be mixed with saline or other liguid to form a putty before use, and all are provided sterile in a single unit mixing/deliverv applicator (syringe) for single patient use.
Any differences in the technological characteristics between the subject and predicate devices do not raise new issues of safety or efficacy.
Data submitted, referenced, or relied upon to demonstrate substantial equivalence included: material characterization by Fourier transform infrared spectroscopy, x-ray diffraction, and inductively coupled plasma optical emission spectrometry; physical characterization by gravimetric methods, mechanical sieve separation, scanning electron microscopy; performance characteristics including setting time and reaction temperature measurements, compressive mechanical properties, pH. dissolution. device handling under simulated usage conditions, and clinical performance testing.
{2}------------------------------------------------
# 510(k) Summary K121177
The data included in this submission demonstrates substantial equivalence to the predicate devices listed above.
Overall, Bond Apatite™ has the following similarities to the predicate devices:
- has the same intended use. o
- uses the same operating principle, .
- incorporates the same basic design, .
- incorporates the same or very similar materials, and .
- has similar packaging and is sterilized using the same materials and processes. .
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three legs and a circular border. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged around the top half of the circular border.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
#### December 5, 2013
Augma Biomaterials, Limited C/O Kevin A. Thomas, PhD. PaxMed International, LLC 11234 El Camino Real, Suite 200 San Diego, CA 92130
Re: K121177
Trade/Device Name: Bond Apatite™ Regulation Number: 21 CFR 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: II Product Code: LYC Dated: November 26, 2013 Received: December 2, 2013
Dear Dr. Thomas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2 - Dr. Thomas
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Kwame O.
Ulmer-S
for
Erin 1. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K121177
510(k) Summary
Bond Apatite™
# Indications for Use
510(k) Number:
Bond Apatite™ Device Name:
Indications for Use:
Bond Apatite™ is a synthctic osteoconductive, bioresorbable bone grafting material composed of hydroxyapatite and biphasic calcium sulfate in granulated powder form, intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Prescription Use X (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Mary S. Runner -S BA 2013.12.05 14:40:01 -05'00'
Page 15 of 227
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.