A bioabsorbable implantable matrix intended for use as a space-making barrier in the treatment of periodontal defects and maxillofacial guided tissue regeneration procedures, including preservation and conservation of alveolar bone height and volume, ridge and extraction site augmentation, sinus lifts, and treatment of associated cystic defects. It is also intended for use as a grafting material containment matrix.
Device Story
Bioscaff™ Alvelac™ is a synthetic, bioabsorbable implantable matrix; composed of PLGA and PVA. Designed with macro channels and micropores (20-150 um) to facilitate cell adhesion and bone/tissue ingrowth. Used in dental/maxillofacial surgery as a space-making barrier and grafting material containment matrix. Provides temporary structural support during tissue and bone regeneration; resorbs over time. Supplied sterile for single-use. Clinician-operated during surgical procedures to support bone height/volume preservation and site augmentation.
Clinical Evidence
Bench testing confirmed mechanical properties suitable for intended use. Animal studies demonstrated biocompatibility, appropriate resorption rates, and effective bone formation.
Technological Characteristics
Synthetic polymer scaffold; PLGA (85:15 monomer ratio) and PVA. Porosity: 70% air, 30% PLGA; pore size 20-150 um. Supplied sterile for single use. Bio-absorbable.
Indications for Use
Indicated for patients requiring periodontal or maxillofacial guided tissue regeneration, including alveolar bone preservation, ridge/extraction site augmentation, sinus lifts, and cystic defect treatment. Used as a space-making barrier and grafting material containment matrix.
Regulatory Classification
Identification
Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Special Controls
*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for "bio-scaffold International Pte Ltd". The logo has the letters "BSI" in a circle on the left side. To the right of the circle is the word "bio-scaffold" in a sans-serif font, with the word "International Pte Ltd" underneath.
# K680308
i
:
---------
------------ --------------
# 510(k) Summary in accordance with 21 CFR 807.92(c)
The 510(k) Summary of Safety and Effectiveness for Bioscaff™ Alvelac™ follows:
| Device Name: | | Bioscaff™ Alvelac™ | |
|--------------------------------|-------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| Type of 510(k) submission: | | Abbreviated | |
| Date of Submission: | | February 1st, 2008 | |
| Manufacturer/510(k) Submitter: | | Bio-Scaffold International Pte Ltd<br>61 Science Park Road #02-01/10<br>The Galen, Singapore Science Park III<br>Singapore 117525 | |
| | | | Tel: +65 646 55395<br>Fax: +65 646 55394 |
| 510(k) Contact: | | Roger Gray<br>Director, Global Regulatory Affairs<br>Donawa Consulting<br>Piazza Albania, 10<br>00153 Rome<br>Italy | |
| | | | Tel: +39 06 578 2665<br>Fax: +39 06 574 3786 |
| Trade Name: | | Bioscaff™ Alvelac™ | |
| Common Name: | | Dental Bone Grafting Material | |
| Classification Name: | | Bone Grafting Material | |
| Regulation: | | 21 CFR 872.3930 | |
| Product Code: | | LYC | |
| Predicate Devices: | | | |
| 1. | Cytoplast™ Resorb<br>Osteogenics Co<br>3234 64th Street<br>Lubbock, TX 79413<br>USA | 2. | Inion GTR™ Biodegradable Membrane<br>Inion Ltd<br>Laakarinkatu 2<br>FIN-33520, Tampere<br>Finland |
K993610
K033074
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for "bi-scaffold International Pte Ltd". The logo has "BSI" in a circle on the left, followed by "bi-scaffold" in a stylized font. Below "bi-scaffold" is the text "International Pte Ltd".
#### Device Description:
The Bioscaff™ Alvelac™ is a synthetic polymer scaffold synthesized from Poly (lactic-coglycolic) acid (PLGA) and poly vinyl acetate (PVA). The scaffold is designed with macro channels and micropores for cell adhesion, as well as to provide the necessary space for bone and tissues to grow.
#### Intended Use:
A bioabsorbable implantable matrix intended for use as a space-making barrier in the treatment of periodontal defects and maxillofacial guided tissue regeneration procedures, including preservation and regeneration of alveolar bone height and volume, ridge and extraction site augmentation, sinus lifts, and treatment of associated cystic defects. It is also intended for use as a grafting material containment matrix,
#### Technological Characteristics:
The Bioscaff™ Alvelac™ uses PLGA with a monomer ratio of 85:15 (85% lactic acid and 15% glycolic acid), formed into one of a range of devices of different shapes and sizes. The matrix is approximately 30% PLGA, 70% air, with a pore size of 20-150 um, suited to promoting bony in-growth, whilst retaining the necessary configuration, shape and strength for clinical use.
The predicate devices use similar biocompatible copolymers to produce bio-absorbable membranes or mesh that is used to provide space-making barriers after dental or maxillofacial surgery, to provide temporary support during the tissue and bone regeneration process.
The different types of copolymer used result in different resorption rates, to suit the intended clinical objectives.
All three devices are supplied sterile for single use.
#### Performance Data:
The mechanical properties of the Bioscaff™ Alvelac™ have been determined to be suitable for the intended use. Animal studies have confirmed biocompatibility, resorption rate, and good bone formation.
#### Conclusions:
Based on the information contained within this submission, it is concluded that the Bioscaff™ Alvelac™ is substantially equivalent to the predicate devices already in interstate commerce within the USA.
{2}------------------------------------------------
Image /page/2/Picture/11 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES . USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with three wing-like shapes extending from its body.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## APR 1 4 2008
Bio-Scaffold International Pte Limited C/O Mr. Roger Gray Director, Global Regulatory Affairs Donawa Consulting Piazza Albania, 10 00153 Rome ITALY
Re: K080308
Trade/Device Name: Bio-Scaffold Bioscaff Alvelac Resorbable Scaffold Regulation Number: 21 CFR 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: II Product Codes: LYC and NPK Dated: February 1, 2008 Received: February 5, 2008
Dear Mr. Gray:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2- Mr. Gray
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Smitte Y. M, chani Quis
Chiu Lin. Ph.D. Director Division of Anesthesiology, General Hospital. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### Appendix A
#### Indications for Use Statement
See overleaf for Indications for Use Statement in accordance with the format requested by FDA.
#### Indications for Use
K080308 510(k) Number (if known): Not-known-
Device Name: Bio-Scaffold Bioscaff Alvelac Resorbable Scaffold
Indications for Use: Bio-Scaffold Bioscaff Alvelac Scaffold is a bioabsorbable implantable matrix intended for use as a space-making barrier in the treatment of periodontal defects and maxillofacial guided tissue regeneration procedures, including preservation and conservation of alveolar bone height and volume, ridge and extraction site augmentation, sinus lifts, and treatment of associated cystic defects. It is also intended for use as a grafting material containment matrix.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
×
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
#### Concurrence of CDRH, Office of Device Evaluation (ODE)
Susan Raser
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number:
Page 1 of 1
Rev.2 Apr 4, 2008
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.