calc-i-oss is indicated for the filling and/or augmentation of intraoral/maxillofacial osseous defects, such as intrabony periodontal osseous defects, furcation defects, augmentation of bony defects, augmentation of bony defects of the alveolar ridge, filling of tooth extraction sites, and sinus elevation grafting.
Device Story
calc-i-oss is a synthetic, resorbable, osteoconductive bone graft substitute composed of beta-tricalcium phosphate (β-TCP). It is provided as porous, spherical granules (315–1600 μm) with an interconnecting microstructure. The device is used by dental surgeons in clinical settings to fill or augment osseous defects in the oral and maxillofacial region. The material acts as a scaffold for new bone formation, gradually resorbing over time. It is applied directly to the surgical site, often in conjunction with membranes or autologous bone to facilitate guided bone regeneration. The device provides a biocompatible, osteoconductive matrix to support natural healing and bone repair in dental procedures.
Clinical Evidence
Bench testing only. Evidence includes material characterization, chemical purity analysis, and porosity/resorption studies. Biocompatibility is supported by established literature on β-TCP. Sterilization validation performed per ISO 11137 to achieve a sterility assurance level (SAL) of 10^-6.
Indicated for patients requiring bone grafting in intraoral/maxillofacial osseous defects, including periodontal defects, furcation defects, alveolar ridge augmentation, tooth extraction sites, sinus floor elevations, and defects following cyst removal or apicoectomy. Used alone or in combination with autologous bone and membranes.
Regulatory Classification
Identification
Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Special Controls
*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{1}------------------------------------------------
This summary of 510(k) substantial equivalence information is being submitted in accordance with the requirements of 21 CFR 807.92 and supports the conclusion of SE for calc-i-oss noted below.
#### Applicant's Name and Address: 1.
# JUL 1 9 2005
| Ultradent Products, Inc. |
|--------------------------|
|--------------------------|
| 505 West 10200 South |
|----------------------|
|----------------------|
| South Jordan, Utah 84095 |
|--------------------------|
|--------------------------|
| Telephone number: | 801-553-4200 |
|-------------------|-------------------------|
| Fax number: | 801-572-0600 |
| Contact Person: | Tammy Lavery |
| | RA/QA/QC Senior Manager |
#### Name of the Device: 2.
| Tradename: | calc-i-oss |
|-----------------|----------------------------------------------------------------------------------|
| Common Name: | Osteoconductive Bone Void Filler and<br>Synthetic Resorbable Bone Graft Material |
| Classification: | II (21CFR 872.3930) |
#### Legally Marketed Predicate Devices to which Equivalence is Claimed: 3.
Predicate Device Identified: Cerasorb Dental (P800035).
| | calc-i-oss | Legally marketed Cerasorb (Dental) |
|--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | |
| INDICATIONS: | Indicated for defects after removal of cysts,<br>Augmentation of alveolar crest, possibly in<br>combination with autologous bone and<br>membrane (i.e. guided bond regeneration.)<br>Indicated for apicoectomy and extraction<br>defects in combination with membranes.<br>Indicated for filling of defects after surgical<br>removal of retained teeth Sinus floor<br>elevations and for defects after removal of<br>autologous bone. | Indicated for defects after removal of cysts;<br>repair of marginal and periapical periodontal<br>alveolar bony pockets as well as bifurcations<br>and trifurcations of the teeth; augmentation of<br>the atrophied alveolar ridge; alveolar<br>augmentation of mandibular and maxillary<br>ridges; defects after apicoectomy; and filling<br>bone defects after surgical resection of impacted<br>teeth (without implantation). |
| CHEMICAL<br>COMPOSITION<br>& PURITY: | > 99% Phase-pure synthetic β-tricalcium<br>phosphate. | > 99% Phase-pure synthetic β-tricalcium<br>phosphate. |
| FORM: | High purity β-tricalcium phosphate porous<br>spherical granules. | High purity β-tricalcium phosphate porous<br>spherical granules. |
| PARTICLE SIZE<br>AND RANGE<br>(μm): | Granulate size range from 315 – 1600. | Granulate size range 500-1000. |
| POROSITY /<br>RESORPTION: | Interconnecting porous material 58% for high<br>level of resorption. Refer to the enclosed<br>Solubility Report for more information on<br>porosity as well as resorption. | Interconnecting porous material 22% for high<br>level of resorption. Refer to the enclosed<br>Solubility Report for more information on<br>porosity as well as resorption. |
### Predicate Comparison Table
{2}------------------------------------------------
#### Device Description: 4.
calc-i-oss is a synthetic resorbable osteoconductive bone graft substitute composed of tricalcium phosphate. The device is intended for dental intraosseous, oral, and cranio-/maxillofactial bony defects.
#### 5. Intended Use:
calc-i-oss is indicated for the filling and/or augmentation of intraoral/maxillofacial osseous defects, such as intrabony periodontal osseous defects, furcation defects, augmentation of bony defects, augmentation of bony defects of the alveolar ridge, filling of tooth extraction sites, and sinus elevation grafting.
#### Technological Characteristics: 6.
## a. Chemical Composition:
calc-i-oss is a granulate consisting of bioresorbable, medical grade beta tricalcium phosphate and used for the filling of bone defects. (Refer to confidential test reports and data attached.)
## b. Physical Properties:
calc-i-oss has a round macrostructure. Granulate sizes range between 315 and 1600 um. calc-i-oss is characterized by a porous and interconnecting microstructure. (Refer to confidential test reports and data attached.)
#### Risk Analysis and Test Methods: 7.
The risks noted below were identified and considered during the design of the product. Testing and/or actions were identified to mitigate each area of possible concern.
| Identified Risk | Mitigation |
|----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Ineffective Bone Formation | Material Characterization - See attached confidential test reports<br>and technical data of this submission. |
| Adverse Tissue Reaction | See attached confidential test reports and technical data of this<br>submission. Note: β-TCP is known to be highly biocompatible,<br>resorbable and osteoconductive. Numerous articles on safety and<br>effectiveness are available concerning indications for use in<br>dentistry and are noted in the literature. Mitigation concerning<br>infection is also addressed below. |
| Infection | Sterilization (SAL < 10-6) per ISO 11137. See attached report. |
| Improper Use | Labeling - Refer to the section for labeling and instructional use<br>and warnings. |
{3}------------------------------------------------
### Substantial Equivalence: 8.
In conclusion and per the review noted above, the calc-i-oss manufactured and In conclusion and per the review notes are., 2000 South, South Jordan, Utah marketed by Offradent Froduces, the, 500 - Woomarketed device: Cerasorb (Dental)
84095, is substantially equivalent to the legally-marketed device for seme intended 84095, is substantially equivalent to the regarly maniete and material for same intended use.
Tammy Lavery RA/QA/QC Senior Manager Date
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol.
SEP 1 3 2007
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Tammy Lavery Regulatory Affairs Senior Manager Ultradent Products, Incorporated 505 West 10200 South South Jordan, Utah 84095
Re: K042583 Trade/Device Name: Calc-i-oss Regulation Number: 21 CFR 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: II Product Code: LYC Dated: July 6, 2005 Received: July 7, 2005
Dear Ms. Lavery:
This letter corrects our substantially equivalent letter of July 19, 2005.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device mendments or to devices that have been reclassified in accordance with the provisions of Amendinents of to actives and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls approval (1 MI ). provisions stration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your (1 vir 1) an he found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Ms. Lavery
Please be advised that FDA's issuance of a substantial equivalence determination does not Please be advised that FDA s Issualice of a successor players with other requirements
mean that FDA has made a determination that your devices with other Federal agencies mean that FDA has made a decerminations administered by other Federal agencies.
of the Act or any Federal statutes and regulations administered by registration of the Act or any rederal statues and reguirements, including, but not limited to: registration
You must comply with all the Act's requirements and monufacturing practice You must comply with an the Act 3 requirements; massaling
and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice and listing (21 CFR Part 807), labelling (21 CFR Part 807), as 12 CFR Part 820); and if
requirements as set forth in the quality systems (QS) regulations 531 542 of the Ar requirements as set form in the quality systems (20) regarders (sections 531-542 of the Act);
applicable, the electronic product radiation control provisions (sections 531-54 21 CFR 1000-1050.
This letter will allow you to continue marketing your device as described in your Section This letter will allow you to continue marketing your dostantial equivalence of your device to
510(k) premarket notification. The FDA finding of substantial equivaleriae and 510(K) premarket nothleation. The I DA Intentig or successfication for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), 1991
(2001), 1978 (2007), 1949-19-19 10, 2016 n you desire spee of Compliance at (240) 276-0115 please contact the Office of Comphanee at (210) = 0 compulsion structure). Also, please
(http://www.fda.gov/cdrh/organiz.html#OC_for OC organizations bat notification" (21 C (http://www.lda.gov/curinorgamiz.nam/>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>> note the regulation entitled, "Misorananing of 1915. Constitution on your responsibilities under the Part 807.97). You may other general international and Consumer Assistance at its
Act from the Division of Small Manufacturers, International and Consumer address Act from the Drvision of Billan I an (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Signature
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## INDICATIONS FOR USE
KO42583
510(k) Number (if known): Unknown
Device Name:
Indications For Use:
Defects after removal of bone cysts, Periodontal defects in combination with membranes, I crodonal deleces in combination with autologous bone and membrane (Guided Bone Regeneration) Apicoectomy Extraction defects in combination with membranes Defects after surgical removal of retained teeth Sinus floor elevations Defects after removal of autologous bone
X Prescription Use_ (Per 21 801 CFR Subpart D) AND/OR
Over-The-Counter Use (Per 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
A. Betz, DDS for Dr. Susan Remmer
(Division Sign-Off)
Division of Anesthesiology, General Hospital,
Infection Control, Dental Devices
510(k) Number
Page 1 of (Posted November 13, 200
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.