← Product Code [LYC](/submissions/DE/subpart-d%E2%80%94prosthetic-devices/LYC) · K031073

# PERIOGLAS PLUS - SETTABLE BONE GRAFT SUBSTITUTE (K031073)

_Novabone Products, LLC · LYC · Nov 5, 2003 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/LYC/K031073

## Device Facts

- **Applicant:** Novabone Products, LLC
- **Product Code:** [LYC](/submissions/DE/subpart-d%E2%80%94prosthetic-devices/LYC.md)
- **Decision Date:** Nov 5, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3930
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

PerioGlas Plus - Settable Bone Graft Substitute is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. PerioGlas Plus is indicated to be packed into bony voids or gaps to fill and/or augment dental intraosseous, oral and maxillofacial defects. These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone, including: periodontal/infrabony defects; alveolar ridge augmentation (sinusotomy, osteotomy, cystectomy); dental extraction sites (ridge maintenance, implant preparation/placement); craniofacial augmentation; sinus lifts; cystic defects. The product provides a bone void filler that resorbs and is replaced with bone during the healing process.

## Device Story

PerioGlas Plus is a synthetic, resorbable, osteoconductive bone graft substitute. It consists of calcium phospho-silicate (Bioglass) particles and a calcium sulfate hemihydrate binder. Upon mixing with water, the binder converts to calcium sulfate dihydrate, allowing the material to be packed into dental, oral, and maxillofacial bony voids. Used by clinicians in clinical settings, the device acts as a non-structural scaffold. The calcium sulfate binder resorbs within 4-8 weeks, while the Bioglass particles are progressively replaced by host bone over approximately six months. The device does not provide mechanical load-bearing support; clinicians must use standard fixation techniques if stabilization is required. It benefits patients by facilitating natural bone regeneration in non-structural defects.

## Clinical Evidence

Bench testing only. Provided in vitro data on dissolution properties of the calcium sulfate binder and characterization of residual particulate after binder dissolution. Summarized prior in vivo performance data for individual components.

## Technological Characteristics

Synthetic, inorganic, biocompatible, osteoconductive bone graft substitute. Composition: calcium phospho-silicate (Bioglass) particles and calcium sulfate hemihydrate binder. Form factor: settable paste when mixed with water. Mechanism: osteoconductive scaffold; calcium sulfate binder resorbs in 4-8 weeks; Bioglass particles resorb/remodel over ~6 months. Non-load-bearing.

## Regulatory Identification

Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.

## Special Controls

*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- PerioGlas - Synthetic Bone Graft Particulate ([K992416](/device/K992416.md), [K962492](/device/K962492.md), [K930115](/device/K930115.md))
- CAPSET Calcium Sulfate Bone Graft Plaster ([K955096](/device/K955096.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

<031073

# 3 Confidential

510(k) Premarket Notification NovaBone Products, LLC PerioGlas® Plus - Settable Bone Graft Substitute

## NOV = 5 2003

04/03/03

## 510(k) Summary PerioGlas® Plus - Settable Bone Graft Substitute

- Submitter Information: 1.

| Name:      | NovaBone Products, LLC                         |
|------------|------------------------------------------------|
| Address:   | 1 Progress Boulevard, #33<br>Alachua, FL 32615 |
| Telephone: | (386) 462-7660                                 |
| Facsimile: | (386) 418-1465                                 |
| Contact:   | David M. Gaisser                               |

#### 2. Name of Device:

| Trade Name:          | PerioGlas Plus – Settable Bone Graft Substitute                              |
|----------------------|------------------------------------------------------------------------------|
| Common Name:         | Osteoconductive Bone Void Filler<br>Synthetic Resorbable Bone Graft Material |
| Classification Name: | Unknown                                                                      |

#### 3. Legally Marketed Predicate Device:

| Predicate #1:     | PerioGlas - Synthetic Bone Graft Particulate [K992416, |
|-------------------|--------------------------------------------------------|
| K962492, K930115] |                                                        |
| Predicate #2:     | CAPSET Calcium Sulfate Bone Graft Plaster [K955096]    |

#### 4. Device Description

PerioGlas Plus is a synthetic resorbable osteoconductive bone graft substitute composed of a calcium phospho-silicate material and a calcium sulfate binder. The device is intended for dental intraosseous, oral, and maxillofacial bony defects. The inorganic calcium and phosphorous components are thermally incorporated in a sodium silicate network (PerioGlas) designed specifically for its absorbability and osteoconductive nature. The calcium sulfate component binds the PerioGlas particles together at the time of implantation and is absorbed from the graft site over the first several weeks following implantation. On absorption of the calcium sulfate, the PerioGlas particles remain in the graft site and are progressively aborbed and replaced by host bone during the healing process.

- న్. Intended Use
PerioGlas Plus - Settable Bone Graft Substitute is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. PerioGlas Plus is indicated to be packed into bony voids or gaps to fill and/or augment dental intraosseous, oral and maxillofacial defects. These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone, including: periodontal/infrabony defects; alveolar ridge augmentation (sinusotomy,

{1}------------------------------------------------

K031073

osteotomy, cystectomy); dental extraction sites (ridge maintenance, implant preparation/placement); craniofacial augmentation; sinus lifts; cystic defects. The product provides a bone void filler that resorbs and is replaced with bone during the healing process.

- Technological Characteristics 6.
The technological characteristics of PerioGlas Plus, PerioGlas, and CAPSET are similar, although not identical. All three devices are designed to be osteoconductive space-filling materials to be packed into or over osseous defect sites and used as nonstructural scaffolds for the body's natural healing and bone regeneration process. To meet this design, PerioGlas Plus and the predicate devices are similar in nature; all three devices are synthetic, inorganic, biocompatible and osteoconductive materials.

The main technological characteristic difference between PerioGlas Plus and the predicate devices is their composition. PerioGlas is composed of particulate Bioglass® (see description). CAPSET is composed of powdered calcium sulfate hemihydrate which, when combined with an aqueous-based setting solution, is chemically converted to calcium sulfate dihydrate. PerioGlas Plus is composed of particulate Bioglass and powdered calcium sulfate hemihydrate; when mixed with water, the hemihydrate is chemically converted to calcium sulfate dihydrate and acts as a binder for the Bioglass particles. The calcium sulfate in the PerioGlas Plus and CAPSET devices is absorbed between four and eight weeks after implantation, depending on the graft site, size and material used. The particulate Bioglass in the PerioGlas Plus device is identical to that in the PerioGlas predicate, being substantially absorbed within the six-month timeframe normally associated with bone remodeling. For all three devices, bone forms throughout the graft site with the material being absorbed and replaced by new bone tissue.

In vitro data on the dissolution properties of the device calcium sulfate binder are Characterization data on the residual particulate after dissolution of the presented. calcium sulfate binder also is provided. Prior in vivo performance data for the device components are summarized.

- 7. Warnings and Precautions
PerioGlas Plus does not possess sufficient mechanical strength to support load bearing defects prior to tissue ingrowth. In cases where load support is required, standard internal or external stabilization techniques must be followed to obtain rigid stabilization in all planes.

PerioGlas Plus is intended for use by clinicians familiar with bone grafting and internal/external fixation techniques. PerioGlas Plus must not be used to gain screw purchase or to stabilize screw placement.

{2}------------------------------------------------

K031073

510(k) Premarket Notification NovaBone Products, LLC PerioGlas® Plus - Settable Bone Graft Substitute

#### 8. Complications

Possible complications are the same as to be expected of autogenous bone grafting procedures. These may include: superficial wound infection, deep wound infection, deep wound infection with osteomyelitis, delayed union, loss of reduction, failure of fusion, loss of bone graft, graft protrusion and / or dislodgement, and general complications that may arise from anesthesia and / or surgery. Complications specific to oral/dental use are those as may be typically observed for similar bone grafting procedures and may include: tooth sensitivity, gingival recession, flap sloughing, resorption or ankylosis of the treated root, abscess formation

- Conclusion 9.
PerioGlas Plus is claimed to be substantially equivalent to PerioGlas and CAPSET as a non-structural osteoconductive bone void filler for oral and craniofacial osseous defects. Additional supporting in vitro data were supplied.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NOV - 5 2003

Mr. David M. Gaisser Director, Operations NovaBone Products, LLC One Progress Boulevard, #33 Alachua, Florida 32615

Re: K031073

Trade/Device Name: PerioGlas® Plus-Settable Bone Graft Substitute Regulation Number: None Regulation Name: None Regulatory Class: Unclassified Product Code: LYC Dated: August 12, 2003 Received: August 13, 2003

Dear Mr. Gaisser:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

### Page 2 - Mr. Gaisser

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4613. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Patricia Cuente/foe

Chiu Lin. Ph.D. Director Division of Anesthesiology. General Hospital. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

### STATEMENT OF INDICATIONS FOR USE

510(k) Number (if known): K031073

PerioGlas® Plus - Settable Bone Graft Substitute Device Name:

Indications For Use:

PerioGlas Plus - Settable Bone Graft Substitute is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. PerioGlas Plus is indicated to be packed into bony voids or gaps to fill and/or augment dental intraosseous, oral and maxillofacial defects. These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone, including: periodontal/infrabony defects; alveolar ridge augmentation (sinusotomy, osteotomy, cystectomy); dental extraction sites (ridge maintenance, implant preparation/placement); craniofacial augmentation; sinus lifts; cystic defects. The product provides a bone void filler that resorbs and is replaced with bone during the healing process.

Susan Runr

(Division Sian-Off) Division of Anesthesiology, General Hospital, Infection Control. Dental Devices

510(k) Number: k031073

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use \

OR (Per 21 CFR 801.109) Over-The-Counter Use

0 (

1-1

---

**Source:** [https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/LYC/K031073](https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/LYC/K031073)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
