K110198 · Blue Sky Bio, LLC · LYC · Dec 22, 2011 · Dental
Device Facts
Record ID
K110198
Device Name
BLUE SKY BIO TCP
Applicant
Blue Sky Bio, LLC
Product Code
LYC · Dental
Decision Date
Dec 22, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3930
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Blue Sky Bio TCP Bone Graft Substitute is intended for use as a bone grafting material to fill, augment or reconstruct periodontal or oral/maxillofacial defects. - Filling of periodontal/infrabony defects - Ridge augmentation - Filling of extraction sites (implant preparation/placement) - Sinus lifts - Filling of cystic cavities The device gradually resorbs and is replaced with bone during the healing process.
Device Story
Blue Sky Bio TCP Bone Graft Substitute is a porous, resorbable, biocompatible calcium phosphate ceramic (ß-tricalcium phosphate) used as a bone void filler. Provided in sterile, granular form (up to 2000 μm), it is applied by clinicians in dental or oral/maxillofacial surgical settings. The material acts as an osteoconductive scaffold; it is placed into bony defects where it gradually resorbs and is replaced by natural bone during the healing process. It is used to support reconstruction of periodontal defects, ridge augmentation, extraction site filling, sinus lifts, and cystic cavity filling.
Clinical Evidence
Bench testing only. No clinical data provided. Biocompatibility testing (cytotoxicity, intracutaneous irritation, maximization sensitization, pyrogen test, acute toxicity) and material characterization (XRD profiles, dissolution testing) were performed to demonstrate safety and equivalence.
Technological Characteristics
Material: Pure ß-tricalcium phosphate (ASTM F1088-04). Form: Sterile, pyrogen-free granules up to 2000 μm. Principle: Osteoconductive, resorbable ceramic scaffold. Sterilization: Not specified.
Indications for Use
Indicated for patients requiring bone grafting material to fill, augment, or reconstruct periodontal or oral/maxillofacial defects, including infrabony defects, ridge augmentation, extraction sites, sinus lifts, and cystic cavities.
Regulatory Classification
Identification
Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Special Controls
*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services USA logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 6, 2024
Blue Sky Bio, LLC Michele Vovolka Consultant 888 E Belvidere Rd. Suite 212 Grayslake, Illinois 60030
Re: K110198
Trade/Device Name: Blue Sky Bio TCP Bone Graft Substitute Regulation Number: 21 CFR 872.3930 Regulation Name: Bone grafting material Regulatory Class: Class II Product Code: LYC
Dear Michele Vovolka:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated December 22, 2011. Specifically, FDA is updating this SE Letter because FDA has better categorized your device technology under the product code LYC.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Andrew Steen, OHT : Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices, 301-796-6284, or andrew.steen@fda.hhs.gov.
Sincerely.
# Andrew I. Steen -S
Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. Inside the circle is an abstract symbol that resembles an eagle or other bird.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
DEC 2 2 2011
Ms. Vovolka Consultant Blue Sky Bio, LLC 888 E Belvidere Road, Suite 212 Grayslake, Illinois 60030
Re: K110198
Trade/Device Name: Blue Sky Bio TCP Bone Graft Substitute Regulation Number: 21 CFR 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: II Product Code: LPK Dated: December 12, 2011 Received: December 17, 2011
Dear Ms. Vovolka:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Page 2 - Ms. Vovolka
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
hh for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
K110/98
# INDICATIONS FOR USE
510(k) Number (if known):
Device Name: Blue Sky Bio TCP Bone Graft Substitute
Indications for Use: Blue Sky Bio TCP Bone Graft Substitute is intended for use as a bone grafting material to fill, augment or reconstruct periodontal or oral/maxillofacial 100000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000 . " · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · defects.
- Filling of periodontal/infrabony defects .
- Ridge augmentation .
- Filling of extraction sites (implant preparation/placement) .
- Sinus lifts .
- Filling of cystic cavities 0
The device gradually resorbs and is replaced with bone during the healing process.
Susan Russe
(Division Sign-Off)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K110988
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{4}------------------------------------------------
. . .
K110198
#### 510(k) SUMMARY 5
#### OWNER:
Blue Sky Bio, LLC 888 E. Belvidere Suite 212 Grayslake, IL 60030 Telephone: 847-548-8499 Fax: 888-234-3685
#### OFFICIAL CONTACT:
Michele Vovolka Vantage Consulting International P.O. Box 848 Grayslake, IL 60030 Telephone: 815-856-0355 Fax: 888-234-3685 Email: vantagemv(0)comcast.net
#### DEVICE NAME:
Trade Name: Blue Sky Bio TCP Bone Graft Substitute Bone Grafting Material, for Dental Bone Repair Common Name: Bone Grafting Material Bone Grafting Material, Synthetic Classification Name: Class: Class II Regulation Number: 21 CFR 872.3930 LYC Product Code:
{5}------------------------------------------------
## PREDICATE DEVICE(S):
| Predicate 510(k) | Device Name | Indication | Clearance Date | Company |
|------------------|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------|-------------------------------------------|
| K051443 | Cerasorb® M<br>Dental | Cerasorb® M DENTAL is recommended for:<br>• Augmentation or reconstructive treatment of the<br>alveolar ridge<br>• Filling of infrabony periodontal defects<br>• Filling of defects after root resection,<br>apicoectomy, and cystectomy<br>• Filling of extraction sockets to enhance<br>preservation of the alveolar ridge<br>• Elevation of the maxillary sinus floor<br>• Filling of periodontal defects in conjunction with<br>products intended for Guided Tissue<br>Regeneration (GTR) and Guided Bone<br>Regeneration (GBR)<br>• Filling of perio-implant defects in conjunction<br>with products intended for Guided Bone<br>Regeneration (GBR) | 22 July 2005 | Curasan AG,<br>Kleinostheim,<br>Germany |
| K083372 | ArrowBone-β | ArrowBone-β is intended for use in the<br>reconstruction of natural or surgical periodontal<br>defects of the oral and maxillofacial region,<br>including sinus floor elevation and augmentation of<br>the alveolar crest.<br>ArrowBone-β is intended for filling into the site of a<br>bony defect in combination with patient blood,<br>autologous bone, membranes or sterile saline after<br>removal of cysts or surgical removal of retained<br>tooth. | 8 December<br>2009 | BrainBase<br>Corporation,<br>Tokyo, Japan |
## DEVICE DESCRIPTION:
Blue Sky Bio TCP Bone Graft Substitute is porous, resorbable, and biocompatible calcium phosphate ceramic consisting of ß-tricalcium phosphate. ß-tricalcium phosphate is an osteoconductive implant that is biodegradeable. Blue Sky Bio TCP Bone Graft Substitute is provided sterile in granular form, pyrogen-free and available in granule sizes up to 2000 μm.
{6}------------------------------------------------
## STATEMENT OF INTENDED USE:
Blue Sky Bio TCP Bone Graft Substitute is intended for use as a bone grafting material to fill, augment or reconstruct periodontal or oral/maxillofacial defects.
- · Filling of periodontal/infrabony defects
- · Ridge augmentation
- · Filling of extraction sites (implant preparation/placement)
- Sinus lifts
- · Filling of cystic cavities
The device gradually resorbs and is replaced with bone during the healing process.
#### TECHNOLOGICAL CHARACTERISTICS:
The Blue Sky Bio TCP Bone Graft Substitute consists of pure ß-tricalcium phosphate as described in ASTM F1088-04, Standard Specification for ß-tricalcium Phosphate for Surgical Implants.
The following biocompatibility tests have been completed and results support that Blue Sky Bio TCP Bone Graft Substitute is compatible with surrounding tissues.
| Biocompatibility Tests | Results |
|---------------------------------|---------|
| Cytotoxicity | Pass |
| Intracutaneous Irritation Test | Pass |
| Maximization Sensitization Test | Pass |
| Pyrogen Test | Pass |
| Acute Toxicity | Pass |
{7}------------------------------------------------
K110198
#### BASIS FOR SUBSTANTIAL EQUIVALENCE:
Blue Sky Bio TCP Bone Graft Substitute meets the requirements of established standards for materials, biocompatibility, pyrogenicity and sterilization.
Blue Sky Bio TCP Bone Graft Substitute is substantially equivalent to the currently marketed Cerasorb® M DENTAL and BrainBase Corporation ArrowBone-B, as a bone void filler for defects in the oral/maxillary and dental region. These bone graft materials are equivalent in that they consist of ß-tricalcium phosphate with a phase purity of more than 95% and comply with ASTM F 1088-04. The devices are substantially equivalent with regard to materials as demonstrated by comparison of XRD profiles and dissolution testing, intended use, indications for use, anatomical site and performance data.
The information provided in this submission demonstrates that the Blue Sky Bio TCP Bone Graft Substitute is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.