K131376 · Elevate Oral Care · LBH · Aug 27, 2013 · Dental
Device Facts
Record ID
K131376
Device Name
SMARTCOAT 2.5% NAF VARNISH
Applicant
Elevate Oral Care
Product Code
LBH · Dental
Decision Date
Aug 27, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3260
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
SmartCoat 2.5% NaF Varnish is intended for use on sensitive teeth, over exposed dentin and root surface sensitivity, and under temporary restorations or cements where post operative sensitivity is of concern. Rapid Desensitizer. Provides rapid relief from sensitive teeth due to hypersensitivity.
Device Story
SmartCoat 2.5% NaF Varnish is a topically applied, flavored cavity varnish containing sodium fluoride and hydroxyapatite in a shellac resin base. Applied by dental professionals to dentin surfaces, the viscous liquid dries to form a film that occludes open dentinal tubules, providing rapid relief from hypersensitivity. The device is supplied with a dosing and dispensing mechanism for consistent application. By creating a physical barrier over compromised surfaces, it reduces sensitivity and protects against post-operative discomfort. The use of shellac resin as a film-forming agent avoids potential allergic reactions associated with colophony rosin found in other varnishes.
Clinical Evidence
Bench testing only. Subjective light microscopy at 500x magnification evaluated dentin tubule occlusion on human dentin slabs. Specimens were tested at 0, 3, and 6 days post-application after vigorous stirring in de-ionized water. Results showed complete occlusion (rating 6 on a 1-6 scale) at all time points, with tubules remaining covered by the transparent varnish layer.
Technological Characteristics
Composition: Sodium fluoride, hydroxyapatite, shellac resin, and other film-forming ingredients. Principle: Physical occlusion of dentinal tubules via a dried surface film. Form factor: Viscous liquid varnish with dosing/dispensing mechanism. Biocompatibility: Assessed per ISO 10993 and ISO 14971.
Indications for Use
Indicated for patients with sensitive teeth, exposed dentin, root surface sensitivity, or requiring post-operative sensitivity management under temporary restorations/cements.
Regulatory Classification
Identification
Cavity varnish is a device that consists of a compound intended to coat a prepared cavity of a tooth before insertion of restorative materials. The device is intended to prevent penetration of restorative materials, such as amalgam, into the dentinal tissue.
Special Controls
*Classification.* Class II (special controls). The device, when it is an external cleaning solution, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for Elevate Oral Care. The logo consists of a stylized globe on the left, followed by the word "elevate" in a lowercase, sans-serif font. Below "elevate" are the words "ORAL CARE" in a smaller, sans-serif font.
# 5. 510(k) Summary
### Submitter:
| Company: | Elevate Oral Care |
|------------------------|-----------------------------|
| Street: | 346 Pike Road, Suite 5 |
| City, State Zip: | West Palm Beach, FL 33411 |
| Country: | USA |
| Estab. Registration #: | 3009603151 |
| Correspondent: | Steve Pardue |
| | Managing Member |
| Phone: | 877-866-9113 |
| Fax: | 561-244-1927 |
| Email: | spardue@elevateoralcare.com |
| Date: | May 10, 2013 |
ww.elevateoralcare.com
346 Pike Road, Suite 5 . West Palm Beach, FL 3341 1 Phone: 877-866-9113 · Fox: 561-244-1927
AUG 27 2013
#### Name of Device
| Proprietary Name: | SmartCoatTM 2.5% NaF Varnish |
|----------------------|------------------------------------|
| Classification Name: | Cavity Varnish, Dental |
| | 21 CFR 872.3260 as Class II device |
| Common Name: | Cavity Varnish |
#### Predicate Devices
| Device | 510(k) |
|-------------------------|---------|
| Vanish 5% NaF Varnish | K092141 |
| VivaSens | K030922 |
| Duraphat 5% NaF Varnish | K945794 |
#### Description
SmartCoat 2.5% NaF Varnish is classified as a Cavity Varnish (21 CFR 872.3260) because it is a device that provides relief from tooth surface hypersensitivity when applied to dentin surfaces by forming a film that facilitates occlusion of compromised surfaces including open dentinal tubules.
SmartCoat is a topically applied, flavored cavity varnish containing sodium fluoride in a shellac resin based preparation. The varnish is an insoluble viscous liquid that dries once applied to the teeth to form a surface film. The packaging provides an accurate dosing and uniform dispensing mechanism for consistent product mixing and delivery.
# For Total Body Health
{1}------------------------------------------------
## Indications for Use
- SmartCoat 2.5% NaF Varnish is intended for use on sensitive teeth, over exposed dentin . and root surface sensitivity, and under temporary restorations or cements where post operative sensitivity is of concern.
- . Rapid Desensitizer
- . Provides rapid relief from sensitive teeth due hypersensitivity
### Technological Characteristics
The chemical composition of SmartCoat is similar to the predicate devices and other fluoride varnishes which have been in use for many decades. These products use resins, rosins, modified rosins, hydroxypropyl cellulose, polyurethane, methacrylates, polyethyleneglycol dimethacrylate, and other film forming ingredients to cover dentinal tubules and provide relief from dentinal hypersensitivity.
Shellac resin was chosen as the film forming agent for SmartCoat Varnish due to its extensive safe history in dental and food applications as a film forming barrier and properties similar to existing resins in the Cavity Varnish category. The use of this ingredient rather than colophony rosin eliminates the concern of colophony rosin allergies experienced by some varnishes. Shellac has been used in denture trays and base plates for many years; these trays and plates are currently considered 510(k) exempt (regulation numbers 872.3670, 872.6200).
Many cavity varnishes contain sodium fluoride, calcium phosphates or other mineral ingredients to help occlude tubules. SmartCoat contains sodium fluoride and hydroxyapatite to help occlude tubules.
#### Summary of Physical Tests
Smart coat was tested using subjective light microscopy analysis at 500x magnification to determine dentin tubule occlusion. Human dentin slabs were prepared as described in the bench performance testing section (Section 18) and varnish was applied to the sample. The varnish was allowed to cure in an incubator and then placed in de-ionized water and stirred vigorously for 6 days. Images were taken immediately after application, and after 3 and 6 days. The specimens were examined for tubule occlusion by image analysis.
All dentinal tubules that were initially covered at time 0 remained covered through the entire treatment period well enough that nearly no dentinal tubules were visible in the coated area. The only visible tubules were still covered by a layer of SmartCoat, however the layer was transparent in a small area near the border. The amount of remaining tubule covering was ranked on a scale of 1-6 with 1 = no occlusion, 2 = 25-50% or less occlusion, 3 = 25-50% occlusion, 4 = 50-75% occlusion, 5 = more than 75% occlusion, and 6 = complete occlusion. At all measured time points occlusion ratings for SmartCoat Varnish were 6.
{2}------------------------------------------------
It was concluded that SmartCoat varnish will occlude exposed tubules and therefore the varnish should provide relief from tooth hypersensitivity.
# Description of Safety and Substantial Equivalence
The chemical components in SmartCoat Varnish have been used in predicate devices, are listed as GRAS ingredients, or both. Shellac and ammonium phosphate have extensive use in food preparation and coatings as well as dental applications that provide significant history of safe use. We believe these facts well support the compatibility of SmartCoat Varnish, and the safety of the applicant device is substantially equivalent to the predicate devices in properties, intended use and composition.
Information provided in this submission confirms the substantial equivalence to the predicate devices with common indications. The data provided in this 510(k) submission also shows that the composition is safe for its intended use based on the biocompatibility assessment and risk assessment conduced according to ISO 10993 and ISO 14971.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
#### August 27, 2013
Elevate Oral Care Mr. Steve Pardue Managing Member 346 Pike Road, Suite 5 WEST PALM BEACH, FL 33411
Re: K131376
Trade/Device Name: SmartCoat™ 2,5% NaF Varnish Regulation Number: 21 CFR 872.3260 Regulation Name: Cavity Varnish Regulatory Class: II Product Codes: LBH Dated: June 10, 2013 Received: June 11, 2013
Dear Mr. Pardue:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Pardue
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Image /page/4/Picture/7 description: The image shows the name "Mary S. Runner -S" in a simple, sans-serif font. The text is horizontally oriented and appears to be a name or title. The letters are black against a white background, providing a clear contrast.
Kwame Ulmer M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# 4. Indications for Use Statement
Indications for Use
510(k) Number (if known):
Device Name: SmartCoat 2.5% NaF Varnish
Indications for Use:
- . SmartCoat 2.5% NaF Varnish is intended for use on sensitive teeth, over exposed dentin and root surface sensitivity, and under temporary restorations or cements where post operative sensitivity is of concern.
- Rapid Desensitizer .
- Provides rapid relief from sensitive teeth duc to hypersensitivity .
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)
. AND/OR Over-The-Counter Use __ (21 CFR 801 Subpart C)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| | Sheena A. Green -S<br>2013.08.27 15:30:24 -04'00' |
|-----------------------------------------------------------------------------------|---------------------------------------------------|
| | for M. Susan Runner, DDS, MA |
| | Page __ of __ |
| (Division Sign-Off) | |
| Division of Anesthesiology, General Hospital<br>Infection Control, Dental Devices | |
| 510(k) Number: | K131374 |
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.