← Product Code [LBH](/submissions/DE/subpart-d%E2%80%94prosthetic-devices/LBH) · K112946

# ACTIV VARNISH (K112946)

_Pulpdent Corporation · LBH · Dec 9, 2011 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/LBH/K112946

## Device Facts

- **Applicant:** Pulpdent Corporation
- **Product Code:** [LBH](/submissions/DE/subpart-d%E2%80%94prosthetic-devices/LBH.md)
- **Decision Date:** Dec 9, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3260
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

Activ Varnish is a resin-based varnish, containing 17% Xylitol and less than 20% ethanol, that is applied to enamel or dentin for professional treatment of dental hypersensitivity. Activ Varnish treats hypersensitivity by releasing fluoride, calcium and phosphorus ions that precipitate on and, thereby, occlude dentinal tubules and fill superficial, non-carious enamel lesions.

## Device Story

Activ Varnish is a resin-based dental varnish applied to enamel or dentin by dental professionals. It functions by releasing fluoride, calcium, and phosphorus ions, which precipitate to occlude dentinal tubules and fill superficial, non-carious enamel lesions, thereby treating dental hypersensitivity. The device is available in unit-dose or bulk packaging.

## Clinical Evidence

No clinical data provided; substantial equivalence is based on design, composition, and performance similarities to legally marketed predicate devices with established safety and effectiveness profiles.

## Technological Characteristics

Resin-based varnish containing 17% Xylitol and <20% ethanol. Active ingredients include sodium fluoride, calcium, and phosphate salts. Formulated for professional topical application to enamel or dentin. Available in unit-dose or bulk packaging.

## Regulatory Identification

Cavity varnish is a device that consists of a compound intended to coat a prepared cavity of a tooth before insertion of restorative materials. The device is intended to prevent penetration of restorative materials, such as amalgam, into the dentinal tissue.

## Special Controls

*Classification.* Class II (special controls). The device, when it is an external cleaning solution, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9.

## Predicate Devices

- Pulpdent Fluoride Varnish ([K093620](/device/K093620.md))
- Ultradent Flor-Opal Varnish White ([K080249](/device/K080249.md))
- 3M Vanish 5% NaF White Varnish ([K880741](/device/K880741.md))
- Colgate Duraphat ([K945794](/device/K945794.md))
- Voco ReminPro ([K101104](/device/K101104.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# R11294

## EXHIBIT 2

DEC - 9 2011

#### 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS DATA

Kenneth J. Berk 80 Oakland Street PO Box 780 Watertown, MA 02472 USA

DEVICE

.

Trade Name: Activ Varnish Class: Il

Telephone: 617-926-6666 617-926-6262 Fax: ken@pulpdent.com

Classification Name: Cavity Varnish FDA Product Code: 76LBH, 21 CFR Part 872.3260

#### PREDICATE DEVICES:

Pulpdent Fluoride Varnish Ultradent Flor-Opal Vamish White 3M Vanish 5% NaF White Vamish Colgate Duraphat Voco ReminPro

#### DESCRIPTION AND INTENDED USE:

Activ Vamish is a resin-based varnish, containing 17% Xyliol and less than 20% ethanol, that is applied to enamel or dentin for professional treatment of dental hypersensitivity. Activ Varnish treats hypersensitivity by releasing fluoride, calcium and phosphorus ions that precipitate on and, thereby, occlude dentinal tubules and fill superficial, non-carious enamel lesions. Activ Vamish is available in unit-dose and bulk packaging.

#### COMPARISON WITH PREDICATE PRODUCTS:

Activ Varnish is substantially equivalent in design, composition, performance, and intended use to the predicate products:

| Product                              | 510(k) Number | Description                              | Intended Use                                     | Composition                                                                                   |
|--------------------------------------|---------------|------------------------------------------|--------------------------------------------------|-----------------------------------------------------------------------------------------------|
| Pulpdent Activ Varnish               |               | Resin-based fluoride<br>varnish          | To treat tooth<br>hypersensitivity               | Resin, Ethyl alcohol, Xylitol,<br>Sodium fluoride, Calcium and<br>Phosphate salts, Flavorant. |
| Pulpdent Fluoride Varnish            | K093620       | Resin-based fluoride<br>varnish          | To treat tooth<br>hypersensitivity               | Rosin, Ethyl lactate FCC, Sodium<br>fluoride, Flavorant                                       |
| Ultradent<br>Flor-Opal Varnish White | K080249       | 5% sodium fluoride in a<br>resin carrier | To treat tooth<br>hypersensitivity               | Natural resin, Alcohol, Xylitol                                                               |
| 3M Vanish<br>5% NaF White Varnish    | K880741       | Resin-based fluoride<br>varnish          | To treat tooth<br>hypersensitivity               | Rosin ester, n-Hexane, Ethyl<br>alcohol, Sodium fluoride,<br>Thickener, Flavor and Color      |
| Colgate Duraphat                     | K945794       | Rosin-based fluoride<br>varnish          | To treat tooth<br>hypersensitivity.              | Rosin, Ethyl alcohol, Sodium<br>fluoride, Water, Flavorant                                    |
| Voco ReminPro                        | K101104       | Hydroxyapatite and<br>fluoride paste     | Prevention and<br>control of<br>hypersensitivity | Calcium phosphate, Sodium<br>fluoride, Xylitol                                                |

#### SAFETY AND EFFECTIVENESS:

Activ Vamish is substantially equivalent in design, composition, performance, intended use, safety and effectiveness to the predicate products listed above that have been on the market and used successfully by dental professionals for more than 15 years with no serious safety or effectiveness problems.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wing. The eagle is facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Mr. Kenneth J. Berk Director Pulpdent Corporation 80 Oakland Street P.O. Box 780 Watertown, Masachussets 02472

Re: K112946

Trade/Device Name: Pulpdent Activ Varnish Regulation Number: 21 CFR 872.3260 Regulation Name: Cavity Varnish Regulatory Class: II Product Code: LBH Dated: September 29, 2011 Received: October 6, 2011

Dear Mr. Berk:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

DEC - 9 2011

{2}------------------------------------------------

Page 2 – Mr. Berk

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm11.5809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Anthony V. Amato

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known): K | | 2 946

Device Name: Pulpdent Activ Varnish

### Indications For Use:

Activ Varnish is a resin-based varnish, containing 17% Xylitol and less than 20% ethanol, that is applied to enamel or dentin for professional treatment of dental hypersensitivity. Activ Varnish treats hypersensitivity by releasing fluoride, calcium and phosphorus ions that precipitate on and, thereby, occlude dentinal tubules and fill superficial, non-carious enamel lesions.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Susan Kuener

(Division Sign-Off) (Division of Anesthesiology. General Hospital Infection Control, Dental Devices

Page 1 of 1

510(k) Number: K112946

---

**Source:** [https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/LBH/K112946](https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/LBH/K112946)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
