FLOR-OPAL VARNISH WHITE; MODELS 4503, 4504, 4506, S4505

K080249 · Ultradent Products, Inc. · LBH · Feb 8, 2008 · Dental

Device Facts

Record IDK080249
Device NameFLOR-OPAL VARNISH WHITE; MODELS 4503, 4504, 4506, S4505
ApplicantUltradent Products, Inc.
Product CodeLBH · Dental
Decision DateFeb 8, 2008
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 872.3260
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Flor-Opal Varnish White is a 5% sodium fluoride in a varnish carrier which produces a mechanical/chemical occlusion of the dentinal tubules in the treatment of tooth hypersensitivity.

Device Story

Flor-Opal Varnish White is a flavored, xylitol-sweetened 5% sodium fluoride varnish in a resin carrier. It is applied by dental professionals to tooth surfaces to treat hypersensitivity. The varnish adheres to wet or dry tooth surfaces, tolerating saliva and food, while physically and chemically occluding dentinal tubules. The device is delivered via a syringe-to-syringe mixing system. It functions as a topical treatment to provide relief from dental sensitivity.

Clinical Evidence

Bench testing only. Biocompatibility testing was performed to demonstrate safety. No clinical trials were required for this 510(k) submission.

Technological Characteristics

5% neutral sodium fluoride in an alcohol and natural resin suspension. 1ml contains 50mg sodium fluoride. Formulated as a varnish for topical application. Adheres to wet/dry tooth surfaces. White color when dry.

Indications for Use

Indicated for the treatment of tooth hypersensitivity in patients requiring mechanical/chemical occlusion of dentinal tubules.

Regulatory Classification

Identification

Cavity varnish is a device that consists of a compound intended to coat a prepared cavity of a tooth before insertion of restorative materials. The device is intended to prevent penetration of restorative materials, such as amalgam, into the dentinal tissue.

Special Controls

*Classification.* Class II (special controls). The device, when it is an external cleaning solution, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo for Ultra Raven Products, Inc. The logo is black and white and features a stylized "U" followed by the words "ULTRA RAVEN" in a bold, sans-serif font. Below that, the words "PRODUCTS, INC." are written in a smaller, thinner font. Special 510(k) Flor-Opal Varnish White Image /page/0/Picture/2 description: The image is a black and white photograph that appears to be a landscape. The photo is vertically oriented and shows a progression of terrain from what looks like a rocky foreground to a darker, possibly mountainous, background. There are a few bright, circular spots scattered throughout the image, which could be highlights or some other feature of the landscape. SPECIAL 510(K) SUMMARY # Flor-Opal Varnish White This summary of the Traditional 510(k) substantial equivalence information is being submitted in accordance with the requirements of 21 CFR 807 for Flor-Opal Varnish White ### Applicant's Name and Address Ultradent Products, Inc. 505 West 10200 South South Jordan, UT 84095 Contact Person: Title: Telephone: FAX: Date Summary Prepared: Diane Rogers Regulatory Affairs Product Specialist 800-552-5512 x4491, 801-553-4491 801-553-4609 January 23, 2008 FEB - 8 . ... ## Name of the Device Trade Name: Common Name: Device Classification: Classification Product Code: Flor-Opal Varnish White Cavity Varnish ll LBH #### Legally Marketed Predicate Devices to Which Equivalence is Claimed The predicate device is Flor-Opal Varnish (K051750) This device is manufactured and distributed by Ultradent Products, Inc., 505 West, 10200 South, South Jordan, Utah 84095. Product Description: Flor-Opal Varnish White is a flavored, xylitol-sweetened 5% sodium fluoride in a resin carrier, delivered in a .5ml unit-dose syringe-to-syringe mixing system. Indications for Use: Flor-Opal Varnish White is a 5% sodium fluoride in a varnish carrier which produces a mechanical/chemical occlusion of the dentinal tubules in the treatment of tooth hypersensitivity. {1}------------------------------------------------ Table 1: Product Comparison | Property | Predicate:<br>Flor-Opal Varnish (K051750) | Flor-Opal Varnish<br>White | |-------------------------|-------------------------------------------------------------------------------------------------------------------|----------------------------| | Intended Use | Cavity Varnish | Same | | Type of material | Resin based | Same | | Characteristics | Yellow-white color | White | | Human factors | Syringe-to syringe mixing<br>system | Same | | Biocompatibility/Safety | Cytotoxicity testing passed.<br>Literature and testing to<br>demonstrate product is safe<br>when used as directed | Same | #### Technological Characteristics A pleasantly flavored 5% neutral sodium fluoride in a varnish carrier for relief in dental hypersensitivity. 1ml of varnish contains 50mg of sodium fluoride in an alcohol and natural resin suspension. The varnish adheres well to a wet or dry tooth surface and is white in color when dry. It tolerates saliva, and food while occluding the dentinal tubules. #### Brief Description of Testing Performed Each lot of product must pass internal test specifications prior to release. The results of biocompatibility testing demonstrate that Flor-Opal Varnish WHITE is safe and effective when used according to the Instructions for Use. #### Conclusion and Substantial Equivalence In conclusion, Flor-Opal Varnish WHITE, to be manufactured and marketed by Ultradent Products, Inc., 505 West 10200 South, South Jordan, UT 84095, is substantially equivalent to the Flor-Opal Varnish (K051750), also manufactured by Ultradent Products, Inc. The two products are composed of nearly the same materials, have the same intended use and technological characteristics, and both are safe and effective when used for the indications described. {2}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with its wings spread, facing to the right. The eagle is composed of thick, black lines. Surrounding the eagle is a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in capital letters. The text is arranged along the circumference of the circle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 FEB - 8 2008 Ms. Diane Rogers Regulatory Affairs Product Specialist Ultradent Products, Incorporated 505 West 10200 South South Jordan, Utah 84095 Re: K080249 Trade/Device Name: Flor-Opal Varnish White Regulation Number: 21 CFR 872.3260 Regulation Name: Cavity Varnish Regulatory Class: II Product Code: LBH Dated: January 31, 2008 Received: February 4, 2008 Dear Ms. Rogers: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2- Ms. Rogers Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Quc Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Statement of Indications for Use 60861 4 9 510(k) Number (if known): Flor-Opal Varnish White Device Name: Indications for Use: Flor-Opal Varnish White is 5% sodium fluoride in a varnish carrier which produces a mechanical/chemical occlusion of the dentinal tubules in the treatment of tooth hypersensitivity. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) WDA (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: K080247 Page _1_ of _1 (Posted November 13, 2003)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...