← Product Code [KLE](/submissions/DE/subpart-d%E2%80%94prosthetic-devices/KLE) · K964525

# PRIME & BOND 2.1 M.P. DENTIN/ENAMEL BONDING AGENT W/ACTIVATOR (K964525)

_Dentsply Intl. · KLE · Jan 22, 1997 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/KLE/K964525

## Device Facts

- **Applicant:** Dentsply Intl.
- **Product Code:** [KLE](/submissions/DE/subpart-d%E2%80%94prosthetic-devices/KLE.md)
- **Decision Date:** Jan 22, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3200
- **Device Class:** Class 2
- **Review Panel:** Dental

## Intended Use

Prime & Bond™ 2.1 Multipurpose Dentin/Enamel Bonding Agent is a one-component, no-mix visible light curable dental bonding agent that contains fluoride. A new accessory, Prime & Bond™ Self-Curing Activator, can be used with Prime & Bond™ 2.1 Multipurpose Dentin/Enamel Bonding Agent, enabling the practitioner to use self-cure bonding techniques for the current Indications for Use of the Bonding Agent where a dual-cured adhesive may be desired. Also because of the self-cure capability, use of the Bonding Agent and Activator together allows adhesion of amalgam restorations to tooth structures, which is a NEW INDICATION FOR USE.

## Device Story

One-component, no-mix visible light curable dental bonding agent containing fluoride; used with new Prime & Bond™ Self-Curing Activator (acetone/ethanol solution of aromatic sulfinic acid salt). Activator initiates self-cure free radical polymerization of monomers/oligomers in bonding agent. Used by dental practitioners in clinical settings. Enables dual-cured adhesive techniques and bonding of amalgam restorations to tooth structures. Benefits include expanded clinical utility for self-cure applications and improved restorative adhesion.

## Clinical Evidence

Bench testing only; includes performance data for the bonding agent/activator combination and cytotoxicity test results for the activator's active ingredient.

## Technological Characteristics

One-component, no-mix visible light curable resin bonding agent; contains fluoride. Activator is an acetone/ethanol solution of an aromatic sulfinic acid salt. Initiates free radical polymerization.

## Regulatory Identification

A resin tooth bonding agent is a device material, such as methylmethacrylate, intended to be painted on the interior of a prepared cavity of a tooth to improve retention of a restoration, such as a filling.

## Predicate Devices

- Scotchbond M.P. Plus Dental Adhesive System ([K942493](/device/K942493.md))
- Prime & Bond™ 2.1 Multipurpose Dentin/Enamel Bonding Agent ([K962348](/device/K962348.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

JAN 22 1997 510(k) SUMMARY K964525 Page 1 of 2

DENTSPLY

NAME &amp; ADDRESS: DENTSPLY International
570 West College Avenue
P.O. Box 872
York, PA 17405-0872
(717) 845-7511
Fax (717) 854-2343
P. J. Lehn Telefax (717) 849-6343

CONTACT: P. Jeffery Lehn    DATE PREPARED: November 11, 1996

TRADE NAME: PRIME &amp; BOND™ 2.1 MULTIPURPOSE DENTIN/ENAMEL BONDING AGENT
WITH ACTIVATOR

COMMON OR USUAL NAME: Bonding agent

CLASSIFICATION NAME: Resin tooth bonding agent 21 CFR, 872.3200

PREDICATE DEVICES: Scotchbond M.P. Plus Dental Adhesive System K942493
Prime &amp; Bond™ 2.1 Multipurpose Dentin/Enamel Bonding Agent K962348

DESCRIPTION AND INTENDED USE: Prime &amp; Bond™ 2.1 Multipurpose Dentin/Enamel
Bonding Agent is a one-component, no-mix visible light curable dental bonding
agent that contains fluoride. No changes have been made to Prime &amp; Bond™ 2.1
Multipurpose Dentin/Enamel Bonding Agent in design, material, chemical
composition or manufacturing process.

A new accessory, Prime &amp; Bond™ Self-Curing Activator, can be used with Prime &amp;
Bond™ 2.1 Multipurpose Dentin/Enamel Bonding Agent, enabling the practitioner
to use self-cure bonding techniques for the current Indications for Use of the
Bonding Agent where a dual-cured adhesive may be desired.

Also because of the self-cure capability, use of the Bonding Agent and Activator
together allows adhesion of amalgam restorations to tooth structures, which is
a NEW INDICATION FOR USE.

Prime &amp; Bond™ Self-Curing Activator is an acetone/ethanol solution of an
aromatic sulfinic acid salt. When applied in combination with Prime &amp; Bond™ 2.1
Multipurpose Dentin/Enamel Bonding Agent, the Activator initiates a self-cure,
free radical polymerization of the monomers and oligomers present in the Prime
&amp; Bond™ 2.1 Multipurpose Dentin/Enamel Bonding Agent.

{1}

510(k) SUMMARY (cont'd.)
Page 2 of 2

# TRADE NAME: PRIME &amp; BOND™ 2.1 MULTIPURPOSE DENTIN/ENAMEL BONDING AGENT WITH ACTIVATOR

TECHNOLOGICAL CHARACTERISTICS: No changes have been made to Prime &amp; Bond™ 2.1 Multipurpose Dentin/Enamel Bonding Agent in design, material, chemical composition or manufacturing process.

All of the components of the Prime &amp; Bond™ Self-Curing Activator have been used in legally marketed devices or have been proven to be safe for dental use.

We believe the fact that Prime &amp; Bond™ 2.1 Multipurpose Dentin/Enamel Bonding Agent has not been changed, the prior use of the Prime &amp; Bond™ Self-Curing Activator components in predicate devices, the performance data presented for the use of Prime &amp; Bond™ 2.1 Multipurpose Dentin/Enamel Bonding Agent with the Prime &amp; Bond™ Self-Curing Activator, and the cytotoxicity test results of the Activator active ingredient support the safety and effectiveness of Prime &amp; Bond™ 2.1 Multipurpose Dentin/Enamel Bonding Agent when used with Prime &amp; Bond™ Self-Curing Activator for the new indications for use.

---

**Source:** [https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/KLE/K964525](https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/KLE/K964525)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
