Adhese 2

K252450 · Ivoclar Vivadent, Inc. · KLE · Mar 5, 2026 · Dental

Device Facts

Record IDK252450
Device NameAdhese 2
ApplicantIvoclar Vivadent, Inc.
Product CodeKLE · Dental
Decision DateMar 5, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3200
Device ClassClass 2
AttributesTherapeutic

Intended Use

Missing tooth structure Defective restorations Areas of application: a. Pure self-curing of Adhese 2 only in combination with VivaPen DC cannulas and DC applicators - Adhesive cementation of indirect restorations and endodontic posts using Variolink® Esthetic DC in combination with all etching techniques. - Adhesive cementation of indirect restorations and endodontic posts using light- and dual-curing luting composites in combination with the etch & rinse technique. b. Light-curing of Adhese 2 at 500 to 1400 mW/cm2 (10 seconds) or at 1800 to 2200 mW/cm2 (5 seconds) - Adhesive cementation of indirect restorations with light- and dualcuring luting composites - Directly placed reconstructive build-ups and core build-ups made with light-, self- and dual-curing composites - Directly placed light-curing composite and compomer restorations - Repair of fractured composite and compomer restorations - Desensitization of hypersensitive cervical areas (Note: Provide additional cooling with an air syringe when applying the product.) - Sealing of prepared tooth surfaces before temporary/permanent cementation of indirect restorations c. Light-curing of Adhese 2 using Bluephase PowerCure in the 3sCure curing mode - Only restorations in the posterior region of permanent dentition (Class I and II, including the replacement of individual cusps) when light-cured from the occlusal aspect.

Device Story

Adhese 2 is a dual-curing, single-component dental adhesive for enamel and dentin. It is used by dentists in clinical settings for restorative procedures and cementation. The device is applied to prepared tooth surfaces following isolation and conditioning with phosphoric acid gel. It is delivered via VivaPen, bottles, or single-dose units. For self-curing, it is used with initiator-coated cannulas/applicators. For light-curing, it is activated by dental curing lights at specified intensities and durations. The adhesive facilitates the bonding of restorative or luting composites to tooth structure. By providing a reliable bond, it enables the restoration of missing tooth structure and repair of defective restorations, benefiting patients through improved structural integrity and reduced sensitivity.

Clinical Evidence

Bench testing only. Performance evaluated via shear bond strength testing per EN ISO 29022:2013. Biocompatibility evaluated per ISO 10993, ISO 7405, ISO 21726:2019, and ISO 14971:2019, concluding no clinical cytotoxic, pyrogenic, or toxicological risk when used according to instructions.

Technological Characteristics

Dual-curing, single-component resin adhesive. Materials include initiator vanadyloxalat in VivaPen cannulas. Compatible with self-etch, selective-enamel-etch, and etch & rinse techniques. Standards: EN 1641, ISO 29022. Delivery: VivaPen, bottle, or single-dose units. Curing: Light-curing (500-3300 mW/cm2) or self-curing via initiator-coated applicators.

Indications for Use

Indicated for patients with missing tooth structure or defective restorations requiring adhesive cementation of indirect restorations, endodontic posts, or direct restorative procedures including build-ups, composite/compomer restorations, repair of fractured restorations, and desensitization of hypersensitive cervical areas. Contraindicated in patients with known allergies to any ingredients or where the stipulated working technique cannot be employed; not for direct pulp capping.

Regulatory Classification

Identification

A resin tooth bonding agent is a device material, such as methylmethacrylate, intended to be painted on the interior of a prepared cavity of a tooth to improve retention of a restoration, such as a filling.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION March 5, 2026 Ivoclar Vivadent, Inc. Anderjeet Gulati Sr. Manager of QA & Regulatory Affairs 175 Pineview Drive Amherst, New York 14228 Re: K252450 Trade/Device Name: Adhese 2 Regulation Number: 21 CFR 872.3200 Regulation Name: Resin Tooth Bonding Agent Regulatory Class: Class II Product Code: KLE Dated: July 24, 2025 Received: August 4, 2025 Dear Anderjeet Gulati: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K252450 - Anderjeet Gulati Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K252450 - Anderjeet Gulati Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, MICHAEL E. ADJODHA -S Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. 510(k) Number (if known) K252450 Device Name Adhese 2 Indications for Use (Describe) - Missing tooth structure - Defective restorations Areas of application: a. Pure self-curing of Adhese 2 only in combination with VivaPen DC cannulas and DC applicators - Adhesive cementation of indirect restorations and endodontic posts using Variolink® Esthetic DC in combination with all etching techniques. - Adhesive cementation of indirect restorations and endodontic posts using light- and dual-curing luting composites in combination with the etch & rinse technique. b. Light-curing of Adhese 2 at 500 to 1400 mW/cm2 (10 seconds) or at 1800 to 2200 mW/cm2 (5 seconds) - Adhesive cementation of indirect restorations with light- and dualcuring luting composites - Directly placed reconstructive build-ups and core build-ups made with light-, self- and dual-curing composites - Directly placed light-curing composite and compomer restorations - Repair of fractured composite and compomer restorations - Desensitization of hypersensitive cervical areas (Note: Provide additional cooling with an air syringe when applying the product.) - Sealing of prepared tooth surfaces before temporary/permanent cementation of indirect restorations c. Light-curing of Adhese 2 using Bluephase PowerCure in the 3sCure curing mode - Only restorations in the posterior region of permanent dentition (Class I and II, including the replacement of individual cusps) when light-cured from the occlusal aspect. Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (6/20) Page 1 of 1 PSC Publishing Services (301) 443-6740 {4} 510(K) SUMMARY ivoclar 1 K252450 Contact: Anderjeet Gulati, Sr. Manager QA & Regulatory Affairs Ivoclar Vivadent, Inc. 175 Pineview Drive Amherst, New York 14228 716-264-2046 Anderjeet.gulati@ivoclar.com Company: Ivoclar Vivadent, AG Bendererstrasse 2, Schaan, FL-9494, Liechtenstein +423-235-3535 Date Prepared: August 1, 2025 Proprietary Name: Adhese® 2 Classification Name: Agent, Tooth Bonding, Resin (872.3200) (Classification Code KLE) Common Name: Resin tooth bonding agent Predicate Device: Adhese Universal DC (K210804) by Ivoclar Vivadent AG, Liechtenstein Device Description: Adhese® 2 is a dual-curing single-component dental adhesive for enamel and dentin that is compatible with all etching techniques (self-etch, selective-enamel-etch and etch & rinse techniques). The areas of application include adhesive cementation of indirect restorations and direct restorative procedures. Adhese 2 can either be light-cured for 10 s at a light intensity of 500 to 1,400 mW/cm2 or for 5 s at a light intensity of 1,800 to 2,200 mW/cm2. In the restoration of Class I and II cavities, Adhese 2 can be light-cured in only 3 s from the occlusal aspect at a light intensity of 2,700 to 3,300 mW/cm2. Adhese 2 is available in the VivaPen®, in bottles or in Free Stand® Single Dose units. Principles of operation: 1. Isolation 2. Pulp protection/cavity liner 3. Conditioning with phosphoric acid gel 4. Handling of the VivaPen, bottle and Free Stand Single Dose unit 5. Application of the adhesive 6. Application of the restorative or luting composite 510K Summary -1 {5} 510(K) SUMMARY ivoclar 2 # Indications for Use: - Missing tooth structure - Defective restorations # Areas of application: a. Pure self-curing of Adhese 2 only in combination with VivaPen DC cannulas and DC applicators - Adhesive cementation of indirect restorations and endodontic posts using Variolink® Esthetic DC in combination with all etching techniques. - Adhesive cementation of indirect restorations and endodontic posts using light- and dual-curing luting composites in combination with the etch & rinse technique. b. Light-curing of Adhese 2 at 500 to 1400 mW/cm2 (10 seconds) or at 1800 to 2200 mW/cm2 (5 seconds) - Adhesive cementation of indirect restorations with light- and dual-curing luting composites - Directly placed reconstructive build-ups and core build-ups made with light-, self- and dual-curing composites - Directly placed light-curing composite and compomer restorations - Repair of fractured composite and compomer restorations - Desensitization of hypersensitive cervical areas (Note: Provide additional cooling with an air syringe when applying the product.) - Sealing of prepared tooth surfaces before temporary/permanent cementation of indirect restorations c. Light-curing of Adhese 2 using Bluephase # PowerCure in the 3sCure curing mode - Only restorations in the posterior region of permanent dentition (Class I and II, including the replacement of individual cusps) when light-cured from the occlusal aspect. 510K Summary -2 {6} 510(K) SUMMARY ivoclar Comparison to Predicate: The primary predicate device to which Adhese 2 has been compared is Ivoclar Vivadent AG's Adhese Universal DC (K210804). Comparison Matrix : | | Predicate Device: | Proposed Device: | Deviation | | | --- | --- | --- | --- | --- | | Device | Ivoclar Vivadent AG: Adhese Universal DC (K210804) | Ivoclar Vivadent AG: Adhese 2 | Yes | No | | Indications for Use | – Missing tooth structure – Defective restorations | – Missing tooth structure – Defective restorations Areas of application: a. Pure self-curing of Adhese 2 only in combination with VivaPen DC cannulas and DC applicators – Adhesive cementation of indirect restorations and endodontic posts using Variolink® Esthetic DC in combination with all etching techniques. – Adhesive cementation of indirect restorations and endodontic posts using light- and dual-curing luting composites in combination with the etch & rinse technique. b. Light-curing of Adhese 2 at 500 to 1400 mW/cm2 (10 seconds) or at 1800 to 2200 mW/cm2 (5 seconds) – Adhesive cementation of indirect restorations with light- and dual-curing luting composites – Directly placed reconstructive build-ups and core build-ups made with light-, self- and dual-curing composites – Directly placed light-curing composite and compomer restorations – Repair of fractured composite and compomer restorations – Desensitization of hypersensitive cervical areas (Note: Provide additional cooling with an air syringe when applying the product.) – Sealing of prepared tooth surfaces before temporary/permanent cementation of indirect restorations c. Light-curing of Adhese 2 using Bluephase PowerCure in the 3sCure curing mode – Only restorations in the posterior region of permanent dentition (Class I and II, including the replacement of individual cusps) when light-cured from the occlusal aspect. | ☑ | ☐ | 510K Summary -3 {7} 510(K) SUMMARY ivoclar 4 | | | | | | | --- | --- | --- | --- | --- | | Precaution Measures/Contraindications / Processing restrictions/ Side effects | Contraindications: Do not use Adhesе Universal DC if the patient is known to be allergic to any of the materials' ingredients or if the stipulated working technique cannot be employed. Limitations for use: - Direct pulp capping - Do not use Adhesе Universal DC as a primer for ceramic restorative materials. A suitable ceramic primer must be used (e.g. Monobond Etch & Prime). | Contraindication - The use of the product is contraindicated if the patient is known to be allergic to any of its ingredients. - Direct pulp capping Limitations of use - Do not use Adhesе 2 as a primer for ceramic restorative materials. In such cases, suitable ceramic primers must be used. - Adhesе 2 must not be used if the stipulated working technique cannot be employed. | ☐ | ☑ | | Summary of Indications, Precaution Measures/Contraindications / Processing restrictions/ Side effects | Both products are dental adhesives. The indications and contraindications are basically the same. The area of application has been added for the new device Adhesе 2 under section "indication" to clarify how the new adhesive can be used with which light-intensities. Therefore, the products are substantially equivalent | | | | | Working Principle | Adhesе Universal DC is a dual-curing, single component dental adhesive for enamel and dentin that is compatible with all etching techniques (self-etch, selective enamel-etch, etch & rinse techniques). The areas of application include adhesive cementation of indirect restorations and direct restorative procedures. Adhesе Universal DC is available in bottles and Free Stand Single Dose units. The applicator is coated with co-initiators required for the self-curing reaction. The initiator and adhesive are mixed together by dipping the applicator into the single dose unit containing the liquid Adhesе Universal DC. | Adhesе 2 is a dual-curing single-component dental adhesive for enamel and dentin that is compatible with all etching techniques (self-etch, selective enamel-etch and etch & rinse techniques). The areas of application include adhesive cementation of indirect restorations and direct restorative procedures. Adhesе 2 is available in the VivaPen, in bottles or in Free Stand Single Dose units. The applicator is coated with co-initiators required for the self-curing reaction. The initiator and adhesive are mixed together by dipping the applicator into the single dose unit containing the liquid Adhesе 2. To enable the function for the VivaPen, the cannulas adapted to the VivaPen are coated with the initiator. | ☑ | ☐ | | Summary of Working Principle | As Adhesе 2 is also available in the VivaPen, cannulas coated with the initiator are required. This application form (VivaPen) is new compared to the predicate device Adhesе Universal DC. However, there are no significant differences. Therefore, both devices are considered to be substantially equivalent. | | | | | Delivery forms/dosage | Single Dose 0.1 g Bottle with 6 g with a separate brush | VivaPen 0.5ml, 2 ml Single Dose 0.1 g Bottle 6g | ☑ | ☐ | | Summary of Delivery forms/dosage | The new device Adhesе 2 is also available in the VivaPen in addition to the single dose and bottle compared to the predicate Adhesе Universal DC. To enable the function for the VivaPen, the cannulas are coated with the initiator vanadyloxalat. However, there are no significant differences and the products are considered to be substantially equivalent. | | | | | Storage Conditions | 24 months at 2-28 °C / 36-82 °F | 24 months at 2-28 °C / 36-82 °F | ☐ | ☑ | | Summary of Storage Conditions | No difference. | | | | | Principles of Operation | Step-by-step application: 1. Isolation 2. Pulp protection/cavity liner 3. Conditioning with phosphoric acid gel 4. Activation of the Free Stand Single Dose | Step-by-step application: 1. Isolation 2. Pulp protection/cavity liner 3. Conditioning with phosphoric acid gel | ☐ | ☑ | 510K Summary -4 {8} 510(K) SUMMARY ivoclar 5 | | 5. Application of the Adhesive 6. Curing the adhesive 7. Application of the restorative or luting composite | | | 4. Handling of the VivaPen, bottle and Free Stand Single Dose unit 5. Application of the adhesive 6. Application of the restorative or luting composite | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Summary Principles of operation | No difference. | | | | | | | | | Summary of Chemical Composition | The chemical composition slightly differs between Adhese Universal DC and the new device Adhese 2. The results of the Biocompatibility Assessment for Adhese 2 is substantially equivalent to the results of the Biocompatibility Assessment for the predicate device Adhese Universal DC. | | | | | | | | | Finished Device Specification | Applicable Standards: EN 1641:2009 – Dentistry – Medical devices for dentistry – Materials ISO 29022 – Dentistry – Adhesive – Notched-edge sheer bond strength test | | | Applicable Standards: EN 1641:2009 – Dentistry – Medical devices for dentistry – Materials ISO 29022 – Dentistry – Adhesive – Notched-edge sheer bond strength test | | | ☐ | ☐ | | Summary of Finished Device Specification | No difference. | | | | | | | | | Device Specification | Performance | Value | | Unit | Value | | Unit | | | | Specification Adhesion | --- | --- | --- | Shear bond strength (on Dentin) light-curing (LC) | ≥ 25 | MPa | ☐ | | | | Shear bond Strength (Dentin) self curing | ≥ 20 | MPa | Shear bond strength (on Dentin) dual-curing (DC) | ≥ 20 | MPa | ☑ | | Summary of Performance Specification | The specification for the new device Adhese 2 and the predicate show the same specification for the self-curing/dual-curing mode. The specification for shear bond strength LC for Adhese 2 is higher (25 MPa) as the light-curing mode generally shows higher values. The specification for both devices is substantially equivalent. | | | | | | | | | Sterilization | Not applicable. No sterilization recommendation. | | | | Not applicable. No sterilization recommendation. | | | ☐ | | Single use | Consumable material | | | | Consumable material | | | ☐ | ## Substantial Equivalence to the predicate: The indications of Adhese Universal and Adhese Universal DC are basically the same regarding the curing times for adhese 2. The biocompatibility of the new formulation was fully assessed and is substantially equivalent to Adhese Universal DC. The device performance was tested according to ISO 29022:2013 and was found to meet the relevant performance criteria. Therefore, Adhese 2 is substantially equivalent to the predicate device. ## Differences: Indication differs in that Adhese 2 indications include additional information regarding the curing times. The Working Principle is the same for both devices, but as Adhese 2 is also available in the VivaPen, cannulas coated with the initiator are required. This application form (VivaPen) is new compared to the predicate device Adhese Universal DC. However, there are no significant differences. 510K Summary -5 {9} 510(K) SUMMARY ivoclar 6 Adhese 2 is available in the VivaPen Delivery forms/dosage that the predicate device is not available in. To enable the function for the VivaPen, the cannulas are coated with the initiator vanadyloxalat. However, there are no significant differences, and the products are considered to be substantially equivalent. The Chemical Composition in Adhese 2 is slightly different, but the biocompatibility of the new formulation was assessed and is substantially equivalent to Adhese Universal DC. The device specification for shear bond strength LC for adhese 2 is higher (25 MPa) as the light-curing mode generally shows higher values. ## Non-clinical performance testing: Bench testing was performed to test the physical properties included in the Finished Device Specification for the subject device including: Shear bond strength according to EN ISO 29022:2013 Dentistry – Adhesion- Notched-edge shear bond strength test ## Biocompatibility: The subject devices were evaluated for Biocompatibility according to ISO 10993, ISO 7405, ISO 21726:2019 and ISO 14971:2019. The biological evaluation was performed based on toxicological data on relevant component materials / compounds, information on prior use of relevant component materials / compounds, data from biological tests, data on the history of clinical use or human exposure. Based on the information included in the biological evaluation reports it can be concluded that the products under evaluation are acceptable in relation to its clinical benefit. The product under assessment Adhese 2 has no clinical cytotoxic risk, holds no potential for material-mediated pyrogenicity, presents no risks for acute, sub-acute, sub-chronic or chronic toxicity, is not an inducer of adverse implantation effects, is not genotoxic, is not carcinogenic, and is not causing reproductive and development toxicity. While Adhese 2 can cause irritation and sensitization mostly likely in its unpolymerized state. If proper safety measures (according to the IFU) during handling and use are employed, health risks can be avoided. The products do not represent a toxicological risk for the patient and the user. On the basis of the toxicological evaluation of the products and the longstanding worldwide clinical use of similar materials it can be concluded that the benefits provided by the final product will exceed any potential risks produced by device materials providing that the instructions for use have been followed. ## Conclusion: Adhese Universal DC and the new device Adhese 2 are both dental adhesives. The main difference between the two is that Adhese 2 is available in the VivaPen beside the delivery forms single dose and bottle. In addition, the recipe is slightly changed for the new device. Therefore, Adhese 2 is substantially equivalent to the predicate device, Adhese Universal DC. 510K Summary -6
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...