← Product Code [KLE](/submissions/DE/subpart-d%E2%80%94prosthetic-devices/KLE) · K252151

# els unibond (K252151)

_Saremco Dental AG · KLE · Dec 5, 2025 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/KLE/K252151

## Device Facts

- **Applicant:** Saremco Dental AG
- **Product Code:** [KLE](/submissions/DE/subpart-d%E2%80%94prosthetic-devices/KLE.md)
- **Decision Date:** Dec 5, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3200
- **Device Class:** Class 2
- **Review Panel:** Dental

## Indications for Use

els unibond promotes adhesion of direct or indirect prosthesis for reconstruction or correction of functionally compromised natural dentition (e.g., deficient teeth).

## Device Story

els unibond is an extra low shrinkage, light-curing, single-component, self-etching dental adhesive. Used by dental professionals in clinical settings to create permanent, marginal-gap-free adhesion between tooth structure and light-curing filling or fixing materials (e.g., composites). The adhesive is applied to the tooth surface and light-cured to facilitate bonding of direct restorations or indirect prosthetics like inlays, onlays, crowns, and bridges. By ensuring a secure bond, the device aids in the reconstruction or correction of functionally compromised natural dentition. It is supplied as a yellow liquid that cures into a colorless solid.

## Clinical Evidence

No human clinical testing performed. Substantial equivalence supported by bench testing, including shear bond strength measurements on enamel and dentin (self-etch and total-etch modes) and biocompatibility testing per ISO 10993-1:2018 and ISO 7405:2018 (cytotoxicity, irritation, intracutaneous reactivity, and delayed type hypersensitivity).

## Technological Characteristics

Resin-based, single-component, light-cured dental adhesive. Composition includes monomers, film formers, fillers, solvents, initiators, silanization agents, and stabilizers. Non-sterile. Rx only. Form factor: liquid in bottle.

## Regulatory Identification

A resin tooth bonding agent is a device material, such as methylmethacrylate, intended to be painted on the interior of a prepared cavity of a tooth to improve retention of a restoration, such as a filling.

## Predicate Devices

- K-Bond Universal ([K241445](/device/K241445.md))
- Scotchbond Universal ([K110302](/device/K110302.md))

## Reference Devices

- Adhese Universal DC ([K210804](/device/K210804.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

December 5, 2025

Saremco Dental AG
% Nevine Erian
Regulatory Consultant
BQC Consulting, LLC
24341 Barbados Dr.
Dana Point, California 92629

Re: K252151
Trade/Device Name: els unibond
Regulation Number: 21 CFR 872.3200
Regulation Name: Resin Tooth Bonding Agent
Regulatory Class: Class II
Product Code: KLE
Dated: November 29, 2025
Received: December 1, 2025

Dear Nevine Erian:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K252151 - Nevine Erian
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

{2}

K252151 - Nevine Erian
Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

MICHAEL E. ADJODHA -S

Michael E. Adjodha, MChE, RAC, CQIA
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

els unibond
Page 7 of 34

|  Indications for Use  |   |   |
| --- | --- | --- |
|  Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K252151 | ?  |
|  Please provide the device trade name(s). |   | ?  |
|  els unibond  |   |   |
|  Please provide your Indications for Use below. |   | ?  |
|  - Direct restorations with light-curing composite-based materials.
- Indirect restorations: at the use of light-curing composite cements to fix inlays, onlays, crowns and bridges.  |   |   |
|  Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) | ?  |
|   |  ☐ Over-The-Counter Use (21 CFR 801 Subpart C)  |   |

{4}

els unibond - 510(k) Premarket Notification
Page 1

# 510(k) Summary - K252151

## Submitter

Saremco Dental AG
Gewerbestrasse 4
CH-9445 Rebstein, Switzerland
Establishment Reg. No. 3008582054

## Contact

Dr. Christoph Evers
Head of R&amp;D
Phone +41 71 775 80 90
Fax +41 71 775 80 99
E-Mail christoph.evers@saremco.ch

## Official Correspondent

Nevine Erian
Regulatory Consultant
BQC Consulting, LLC
24341 Barbados Dr.
Dana Point, CA 92629
Phone (949) 370-7155
E-mail nevine@bqcconsulting.com

## Date Prepared

December 3, 2025

|  Trade/Device Names | els unibond  |
| --- | --- |
|  Common Name | Dental Adhesive  |
|  Classification Name & Regulation Number | Resin Tooth Bonding Agent – 21 CFR 872.3200  |
|  Product Code | KLE  |

## Predicate Devices

K-Bond Universal (Spident Co., Ltd.) – K241445 – Primary Predicate
Scotchbond Universal (3M ESPE AG) – K110302 – Secondary Predicate
Adhese Universal DC (Ivoclar Vivadent, AG) – K210804 – Reference Device

{5}

els unibond - 510(k) Premarket Notification
Page 2

# Device Description

els unibond is an extra low shrinkage, light-curing, single component, self-etching adhesive, used to create a permanent marginal-gap-free adhesion between the tooth structure and the light-curing filling/fixing material, e.g. els extra low shrinkage® composite and els extra low shrinkage® flow.

# Intended Use

els unibond promotes adhesion of direct or indirect prosthesis for reconstruction or correction of functionally compromised natural dentition (e.g., deficient teeth).

# Statement of Indication for Use

- Direct restorations with light-curing composite-based materials.
- Indirect restorations: at the use of light-curing composite cements to fix inlays, onlays, crowns and bridges.

# Material Composition

els unibond is composed of monomers, film formers, fillers, solvents, initiators, silanization agents and stabilizers.

# Technological Characteristics

els unibond is a light cured extra low shrinkage adhesive.

# Non-Clinical Performance Testing

els unibond was tested for shear bond strength.

Bench test results allowed us to conclude that els unibond meets its intended uses.

# Biocompatibility

Els unibond was tested for cytotoxicity, irritation and intracutaneous reactivity, and delayed type hypersensitivity. A comprehensive biocompatibility risk assessment was conducted by a licensed toxicologist and concluded that els unibond meets the biocompatibility requirements of the following standards:

- ISO 10993-1:2018 – Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process
- ISO 7405:2018 Dentistry – Evaluation of Biocompatibility of Medical Devices Used in Dentistry

{6}

# Clinical Performance Data

The performance of methacrylate-based adhesives in the clinical environment has been well established. No human clinical testing was performed to support the substantial equivalence of els unibond.

# Substantial Equivalence

The technical characteristics of els unibond are substantially equivalent to the predicate devices.

# Material

els unibond is a resin-based adhesive as the predicate devices.

# Physical Properties

els unibond has similar physical properties as the predicate devices.

# Comparison of els unibond to Predicate Devices

|  Attribute | els unibond | K-Bond Universal | Scotchbond Universal | Adhese Universal DC  |
| --- | --- | --- | --- | --- |
|  Indications  |   |   |   |   |
|  Direct restorations with light-curing composite-based materials. | Yes | Yes | Yes | Yes  |
|  Indirect restorations: at the use of light-curing composite cements to fix inlays, onlays, crowns and bridges. | Yes | Yes | Yes | Yes  |
|  Physical Properties  |   |   |   |   |
|  Packaging | Bottle | Bottle | Bottle | Bottle  |
|  Uncured Physical State | Yellow Liquid | Yellow Liquid | Yellow Liquid | Yellow Liquid  |
|  Cured Physical State | Colorless Solid | Colorless Solid | Colorless Solid | Colorless Solid  |
|  Shear Bond Strength Self-etch enamel [MPa] | 15.1 | 15.9 | 26.4 | 25  |

els unibond - 510(k) Premarket Notification

{7}

|  Attribute | els unibond | K-Bond Universal | Scotchbond Universal | Adhese Universal DC  |
| --- | --- | --- | --- | --- |
|  Shear Bond Strength Self-etch dentin [MPa] | 25.6 | 24.2 | 35.8 | 38  |
|  Shear Bond Strength Total etch enamel [MPa] | 27.8 | 20.6 | 32.4 | 33  |
|  Shear Bond Strength Total etch dentin [MPa] | 34.2 | 29.4 | 31.8 | 38  |
|  Material Comparison  |   |   |   |   |
|  Chemical Composition | Monomers, fillers, solvents, initiators, silanization agents and stabilizers | Monomers, fillers, solvents, initiators, silanization agents and stabilizers | Monomers, fillers, solvents, initiators, silanization agents and stabilizers | Solvents, (adhesive) monomers, film formers, fillers, initiators, and stabilizers  |
|  Technical Attributes  |   |   |   |   |
|  Product Code(s) | KLE | KLE | KLE | KLE, LBH  |
|  One-component dental adhesive | Yes | Yes | Yes | Yes  |
|  Fabrication Method | Light Cured | Light Cured | Light Cured | Dual Cured  |
|  Sterile | No | No | No | No  |
|  Rx or OTC | Rx | Rx | Rx | Rx  |

## Differences between els unibond and Predicate Devices

The differences in the physical properties and chemical compositions between els unibond and the predicate devices do not impact safety and effectiveness, as the finished clinical product is a biocompatible dental adhesive, regardless of the material variation.

## Conclusion

Information provided in this application demonstrates that els unibond is substantially equivalent to the predicate devices. els unibond shares the same indications, similar material composition, similar physical properties and technological characteristics as the predicate devices.

els unibond - 510(k) Premarket Notification

---

**Source:** [https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/KLE/K252151](https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/KLE/K252151)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
