Ultra SEP is a self etching primer intended to be used for the preparation of an enamel surface prior to bonding with a light cure orthodontic adhesive.
Device Story
Ultra SEP is a self-etching primer used in orthodontic procedures to prepare enamel surfaces for bonding brackets. The liquid material is dispensed from a bottle into a light-impervious mixing well and applied to teeth using a microbrush. It combines etching and priming into a single step, creating a surface for light-cure adhesive paste. Used in dental/orthodontic clinics by professionals. The output is a chemically prepared tooth surface that facilitates bracket adhesion, improving the efficiency of orthodontic treatment workflows.
Clinical Evidence
No clinical data. Evidence consists of bench testing, specifically shear bond strength comparisons against the predicate and reference devices, and oral toxicity testing. Biocompatibility was demonstrated via ISO 7405 and ISO 10993-5 evaluations.
Technological Characteristics
Self-etching primer consisting of acidic monomer and solvent. Non-sterile. Delivered via Boston round bottle with dropper cap, light-impervious mixing well, and microbrushes. Class II device (21 CFR 872.3200).
Indications for Use
Indicated for preparation of enamel surfaces prior to bonding with light cure orthodontic adhesives in adults and pediatric patients. For professional use by orthodontic or dental office personnel.
Regulatory Classification
Identification
A resin tooth bonding agent is a device material, such as methylmethacrylate, intended to be painted on the interior of a prepared cavity of a tooth to improve retention of a restoration, such as a filling.
Predicate Devices
3M Unitek Transbond™ Plus Self Etching Primer (K984246)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 8, 2021
Reliance Orthodontic Products, Inc Brian Dean Biomedical Engineering Consultant Cook Device Solutions 7640 Delaine Ct Indianapolis, Indiana 46254
Re: K210945
Trade/Device Name: Ultra SEP Regulation Number: 21 CFR 872.3200 Regulation Name: Resin Tooth Bonding Agent Regulatory Class: Class II Product Code: KLE Dated: May 12, 2021 Received: May 13, 2021
Dear Brian Dean:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# SECTION 6.0
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K210945
Device Name Ultra SEP
Indications for Use (Describe)
Ultra SEP is a self etching primer intended to be used for the preparation of an enamel surface prior to bonding with a light cure orthodontic adhesive.
Type of Use (Select one or both, as applicable)
[X] Prescription Use (Part 21 CFR 801 Subpart D) [] Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW. *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs .gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid 0MB number."
Form Approved: 0MB No. 0910-0120
Expiration Date: 06/30/2020
See PRA Statement below.
{3}------------------------------------------------
SECTION 5.0 - 510 (k) Summary
# K210945
Image /page/3/Picture/2 description: The image shows the logo for Reliance Orthodontic Products. The logo consists of a circular icon with the letter "R" inside, followed by the word "RELIANCE" in bold, sans-serif font. Below the word "RELIANCE" are the words "ORTHODONTIC PRODUCTS" in a smaller, sans-serif font.
Reliance Orthodontic Products Inc. 1540 West Thorndale Ave. Itasca, IL 60143 USA Phone: (630) 773-4009 | Website: www.relianceorthodontics.com
#### Submitter Information:
Date of 510(k) Summary: 3/19/2021 Submitter: Reliance Orthodontics Products Inc. Preparer and Contact Name: Brian Dean, BME Consultant 510(k) Owner: Paul Gange, President Address: 1540 West Thorndale Ave. Itasca, IL 60143 USA Phone: 630-773-4009 / Fax: 630-250-7704
Email: bdean@cookds.com Email: pgange(@relianceorthodontics.com
### Device Name and Classification:
| Common Name of Device: | Self Etching Primer |
|--------------------------|---------------------------|
| Device Proprietary Name: | Ultra SEP |
| Classification Panel: | Dental |
| Classification Number: | 872.3200 |
| Classification Name: | Resin Tooth Bonding Agent |
| Class: | II |
| Product Code: | KLE |
#### Substantial Equivalence:
Legally marketed devices to with equivalence is claimed:
- 3M Unitek Transbond™ Plus Self Etching Primer 510(k) (Submitted to FDA as K984246 Prompt L-. Pop by Espe Dental AG)
#### Device Description
Ultra SEP is a self etching primer which can etch up to 6 teeth with a single drop. The material is dispensed from a bottle into a light impervious mixing well, and transferred to teeth by a saturated microbrush. Ultra SEP creates an ideal surface for a light cure paste to adhere teeth and brackets for orthodontic treatment.
Ultra SEP is used in combination with a light impervious mixing well (SEPW) and microbrushes (MB).
### Indications for Use and Population:
Intended Use: Ultra SEP is a self etching primer intended to be used for the preparation of an enamel surface prior to bonding with a light cure orthodontic adhesive.
Diseases/Conditions for diagnosis, treatment, prevention, cure, or mitigation: None
Population: Adults and Pediatrics. Orthodontic or Dental Office professional use.
# Predicate Device:
- 3M Unitek Transbond 100 Plus Self Etching Primer 510(k) K984246 dated 11/17/1998 is similar in intended use, ● handling and technology compared to the device described in this 3M product is considered the primary predicate.
- Reliance S.E.P. 510(k) K071688 dated 08/24/2007 is similar in intended use, handling and technology compared . to the device described in this submission; this Reliance Orthodontics product is considered a reference.
{4}------------------------------------------------
# SECTION 5.0 - 510 (k) Summary
Image /page/4/Picture/1 description: The image shows the logo for Reliance Orthodontic Products. The logo consists of a circular symbol with the letter "R" inside, followed by the word "RELIANCE" in bold, uppercase letters. Below the word "RELIANCE" are the words "ORTHODONTIC PRODUCTS" in smaller, uppercase letters.
Reliance Orthodontic Products Inc. 1540 West Thorndale Ave. Itasca, IL 60143 USA Phone: (630) 773-4009 | Website: www.relianceorthodontics.com
Comparison of Technological Characteristics with the Predicate Device:
| Submission | Reliance Orthodontics<br>Ultra SEP | 3M Unitek TransbondTM Plus<br>Self Etching Primer | Reliance Orthodontics<br>S.E.P. |
|-------------------------|-------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|
| Classification | Class II, KLE, 21CFR872.3200 | Class II, KLE, 21CFR872.3200 | Class II, KLE, 21CFR872.3200 |
| Intended Use | Self Etching Primer, Single Step<br>Delivery for use with light cure<br>bonding orthodontic adhesives | Self Etching Primer, Single Step<br>Delivery for use with light cure<br>bonding orthodontic adhesives | Self Etching Primer, Single Step<br>Delivery for use with light cure<br>bonding orthodontic adhesives |
| Device<br>Description | • Combines etching and priming<br>in one easy step<br>• Effective bond strength | • Combines etching and priming<br>in one easy step<br>• Effective bond strength | • Combines etching and priming<br>in one easy step<br>• Effective bond strength |
| Chemical<br>Composition | Acidic Monomer / Solvent | Acidic Monomer / Solvent | Acidic Monomer / Solvent |
| Dispensing | Boston Round Bottle with<br>Dropper Cap | A single-patient use foil pack<br>contains pre-measured etchant and<br>light cure primer. | A multi-patient use hand-held<br>dispenser with SEP filled cartridge. |
| Sterilization | Non-Sterile | Non-Sterile | Non-Sterile |
#### Non-Clinical Data Submitted
Shear bond strength testing comparisons conducted between the Reliance Orthodontics Ultra SEP, the predicate device 3M Unitek Transbond™ Plus Self Etching Primer, and the reference Reliance Orthodontics S.E.P. demonstrate that the applicant device is equivalent to the legally marketed device in terms of effectiveness.
Additionally, toxicity testing was conducted using an Oral Toxicity Test method and was found to be non toxic.
#### Conclusions
Clinical testing was determined to be unnecessary with the approval of all non-clinical testing, based on the nonclinical testing conducted and equivalency in characteristics demonstrated between Reliance Ultra SEP and 3M Unitek Transbond 110 Plus Self Etching Primer devices. In conjunction with biocompatibility demonstrated via ISO 7405 and ISO 10993-5 evaluations, evidence has been submitted to demonstrate Reliance Ultra SEP is safe and equivalent to or better than the 3M Unitek Transbond™ Plus Self Etching Primer (K984246) device in terms of performance as an Orthodontic Self Etching Primer.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.