← Product Code [KLE](/submissions/DE/subpart-d%E2%80%94prosthetic-devices/KLE) · K192961

# ADH19 (K192961)

_3M Deutschland GmbH · KLE · Oct 31, 2019 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/KLE/K192961

## Device Facts

- **Applicant:** 3M Deutschland GmbH
- **Product Code:** [KLE](/submissions/DE/subpart-d%E2%80%94prosthetic-devices/KLE.md)
- **Decision Date:** Oct 31, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3200
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** 3rd-Party Reviewed

## Indications for Use

A material primarily intended to be used as a bonding-promoting substance between tooth substance and dental restorations. It may also be used as a dentin sealant and as a bonding agent for repair of restorations.

## Device Story

ADH19 is a one-component dental adhesive used by dentists in clinical settings to bond restorative materials to tooth structure. It functions via self-etch, selective enamel etch, or total-etch modes. Unlike its predicate, it requires no separate dual-cure activator for self- or dual-cure materials. The clinician applies the adhesive to the tooth surface or restoration substrate; it acts as a bonding agent for direct and indirect procedures, including fillings, veneers, and core build-ups. By facilitating adhesion between tooth substance and restorative materials (metals, ceramics, composites), it improves restoration retention and seals dentin, potentially reducing sensitivity and preventing microleakage.

## Clinical Evidence

Bench testing only. In vitro studies compared shear bond strength of ADH19 to Scotchbond Universal across multiple substrates (enamel, dentin, titanium, zirconia, NPM alloy, composite, high gold alloy, feldspathic glass ceramic, lithium disilicate) using various curing modes. Results confirmed ADH19 performance is comparable to the predicate.

## Technological Characteristics

One-component dental adhesive containing monomers, fillers, solvents, initiators, silanization agents, and stabilizers. Delivery via vial or L-Pop unit dose. Supports self-etch, selective enamel etch, and total-etch modes. Complies with ISO 29022:2013 for notched-edge shear bond strength testing. Biocompatibility assessed per ISO 10993 series and ISO 7405.

## Regulatory Identification

A resin tooth bonding agent is a device material, such as methylmethacrylate, intended to be painted on the interior of a prepared cavity of a tooth to improve retention of a restoration, such as a filling.

## Predicate Devices

- Scotchbond Universal ([K110302](/device/K110302.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".

October 31, 2019

3M Deutschland GmbH % Prithul Bom Responsible Third Party Official Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114

Re: K192961

Trade/Device Name: Adh19 Regulation Number: 21 CFR 872.3200 Regulation Name: Resin tooth bonding agent Regulatory Class: Class II Product Code: KLE Dated: October 21, 2019 Received: October 22, 2019

# Dear Prithul Bom:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-

542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

for Srinivas Nandkumar, Ph.D. Acting Assistant Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known) K192961

Device Name ADH19

Indications for Use (Describe)

Direct Indications:

· Bonding for all methacrylate-based light-, dual-, and self-cure composite or compomer filling materials

·Root surface desensitization

· Bonding of methacrylate-based fissure sealants

· Protective varnish for glass ionomer fillings

·Repair of composite and compomer fillings

· Sealing of cavities prior to placement of amalgam restorations

Indirect Indications:

· Cementation of indirect restoration with Suglue 3 and other resin cements (follow applicable Instructions for Use)

· Bonding for all methacrylate-based light-, self-, and dual-cure core build-up materials and cements

· Cementation of veneers when combined with RelyX Veneer Cement

· Intraoral repair of composite restorations, porcelain fused to metal, and all-ceramic restorations without extra primer

•Sealing of cavities and preparation of tooth stumps prior to temporary cementation of indirect restorations

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (7/17)

Page 1 of 1

PSC Publishing Services (301) 443-6740 EF 

{3}------------------------------------------------

3M Deutschland GmbH

Carl-Schurz-Straße 1 41453 Neuss Germany

Postal Address: ESPE Platz 82229 Seefeld Germany

Tel.:+49 (0)8152 700-0 Fax+49 (0)8152 700-1366 Internet: www.3MESPE.de E-Mail: info3MESPE@3M.com WEEE-Req.-Nr. DE 36963167 VAT-ID: DE 120679179

# 510(k) Summary

| Submitter:           | 3M Deutschland GmbH<br>ESPE Platz<br>82229 Seefeld<br>Germany<br>Establishment Registration Number: 9611385                              |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person       | Ruediger Franke<br>Regulatory Affairs Specialist<br>Phone: +49-8152-700 1802<br>Fax: +49-8152-700 1869<br>e-mail: ruediger.franke@3M.com |
| Date:                | August 21, 2019                                                                                                                          |
| Trade Name:          | ADH19                                                                                                                                    |
| Common Name:         | Adhesive                                                                                                                                 |
| Classification Name: | Resin Tooth Bonding Agent<br>(21 CFR 872.3200, product code KLE)                                                                         |
| Device Class:        | Class II                                                                                                                                 |

Predicate Devices ................................... Scotchbond Universal (K110302)

## Description of Device

ADH19 is a one-component dental adhesive that is based on the chemistry of Scotchbond Universal (by 3M Deutschland GmbH, K110302, cleared as ADHESIVE EXL-759). It can be used in a self-etch mode, selective enamel etch mode or in a total-etch mode for both direct and indirect dental restorative procedures. It is intended to bond methacrylate-based restorative, cement and sealant materials to dentin, enamel, glass ionomer and various indirect restorative substrates (metals, glass ceramics, alumina and zirconia). The primary use will be with lightcured materials, however it will have the capability to also bond self- or dual-cure composite and cement materials. In contrast to Scotchbond Universal, no separate Dual Cure Activator will be necessary in this case.

Sitz: 41453 Neuss · Handelsregister: B 1878 Amtsgericht Neuss Geschäftsführer: Christiane Grün, Manfred Hinz, Oliver Leick Vorsitzende der Geschäftsführung: Christiane Grün: Vorsitzender des Aufsichtsrates: Günter Gressler

Image /page/3/Picture/11 description: The image is a logo for "Deutsche Dental Industrie". The logo consists of a stylized letter "G" inside of three concentric circles. To the right of the circles is the text "Deutsche Dental Industrie" stacked vertically.

3M Deutschland GmbH

{4}------------------------------------------------

#### Applicable Standards for Product Tests

- . ISO 29022: 2013: Dentistry - Adhesive - Notched-edge sheer bond strength test
#### Indications for ADH19

## Direct Indications:

- Bonding for all methacrylate-based light-, dual-, and self-cure composite or compomer filling materials
- Root surface desensitization .
- . Bonding of methacrylate-based fissure sealants
- . Protective varnish for glass ionomer fillings
- Repair of composite and compomer fillings .
- Sealing of cavities prior to placement of amalgam restorations .

## Indirect Indications:

- Cementation of indirect restorations in combination with Suglue 3 and other resin cements . (follow applicable Instructions for Use)
- Bonding for all methacrylate-based light-, self-, and dual-cure core build-up materials and . cements
- . Cementation of veneers when combined with RelyX Veneer Cement
- . Intraoral repair of composite restorations, porcelain fused to metal, and all-ceramic restorations without extra primer
- Sealing of cavities and preparation of tooth stumps prior to temporary cementation of indirect restorations

#### Comparison

ADH19 was compared to Scotchbond Universal regarding indications for use, intended use, composition, technology and physical and mechanical properties. The tables below summarize the indications and technology of ADH19 and predicate devices:

| Indications Comparison                                                                                        | ADH19 | Scotchbond Universal<br>(K110302)       |
|---------------------------------------------------------------------------------------------------------------|-------|-----------------------------------------|
| Indications for Use<br>(from the labeling)                                                                    |       |                                         |
| Bonding for all methacrylate-based light-,<br>dual-, and self-cure composite or<br>compomer filling materials | X     | X<br>partly with Dual Cure<br>Activator |
| Root surface desensitization                                                                                  | X     | X                                       |
| Bonding of methacrylate-based fissure<br>sealants                                                             | X     | X                                       |
| Protective varnish for glass ionomer fillings                                                                 | X     | X                                       |

{5}------------------------------------------------

| Indications Comparison                                                                                                                                                                                                                               | ADH19 | Scotchbond Universal<br>(K110302)       |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------|-----------------------------------------|
| Repair of composite and compomer filling                                                                                                                                                                                                             | ×     | ×                                       |
| Sealing of cavities prior to placement of<br>amalgam restorations                                                                                                                                                                                    | ×     | ×                                       |
| Cementation of indirect restorations in<br>combination with Suglue 3 for ADH19 or<br>Suglue 10 for Scotchbond Universal and<br>other resin cements (follow applicable<br>Instructions for Use)                                                       | ×     | ×<br>partly with Dual Cure<br>Activator |
| Bonding for all methacrylate-based light-,<br>self-, and dual-cure core build-up materials<br>and cements                                                                                                                                            | ×     | ×<br>partly with Dual Cure<br>Activator |
| Cementation of veneers when combined<br>with RelyX Veneer Cement                                                                                                                                                                                     | ×     | ×                                       |
| Intraoral repair of composite restorations,<br>porcelain fused to metal, and all-ceramic<br>restorations without extra primer                                                                                                                        | ×     | ×                                       |
| Sealing of cavities and reparation of tooth<br>stumps prior to temporary cementation of<br>indirect restorations                                                                                                                                     | ×     | ×                                       |
| Intended Use<br>A material primarily intended to be used as<br>a bonding-promoting substance between<br>tooth substance and dental restorations. It<br>may also be used as a dentin sealant and<br>as a bonding agent for repair of<br>restorations. | ×     | ×                                       |

Table Comparison of indications

| Technology                                                                                                                             | ADH19        | Scotchbond Universal<br>(K110302) |
|----------------------------------------------------------------------------------------------------------------------------------------|--------------|-----------------------------------|
| Vial for multiple dosing and the L-Pop<br>delivery device for unit dosed dispensing                                                    | X            | X                                 |
| One-component dental adhesive                                                                                                          | X            | X                                 |
| Self-etch mode, selective enamel etch<br>mode or in a total-etch mode for both<br>direct and indirect dental restorative<br>procedures | X            | X                                 |
| Need to incorporate a separate ceramic<br>or metal primer                                                                              | Not required | Not required                      |
| No separate Dual Cure Activator required<br>to bond self- or dual-cure composite and<br>cement materials                               | Not required | Required                          |

Table Comparison to Predicate Technology

ADH19 and Scotchbond Universal are dental adhesive materials containing monomers, fillers, solvents, initiators, silanization agents and stabilizers. 3M Deutschland GmbH is providing

{6}------------------------------------------------

information to the Agency regarding information to FDA about the composition of ADH19 and compared this to the predicate device.

In vitro testing was conducted to show the performance of ADH19 compared to the predicate device Scotchbond Universal regarding shear bond strength to enamel (self-etch & total-etch), to dentin (self-etch & total-etch), to restoration materials titanium, zirconia, NPM alloy, composite, high gold alloy, feldspathic glass ceramic, lithium disilicate, with RelyX Ultimate to enamel and dentin, and with different self cure and dual cure composites to enamel and dentin. The results of ADH19 are comparable to Scotchbond Universal. In summary, 3M Deutschland GmbH concludes that ADH19 is substantially equivalent to the predicate devices regarding performance and physical and mechanical properties.

#### Biocompatibility

The biocompatibility assessment for the products was conducted in accordance with the following guidance:

| Guidance                                                                     | Edition       | Title                                                                                                                                                                                                                         |
|------------------------------------------------------------------------------|---------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| US FDA Docket<br>Number FDA-2013-<br>D-0350. CDRH<br>Document Number<br>1811 | June 16, 2016 | Use of International Standard ISO 10993-1, Biological<br>evaluation of medical devices - Part 1: Evaluation and testing<br>within a risk management process - Guidance for Industry<br>and Food and Drug Administration Staff |
| ISO 10993-1                                                                  | 2018          | Evaluation and testing within a risk management process                                                                                                                                                                       |
| ISO 10993-3                                                                  | 2014          | Tests for genotoxicity, carcinogenicity and reproductive<br>toxicity                                                                                                                                                          |
| ISO 10993-5                                                                  | 2009          | Tests for in vitro cytotoxicity                                                                                                                                                                                               |
| ISO 10993-6                                                                  | 2016          | Tests for local effects after implantation                                                                                                                                                                                    |
| ISO 10993-10                                                                 | 2010          | Tests for irritation and skin sensitization                                                                                                                                                                                   |
| ISO 10993-11                                                                 | 2017          | Tests for systemic toxicity                                                                                                                                                                                                   |
| ISO 10993-12                                                                 | 2012          | Sample preparation and reference materials                                                                                                                                                                                    |
| ISO 10993-18                                                                 | 2005          | Chemical characterization of materials                                                                                                                                                                                        |
| ISO/TR 10993-22                                                              | 2017          | Guidance on nanomaterials                                                                                                                                                                                                     |
| ISO 7405                                                                     | 2018          | Evaluation of biocompatibility of medical devices used in<br>dentistry                                                                                                                                                        |

Table Guidances for biocompatibility assessment

The biocompatibility of ADH19 has been assessed by a board-certified toxicologist according to recommendations in FDA guidance and internationally recognized standards for medical and dental devices. The conclusion of the assessment is ADH19 is safe for its intended use.

## Conclusion

Comparisons of the indications for use/intended use, composition, technology, and physical and mechanical properties showed that ADH19 does not raise any new questions about safety and effectiveness and is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/KLE/K192961](https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/KLE/K192961)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
