Peak® Z is intended for use as a surface treatment for restorations made of the following materials: Zirconia . Aluminum oxide . Metal/Alloy . Titanium .
Device Story
Peak® Z is a single-component resin tooth bonding agent containing 12-methacryloyldodecylphosphate (MDP). It functions as a chemical primer to enhance adhesion between resin-based materials and restorative surfaces (zirconia, alumina, metal/alloys, titanium). The device is supplied in a pre-mixed syringe or bottle/brush delivery system. It is used by dentists in a clinical setting to prepare surfaces for restorations or repairs. By creating a chemical bond, the primer increases adhesion, potentially improving the longevity and stability of dental restorations. The device is applied directly to the restoration surface prior to resin application.
Clinical Evidence
No clinical trials were conducted. Evidence consists of bench testing (shear bond strength, artificial aging, stability) and a literature review supporting safety and effectiveness for all ages when prescribed by a dentist. Bench testing compared Peak® Z to the predicate device (K091705) and a control without primer, demonstrating comparable or superior performance. Biocompatibility testing (cytotoxicity, sensitization, genotoxicity) was performed per ISO 10993-1.
Technological Characteristics
Single-component resin bonding agent containing 12-methacryloyldodecylphosphate (MDP). Supplied in pre-mixed syringe or bottle/brush delivery. Biocompatibility tested per ISO 10993 (cytotoxicity, sensitization, genotoxicity).
Indications for Use
Indicated for surface treatment of dental restorations made of zirconia, aluminum oxide, metal/alloys, or titanium in patients of all ages requiring dental restoration or repair, as prescribed by a dentist.
Regulatory Classification
Identification
A resin tooth bonding agent is a device material, such as methylmethacrylate, intended to be painted on the interior of a prepared cavity of a tooth to improve retention of a restoration, such as a filling.
{1}------------------------------------------------
# Section 5
# Traditional 510(k) Summary
This summary of the Traditional 510(k) substantial equivalence information is being submitted in accordance with the requirements of 21 CFR 807 .92.
#### Applicant's Name and Address
Ultradent Products, Inc. 505 West 10200 South South Jordan, UT 84095
SEP 06 2013
Contact Person: Title: Telephone: FAX: Date Summary Prepared: Diane Rogers Manager of Regulatory and Global Affairs (800) 552-5512 x4491, (801) 553-4491 (801) 553-4609 May 10, 2013
#### Name of the Device
| Trade Name: | Peak® Z |
|------------------------------|---------------------------|
| Common Name: | Resin tooth bonding agent |
| Device Classification: | Class II |
| Classification Product Code: | KLE |
#### Legally Marketed Predicate Devices to Which Equivalence is Claimed
The predicate device is: K091705 Prime Plus (Z Prime Plus) by Bisco Inc. located at 1100 W. Irving Park Road, Schaumburg IL 60193.
Peak® Z is very similar to our predicate device in that both devices are intended to be used as resin tooth bonding agents.
Indications for Use: Peak® Z is intended for use as a surface treatment for restorations made of the following materials:
- Zirconia .
- Aluminum oxide �
- Metal/Alloy .
- Titanium �
{2}------------------------------------------------
Product Description: Peak® Z is a single component surface treatment that enhances bond values when applied to zirconia, alumina, and metal/alloys. It contains 12methacryloyIdoeceylphosphate (MDP), which produces a chemical bond that significantly increases adhesion between a resin-based material and the bonding surface of the restoration. Peak® Zr Primer can be applied through a syringe delivery or brush and bottle.
## Technological characteristics
| Product | 510(k) Number | Indications for Use |
|------------------------|----------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Peak® Z | | Peak® Z is intended for use as a surface treatment for restorations made of the following materials:<br>Zirconia Aluminum oxide Metal/Alloy Titanium |
| Primer plus ( Z-Prime) | K091705 | The principal uses of the Primer Plus are:<br>1. Indirect restorations (such as composite, endodontic posts, metal/metal alloys, porcelain, zirconia, alumina, ceramics, and hybrid ceramics)<br>2. Intraoral repairs of fractured crowns and bridges (such as metals/metal alloys, porcelain, zirconia, alumina, ceramics, or composite resin). |
| Characteristics | Surface treatment to enhance<br>restorations and repairs | Surface treatment to enhance<br>restorations and repairs |
| Human Factors | Syringe delivery or brush and<br>bottle | Brush and bottle |
{3}------------------------------------------------
| Biocompatibility | ISO 10993 | ISO 10993 |
|------------------|---------------|---------------|
| | Cytotoxicity | Cytotoxicity |
| | Sensitization | Sensitization |
| | Irritation | Irritation |
| | Genotoxicity | Genotoxicity |
Peak® Z is supplied in a pre-mixed syringe for easy delivery and no waste, and also in bottles with brushes that can be directly applied to surfaces.
Primer plus is supplied in single bottles that can be directly applied to surfaces.
The patient population is intended for all ages that need a restoration or repair as prescribed by a dentist. The device is a primer that is intended to prepare surface treatments of restorations and repairs.
### Brief Description of Testing Performed
The following bench tests were conducted during the R & D phase on Peak ® Z and compared to K091705 Primer Plus) by Bisco Inc. Final test results are in Section 18 "Bench Testing".
Shear bond: This in-vitro test determines how well the material enhances the bonding to various surfaces compared to both the Bisco's system and the Ultradent system without a primer. This shows whether or not a primer is needed and if the R&D formulation, Peak® Z, (PAZR41) is comparable to a product that is currently on the market and is well-received.
Shear bond with artificial aging and conditions: This in-vitro test artificially replicates a possible extreme condition that may occur in the clinical setting. This will indicate whether or not the material is robust enough compared to a product currently on the market.
Stability: The material was tested at various data points to verify that the function of the product did not degrade under normal and accelerated conditions.
### Clinical Summary
A complete Clinical Summary of Peak® Z is included in Section 20: We conducted a literature study to show safety and effectiveness of this product can be used on any age patient when treatment is prescribed by a dentist. The device has the same technological characteristics compared to K091705 Primer Plus (Z Prime Plus) by Bisco inc.
Primer Plus has been widely used by numerous dentists in the dental industry.
{4}------------------------------------------------
The efficacy or suitability to the intended purpose of Peak® Z has been demonstrated by a combination of in-house testing and side-by-side comparisons to a predicate device currently on the market. Results of our bench testing indicates that Peak® Z performs as well or better than the predicate device currently on the market.
### Summary Risk/Benefit Review
Considering the safe history of our predicate, K091705 Primer Plus (Z Prime Plus) by Bisco Inc., Peak® Z is substantially equivalent and considered to be a safe medical device. Our research indicates that our predicate has been used by many dentists and large group practices in the United States and purchased by a large number of international distributors. To date, there have been no reported complaints of local or systemic adverse effects associated with the use of the predicate product.
Peak® Z was tested for biocompatibility in Cytotoxicity, Sensitization and Genotoxicity tests according to ISO 10993-1. An abstract of the testing along with signed test reports are included in Section 15 "Biocompatibility" of this submission.
In conclusion, Peak® Z has been designed and manufactured with the intended use and claims for the product in mind. Scientific literature, etc. has been collected and evaluated to determine safety and efficacy of similar products used for the same indication. Following the clinical review as documented above, Ultradent Products, Inc. deems that when this device is used under the conditions and for the purposes intended, it will not compromise the clinical condition or the safety of the patient and the association with its use constitutes acceptable risks when weighed against to the patient. Therefore, the product is compatible with a high level of protection of health and safety and may be released to the market.
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an emblem of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring MD 20993-0002
#### September 6, 2013
Ultradent Products, Incorporated Ms. Diane Rogers Manager of Regulatory and Globel Affairs 505 West 10200 South SOUTH JORDAN UT 84095
Re: K131357
Trade/Device Name: PEAK® Z Regulation Number: 21 CFR 872.3200 Regulation Name: Resin Tooth Bonding Agent Regulatory Class: II Product Code: KLE Dated: June 6, 2013 Received: June 10, 2013
Dear Ms. Rogers:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
#### Page 2 - Ms. Rogers
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Image /page/5/Picture/7 description: The image shows the name "Mary S. Runner - S" in a bold, sans-serif font. The text is horizontally oriented and appears to be a signature or label. The letters are clearly legible, with consistent spacing between the words and initials.
Kwame Ulmer M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
## Statement of Indications for Use
K131357 510(k) Number (if known): __
Device Name: Peak® Z
Indications for Use:
Peak® Z is intended for use as a surface treatment for restorations made of the following materials:
- Zirconia .
- Aluminum oxide .
- Metal/Alloy �
- . Titanium
Prescription Use _ X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page _1_ of _1_
# Andrew I. Steen -S 2013.09.06 09:32:53 -04'00'
. '
(Posted November 13, 2003)
(Division Sign-Off) Tivision of Anesthesiology, General Hospital ifection Control, Dental Devices
10(k) Number: K13357
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.