CLEARFIL TRI-S BOND PLUS SINGLE DOSE STANDARD / VALUE / INTRODUCTORY PACK

K111975 · Kuraray Medical, Inc. · KLE · Oct 14, 2011 · Dental

Device Facts

Record IDK111975
Device NameCLEARFIL TRI-S BOND PLUS SINGLE DOSE STANDARD / VALUE / INTRODUCTORY PACK
ApplicantKuraray Medical, Inc.
Product CodeKLE · Dental
Decision DateOct 14, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3200
Device ClassClass 2

Indications for Use

Direct restorations using light cured composite resin Cavity sealing as a pretreatment for indirect restorations Treatment of exposed root surfaces Intraoral repairs of fractured restorations made of porcelain, ceramic or composite resin Surface treatment of non-precious metal posts Post cementation and core build·ups using dual·cured composite resin CLEARFIL DC CORE PLUS Core build-ups using light-cured composite resin

Device Story

Single-component, light-cured dental bonding agent; supplied in single-dose containers with applicator tips for direct cavity placement. Used by dental professionals for bonding dentin and enamel. Facilitates direct composite restorations, cavity sealing, root surface treatment, and repair of fractured restorations. Biocompatibility verified per ISO 10993 and ISO 7405. Effectiveness validated via tensile bond strength and fluoride-releasing property testing.

Clinical Evidence

Bench testing only. Effectiveness validated through comparative tensile bond strength testing and fluoride-releasing property analysis against predicate devices. Biocompatibility of new chemical ingredients confirmed via ISO 10993 and ISO 7405 standards.

Technological Characteristics

Single-component, light-cured resin bonding agent. Contains two new chemical ingredients evaluated for biocompatibility per ISO 10993 and ISO 7405. Delivered in single-dose containers with applicator tips. No international performance standards exist for this device type; performance validated via comparative bench testing.

Indications for Use

Indicated for dental patients requiring direct light-cured composite restorations, cavity sealing for indirect restorations, treatment of exposed root surfaces, intraoral repair of fractured porcelain/ceramic/composite restorations, surface treatment of non-precious metal posts, and post cementation/core build-ups.

Regulatory Classification

Identification

A resin tooth bonding agent is a device material, such as methylmethacrylate, intended to be painted on the interior of a prepared cavity of a tooth to improve retention of a restoration, such as a filling.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 510(k) Summary OCT 1 4 2011 ## 3-1. 510(k) owner (submitter) | 3-1. 510(k) owner (submitter) | | | | |-------------------------------|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------| | | 1) Name | KURARAY MEDICAL INC. | | | | 2) Address | 1621 Sakazu, Kurashiki, Okayama 710-0801, Japan | | | | 3) Contact person | Michio Takigawa<br>Quality Assurance Department | | | | 4) Contact person in US | Kiyoyuki Arikawa<br>KURARAY AMERICA INC.<br>600 Lexington Avenue, 26th Floor<br>New York, NY 10022<br>Tel: (212)-986-2230 (Ext. 115) or (800)-879-1676<br>Fax: (212)-867-3543 | | | 3-2. Name of Device | | | | | | 1) Trade / Proprietary name | CLEARFIL TRI-S BOND PLUS Single Dose | | | | 2) Classification name | Resin Tooth Bonding Agent<br>(21 CFR section 872.3690. Product code: KLE) | | | | 3) Common name | Dental bonding agent | | | 3-3. Predicate device | | | | | | 1) CLEARFIL TRI-S BOND | 510(k) Number:<br>Product Code:<br>21 CFR Section:<br>Applicant: | K042913<br>KLE<br>872.3690<br>KURARAY MEDICAL INC. | | | 2) CLEARFIL DC BOND | 510(k) Number:<br>Product Code:<br>21 CFR Section:<br>Applicant: | K062382<br>KLE<br>872.3200<br>KURARAY MEDICAL INC. | | | 3) PANAVIA F 2.0 | 510(k) Number:<br>Product Code:<br>21 CFR Section:<br>Applicant: | K032455<br>EMA<br>872.3275<br>KURARAY MEDICAL INC. | | | 4) CLEARFIL MAJESTY Flow | 510(k) Number:<br>Product Code:<br>21 CFR Section:<br>Applicant: | K063593<br>EBF<br>872.3690<br>KURARAY MEDICAL INC. | ・・ : {1}------------------------------------------------ #### 3-4. Device Description . l ) The subject device is a single-component. light-cured bonding agent that allows simultaneous treatment of both dentin and enamel. A single dose of it comes in a container with an applicator tip for fast, direct placement into the cavity. #### 3-5. Substantial Equivalence Discussion - l) Intended uses The intended uses of the subject device were written up based on those of CLEARFIL TRI-S BOND and CLEARFIL DC BOND, the predicate devices. Therefore, the intended uses of the subject device are substantially equivalent to those of the predicate devices. #### 2) Chemical ingredients / Safety Except for 2 new chemical ingredients, all ingredients in the subject device have been used in the predicate devices. Regarding the predicate devices, there have not been any reported problems or recalls according to the post market adverse event reporting requirements in the US. And 2 new ingredients have been evaluated referring to ISO 10993 series and ISO 7405. As the result, it was confirmed that these substances were biocompatible. In conclusion, it can be said that the safety of the subject device is substantially equivalent to that of the predicate devices. - 3) Technological characteristics /Effectiveness and Performance Since there have not been any international standards concerning performance of this type of device, certain tests were performed on this device, in comparison with the predicate device and it was confirmed that this device was substantially equivalent to the predicate device in terms of the effectiveness and performance. Tensile bond strength test and Fluoride releasing property were performed to validate the substantial equivalence of the subject device with the predicate device in terms of effectiveness and performance for the intended uses. The test has exhibited almost the same results for the subject device indicating that the subject device was as effective as and performs as good as or better than the predicate device. #### 3-6. Biocompatibility Except for 2 new chemical ingredients, all ingredients in the subject device have been used in the predicate devices. Regarding the predicate devices, there have not been any reported problems or recalls according to the post market adverse event reporting requirements in the US. And 2 new ingredients have been evaluated referring to ISO 10993 series and ISO 7405. As the result, it was confirmed that these substances were biocompatible. In conclusion, it was concluded that the biocompatibility of the subject device could be assured. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 OCT 1 4 2011 Kuraray Medical, Incorporated C/O Mr. Kiyoyuki Arikawa General Manger Kuraray America, Incorporated 600 Lexington Avenue, 26th Floor New York, NY 10022 Re: K111975 Trade/Device Name: Clearfil Tri-S Bond Plus Single Dose Regulation Number: 21 CFR 872.3200 Regulation Name: Resin Tooth Bonding Agent Regulatory Class: II Product Code: KLE Dated: July 6, 2011 Received: July 12, 2011 Dear Mr. Arikawa: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class 111 (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ #### Page 2 – Mr. Arikawa Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htmg for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours. hr for Anthony D. Watson, B.S.. M.S.. M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health #### Enclosure {4}------------------------------------------------ ### Indications for Use 510(k) Number (if known): Device Name: CLEARFIL TRI-S BOND PLUS Single Dose Indications for Use: - [1] Direct restorations using light cured composite resin - [2] Cavity sealing as a pretreatment for indirect restorations - [3] Treatment of exposed root surfaces - [4] Intraoral repairs of fractured restorations made of porcelain, ceramic or composite resin - [5] Surface treatment of non-precious metal posts - [6] Post cementation and core build·ups using dual·cured composite resin CLEARFIL DC CORE PLUS - [7] Core build-ups using light-cured composite resin Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use N/A (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Susan Kuaiep (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...