K063780 · Bisco, Inc. · KLE · Jan 19, 2007 · Dental
Device Facts
Record ID
K063780
Device Name
ACE BOND SE
Applicant
Bisco, Inc.
Product Code
KLE · Dental
Decision Date
Jan 19, 2007
Decision
SESE
Submission Type
Special
Regulation
21 CFR 872.3200
Device Class
Class 2
Indications for Use
As a Universal Self-Etching Adhesive System, ACE BOND SE, as the name implies, is the only product one needs to etch, prime, and bond to a tooth structure and dental substrates. ACE BOND SE is used for: 1. Direct restorations (composite, amalgam) 2. All indirect restorations (composite, metal, porcelain) 3. Desensitization of crown preparations prior to impression making/provisionalization 4. Composite core build-ups 5. Composite to metal/set amalgam (direct veneering) 6. Root desensitization 7. New amalgam to existing amalgam 8. Repairs (composite-to-composite, composite-to-porcelain)
Device Story
ACE BOND SE is a light-cured, two-component, ethanol/water-based, single-step dental adhesive system. It functions as a self-etching agent that simultaneously etches, primes, and bonds to tooth structures and various dental substrates. Used by dental professionals in clinical settings, the adhesive is applied to prepared tooth surfaces or restorative materials to facilitate bonding for direct and indirect restorations, core build-ups, and desensitization. The low-viscosity, pink-colored resin is cured via light activation. By integrating etching and bonding into one step, the device simplifies the restorative workflow, potentially reducing procedure time and improving clinical efficiency for the practitioner.
Clinical Evidence
Bench testing only. Oral toxicity testing was performed, confirming the device is non-toxic. No clinical data provided.
Technological Characteristics
Light-cured, unfilled, multifunctional methacrylate-based resin. Ethanol/water-based, two-component, single-step system. Low viscosity. Formulated for self-etching, priming, and bonding. No specific ASTM standards or software components described.
Indications for Use
Indicated for dental patients requiring etching, priming, and bonding for direct/indirect restorations (composite, amalgam, metal, porcelain), composite core build-ups, tooth desensitization (dentin/enamel/root surfaces), and dental repairs.
Regulatory Classification
Identification
A resin tooth bonding agent is a device material, such as methylmethacrylate, intended to be painted on the interior of a prepared cavity of a tooth to improve retention of a restoration, such as a filling.
{0}------------------------------------------------
Ko63780
Page 6-1
# Section 6 SPECIAL 510 (k) SUMMARY
| Applicant: | Bisco, Inc.<br>1100 W. Irving Park Road<br>Schaumburg IL, 60193 | JAN 1 9 2007 |
|----------------------|------------------------------------------------------------------|--------------|
| Contact Person: | Benjamin Lichtenwalne-<br>Tel: 847-534-6146<br>Fax: 847-534-6111 | |
| Date Prepared: | December 20, 2006 | |
| Trade Name: | ACE BOND SE | |
| Common Name: | Dental Adhesive | |
| Classification/Name: | Resin Tooth Bonding Agent Class II per 21 CFR 872.3200 | |
### Description of Applicant Device:
ACE BOND SE is a light-cured all-in-one bending agent that combines etching, priming and bonding into one single step. ACE BOND SE is an ethanol/water-based, two-component, single step adhesive system.
### Intended uses of Applicant Device:
The principle uses of the ACE BOND SE adhesive is to etch, prime, and bond for direct and indirect restorations involving dentin, enamel, light-cure, and dual-cure composites. amalgam/metals, porcelain, and core build-ups. ACE BOND SE can also be used for desensitization of tooth structures such as hyper ensitive dentin/enamel and exposed root surfaces.
#### Predicate Devices:
ALL-IN-ONE from Bisco, Inc, cleared under K 50647 dated April 20, 2006.
# Significant Performance Characteristics:
#### ALL-IN-ONE to ACE BOND SE
| Property | ALL-IN-ONE | ACE BOND SE |
|------------------------------------|----------------------------------------------------------------------------------------|-------------------------------------------------------------------------------|
| Intended use | Self -Etching, Single Step Dental<br>Adhesive | Self-Etching, Single Step<br>Dental Adhesive |
| Chemical composition | Light-cure, unfilled,<br>multifunctional methacrylate<br>based resin. | Light-cure, unfilled.<br>multifunctional methacrylate<br>based resin. |
| Mechanical /physical<br>properties | Medium viscosity, light pink<br>colored dental etching, priming,<br>and bonding agent. | Low viscosity, pink colored<br>dental etching, priming, and<br>bonding agent. |
Side by side comparisons of ACE BOND SE to the device ALL-IN-ONE clearly demonstrates that the applicant device is substantially equivalent to the legally marketed device. ACE BOND SE was tested for oral toxicity and was found to be non-toxic. It is concluded that the information supplied in this submission has proven the safety and efficacy of ACE BOND SE.
{1}------------------------------------------------
Public Health Service
DEPARTMENT OF HEALTH & HUMAN SERVICES, USA
od and Drug Administration 200 Corporate Boulevard Rockville MD 20850
Mr. Benjamin Lichtenwalner Regulatory Affairs Manager Bisco, Incorporated 1100 West Irving Park Road Schaumburg, Illinois 60193
JAN 1 9 2007
Re: K063780
Trade/Device Name: ACE BOND SE Regulation Number: 872.3200 Regulation Name: Resin Tooth Bonding Agent Regulatory Class: II Product Code: KLE Dated: December 20, 2006 Received: December 26, 2006
Dear Mr. Lichtenwalner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adultcration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Page 2 – Mr. Lichtenwalner
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Syrite H. Michael PhD.
Chiu Lin. Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
06378V 510 (k) Number (if known): / 1
Device Name: ACE BOND SE
Indications for Use:
As a Universal Self-Etching Adhesive System, ACE BOND SE, as the name implies, is the only product one needs to etch, prime, and bond to a tooth structure and dental substrates.
ACE BOND SE is used for:
- 1. Direct restorations (composite, amalgam)
- 2. All indirect restorations (composite, metal, porcelain)
- 3. Desensitization of crown preparations prior to impression making/provisionalization
- 4. Composite core build-ups
- 5. Composite to metal/set amalgam (direct veneering)
- 6. Root desensitization
- 7. New amalgam to existing amalgam
- 8. Repairs (composite-to-composite, composite-to-porcelain)
Prescription Use ✓ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Suan Runner
(Division Spr-0.
Division of A
Lateral Hospital,
Infection Conces
510(k) Number: K063780
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.