The product is used for indirect pulp capping, direct pulp capping, root end filling and Repair of perforation.
Device Story
ENDOCEM MTA PREMIXED REGULAR is a root canal filling material supplied in a pre-loaded syringe. It is a paste-type material that does not require pre-mixing; it sets by absorbing moisture from the root canal environment. Used by dental professionals in clinical settings for pulp capping, root end filling, and perforation repair. The device functions as a biocompatible filler; it is placed into the root canal to seal and repair dental structures. The material's physical properties and biocompatibility support its use in permanent root canal applications. Benefits include ease of use due to the pre-mixed delivery system and clinical utility in endodontic procedures.
Clinical Evidence
Bench testing only. Testing included visual inspection, volume, setting time (4.2 min), solubility, radiopacity, and biocompatibility (ISO 10993-5 cytotoxicity, ISO 10993-10 skin sensitization, ISO 10993-11 acute systemic toxicity, and ISO 10993-3 genotoxicity). All results met preset criteria.
Technological Characteristics
Root canal filling material; paste form; pre-loaded syringe. Composition: Zirconium dioxide, calcium silicate, DMSO, calcium aluminate, calcium sulfate, bentonite, hydroxypropyl methyl cellulose, silica, lithium carbonate, water, ethanol. Sets via moisture absorption. Conforms to ISO 6876. Non-sterile. Shelf life: 2 years.
Indications for Use
Indicated for dental patients requiring indirect pulp capping, direct pulp capping, root end filling, or repair of perforation.
Regulatory Classification
Identification
A root canal filling resin is a device composed of material, such as methylmethacrylate, intended for use during endodontic therapy to fill the root canal of a tooth.
Predicate Devices
iRoot BP Injectable Root Canal Repair Filling Material (K082943)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 14, 2023
Maruchi Sungwook Jang CEO 2-208, Medical Industry Complex Bldg., 42-10, Taejanggongdan-gil Wonju-si. Gangwon-do 26311 SOUTH KOREA
Re: K213757
Trade/Device Name: Endocem MTA Premixed Regular Regulation Number: 21 CFR 872.3820 Regulation Name: Root Canal Filling Resin Regulatory Class: Class II Product Code: KIF Dated: May 20, 2022 Received: May 26, 2022
Dear Sungwook Jang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Michael E. Adjodha -S
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213757
Device Name ENDOCEM MTA PREMIXED REGULAR
Indications for Use (Describe)
The product is used for indirect pulp capping, direct pulp capping, root end filling and Repair of perforation.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.qov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image contains the logo for Maruchi. The logo consists of a stylized flower-like graphic on the left and the word "MARUCHI" in a sans-serif font on the right. The flower graphic is multi-colored, with petals in shades of blue, green, purple, pink, and orange. The word "MARUCHI" is in a dark gray or black color.
#### 510(k) Summary
Date: February 13, 2023
### 1. SUBMITTER
#### MARUCHI
2-208, Medical Industry Complex Bldg., 42-10, Taejanggongdan-gil, Wonju-si, Gangwon-do, 26311, Republic of Korea TEL : +82-33-734-0330 FAX : +82-33-746-2804 Contact Name: Sung Wook, Jang Email: ra@endocem.com
### 2. DEVICE
·Trade Name: ENDOCEM MTA PREMIXED REGULAR ·Common Name: Root Filling Material · Classification Name: Resin, Root canal filling · Regulation Number 872.3820 ·Class: 2 ·Classification Product Code: KIF
### 3. PREDICATE DEVICE
K082943, iRoot BP Injectable Root Canal Repair Filling Material, Innovative BioCeramix Inc
K102867, iRoot FS, Innovative BioCeramix Inc
# 4. DEVICE DESCRIPTION
ENDOCEM MTA PREMIXED REGULAR is a root canal filling material conforming to ISO 6876, in a pre-loaded syringe that does not require any pre-mixing and is set by absorbing moisture from the root canal environment.
# 5. INDICATIONS FOR USE
The product is used for indirect pulp capping, direct pulp capping, root end filling and
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for Maruchi. The logo consists of a colorful, stylized flower-like design with rounded petals in shades of blue, green, purple, and pink. To the right of the floral design is the word "MARUCHI" in bold, black sans-serif letters. The overall design is clean and modern.
Repair of perforation.
# 6. PERFORMANCE TESTING (NON-CLINICAL)
The following test articles were tested based on the referenced standard. All the test results met the preset test criteria.
- Testing institution's method Visual, Volume test, Setting time, pH .
- ISO 6876 - Package, Extraneous matter, Setting time, Solubility, Radioopacity
- ISO 10993-5 Cytotoxicity (MTT) .
- ISO 10993-10 - Guinea Pig Maximization Test for Skin sensitization (GPMT)
- ISO 10993-11 - Acute systemic toxicity
- ISO 10993-3 Genotoxicity (Mammalian chromosome aberration test & . Bacterial Reverse Mutation)
| | Proposed Device | Predicate Device | Reference Device | Discuss/Justify the Differences |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------|
| 510(k) Number | New | K082943 | K102867 | - |
| Trade Name | ENDOCEM MTA<br>PREMIXED<br>REGULAR | iRoot BP Injectable<br>Root Canal Repair<br>Filling Material | iRoot FS | - |
| Manufacturer | MARUCHI | Innovative<br>BioCeramix Inc | Innovative<br>BioCeramix Inc | - |
| Common Name | Root Filling<br>Material | Root Filling Material | Root Filling Material | Equivalent |
| Classification<br>Name | Resin, Root canal filling | Resin, Root canal filling | Resin, Root canal filling | Equivalent |
| Device Class | 2 | 2 | 2 | Equivalent |
| Product Code | KIF | KIF | KIF | Equivalent |
| Device<br>Description | ENDOCEM MTA<br>PREMIXED<br>REGULAR is a<br>root canal filling<br>material<br>conforming to ISO<br>6876, in a pre-<br>loaded syringe that<br>does not require<br>any pre-mixing and<br>is set by absorbing<br>moisture from the<br>root canal<br>environment. | iRoot BP Injectable<br>Root Canal Repair<br>Filling Material<br>(iRoot BP) is a<br>convenient ready-to-<br>use white hydraulic<br>premixed injectable<br>BioAggregate paste<br>developed for<br>permanent root canal<br>repair and filling<br>applications. iRoot<br>BP is an insoluble,<br>radiopaque and<br>aluminum-free<br>material based on a<br>calcium silicate<br>composition, which<br>requires the presence<br>of water to set and<br>harden. iRoot BP<br>does not shrink<br>during setting and<br>demonstrates<br>excellent physical<br>properties. iRoot BP<br>is packaged in a<br>preloaded syringe<br>and is supplied with<br>disposable tips. | iRoot FS Fast Set<br>Root Repair Material<br>(iRoot FS) is a<br>convenient ready-to-<br>use fast setting white<br>hydraulic premixed<br>bioceramic paste<br>developed for<br>permanent root canal<br>repair of root<br>perforation and<br>root canal resorption,<br>and root end filling,<br>apexification and<br>pulp capping<br>applications. iRoot<br>FS is an insoluble,<br>radiopaque and<br>aluminum-free<br>material based on a<br>calcium silicate<br>composition, which<br>requires the presence<br>of water to set and<br>harden. iRoot FS<br>does not shrink<br>during setting and<br>demonstrates<br>excellent physical<br>properties. iRoot FS<br>is available as a | |
| Indications for<br>Use | The product is used<br>for indirect pulp<br>capping, direct pulp<br>capping, root end<br>filling and Repair<br>of perforation. | Repair of Root<br>Perforation<br>Repair of Root<br>Resorption<br>Root End Filling<br>Apexification<br>Pulp Capping | preloaded syringe<br>with disposable tips<br>and a preloaded<br>container.<br>Repair of Root<br>Perforation<br>Repair of Root<br>Resorption<br>Root End Filling<br>Apexification<br>Pulp Capping | Equivalent |
| Intended user | Dental professional | Dental professional | Dental professional | Equivalent |
| Standards | ISO 6876 | ISO 6876 | ISO 6876 | Equivalent |
| Setting times | 4.2 min | Minimum of 2 hours | Approximately 20<br>minutes | Differences |
| Chemical<br>Composition | Zirconium dioxide<br>Calcium silicate<br>DMSO<br>(Dimethylsulfoxide<br>)<br>Calcium aluminate<br>Calcium sulfate<br>Bentonite<br>Hydroxypropyl<br>Methyl Cellulose<br>Silica<br>Lithium carbonate<br>Water<br>Ethanol | Zirconium oxide<br>Calcium silicates<br>Tantalum pentoxide<br>Calcium phosphate<br>monobasic and filler<br>agents. | Zirconium oxide<br>Dicalcium silicate<br>Tantalum pentoxide<br>Tricalcium silicate<br>Calcium Sulfate and<br>filler agents | Differences |
| Liquid Formula | Paste type | Paste type | Paste type | Equivalent |
| Packaging | Pre-loaded syringe | Pre-loaded syringe | Pre-loaded syringe | Equivalent |
| Sterile | Non-sterile | Non-sterile | Non-sterile | Equivalent |
| Shelf Life | 2 years | 2 years | 2 years | Equivalent |
# 7. SUBSTANITAL EQUIVALENCE
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the Maruchi logo. The logo consists of a colorful flower-like graphic on the left and the word "MARUCHI" in black, sans-serif font on the right. The graphic has petals in shades of blue, green, pink, and yellow.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the Maruchi logo. The logo consists of a colorful flower-like graphic on the left and the word "MARUCHI" in a simple, sans-serif font on the right. The flower graphic has multiple petals in different colors, including blue, green, pink, and purple.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Maruchi. The logo consists of a colorful flower-like graphic on the left and the word "MARUCHI" in a sans-serif font on the right. The graphic is made up of several petals in different colors, including blue, green, purple, and pink.
## 8. SUBSTANTIAL EQUIVALENCE DISCUSSION
ENDOCEM MTA PREMIXED REGULAR has the same Indications for Use and the principle of operations as the predicate device and reference device. It is intended purpose as they are placed into the root canal as a root filling materials which met the requirement according to ISO 6876. It has similar physical and biocompatible properties and demonstrates comparable performance specifications to the predicate devices.
The chemical compositions might slightly different from the predicate devices, both are used calcium silicate as base material, and additional components are used to improve flowability.
The bench and biocompatibility testing performed demonstrates that any differences in their technological characteristics do not raise any new questions as to safety and effectiveness. Therefore, it is concluded that ENDOCEM MTA PREMIXED REGULAR is substantially equivalent to the predicate devices. Hence, its equivalent is acceptable.