Sealapex tubes, Sealapex bulk package, Sealapex Xpress dual-barrel syringes

K152959 · Sybron Dental Specialties · KIF · Apr 14, 2016 · Dental

Device Facts

Record IDK152959
Device NameSealapex tubes, Sealapex bulk package, Sealapex Xpress dual-barrel syringes
ApplicantSybron Dental Specialties
Product CodeKIF · Dental
Decision DateApr 14, 2016
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 872.3820
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Sealapex/Sealapex Xpress is a calcium hydroxide, polymeric resin, root canal filling material that is used in conjunction with gutta percha or silver endodontic points.

Device Story

Sealapex/Sealapex Xpress is a non-eugenol, radiopaque, calcium hydroxide polymeric resin root canal sealer. It functions as a two-part base/catalyst paste system that hardens upon mixing in the presence of root canal moisture. The device is used by licensed dental professionals during endodontic procedures to fill microscopic gaps between the obturator and canal walls. It is delivered either via hand-mixed tubes or a dual-barrel syringe with single-use automix tips. The sealer acts as a barrier against blood, exudates, and infection, facilitating successful root canal obturation.

Clinical Evidence

Bench testing only. Performance evaluated per ISO 6876:2012 for consistency (flow), film thickness, and radiopacity. Results showed flow of 23.1 mm (vs 22.8 mm predicate), film thickness of 33.85 µm (vs 20.3 µm predicate), and radiopacity of 6.0 mm Al (vs >3.3 mm Al predicate). All results met ISO specifications.

Technological Characteristics

Two-part paste/paste system (base/catalyst). Base contains calcium oxide, resin, titanium dioxide, and silica. Catalyst contains salicylate moieties, filler, titanium dioxide, and silica. Fumed silica acts as a rheological modifier. Hardens via moisture-activated setting reaction. Delivered via laminate tubes or polypropylene dual-barrel syringes with automix tips. Complies with ISO 6876:2012.

Indications for Use

Indicated for use as a root canal sealing and filling material in conjunction with gutta percha or silver endodontic points to seal prepared root canal apical foramen and tubules from blood, exudates, and infection.

Regulatory Classification

Identification

A root canal filling resin is a device composed of material, such as methylmethacrylate, intended for use during endodontic therapy to fill the root canal of a tooth.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 April 14, 2016 Sybron Dental Specialties Jennifer Dzidrums 1332 S. Lone Hill Ave. Glendora, CA 91740 Re: K152959-S001 Trade/Device Name: SEALAPEX/ SEALAPEX XPRESS Regulation Number: 21 CFR 872.3820 Regulation Name: Resin, Root Canal Filling Regulatory Class: Class II Product Code: KIF Dated: March 07, 2016 Received: March 08, 2016 Dear Jennifer Dzidrums: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {1}------------------------------------------------ the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Erin I. Keith -S Erin Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control. and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use 510(k) Number (if known) Device Name Sealapex/Sealapex Xpress Indications for Use (Describe) Sealapex / Sealapex Xpress is a calcium hydroxide, polymeric resin, root canal filling material that is used in conjunction with gutta percha or silver endodontic points. | Type of Use (Select one or both, as applicable) | |-----------------------------------------------------------------------------------------------------------------------------| | <div> <span> <span style="font-size: 12px;"> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </span> </span> </div> | | <div> <span> <span style="font-size: 12px;"> ☐ Over-The-Counter Use (21 CFR 801 Subpart G) </span> </span> </div> | | X | Prescription Use (Part 21 CFR 801 Subpart D) |_ | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # SECTION 6. 510(k) SUMMARY FOR SEALAPEX/ SEALAPEX XPRESS Image /page/3/Picture/2 description: The image shows the logo for Sybron Dental Specialties. The logo features the letters "sds" in a bold, sans-serif font, with a curved line above the letters. Below the letters, the words "Sybron Dental Specialties" are written in a smaller, sans-serif font. The logo is blue and white. ## Sealapex/ Sealapex Xpress - 1. Submitter Information: Sybron Dental Specialties 1717 W. Collins Ave. Orange CA, 92687 Contact Person: Jennifer Dzidrums Telephone Number: 909-962-5650 909-962-5694 Fax Number: Date Prepared: April 8, 2016 - 2. Device Name: | · Proprietary Name: | Sealapex and Sealapex Xpress | |------------------------|------------------------------| | · Classification Name: | Resin, Root Canal Filling | - 872.3820 • CFR Number: - · Device Class: II - Product Code: KIF ### 3. Predicate Device: The proposed Sealapex Sealapex Xpress is substantially equivalent to the legally marketed device, Sealapex 4 (K010940), cleared on June 20, 2001, product code KIF. ### 4. Description of Device: Sealapex/Sealapex Xpress is a non-eugenol, radiopaque, calcium hydroxide polymeric resin root canal filling material. It is indicated for use as a root canal sealing and filling material, and is used during an endodontic procedure to seal off the prepared root canal apical foramen and tubules from blood, exudates, and infection. The proposed is a twopart, base/catalyst - paste/paste system that is mixed in equal portions. Then, the mixture is carried to the root canal with endodontic obturation points or directly dispensed in the root canal depending on delivery method, e.g. tube or dual-barrel syringe. The proposed is available packaged in tubes (Sealapex) and dual-barrel syringes (Sealapex Xpress) used with single-use automix tips. The tubes are dispensed onto a mixing pad, hand mixed and then applied to the canal with an endodontic point. Sealapex Xpress has the same formulation and indications for use as Sealapex, but is branded as "Sealapex Xpress" since it offers the convenience of single-use automix tips that remove {4}------------------------------------------------ the need for hand mixing. The dual-barrel and single-use automix tips allow the dental professional to deliver the desired volume of Sealapex Xpress directly into the prepped root canal. | Accessories Used with Sealapex<br>Xpress | Manufacturer of Accessory | |------------------------------------------|------------------------------------------------------------------------------| | Dual-barrel syringe | Sulzer Mixpac AG<br>Ruetistrasse 7<br>Haag Sankt gallen, Switzerland<br>9469 | | Auto-mix tips | Sulzer Mixpac AG<br>Ruetistrasse 7<br>Haag Sankt gallen, Switzerland<br>9469 | ### Table 6.1 Description of Accessories used with the Proposed Sealapex Xpress # 5. Statement of Indications for Use: Sealapex/Sealapex Xpress is a calcium hydroxide, polymeric resin, root canal filling material that is used in conjunction with gutta percha or silver endodontic points. - 6. Technological Characteristics: Sealapex/Sealapex Xpress is intended to fill the microscopic gaps between the canal filler/obturator and canal walls. Reactive components that let Sealapex Xpress sealer set in root-canal are Calcium Oxide in base paste and Salicylate moieties in accelerator (catalyst) paste. Material hardens upon mixing of the base and accelerator pastes in the presence of moisture in the root canal. The paste base formulation of the proposed Sealapex Xpress and the predicate device, Sealapex 4 (K010940), is very similar, the only difference being an increase in the amount of Resin and Titanium Dioxide, along with a slight addition of a Silica in the proposed Sealapex/Sealapex Xpress base formulation. The paste catalyst formulation of the proposed Sealapex Xpress and the predicate device, Sealapex 4 (K010940), is very similar, the only difference being a decrease in amount of filler, a decrease in Titanium Dioxide, along with the slight addition of a Silica in the proposed Sealapex/Sealapex Xpress catalyst formulation. Titanium Dioxide (USP) is a common non-toxic white pigment approved by FDA as a food additive, and the minor change does not affect performance or biocompatibility of the paste. Fumed silica filler serves as a rheological modifier. It does not participate in the setting reaction. Both fillers in the proposed base and catalyst formulations are similar in physical characteristics and surface treated to make the filler particle more hydrophobic and to improve the compatibility with the formulation. The Resin is an inert, water-insoluble material that helps thicken the paste and does not participate in setting reaction. The slight increase in the amount of Resin should stabilize the consistency of the paste and does not adversely affect the equivalence and performance of the product. {5}------------------------------------------------ The slightly modified formulation for the proposed Sealapex Xpress has successfully been on the EU market for several years. Performance testing on stability, work time, set time and slump were performed on the modified formula to validate that the proposed products meet the current product specifications and are substantially equivalent to the predicate Sealapex 4 (K010940). - 7. Non-Clinical Performance Data: Verification activities were performed in accordance with design control requirements as specified in 21 CFR 820.30 and ISO 13485 Medical Devices- Quality Management Systems and the results demonstrated substantial equivalence to the predicate. - 8. Summary of Performance Testing Performance Testing Bench Performance testing for the proposed Sealapex/ Sealapex Xpress included testing for Consistency (Flow), Film Thickness, and Radiopacity per ISO 6876:2012 - Dentistry -Root Canal Sealing Materials. In the Physical and Mechanical Properties of Table 6.2, the film thickness has a 13.5 micron (um) difference between the predicate and proposed devices. This difference does not affect the substantial equivalence of the subject and predicate devices. The sealer was tested to meet ISO 6876:2012 standard, which states that a sealer shall have a film thickness of not more than 50 um. The proposed Sealapex and Sealapex Xpress are below the maximum 50 um allowance. Therefore, the predicate and proposed sealers will flow sufficiently into the root canal and adapt to canal shape to fill voids. Table 6.2 displays both an internal and the ISO 6876:2012 test method for determining Flow. The internal test method for Flow differs from the current ISO 6876:2012 test procedure due to the amount of sample used for testing. The predicate and proposed devices produced Flow results of 48.5 mm and 51.0 mm, respectively, based on internal test methods. This data shows no substantial difference between the predicate and proposed devices. Per ISO 6876:2012, the predicate and proposed devices show no substantial difference based on Flow test results of 22.8 mm and 23.1 mm, respectively. The data analyzed from the various tests substantiate that the proposed Sealapex/ Sealapex Xpress is substantially equivalent to the predicate Sealapex 4 (K010940). - 9. Substantial Equivalence: Sealapex 4 (K010940) is an existing device which was granted market clearance by FDA in 2001. Sybron Dental Specialties, on behalf of SybronEndo, seeks only to slightly modify the formula of the existing device cleared under K010940. Additionally, the device would be produced at a sister facility in Scafati, Italy; and the manufacturer name will change to SybronEndo. Table 6.2 depicts the comparison of performance and mechanical properties and modification associated with the proposed Sealapex Xpress. #### Table 6.2: Predicate Sealanex 4 (K010940) and Proposed Sealapex Xpress Comparison Table {6}------------------------------------------------ | Element | Predicate Device- Sealapex 4 | Proposed Sealapex/Sealapex Xpress | | |---------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------| | 510(k) | K010940 | K152959 | | | Trade Name | Sealapex 4 | Sealapex and Sealapex Xpress | | | Target Users | Licensed dental professionals | Licensed dental professionals | | | Indications for Use | Sealapex 4 is a calcium<br>hydroxide, polymeric resin, root<br>canal filling material that is<br>used in conjunction with gutta<br>percha or silver endodontic<br>points | Sealapex/Sealapex Xpress is a<br>calcium hydroxide, polymeric<br>resin, root canal filling material<br>that is used in conjunction with<br>gutta percha or silver endodontic<br>points | | | Device<br>Description | Sealapex 4 is a non-eugenol,<br>radiopaque, calcium hydroxide<br>polymeric resin root canal<br>filling material. Sealapex 4 is a<br>two-part, base/catalyst –<br>paste/paste system. The two-<br>part system is packaged in tubes<br>that is hand mixed, then placed<br>into the prepped root canal. | Sealapex/Sealapex Xpress is a<br>non-eugenol, radiopaque,<br>calcium hydroxide polymeric<br>resin root canal filling material.<br>Sealapex/Sealapex Xpress is a<br>two-part, base/catalyst –<br>paste/paste system. The two part<br>system is packaged in tubes that<br>are hand mixed, then placed into<br>the prepped root canal, or in a<br>dual-barrel syringe (Sealapex<br>Xpress) that does not require<br>hand mixing. | | | Common<br>Name | Root Canal Sealer | Root Canal Sealer | | | Classification<br>Name | Resin, Root Canal Filling per<br>CFR § 872.3820 | Resin, Root Canal Filling per<br>CFR § 872.3820 | | | Class | II | II | | | Product Code | KIF | KIF | | | Base Formula | Contains:<br>Resin<br>Titanium Dioxide | Contains:<br>Increased range of Resin<br>Increased range of Titanium<br>Dioxide<br>Added: Silica | | | Catalyst<br>Formula | Contains:<br>Filler<br>Titanium Dioxide | Contains:<br>Decreased range of Filler<br>Increased range of Titanium<br>Dioxide<br>Added: Silica | | | Legal<br>Manufacturer/<br>Specification<br>Developer | Kerr Corporation<br>28200 Wick Road<br>Romulus, MI 48174<br>USA | SybronEndo<br>1332 S. Lone Hill Ave.<br>Glendora, CA 91740<br>USA | | | Manufacturing<br>Location | Kerr Corporation<br>28200 Wick Road<br>Romulus, MI 48174<br>USA | Kerr Italia, S.R.L.<br>Via Passanti, 332<br>Scafati Salerno, I-84018<br>ITALY | | | Packaging | Laminate Tubes | Laminate Tubes and | | | | | | Polypropylene dual-barrel<br>syringes used with single-use<br>automix tips. | | | <b>Physical and Mechanical Properties</b> | | | | Form | Two Pastes<br>(Catalyst & Base) | Two Pastes<br>(Catalyst & Base) | | | Mix Ratio | Equal Volumes | Equal Volumes | | | Work Time<br>@ 37°C<br>(98.6° F),<br>100% RH | > 60 minutes | > 60 minutes | | | Setting time<br>@ 37°C<br>(98.6° F),<br>100% RH | < 24 hours | < 24 hours | | | Consistency<br>(Flow) per<br>Internal Test<br>Method and<br>Specification | 48.5 mm | 51.0 mm | | | Consistency<br>(Flow) per<br>ISO<br>6876:2012<br>Method | 22.8 mm | 23.1 mm | | | Film<br>Thickness | 20.3 µm | 33.85 µm | | | Solubility and<br>Disintegration | N/A | N/A | | | Radiopacity | > 3.3mm Al | 6.0 mm Al | | {7}------------------------------------------------ # 10. Conclusion as to Substantial Equivalence: The slight modifications of Sealapex Xpress base and catalyst formulations and the move of the manufacturing location to Scafati, Italy does not affect the intended use of the device nor does it alter the fundamental scientific technology of the device. The nonclinical testing demonstrates that the proposed Sealapex Xpress performs as well as the predicate device based on the specifications set by ISO 6876:2012 testing requirements. The proposed Sealapex/Sealapex Xpress are substantially equivalent to the predicate Sealapex 4 (K010940).
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...