K140955 · Avalon Biomed, Inc. · KIF · Jun 11, 2014 · Dental
Device Facts
Record ID
K140955
Device Name
MTA2.1 MATERIAL
Applicant
Avalon Biomed, Inc.
Product Code
KIF · Dental
Decision Date
Jun 11, 2014
Decision
SESE
Submission Type
Special
Regulation
21 CFR 872.3820
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The MTA2.1 MATERIAL is intended for use for dental procedures that contact pulp and periradicular tissues, as well as obturation and sealing of root canals.
Device Story
MTA2.1 Material is a dental restorative material consisting of a hydraulic tricalcium silicate powder and a water-based gel. Clinicians mix the components and apply the material to the site of dental procedures, including pulp capping, pulpotomies, root-end filling, and root canal obturation. The material hydrates to form a hard substance containing calcium hydroxide, which induces hydroxyapatite crystal precipitation. Used by dentists in clinical settings, the material provides a stable, radiopaque seal for vital pulp and periradicular tissues. It is designed to be color-stable over time, offering an improvement over previous formulations regarding discoloration resistance.
Clinical Evidence
No clinical tests were performed. Evidence is based on bench testing and design validation, including ADA 57 standard testing for radiopacity, solubility, dimensional stability, film thickness, and flow. Compressive strength, leaching (arsenic/lead), washout, and color stability were evaluated per ISO 9917. Biocompatibility was assessed via cytotoxicity testing.
Technological Characteristics
Hydraulic tricalcium silicate powder and water-based gel. Radiopaque. Sets via hydration into a hard substance containing calcium hydroxide. Conforms to ADA 57 and ISO 9917 standards for physical properties. No software or electronic components.
Indications for Use
Indicated for dental procedures contacting vital pulp and periradicular tissues, including pulp capping, cavity lining, base material, pulpotomies, root-end filling, apexification, perforation repair, root resorption, and obturation (pulpectomy) in primary or secondary teeth.
Regulatory Classification
Identification
A root canal filling resin is a device composed of material, such as methylmethacrylate, intended for use during endodontic therapy to fill the root canal of a tooth.
K980332 — MTA MATERIAL · Dentsply Intl. · Feb 24, 1998
K132825 — RETROMTA-ORTHOMTA II · Biomta · Apr 10, 2014
Submission Summary (Full Text)
{0}------------------------------------------------
Kiuoass.
## Avalon Biomed Inc.™
Carolyn Primus, President
JUN 1 1 2014
April 4, 2014
## 510(k) Summary
per 21CFR807.92
## CONTACT:
Carolyn Primus, President Avalon Biomed Inc. 7282 55th Ave E # 227 Bradenton, FL 34203 USA www.AvalonBiomed.com Phone: (941) 753-9737 or 896-9948 Fax: (941) 753-8778 or 896-9950 E-mail: info@avalonbiomed.com Web: http://www.avalonbiomed.com/
DATE PREPARED:
May 23, 2014
MTA2.1 MATERIAL
TRADE OR PROPRIETARY NAME:
CLASSIFICATION NAME:
Root Canal Filling Resin 872.3820
PREDICATE DEVICE:
K122892, MTA2 ROOT AND PULP MATERIALS
DEVICE DESCRIPTION: The MTA2.1 MATERIAL is designed and developed for dental clinicians to use in contact with vital pulp tissue and periradicular tissue, including sealing and obturation of root canals. This material
can be used for dental procedures contacting pulpal or periradicular tissue such as: Pulp capping, Cavity lining, Base material in a cavity, Pulpotomies, Root-end filling, Apexification, Perforation repair, Root resorption, and Obturation (pulpectomy) in primary or secondary teeth.
The dentist will mix the powder and water-based gel and place the mixed MTA2.1 MATERIAL into the space created by the procedure. The procedure may be part of caries treatment, root canal treatment, or periapical surgery. These procedures are performed on primary and secondary dentition and are color stable over time. MTA2.1 MATERIAL is tinted yellow, pink and gray.
INTENDED USE: The MTA2.1 MATERIAL is intended for use for dental procedures that contact pulp and periradicular tissues, as well as obturation and sealing of root canals.
TECHNOLOGICAL CHARACTERISTICS: The MTA2.1 MATERIAL is primarily hydraulic tricalcium silicate powder that is substantially similar to the MTA2 ROOT AND PULP
Premarket Notification (Special)
- MTA2.1 MATERIAL
Avalon Biomed Inc.Page 17 of 53
{1}------------------------------------------------
MATERIAL (K122892). The tricalcium silicate powders are known to set with water into a hard substance, containing calcium hydroxide dispersed among the hydrated particles of tricalcium silicates. Both materials are radiopaque.
COMPARISONS TO PREDICATE: We believe the MTA2.1 MATERIAL is substantially equivalent to the MTA2 ROOT AND PULP MATERIAL (K122892) when used for contact with pulp tissue and periapical tissue, and as a root canal sealer.
The only modification to the predicate MTA2 ROOT AND PULP MATERIAL (K122892) is a change in a radiopacity component and inorganic pigments. This modification does not affect indications for use or technology. It does not raise any new questions of safety or effectiveness.
Similarities: The predicate and the MTA2.1 MATERIAL are based on an inorganic powder composed of primarily tricalcium silicate, dicalcium silicate, and a radiopaque inorganic oxide. The predicates and the MTA2 ROOT AND PULP MATERIALS rely on water to hydrate the calcium silicate phases, and cause setting into a hard substance containing hydrated silicates and some calcium hydroxide.
Both the MTA2.1 MATERIAL and the predicate are used in root canals or on vital pulp tissue. Both the MTA2.1 MATERIAL and the predicate MTA2 ROOT AND PULP MATERIALS are used for sealing and obturation of root canals. The MTA2.1 MATERIAL and the predicate MTA2 ROOT AND PULP MATERIALS are similar because both kits contain the same gel.
The new and predicate materials had similar compositions, radiopacity, film thickness, flow and compressive strengths, when mixed at similar powder to liquid/gel ratios. Both MTA2.1 MATERIAL and the predicate material induce the precipitation of hydroxyapatite crystals in synthetic body fluid because of the tricalcium silicate powder component.
Differences: MTA2.1 material will be slightly different in color than the MTA2 ROOT AND PULP MATERIALS, which are white and gray in color. Different radiopaque and pigment powders are included in MTA2.1 than in the predicate. MTA2.1 material will be more stable to potential discoloration over time.
NON-CLINICAL PERFORMANCE: As part of design validation, MTA2.1 MATERIAL meet the ADA 57 standard for radiopacity, solubility, dimensional stability, film thickness, and flow, The ADA 57 tests were also performed to measure the working time and setting time of MTA2.1 compared to the predicates. Tests for compressive strength and leaching of arsenic and lead were performed to shown conformance to ISO 9917. Washout tests confirm the stability of the material. Discoloration tests confirm the color stability of the material.
The MTA2.1 MATERIAL was evaluated for biocompatibility with its gel in cytotoxicity test as part of design validation.
CLINICAL PERFORMANCE: No clinical tests were performed in the development of the MTA2.1 MATERIAL.
Premarket Notification (Special)
ੱ
K140955: MTA2.1 MATERIAL
Avalon Biomed Inc.Page 18 of 53
{2}------------------------------------------------
SUBSTANTIAL EQUIVALENCE: Bench testing was performed as part of design validation to demonstrate continued conformance with the requirements were achieved to conform to the FDA recognized standards ADA 57 and appropriate requirements of ISO 9917.
We believe that the performance data provided herein demonstrate that MTA2.1 ROOT AND PULP MATERIALS are substantially equivalent to the predicate MTA2 (K122892) in design, principle of performance, technology, and composition. We believe the MTA2.1 MATERIAL performs as well as or better than the predicate device.
Premarket Notification (Special)
K140955: MTA2.1 MATERIAL
Avalon Biomed Inc.Page 19 of 53
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" is arranged in a circular fashion around the bird-like figure.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 11, 2014
Avalon Biomed Incorporated Dr. Carolyn Primus President 7282 550 Avenue East #227 Bradenton, FL 34203
Re: K140955
Trade/Device Name: MTA2.1 Material Regulation Number: 21 CFR 872.3820 Regulation Name: Resin, Root Canal Filling Regulatory Class: II Product Code: KIF Dated: May 23, 2014 Received: May 30, 2014
Dear Dr. Primus:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (2 I CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 - Dr. Primus
forth in the quality systems (OS) regulation (2) CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ..
. .
Sincerely yours,
Mary S. Runner -S
· Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known): K140955
Device Name: MTA2.1 MATERIAL
Indications For Use:
The MTA2.1 MATERIAL is indicated for dental procedures that contact pulp and periradicular tissues, as well as obturation and sealing of root canals.
Prescription Use X
(Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF
NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Sheena A. Green -S
2014.06.10 14:27:48 -04'00'
Premarket Notification (Special)
K140955: MTA2.1 MATERIAL
Avalon Biomed Inc. Page 15 of 53
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.