ENDOCEM ZR

K133042 · Maruchi · KIF · May 12, 2014 · Dental

Device Facts

Record IDK133042
Device NameENDOCEM ZR
ApplicantMaruchi
Product CodeKIF · Dental
Decision DateMay 12, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3820
Device ClassClass 2
AttributesTherapeutic

Indications for Use

A root-end filling material For the repair of root canals as an apical plug during apexification For repair of root perforations during root canal therapy As a consequence of internal resorption As a pulp capping material For the filling of pulpotomy of deciduous tooth

Device Story

ENDOCEM Zr is a dental root filling material composed of natural pure cement and zirconium dioxide. Supplied as a powder to be mixed with water, it forms a putty that hardens in the oral environment. Used by dentists during endodontic procedures, including root-end filling, apexification, perforation repair, internal resorption treatment, pulp capping, and deciduous pulpotomy. The device provides a biocompatible seal for root canals. Zirconium dioxide is utilized to provide radiopacity, replacing the bismuth oxide found in the predicate device, thereby improving radiopacity while maintaining biocompatibility.

Clinical Evidence

Bench testing only. Performance evaluated per ISO 6876 (visual, packaging, setting time, solubility, radiopacity), ISO 7405 (cytotoxicity), ISO 10993-10 (delayed hypersensitivity), and ISO 10993-11 (acute systemic toxicity). All results met preset criteria.

Technological Characteristics

Composition: Natural pure cement and zirconium dioxide. Form: Powder/water mixture (putty). Radiopacity agent: Zirconium dioxide. Standards: ISO 6876, ISO 7405, ISO 10993-10, ISO 10993-11. Non-electronic, non-software device.

Indications for Use

Indicated for dental patients requiring root-end filling, apical plug during apexification, repair of root perforations, treatment of internal resorption, pulp capping, or pulpotomy of deciduous teeth.

Regulatory Classification

Identification

A root canal filling resin is a device composed of material, such as methylmethacrylate, intended for use during endodontic therapy to fill the root canal of a tooth.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ K13304a # 510(k) Summary · · This summary of 510(k)-safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92. Date: 04/09/2014 ### 1. Submission Applicant / Submitter: MARUCHI 2-208, 42-10, Taejanggongdan-gil, Wonju-si, Gangwon-do, 220-962, Republic of Korea Tel: +82-33-734-0330 Fax: +82-33-746-2804 ### 2. Submission Correspondent: Priscilla Chung LK Consulting Group USA, Inc. 2651 E. Chapman Ave Ste 110, Fullerton, CA 92833 Phone: 714-202-5789 Fax: 714-409-3357 Email: juhee.c@lkconsultinggroup.com #### 3. Device: | Proprietary Name: | ENDOCEM Zr | |------------------------------|---------------------------| | Common Name: | Root Filling Material | | Classification Name: | Root Canal Filling Resin | | Classification: | Class II, 21 CFR 872.3820 | | Classification Product Code: | KIF | #### 4. Predicate Device: ENDOCEM MTA (Mineral Trioxide Aggregate) (KI 12078) by MARUCHI #### 5. Device Description: The raw materials of the ENDOCEM Zr are Natural Pure Cement and Zirconium dioxide, and it has been showing good sealing ability and biocompatibility. It is prepared as a mixture of powder and water, and is used in a putty form which gradually hardens in the oral environment. 1 {2}------------------------------------------------ ### 6. Intended Use: - A root-end filling material - For the repair of root canals as an apical plug during apexification - For repair of root perforations during root canal therapy - As a consequence of internal resorption - As a pulp capping material - For the filling of pulpotomy of deciduous tooth ### 7. Performance Data(Non-Clinical): The following properties were tested based on the referenced standard. All the test results met the preset test criteria. - · ISO 6876 Visual, Packaging, Setting time, Solubility and Radiopacity - · ISO 7405 Cytotoxicity - · ISO 10993-10 Maximization test for delayed hypersensitivity - · ISO 10993-11 Acute systemic toxicity - · Other bench testing Capacity ### 8. Substantial Equivalence ENDOCEM Zr has similar physical and biocompatible properties, and demonstrates comparable performance specifications to ENDOCEM MTA (Predicate device). In addition, ENDOCEM Zr has a comparable delivery system to the ENDOCEM MTA. The predicate device contains Bismuth Oxide for Radiopacity; whereas, the ENDOCEM Zr contains zirconia oxide instead of bismuth oxide. With the zirconia oxide, the subject device has higher Radiopacity still maintaining biocompatibility. The results of bench and biocompatibility testing performed demonstrate that any differences in their technological characteristics do not raise any new questions as to safety and effectiveness. Therefore, it is concluded that ENDOCEM Zr is substantially equivalent in safety and effectiveness to the predicate device. #### 9. Conclusion: Based on the testing results, MARUCHI concludes that the ENDOCEM Zr is substantially equivalent in safety and effectiveness to the predicate device. {3}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. The logo is black and white. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 May 12, 2014 Maruchi c/o Ms. Priscilla Chung LK Consulting Group USA, Incorporated 2651 E. Chapman Avenue, Suite 110 Fullerton, CA 92833 Re: K133042 Trade/Device Name: ENDOCEM Zr Regulation Number: 21 CFR 872.3820 Regulation Name: Root Canal Filing Resin Regulatory Class: II Product Code: KIF Dated: April 9, 2014 Received: April 16, 2014 # Dear Ms. Chung: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ # Page 2 - Ms. Chung Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. MarySBunner-s Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page. ## 510{k} Number (if known) KI33042 . Device Name ENDOCEM Zr Indications for Use (Describe) · A root-end filling material · For the repair of root canals as an apical plug during apexification - For repair of root perforations during root canal therapy - As a consequence of internal resorption · As a pulp capping material - For the filling of pulpotomy of deciduous tooth Type of Use (Select one or both, as applicable) 2 Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) # PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signalure) Image /page/4/Picture/18 description: The image shows the text "Sheena A. Green S" on the first line and "2014.05.09 13:52:07 -04'00'" on the second line. The text appears to be a timestamp or a record of some kind. The font is simple and clear, making the text easily readable. Form FDA 3881 (9/13) PSC Publishing Servaces (191) 413-6740
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...