K102867 · Innovative Bioceramix, Inc. · KIF · Dec 3, 2010 · Dental
Device Facts
Record ID
K102867
Device Name
IROOT FS
Applicant
Innovative Bioceramix, Inc.
Product Code
KIF · Dental
Decision Date
Dec 3, 2010
Decision
SESE
Submission Type
Special
Regulation
21 CFR 872.3820
Device Class
Class 2
Attributes
Therapeutic
Intended Use
Repair of Root Perforation Repair of Root Resorption Root End Filling Apexification Pulp Capping
Device Story
iRoot FS is a ready-to-use, fast-setting, white hydraulic premixed bioceramic paste; used for permanent root canal repair (perforation, resorption), root end filling, apexification, and pulp capping. Material is insoluble, radiopaque, and aluminum-free; calcium silicate-based composition requires water to set and harden. Device is supplied in preloaded syringes with disposable tips or preloaded containers. Used by dental professionals in clinical settings. Material does not shrink during setting; demonstrates physical properties comparable to predicate bioceramic materials. Benefits include convenient application for endodontic procedures and effective sealing of root canal structures.
Clinical Evidence
No clinical data. Evidence based on bench testing, shelf-life studies, and biocompatibility testing. Bench tests confirmed comparable flowability, working/setting times, dimensional stability, and solubility to predicate iRoot BP. Biocompatibility testing demonstrated the material is non-cytotoxic, non-mutagenic, and well-tolerated by subcutaneous tissue.
Technological Characteristics
Calcium silicate-based bioceramic paste; hydraulic setting mechanism (requires water). Radiopaque, aluminum-free. Form factor: preloaded syringe or container. Delivery: disposable tips. Sterilization: not specified. No software or electronic components.
Indications for Use
Indicated for dental patients requiring root canal repair of root perforation, root resorption, root end filling, apexification, or pulp capping.
Regulatory Classification
Identification
A root canal filling resin is a device composed of material, such as methylmethacrylate, intended for use during endodontic therapy to fill the root canal of a tooth.
K130312 — IROOT FM · Innovative Bioceramix, Inc. · Sep 18, 2014
K082943 — IROOT BP, MODEL IRBP 4609 · Innovative Bioceramix, Inc. · Jan 6, 2009
K092715 — IROOT BP PLUS · Innovative Bioceramix, Inc. · Sep 28, 2009
K222992 — iRoot BP Root Repair Material BioAggregate Paste;iRoot FS Fast Set Root Repair Material ;iRoot BP Plus Root Repair Material · Innovative Bioceramix, Inc. · Nov 22, 2022
K231259 — iRoot SP Plus · Innovative Bioceramix, Inc. · May 31, 2023
Submission Summary (Full Text)
{0}------------------------------------------------
INNOVATIVE BIOCERAMIX INC.
Image /page/0/Picture/1 description: The image shows a logo with the letters "IBC" stacked vertically. The letters are bold and white, set against a black background. The logo has a slightly distressed or textured appearance, giving it a vintage or worn look.
1628 West 75th Avenue Vancouver, BC V6P 6G2 Canada Tel: 604-221-6800 Fax: 604-677-6129
### 510(k) SUMMARY
This summary of 510(k) is being submitted in accordance with 21 CFR 807.92
SUBMITTER: Innovative BioCeramix Inc. 1628 West 75th Avenue Vancouver, BC DEC - 3 2010 V6P 6G2 Canada Tel: 604-221-6800 Fax: 604-677-6129 CONTACT: Quanzu Yang October 27, 2010 SUMMARY PREPARED: iRoot FS TRADE NAME: Fast Set Root Repair Material COMMON NAME: Resin, Root Canal Filling (21 CFR 872.3820, Product Code: KIF) CLASSIFICATION NAME: (K082943) iRoot BP PREDICATE DEVICES: . (K092715) iRoot BP Plus . (K080917) . iRoot SP (K851606) D-PBS. Products No. D5652 . iRoot FS Fast Set Root Repair Material (iRoot FS) is a convenient ready-to-DEVICE use fast setting white hydraulic premixed bioceramic paste developed for DESCRIPTION: permanent root canal repair of root perforation and root resorption, and root end filling, apexification and pulp capping applications. iRoot FS is an insoluble, radiopaque and aluminum-free material based on a calcium silicate composition, which requires the presence of water to set and harden. iRoot FS does not shrink during setting and demonstrates excellent physical properties. iRoot FS is available as a preloaded syringe with disposable tips and a preloaded container. Repair of Root Perforation INTENDED USE: . Repair of Root Resorption . Root End Filling . Apexification . Pulp Capping .
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo and contact information for Innovative BioCeramix Inc. The address is listed as 1628 West 75th Avenue, Vancouver, BC, V6P 6G2 Canada. The telephone number is 604-221-6800.
# Fax: 604-677-6129
#### TECHNOLOGICAL CHARACTERISTICS:
iRoot FS is a modification of iRoot BP. iRoot FS and iRoot BP are designated for the equivalent dental applications, and have comparable chemical and physical properties, and performance specifications. The packaging for iRoot FS includes syringes, jars and unit dose cups.
Additional predicate devices include: iRoot SP and D-PBS each contains specific chemical components that are equivalent to those found in iRoot FS; providing evidence that these materials are safe and effective for medical Furthermore, iRoot FS and iRoot BP Plus have equivalent device use. packaging containers and delivery systems.
#### NON-CLINICAL TESTS PERFORMED:
iRoot FS has undergone shelf life, bench and biocompatibility testing to provide evidence that iRoot FS's chemical and physical properties are substantially equivalent to iRoot BP. Both iRoot FS and iRoot BP have comparable flowability, working times, setting times, dimensional change following setting, solubility, shelf life and biocompatibility properties.
Biocompatibility test results determined that iRoot FS is non-cytotoxic. Since iRoot FS's chemical composition is based on the principal chemical components in iRoot BP, the biocompatibility test data for iRoot BP provides biocompatibility evidence that iRoot FS is non-mutagenic, does not cause an allergenic potential after multiple uses and has a good tolerance by subcutaneous tissue.
Consequently, the shelf life, bench and biocompatibility testing, provided evidence that iRoot FS's chemical and physical properties are substantially equivalent to iRoot BP.
#### iRoot FS has the equivalent indications for use, comparable chemical CONCLUSIONS: composition, physical properties and performance specifications to iRoot BP. The additional chemical components found in iRoot FS were found to be safe and effective in iRoot SP and D-PBS. In addition, iRoot FS has comparable packaging containers and delivery system to iRoot BP Plus. Therefore, it is concluded that iRoot FS is safe, effective and substantially equivalent to the predicate devices.
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features the HHS logo, which consists of a stylized caduceus-like symbol with three horizontal lines curving downwards. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" is arranged in a circular pattern around the logo. The seal is in black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
DEC - 3 2010
Mr. Quanzu Yang President/CEO Innovative BioCeramix Incorporated 1628 West 75th Avenue Vancouver, British Columbia Canada V6P 6G2
Re: K102867
Trade/Device Name: iRoot FS Regulation Number: 21 CFR 872.3820 Regulation Name: Root Canal Filling Resin Regulatory Class: II Product Code: KIF Dated: October 27, 2010 Received: November 3, 2010
#### Dear Mr. Yang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Eederal the state the state of the state of the states of the states Register. Register
{3}------------------------------------------------
Page 2- Mr. Yang
Enclosure
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting . (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mr for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{4}------------------------------------------------
## K 102867
## INDICATIONS FOR USE
| 510(k) Number (if Known): | |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | iRoot FS |
| Indications for Use: | <ul><li>Repair of Root Perforation</li><li>Repair of Root Resorption</li><li>Root End Filling</li><li>Apexification</li><li>Pulp Capping</li></ul> |
| Prescription Use (Part 21 CFR 801 Subpart D) | <div> <div style="display:inline-block;">√</div> <div style="display:inline-block;">AND/OR</div> <div style="display:inline-block;">Over-The-Counter Use (21 CFR 801 Subpart C)</div> </div> |
|----------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|----------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Anesthesiology, General Hospital
Infection Control, Dental Devices
510(k) Number: K102867
Page 1 of 1
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.