The sponsor provided clinical evidence from an ongoing three-year investigation in Europe to demonstrate ease of use, handling characteristics, and clinical outcomes (periapical tissue healing) in a real-world setting. Feedback from practicing dentists over three years was also used to support radiopacity and clinical utility.
Ongoing, controlled three-year investigation; Controlled clinical investigation; Follow-up/Duration: Three years (data reported at one year); Study Period: Ongoing (as of 2010)
Patients undergoing root canal obturation
Not applicable for this study
Ease of use, handling characteristics, periapical tissue healing
Indications for Use
The DRFP ProSmart Root Canal Obturation System is designed for permanent sealing of root canals following established endodontic procedures by qualified healthcare professionals. The ProSmart Root Canal Obturation System is intended for Prescription Use.
Device Story
ProSmart Root Canal Obturation System provides permanent root canal sealing. System consists of ProPoints (radiopaque core with hydrophilic coating) and ProRes Sealer (epoxy-amine resin paste with active powder). Clinician applies ProRes paste to ProPoints before insertion into prepared root canal. Hydrophilic components absorb moisture, causing radial expansion into dentine tubules and irregularities to create mechanical seal. No heating or compression equipment required. Used in dental clinics by dentists. Benefits include tight seal, dimensional stability, and ease of revision without softeners. Output is physical obturation of canal, verified by clinician via X-ray.
Clinical Evidence
Clinical evaluation based on an ongoing, controlled three-year investigation in Europe. Data from one-year follow-up of treated teeth showed 33% healed, 35% improved, 25% stable periapical tissue, and 6% widening of periodontal ligament. All treated teeth were symptom-free and functional; no extractions or retreatment required. Non-clinical bench testing included ISO 6876/6877 compliance, expansion testing, resistance to tooth cracking, and exposure to sodium hypochlorite/EDTA.
Technological Characteristics
System uses hydrophilic polymer technology. ProPoints: radiopaque core, hydrophilic translucent coating. ProRes Sealer: two-component epoxy-amine resin paste with active powder. Dimensions: 4%, 6%, PT, S tapers. Connectivity: N/A. Sterilization: Clean conditions packaging. Standards: ISO 6876:2001, ANSI/ADA Specification #57, ISO 6877:2006, ADA/ANSI Specification #78, ISO 10993, ISO 7405.
Indications for Use
Indicated for permanent sealing of root canals following endodontic procedures in patients requiring root canal therapy. Contraindicated in pregnant women and nursing mothers.
Regulatory Classification
Identification
A root canal filling resin is a device composed of material, such as methylmethacrylate, intended for use during endodontic therapy to fill the root canal of a tooth.
Obturation Points Class I Gutta Percha (EKM CFR 872.3850)
Submission Summary (Full Text)
{0}------------------------------------------------
# DRFP Ltd.
!
.
# ProSmart Root Canal Obturation System
# K 100248
| <b>2.</b> | 510(k) SUMMARY of Safety and Effectiveness<br>[807.92 (c)] | |
|------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| <b>2.1</b> | Submitter [807.92 (a)(1)]<br>DRFP Ltd.<br>Villa Farm<br>Jack Haws Lane<br>BARNACK, Stamford<br>Lincs.<br>PE9 3DY<br>United Kingdom<br>Telephone 44 – 1780 – 740 574<br>Telefax 44 – 1780 – 749 168<br>e-mail info@smart-seal.co.uk<br>web www.smart-seal.co.uk | OCT 2 2 2010 |
| <b>2.2</b> | Submission Correspondent [807.92 (a)(1)]<br>Dagmar Maeser<br>Business Support International<br>Amstel 320-I<br>1017 AP AMSTERDAM<br>Netherlands<br>Telephone 31 – 20 – 428 9591<br>Cell 31 – 651 41 5839<br>Fax 31 – 20 – 201 0175<br>e-mail bsi@xs4all.nl | |
| <b>2.3</b> | Date Summary Prepared [807.92 (a)(1)]<br>October 4, 2010 | |
| <b>2.4</b> | Device Names [807.92 (a)(2)]<br>Proprietary DRFP ProSmart Root Canal<br>Obturation System<br>ProPoints (4%, 6%, PT, S) Obturation Points<br>ProRes Sealer with Active Powder | |
| | Common Root Canal Obturation Points and Sealer | |
| | Classification Names Resin, Root Canal Filling | |
| | Product Code/ CFR KIF Class II | CFR 872.3820 |
| <b>2.5</b> | Reason for Submission [807.81(3)(i)]<br>New Device | |
| <b>2.6</b> | Predicate Devices [807.92(a)(3)]<br>Paste K042769 ADSEAL Root Canal Filling<br>KIF CFR 872.3820<br>(Sponsor: Meta Biomed Company Limited)<br>Obturation Points Class I Gutta Percha<br>EKM CFR 872.3850<br>(FDA-listed by 62 manufacturers) | |
1 | Page e
.
ﺔ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤ
{1}------------------------------------------------
## K 100248
### 2.7 Device Description [807.92(a)(4)]
The ProSmart Root Canal Obturation System is using hydrophilic polymer technology. This causes the material to expand into irregularities and tubules of the root canal to assure a tight mechanical seal with the dentine. The hydrophilic properties of the system allow extraction without softeners in revision procedures.
#### 2.7.1 ProPoint Obturation Points
ProPoints are offered in the traditional 4% and 6% sizes and for use with variable taper files (such as ProTaper™ and Sendoline S5™). The radiopaque core is coated with hydrophilic, translucent material that expands radially up to 20% (6% points) while maintaining dimensional length stability.
The points are packaged in clean conditions and come individually wrapped to prevent cross-contamination. They can be cut to exact tip size and length.
No heating or compression equipment is required. Post insertion can commence immediately after root canal has been filled.
ProPoints are coated with ProRes paste prior to insertion into the root canal.
#### ProRes Paste 2.7.2
Is a two-component base:catalyst paste based on epoxy-amine resin chemistry. It is packaged in a dual-syringe for easy and consistent mixing.
The paste comes with a carefully formulated 'Active Powder'. Mixing a tiny scoop of this material into the prepared ProRes will provide the paste with hydrophilic properties that uses the moisture present in the root canal to expand into crevices and dentine tubules for a tight mechanical seal with the tooth dentine.
Image /page/1/Picture/13 description: The image shows the number 2.8 in bold font. The number is positioned at the top of the image. There is a curved line below the number.
### Statement of Intended Use [807.92(a)(5)]
The DRFP ProSmart Root Canal Obturation System is designed for permanent sealing of root canals following established endodontic procedures by qualified healthcare professionals.
The ProSmart Root Canal Obturation System is intended for Prescription Use.
### Comparison with Predicate Devices [807.92(a)(6)] 2.9
#### 2.9.1 ProPoints
Are functionally substantially equivalent to gutta percha points that have been used in endodontics for more than 100 years.
2 | P ก แ บ บ
{2}------------------------------------------------
# K 100248
### 2.9.2 ProRes Paste
Is substantially equivalent to
K 042769 ADSEAL Root Canal Filling Meta Biomed Company Limited
### 2.9.3 Hydrophilic Materials in Powder and in ProPoint Coating and ProPoint Cores
- Hydrophilic polymers in ProPoint coatings and in ProRes . powder are similar to polymers used in contact and intraocular lenses.
- ProPoint core materials are used in dental implant abutments, ● dental bridges and crown cores as well as in surgical sutures.
### 2.10. Performance Data [807.92(b)]
### 2.10.1 Non-Clinical PerformanceTests (807.92(b)(1)
### Handling & Setting Characteristics ProRes Sealer:
| Parameter | Acceptable Limit<br>(ISO 6876:2001 & ANSI/ADA<br>Specification # 57) | Results |
|----------------|----------------------------------------------------------------------|---------------------|
| Flow | Not less than 20mm | 44mm |
| Working time | Not less than 90% stated by manufacturer | 35 mins at 37°C |
| Setting time | Within range stated by manufacturer | 45 mins at 37°C |
| Film thickness | Not more than 50 $ \mu $ m | 3.3 $ \mu $ m |
| Solubility | Shall not exceed 3% | 0.0324% |
| Radiopacity | Not less than 3mm Al equivalent | 5.4mm Al equivalent |
### ProSmart System:
#### 1) Expansion on Hvdration
Propoints coated with prores expand up to 20%.
- Worst Scenario Bench Test of Resistance to Tooth Cracking 2) Due to Hydrophilic Expansion
Insertion of prores-coated propoints in heat-sterilized extracted teeth (more brittle and desiccated than patients' teeth), showed no sign of cracking or damage due to product expansion.
- Exposure to Sodium Hypochlorite 3)
The manufacturer's recommends to rinse the canal three (3) times prior to insertion of the system. If this regime is followed, the remaining value of 0.01% sodium hypochlorite will have no effect on the safety and effectiveness of the device.
#### Exposure to EDTA 4)
Undiluted 17% EDTA has shown to have no effect on the system.
#### 5) Radiopacity
Extensive testing and continuous interaction with practicing dentists after three (3) years of use in Europe demonstrate adequate radiopacity for the intended purpose. The dentist is able to
- verify that the obturation point is located in the correct . position and has the correct working length
{3}------------------------------------------------
## K 100248
- assure that it has not penetrated the apex and as a . medical record to
- assess healing
### ProPoint Obturation Points:
#### 1) ISO 6877:2006
All applicable performance testing showed the points to be in full compliance with this standard. Since the standard was developed with reference to traditional Gutta Percha points, additional device-specific testing was performed regarding:
- Expansion on Hydration .
- Bonding of Sheath to Core .
- Flexibility .
- Geometry .
- Radiopacity and Analysis of Core Materials .
- In-Vitro Dye Penetration and Sealing Tests .
- CT Scans and X-Ray Photographs
Test results demonstrate that propoints are safe and effective for the intended purpose.
#### 2) ADA/ANSI Specification # 78
DRFP propoints comply with all relevant requirements. Due to device-specific material differences when compared to Gutta Percha, propoints differ in
- Color = white core and translucent coating
- = radiopaque core and radiolucent coating Radiopacity
propoints are visible in the filled root canal when x-rayed so that the endodontist can verify the effectiveness of the obturation
### 2.10.2 Clinical Evaluation (807.92(b)(2)
Radiographic images of an on-going, controlled three-year investigation to evaluate the ease of use by operating dentists, the handling characteristics and the specific root filling techniques of the ProSmart System, demonstrate after one year:
- 33% healed ●
.
- 35% improved .
- 25% same periapical tissue .
- 6% widening of the periodontal ligament (This is an assessment of . the state of tissue surrounding the base of the tooth and not an effect caused by the ProSmart system)
All teeth treated with ProSmart Root Canal Obturation System, were
- symptom-free and in function; .
- No teeth had to be extracted .
- None had to be retreated or required a clinical intervention .
Clinical evidence after close to three (3) years of use in Europe confirms these findings.
Image /page/3/Figure/32 description: The image shows the number 4 followed by the word "Page". The number 4 is on the left side of the image, and the word "Page" is on the right side of the image. The text is in a simple, sans-serif font.
{4}------------------------------------------------
### . ProSmart Root Canal Obturation System Traditional 510(k) ...
.. . . . . . . . . . . . . . . .
K 100248
বিশ্ববিদ্যালয়ের পরিবর্তন বিশ্বকাপ বিশ্বকাপ বিশ্বকাপ বিশ্বকাপ করে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পা
・・・・・・
9
: ---------------------------------------------------------------------------------------------------------------------------------------------------------------------------
### Contraindications
DRFP Ltd.
:
् The manufacturer recommends to not use the system in pregnant women and in nursing mothers.
:
.
...
. .
# Information Bearing on Safety and Effectiveness
(807.92 (b)(3)]
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
r
: The matarials used in the DRFP ProSmart Root Canal Obturation System have a long history of safe and effective use in dental and other medical : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : devices.
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Biocompatibility testing according to ISO 10993 and ISO 7405 has shown the material to be non-toxic, non-carcinogenic and biocompatible with the maturial in There are no characteristics known that should adversely affect the safety and effectiveness of this device :
... . ........................................................................................................................................................................ : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : . Test data demonstrate that the DRFP ProSman Root Canal Obtureting System is substantially equivalent to its predicates. .
The results of design validation and non-clinical and clinical performance testing raise no new issues of safety and effectiven
8/10/10
Date
::
Jerry Watson, BDS
CEO
..............................................................................................................................................................................
: : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : :
::.:::::::::.:...... :::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" in all capital letters. The text is black and the background is white. The text is centered in the image and is the only element present.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
DRFP Limited C/O Ms. Dagmar Maeser Maeser Business Support International, V.O.F. Amstel 320-I 1017 AP Amsterdam Netherlands
OUT 2 2 2010
Re: K100248
Trade/Device Name: DRFP ProSmart Root Canal Obturation System - ProPoint 4%, 6%, PT, S Obturation Points - ProRes Root Canal Sealer ' Regulation Number: 21 CFR 872.3820 Regulation Name: Root Canal Filling Resin Regulatory Class: II Product Code: KIF Dated: October 8, 2010 Received: October 12, 2010
Dear Ms. Maeser:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2- Ms. Maeser
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Parl 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
fim
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation
Center for Devices and
· Radiological Health
Enclosure
{7}------------------------------------------------
### INDICATION-FOR-USE K 100248 510(k) Number (if known): OCT 2 2 2010 Device Name: DRFP ProSmart Root Canal Obturation System ProPoint 4%, 6%, PT, S -Obturation Points ProRes Root Canal Sealer
## INDICATION FOR USE:
The DRFP ProSmart Root Canal Obturation System is designed for permanent sealing of root canals following established endodontic procedures by qualified healthcare professionals.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Susan Rumos
(Division Sign-Off) (Division Sign-Oh)
Division Sign-Only Division of Anoothoon
Intection Control, Dental Devices
510(k) Number: _______________________________________________________________________________________________________________________________________________________________
Concurrence of CDRH, Office of Device Evaluation (ODE)
*
1