K080266 · Vericom Co., Ltd. · KIF · Feb 6, 2008 · Dental
Device Facts
Record ID
K080266
Device Name
WELL-PEX
Applicant
Vericom Co., Ltd.
Product Code
KIF · Dental
Decision Date
Feb 6, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3820
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
Well-Pex™ is a biocompatible temporary or permanent root canal sealer, for use to stimulate healing process due to the mixture of calcium hydroxide and iodoform and the induction effect of these two ingredients. Used to promote healing effects and to help prevent bacterial contamination of the canal, as the two ingredients improve the induction effect for hard tissue induction and deposition. To be used as a medicament for the treatment of infected root canals and as a permanent, low volume additive to the filling process of a treated root canal to assist in the induction and deposition of hard tissue to make the healing process more rapid and complete. For use in the treatment of infected root canals, or following pulpectorny, or for apexegenesis or apexification, and/or for the tip filling of prepared, treated root canals at the time of final filling with gutta-percha.
Device Story
Well-Pex™ is a premixed, non-setting paste containing calcium hydroxide and iodoform; supplied in a syringe with disposable tips. Used by dentists during root canal procedures to stimulate healing, induce hard tissue deposition, and prevent bacterial contamination. The material is applied directly into the root canal; it provides radiopacity and high fluidity for accessibility. It functions as a medicament for infected canals or as an additive during final filling with gutta-percha. The device is stable, requiring no mixing or separation. Benefits include rapid, complete healing and antimicrobial protection.
Clinical Evidence
No clinical data provided. Safety and performance were established through bench testing, including biocompatibility and physical property validation (flow, film thickness, radiopacity, solubility, and disintegration).
Technological Characteristics
Premixed paste containing calcium hydroxide and iodoform. Non-setting, stable formulation. Packaged in a syringe with disposable tips, plastic holder, and tip cap. Provides radiopacity and high fluidity. No electronic or software components.
Indications for Use
Indicated for patients requiring root canal treatment, including infected root canals, post-pulpectomy, apexogenesis, or apexification. Used as a temporary or permanent root canal sealer and as an additive during final gutta-percha filling to promote hard tissue induction and prevent bacterial contamination.
Regulatory Classification
Identification
A root canal filling resin is a device composed of material, such as methylmethacrylate, intended for use during endodontic therapy to fill the root canal of a tooth.
{0}------------------------------------------------
K080266
Vericom Co. Ltd. ンク
Healthy and beautiful teeth with Vericom
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92
Date: ________________________________________________________________________________________________________________________________________________________________________
1. Company making the submission:
| | Submitter |
|----------|----------------------------------------------------------------------------------------------------------------------------|
| Name | VERICOM Co., Ltd. |
| Address | #606, 5th Dongyoung Venturestel<br>199-32, Anyang 7-Dong, Manan-Gu<br>Anyang-Si, Gyeonggi-Do,<br>Republic of Korea 430-817 |
| Phone | +82 31 441-2881 |
| Fax | +82 31 441-2883 |
| Contact | Myung-Hwan Oh |
| Internet | mh-oh@hanmail.net |
FEB - 6
2. Device :
Proprietary Name - Well-Pex™ Common Name - Root canal filling Materials Classification Name - Resin, Root Canal Filling
3. Predicate Device :
Metapex, META BIOMED CO. LTD. K032603
### 4. Description :
··············································································································································································
Well-Pex™ is a temporary root canal filling material after pulpectomy, or for apexogenesis or velifeex is a temporaly root contriming material and the treatment of infected root canals. apexinciation. It is also docu as a good finance shows excellent radiopacity and prevention of microbial contamination. It has high fluidity and excellent accessibility into the root canal. Well-Pex™ is a premixed paste as a non-setting material and is very stable without any soliditication or separation. And it is packaged in a convenient syringe with disposable tips, a plastic holder and disposable tip cap.
# 606,5% Dongyoung Venturestel, 199-32, Anyang 7-dong, Manan-gu, Anyang-si, Gyeonggi-do 430-817, Korea
Image /page/0/Picture/17 description: The image shows the logo for Vericom Dental Materials. The logo consists of the word "VERICOM" in bold, sans-serif font, arched upwards. Below the word "VERICOM" are the words "Dental materials" in a smaller, serif font. The logo is black and white.
{1}------------------------------------------------
Vericom Co. Ltd.
## Healthy and beautiful teeth with Vericom
5. Indication for use :
Well-Pex™ is a biocompatible temporary or permanent root canal sealer, for use to stimulate healing process due to the mixture of calcium hydroxide and iodoform and the induction effect of these two ingredients. Used to promote healing effects and to help prevent bacterial contamination of the canal, as the two ingredients improve the induction effect for hard tissue induction and deposition. To be used as a medicament for the treatment of infected root canals and as a permanent, low volume additive to the filling process of a treated root canal to assist in the induction and deposition of hard tissue to make the healing process more rapid and complete. For use in the treatment of infected root canals, or following pulpectorny, or for apexegenesis or apexification, and/or for the tip filling of prepared, treated root canals at the time of final filing with gutta-percha.
6. Review :
Well-Pex™ has the similar technological characteristics as the predicate device; device design, appearance, main materials and indication for use.
Well-Pex™ has the similar physical properties as the predicate device; flow, film thickness, Radiopacity, slubility and disintegration.
Well-Pex™ has been subjected to extensive safety, performance, and product validations prior to release. Safety tests including biocompatibility have been performed to ensure the devices comply with the applicable International and US regulations.
- 7. Conclusions :
ln accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification Vericom Co., Ltd. concludes that Well-Pex™ is safe and effective and substantially equivalent to predicate devices as described herein.
- 8. Vericom Co., Ltd. will update and include in this summary any other information deemed seasonably necessary by the FDA.
END
# 606,56 Dongyoung Venturestel, 199-32, Anyang 7-dong, Manan-gu, Anyang-si, Gyeonggi-do 430-817, Korea
Image /page/1/Picture/14 description: The image shows a logo for a company called "VERICOM". Below the logo, the words "Dental materials" are printed in a smaller font. The logo is black and white and has a stylized design.
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" written around it. Inside the circle is a stylized image of an eagle with its wings spread.
FEB - 6 2008
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Vericom Company Limited C/O Mr. Morten S. Christensen Responsible Third Party Official Underwriters Laboratories, Incorporated 455 East Trimble Road San Jose, California 95131-1230
Re: K080266
Trade/Device Name: Well-PexTM Regulation Number: 872.3820 Regulation Name: Root Canal Filling Resin Regulatory Class: II Product Code: KIF Dated: January 22, 2008 Received: February 1, 2008
Dear Mr. Christensen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Christensen
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
CluR
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number K ( -
Device Name: Well-Pex™
Indication for use:
Well-Pex™ is a biocompatible temporary or permanent root canal sealer, for use to stimulate healing process due to the mixture of calcium hydroxide and iodoform and the induction effect of these two ingredients. Used to promote healing effects and to help prevent bacterial contamination of the canal, as the two ingredients improve the induction effect for hard tissue induction and deposition. To be used as a medicament for the treatment of infected root canals and as a permanent, low volume additive to the filling process of a treated root canal to assist in the induction and deposition of hard tissue to make the healing process more rapid and complete. For use in the treatment of infected root canals, or following pulpectorny, or for apexegenesis or apexification, and/or for the tip filling of prepared, treated root canals at the time of final filling with gutta-percha.
Prescription Use OR Over-The-Counter Use (Per 21CFR801.109)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Vericom Co., Ltd
4. Indication for use Page # 1 of 1
Suen Ringer
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: Y020666
00-MB-F0031 - Rev. 2.1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.