PANAVIA SA Cement Plus Automix; (Universal(A2), White, Translucent); Value Pack (Universal(A2), Translucent); Automix Trial (Universal(A2), White, Translucent); Mixing tip; Endo tip (SS)
K142625 · Kuraray Noritake Dental, Inc. · EMA · Jan 27, 2015 · Dental
Device Facts
Record ID
K142625
Device Name
PANAVIA SA Cement Plus Automix; (Universal(A2), White, Translucent); Value Pack (Universal(A2), Translucent); Automix Trial (Universal(A2), White, Translucent); Mixing tip; Endo tip (SS)
Applicant
Kuraray Noritake Dental, Inc.
Product Code
EMA · Dental
Decision Date
Jan 27, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3275
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The subject device is indicated for the following uses: - [1] Cementation of crowns, bridges, inlays and onlays - [2] Cementation of prosthetic restorations on implant abutments and frames - [3] Cementation of adhesion bridges and splints - [4] Cementation of posts and cores - [5] Amalgam bonding
Device Story
Dual-cure, fluoride-releasing, radiopaque self-adhesive resin cement; used for permanent cementation of ceramic, composite resin, and metal dental restorations. Supplied in automix delivery system for mixing two components. Used by dental professionals in clinical settings. Applied to tooth structure and restoration surfaces to provide adhesion. Benefits include secure fixation of prosthetic restorations and structural support for posts/cores. Performance verified via shear bond strength testing to various substrates (dentin, enamel, ceramic, metal, titanium, composite, amalgam) and ISO 4049 compliance.
Clinical Evidence
Bench testing only. Evaluated per ISO 4049:2009 for physical/mechanical properties (film thickness, working/setting time, flexural strength, water sorption/solubility, color stability, radio-opacity) and ISO/TS 11405:2003 for shear bond strength to various substrates. Biocompatibility assessed per ISO 10993-1:2009 and ISO 7405:2008. All tests showed compliance and equivalence to predicate devices with no statistically significant differences in bond strength (P>0.05).
Technological Characteristics
Dual-cure (light/self-cure) resin cement. Materials: fluoride-releasing resin matrix. Form factor: automix delivery system. Standards: ISO 4049:2009 (polymer-based restorative materials), ISO 10993-1:2009, ISO 7405:2008, ISO/TS 11405:2003. Radiopaque. No software or electronic components.
Indications for Use
Indicated for cementation of crowns, bridges, inlays, onlays, prosthetic restorations on implant abutments/frames, adhesion bridges, splints, posts, cores, and for amalgam bonding in dental patients.
Regulatory Classification
Identification
Zinc oxide-eugenol is a device composed of zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp. Dental cement other than zinc oxide-eugenol is a device composed of various materials other than zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol representing the human form, depicted as three overlapping profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 27, 2015
Kuraray Noritake Dental Inc. C/O Mr. Goro Asanuma Kuraray America, Inc. 33 Maiden Lane, 6th Floor New York, New York 10038
Re: K142625
Trade/Device Name: PANAVIA SA Cement Plus Automix Regulation Number: 21 CFR 872.3275 Regulation Name: Dental cement Regulatory Class: II Product Code: EMA Dated: December 22, 2014 Received: December 23, 2014
Dear Mr. Asanuma:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina
Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known): K142625
Device Name: PANAVIA SA Cement Plus Automix
Indications for Use:
[1] Cementation of crowns, bridges, inlays and onlays
[2] Cementation of prosthetic restorations on implant abutments and frames
[3] Cementation of adhesion bridges and splints
[4] Cementation of posts and cores
[5] Amalgam bonding
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
January 27th, 2015 Date:
## 510(k) Summary
### 3-1. 510(k) owner (submitter)
| | 1) Name | Kuraray Noritake Dental Inc. | | |
|----|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|--|
| | 2) Address | 1621 Sakazu, Kurashiki, Okayama 710-0801, Japan | | |
| | 3) Contact person | Michio Takigawa<br>Quality Assurance Department | | |
| | 4) Contact person in US | Goro Asanuma<br>KURARAY AMERICA, INC.<br>33 Maiden Lane, 6th Floor, New York, NY 10038<br>Tel: (212)-986-2230 (Ext. 115) or (800)-879-1676<br>Fax: (212)-867-3543 | | |
| | 3-2. Name of Device | | | |
| | 1) Trade / Proprietary name | PANAVIA SA Cement Plus Automix | | |
| | 2) Classification name | Dental cement<br>(21 CFR section 872.3275. Product code: EMA) | | |
| | 3) Common name | Dental adhesive resin cement | | |
| | 3-3. Predicate devices | | | |
| 1) | PANAVIA SA CEMENT Automix | 510(k) Number:<br>Classification:<br>Product Code:<br>21 CFR Section:<br>Applicant: | K120379<br>Dental cement<br>EMA<br>872.3275<br>Kuraray Noritake Dental Inc. | |
| 2) | PANAVIA F 2.0 | 510(k) Number:<br>Classification:<br>Product Code:<br>21 CFR Section:<br>Applicant: | K032455<br>Dental cement<br>EMA<br>872.3275<br>Kuraray Noritake Dental Inc. | |
| 3) | CLEARFIL AP-X | 510(k) Number:<br>Classification:<br>Product Code:<br>21 CFR Section:<br>Applicant: | K012740<br>Tooth shade resin material<br>EBF<br>872.3690<br>Kuraray Noritake Dental Inc. | |
| 4) | CLEARFIL TRI-S BOND | 510(k) Number:<br>Classification:<br>Product Code:<br>21 CFR Section:<br>Applicant: | K042913<br>Resin tooth bonding agent<br>KLE<br>872.3200<br>Kuraray Noritake Dental Inc. | |
| 5) | CLEARFIL DC CORE PLUS | 510(k) Number:<br>Classification:<br>Product Code:<br>21 CFR Section:<br>Applicant: | K111982<br>Tooth shade resin material<br>EBF<br>872.3690<br>Kuraray Noritake Dental Inc. | |
{4}------------------------------------------------
#### 3-4. Device Description
The subject device is a dual-cure (light- and/or self-cure), fluoride releasing, radiopaque self-adhesive resin cement for ceramic (porcelain, lithium disilicate, zirconia, etc.), composite resin, and metal restorations. It is supplied in an automix delivery system which can mix equal amounts of two components.
This is the new registration application for the subject device and there have not been any prior submissions regarding the subject device.
### 3-5. Statement of Intended Use
The subject device is indicated for the following uses:
- [1] Cementation of crowns, bridges, inlays and onlays
- [2] Cementation of prosthetic restorations on implant abutments and frames
- [3] Cementation of adhesion bridges and splints
- [4] Cementation of posts and cores
- [5] Amalgam bonding
### 3-6. Substantial Equivalence Discussion
- 1) Intended uses
Intended uses of the subject device and predicate devices, PANAVIA SA CEMENT Automix which is self-adhesive resin cement and PANAVIA F2.0 which is dental resin cement, are as listed on the following table.
| | Trade name | Intended use |
|----------------------|-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Subject<br>device | PANAVIA SA<br>Cement Plus Automix | [1] Cementation of crowns, bridges, inlays and onlays<br>[2] Cementation of prosthetic restorations on implant abutments<br>and frames<br>[3] Cementation of adhesion bridges and splints<br>[4] Cementation of posts and cores<br>[5] Amalgam bonding |
| Predicate<br>devices | PANAVIA SA<br>CEMENT Automix | [1] Cementation of crowns, bridges, inlays and onlays made of<br>porcelain, ceramic, composite resin or metal<br>[2] Cementation of porcelain, ceramic, composite resin or metal<br>restorations on implant abutments<br>[3] Cementation of metal cores, resin cores, metal posts or glass<br>fiber posts |
| | PANAVIA F2.0 | [1] Cementation of metal crowns and bridges, inlays and onlays<br>[2] Cementation of porcelain crowns, inlays, onlays and veneers<br>[3] Cementation of composite resin crowns, inlays, and onlays<br>[4] Cementation of adhesion bridges<br>[5] Cementation of endodontic cores and prefabricated posts<br>[6] Amalgam bonding |
The intended use of the subject device was written up based on those of the predicate devices. The implant abutments and frames are made of the same materials. Their required adhesive performance is same. The adhesion bridges and splints are also made of the same materials. Their required adhesive performance is same, too.
Therefore, the intended use of the subject device is substantially equivalent to those of the predicate devices.
2) Chemical ingredients/ Safety
Except for 5 chemical ingredients all ingredients in the subject device have been used in the predicate devices. Regarding the predicate devices, there have not been any reported problems or recalls according to the post market adverse event reporting requirements in the US.
The subject device contains new 5 ingredients. Therefore, we evaluated the subject device referring to ISO 10993 series and ISO 7405. As a result of the tests, it was concluded that the subject device is substantially equivalent in biological safety to the predicate device.
3) Technological characteristics/ Effectiveness and Performance
Physical and mechanical properties of the subject device were evaluated according to ISO 4049: 2009 (Dentistry - Polymer-based restorative and materials).
{5}------------------------------------------------
| Section | PANAVIA SA Cement Plus<br>Automix (Subject device) | | PANAVIA SA CEMENT<br>Automix (Predicate device) | |
|-------------------------------------------------------------|----------------------------------------------------|-----------------------|-------------------------------------------------|-----------------------|
| | Shade type:<br>Universal (A2) | Shade type :<br>White | Shade type:<br>Universal (A2) | Shade type :<br>White |
| 5.2.2 Film thickness, luting materials | COMPLIES | COMPLIES | COMPLIES | COMPLIES |
| 5.2.4 Working time, Class 1 and Class<br>3 luting materials | COMPLIES | COMPLIES | COMPLIES | COMPLIES |
| 5.2.6 Setting time, Class 3 materials | COMPLIES | COMPLIES | COMPLIES | COMPLIES |
| 5.2.9 Flexural strength | COMPLIES | COMPLIES | COMPLIES | COMPLIES |
| 5.2.10 Water sorption<br>and solubility | COMPLIES | COMPLIES | COMPLIES | COMPLIES |
| 5.4 Color stability after irradiation and<br>water sorption | COMPLIES | COMPLIES | COMPLIES | COMPLIES |
| 5.5 Radio-opacity | COMPLIES | COMPLIES | COMPLIES | COMPLIES |
The results of comparative study performed according to ISO 4049: 2009 were indicated below
The results indicate that the subject device and the predicate device comply with the requirements of ISO 4049: 2009. From the above, it can be said that comparative study of the subject device is substantially equivalent to that of the predicate device.
The shear bond strengths for the substrates which are matched to each indication referring to ISO/TS 11405: 2003 were indicated below.
| Adherend surface<br>(Material composition) | Criteria | Shear bond strength | | |
|--------------------------------------------|----------------------|---------------------|------------------|-------------------|
| | | Subject device | Predicate device | |
| Tooth dentin | In-house<br>standard | COMPLIES | COMPLIES | |
| Tooth enamel | | COMPLIES | COMPLIES | |
| Ceramic | | COMPLIES | COMPLIES | |
| Metal | | COMPLIES | COMPLIES | PANAVIA SA Cement |
| Titan | | COMPLIES | COMPLIES | |
| Composite resin | | COMPLIES | COMPLIES | |
| Amalgam | | COMPLIES | COMPLIES | PANAVIA F 2.0 |
The shear bond strengths to all substrates of the subject device were not statistically (P>0.05) different from those of the predicate devices.
Therefore, it was concluded that the bonding performance to all substrates of the subject device was equivalent to those of the predicate devices.
Released fluorine ion test was performed to validate the substantial equivalence of the subject device with the predicate device.
It was concluded that amount of released fluorine ion from the subject device is substantially equivalent to those from the predicate device.
#### 3-7. Biocompatibility
The subject device is classified according to ISO 7405:2008 and ISO 10993-1:2009 as an external communicating device that will have permanent (> 30 days) contact with tissues. We performed the biological safety tests of the subject device. As a result of the tests, it was concluded that the subject device is substantially equivalent in biological safety to the predicate device.
#### 3-8. Conclusion
The comparison for intended uses, chemical ingredients/ safety and performance data shows that the subject device is substantially equivalent to the predicate devices.
This submission information including the nonclinical testing provided supports that the subject device is as safe and as effective as the predicate devices.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.