← Product Code [EMA](/submissions/DE/subpart-d%E2%80%94prosthetic-devices/EMA) · K123595

# NEXUS RMGI (K123595)

_Kerr Corporation · EMA · Feb 19, 2013 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/EMA/K123595

## Device Facts

- **Applicant:** Kerr Corporation
- **Product Code:** [EMA](/submissions/DE/subpart-d%E2%80%94prosthetic-devices/EMA.md)
- **Decision Date:** Feb 19, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3275
- **Device Class:** Class 2
- **Review Panel:** Dental

## Indications for Use

Nexus RMGI is intended for use as a permanent dental cement. It is indicated for the following indirect restorations: Cementation of metal-based inlays, onlays, crowns and bridges. Cementation of resin inlays, onlays, crowns and bridges. Cementation of all ceramic inlays. Cementation of high strength (zirconia based) all ceramic crowns and bridges. Cementation of metal, ceramic and fiber posts.

## Device Story

Nexus RMGI is a radiopaque, resin-modified glass ionomer luting cement; supplied in a paste/paste/paste formulation via dual-barrel syringes with automix tips. Used by dental professionals in clinical settings for permanent cementation of indirect restorations (crowns, bridges, inlays, onlays, posts). Features optional tack light-curing for excess cement removal. Provides physical bonding and sealing of restorations to tooth structure. Benefits include improved retention and structural integrity of dental prosthetics.

## Clinical Evidence

No clinical data. Substantial equivalence supported by biocompatibility testing (ISO 10993 MEM Elution, E. coli Reverse Mutation, Intramuscular Implantation, Kligman Maximization, Oral Irritation) and comparative bench testing against the predicate device. Bench metrics included working/set times, flexural/compressive strength, filler loading, translucency, film thickness, radiopacity, solubility, fluoride release, and shear bond strength to various substrates.

## Technological Characteristics

Resin-modified glass ionomer luting cement. Paste/paste/paste formulation dispensed via dual-barrel syringe. Radiopaque. Tested per ISO 9917-2:2010. Characteristics: 36.7 MPa flexural strength, 154 MPa compressive strength, 60.3 wt% filler loading, 9 μm film thickness, 217% Al radiopacity. Biocompatible per ISO 10993 standards.

## Regulatory Identification

Zinc oxide-eugenol is a device composed of zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp. Dental cement other than zinc oxide-eugenol is a device composed of various materials other than zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp.

## Predicate Devices

- RelyX Luting Plus Automix ([K111185](/device/K111185.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

1 < 123595

Image /page/0/Picture/1 description: The image shows a logo for Sybron Dental Specialties. The logo features the letters 'sds' in a stylized font, with a swooping line above the letters. Below the letters, the words 'SYBRON DENTAL SPECIALTIES' are written in a smaller, sans-serif font. The logo is in black and white.

FEB 1 9 2013

510(k) Summary

# Submitter:

Sybron Dental Specialties, Inc. 1717 West Collins Avenue Orange, California 92867 (714) 516-7484 - Phone (714) 516-7472 - Facsimile Colleen Boswell - Contact Person

November 2012 Date Summary Prepared:

- Trade Name Nexus RMG! ●
- Common Name -- Luting Cement ◆
- Classification Name -- Dental Cement, per 21 CFR 872.3275(b) .
- Product Codes EMA ●

Devices for Which Substantial Equivalence is Claimed:

- RelyX Luting Plus Automix, 3M ESPE, K111185 .
## Device Description

Nexus™ RMGI is a radiopaque resin-modified glass ionomer luting cement offered in a paste/paste/paste formulation that provides an option for tack light-curing of excess cement. This paster formalation that problems of a catalyst packaged in dual-barrel syringes. The base and catalyst are dispensed from the dual-barrel syringe through automix tips.

## Indications for Use

Nexus RMG/ is intended for use as a permanent dental cement. It is indicated for the following indirect restorations:

- Cementation of metal-based inlays, onlays, crowns and bridges 1.
- 2. Cementation of resin inlays, onlays, crowns and bridges
- Cementation of all ceramic inlays ന്
- 4. Cementation of high strength (zirconia based) all ceramic crowns and bridges
- 5. Cementation of metal, ceramic and fiber posts

, "

{1}------------------------------------------------

# Summary of Technological Characteristics

| Descriptive Information                        | Nexus RMGI                                                                                                                                                                                                                                                                                                                                                                                                  | RelyX Luting Plus Automix (K111185)                                                                                                                                                                                                                                                                                                                                                                                                                    |
|------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company                                        | Kerr Corporation                                                                                                                                                                                                                                                                                                                                                                                            | 3M ESPE                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Indication for Use                             | Nexus RMGI is intended for use as a permanent dental cement. It is indicated for the following indirect restorations:<br>Cementation of metal-based inlays, onlays, crowns and bridges Cementation of resin inlays, onlays, crowns and bridges Cementation of all ceramic inlays Cementation of high strength (zirconia based) all ceramic crowns and bridges Cementation of metal, ceramic and fiber posts | This device is intended for use as a dental cement. RelyXTM Luting Plus Automix (LEXUS-2) is indicated for Luting:<br>Luting porcelain fused to metal crown and bridges to tooth structure, amalgam, composite or glass ionomer core build ups Luting mental inlays, onlays or crowns Luting pre-fabricated and cast post cementation Luting orthodontic appliances Luting crowns made with all-alumina or all zirconia cores such as Procera AllCeram |
| Gel Time (or Working Time) @ 22.5 ± 1.0°C      | 1'55"                                                                                                                                                                                                                                                                                                                                                                                                       | 3'20"                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Set Time @ 22.5 ± 1.0°C                        | 3'15"                                                                                                                                                                                                                                                                                                                                                                                                       | 6'45"                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Set Time @ 37.0 ± 1.0°C                        | 2'20"                                                                                                                                                                                                                                                                                                                                                                                                       | 4'45"                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Flexural Strength (MPa) as per ISO 9917-2:2010 | 36.7                                                                                                                                                                                                                                                                                                                                                                                                        | 20.7                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Compressive Strength (Mpa)                     | 154                                                                                                                                                                                                                                                                                                                                                                                                         | 103                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Filler Loading, wt.%                           | 60.3                                                                                                                                                                                                                                                                                                                                                                                                        | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Translucency, %                                | 39.1                                                                                                                                                                                                                                                                                                                                                                                                        | 16.1                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Film Thickness, μm                             | 9                                                                                                                                                                                                                                                                                                                                                                                                           | 16                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Descriptive Information                        | Nexus RMGI                                                                                                                                                                                                                                                                                                                                                                                                  | RelyX Luting Plus Automix (K111185)                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Consistency of mixed paste, cm                 | 2.9                                                                                                                                                                                                                                                                                                                                                                                                         | 3.2                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Radiopacity, % Al                              | 217                                                                                                                                                                                                                                                                                                                                                                                                         | 150                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Solubility (7 days), %                         | 0.28                                                                                                                                                                                                                                                                                                                                                                                                        | 0.94                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Cumulative Fluoride Release, ppm               |                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| 1 week                                         | 25.7                                                                                                                                                                                                                                                                                                                                                                                                        | 41                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| 4 week                                         | 48.1                                                                                                                                                                                                                                                                                                                                                                                                        | 81.9                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Shear Bond Strength to various substrates      |                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Dentin, MPa                                    | 20.2                                                                                                                                                                                                                                                                                                                                                                                                        | 15.6                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Enamel (Bovine), MPa                           | 16.2                                                                                                                                                                                                                                                                                                                                                                                                        | 10.5                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Rexillium III, MPa                             | 26.3                                                                                                                                                                                                                                                                                                                                                                                                        | 12.0                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Gold Alloy, MPa                                | 16.8                                                                                                                                                                                                                                                                                                                                                                                                        | 11.8                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Zirconia (Lava), MPa                           | 17.5                                                                                                                                                                                                                                                                                                                                                                                                        | 13.8                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Porcelain (Vitablocs) with HF etch, MPa        | 23                                                                                                                                                                                                                                                                                                                                                                                                          | 14.3                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Composite (Premise Indirect), MPa              | 17.6                                                                                                                                                                                                                                                                                                                                                                                                        | 7.3                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Lithium Disilicate , MPa                       | 17.7                                                                                                                                                                                                                                                                                                                                                                                                        | 14.9                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Alumina, MPa                                   | 14.9                                                                                                                                                                                                                                                                                                                                                                                                        | 13.1                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Titanium, MPa                                  | 25.2                                                                                                                                                                                                                                                                                                                                                                                                        | 13.4                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Post, lbs                                      | 53.9                                                                                                                                                                                                                                                                                                                                                                                                        | 34.8                                                                                                                                                                                                                                                                                                                                                                                                                                                   |

-

.

{2}------------------------------------------------

## Non-Clinical Test Data

Biocompatibility studies were completed which demonstrate that the material is safe for its Blocomputiblicy outdite The following tests: ISO L929 MEM Elution Test, intended abor Nexts Arrenm and Escherichia Coli Reverse Mutation Assay, ISO Intramuscular Implantation Test, ISO Kligman Maximization Test and ISO Oral Irritation Test.

The 510(k) submission also includes data from bench testing used to evaluate performance characteristics of Nexus RMG/ as compared to the predicate device, RelyX Luting Plus Automix currently marketed by 3M ESPE. The characteristics evaluated include Work Time, Set Times, Flexural Strength, Compressive Strength, Filler Loading, Translucency, Film Thickness, Consistency of Mixed Paste, Radiopacity, Solubility, Fluoride Release and Bond Strength.

## Clinical Test Data

1

Clinical testing has not been conducted on this product.

{3}------------------------------------------------

# Conclusion

Based upon the biocompatibility tests and bench testing, the clinical performance of Nexus
 Based of the biocompany of the bit interently Based upon the blocompanisms, testo and benefield be news in the RelyX Luting Plus Automix.

.

:

1 . <

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle symbol. The logo is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

February 19, 2013

Kerr Corporation C/O Ms. Colleen Boswell Director, Regulatory Affairs Sybron Dental Specialties, Incorporated 1717 West Collins Avenue ORANGE CA 92867

Re: K123595

Trade/Device Name: Nexus RMGI Regulation Number: 21 CFR 872.3275(b) Regulation Name: Dental Cement Regulatory Class: II Product Codes: EMA Dated: November 19, 2012 Received: November 21, 2012

## Dear Ms. Boswell:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Image /page/5/Picture/7 description: The image shows the name "Kwame O. Ulmer" with the letters "FDA" in the background. The letters "FDA" are stylized with a line pattern. The name is written in a bold, sans-serif font. 

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

#### Enclosure

{6}------------------------------------------------

## Indications for Use

510(k) Number (if known):

માટિકલ ક

Device Name: Nexus RMGI

#### Indications for Use:

Nexus RMGI is intended for use as a permanent dental cement. It is indicated for the following indirect restorations:

- 1. Cementation of metal-based inlays, onlays, crowns and bridges
- 2. Cementation of resin inlays, onlays, crowns and bridges
- 3. Cementation of all ceramic inlays
- 4. Cementation of high strength (zirconia based) all ceramic crowns and bridges
- 5. Cementation of metal, ceramic and fiber posts

Prescription Use __ X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Mary S. Runner 2013.02.19 15:38:46 ODS BA

05'00': 8 9

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number;

Kerr Corporation - RMGI - 510(k) Submission

Section 004 - Page 2 of 2

---

**Source:** [https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/EMA/K123595](https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/EMA/K123595)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
