# ENA CEM (K062355)

_Micerium S.P.A. · EMA · Dec 27, 2006 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/EMA/K062355

## Device Facts

- **Applicant:** Micerium S.P.A.
- **Product Code:** [EMA](/submissions/DE/subpart-d%E2%80%94prosthetic-devices/EMA.md)
- **Decision Date:** Dec 27, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3275
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

EnaCem is a dual curing radiopaque fluorescent luting composite, available in several dentine colours, for cementation of posts, ceramic and composite inlay, onlay veneers and crowns. It can be used also for core build-up and as a liner.

## Device Story

ENA CEM is a dual-curing, radiopaque, fluorescent luting composite used by dentists for permanent cementation of dental prostheses (posts, ceramic/composite inlays, onlays, veneers, crowns). The high-viscosity version (ENA CEM HF) also functions as a core build-up material and liner. The device is supplied in self-mixing syringes. It is applied by the clinician to the dental appliance or tooth structure during restorative procedures. The material polymerizes via a dual-cure mechanism (light and chemical) to provide structural bonding and restoration support. Benefits include improved wear resistance and versatility in restorative dentistry.

## Clinical Evidence

No clinical data provided. Substantial equivalence is based on material composition, physical characteristics, and intended use compared to existing dental composite cements. Cytotoxicity testing was performed and results are referenced in the appendix.

## Technological Characteristics

Dual-curing radiopaque fluorescent composite. Composition: resin matrix (polymerizable monomers) and glass fillers. Form factor: self-mixing syringes. Conforms to ISO 4049:2000 for polymer-based filling, restorative, and luting materials.

## Regulatory Identification

Zinc oxide-eugenol is a device composed of zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp. Dental cement other than zinc oxide-eugenol is a device composed of various materials other than zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K062355

DEC 2 7 2006

CONFIDENTIAL

### 510(k) Summary

Submitter: Mr. Eugenio Miceli, QA Manager, Micerium SpA. Via Marconi, 83, 16030 Avegno (GE), Italy. Phone: 39 0185 7885 880.

- I. Classification Name and Number: Cement, dental (EMA 872.3275).
- II. Common/Usual Name: Dental cement for post and core, direct and indirect restoration.
- Proprietary Name: ENA CEM. III.
- IV. Registration No .: K062355
- V. "ISO 4049:2000 Dentistry - Polymer - Based Filling, Restorative and Luting materials".
- VI. Description of the Device: The ENA CEM is a luting material available in a flowable consistency (Ena Cem) and also in a high viscous version (Ena Cem HF) that is available in different colours. Both can be used for cementing prosthesis appliances made in laboratory to the tooth. They can be used also for post-cementation. Ena Cem HF can also be used for core build-up and as a liner. The accessories and try-in materials contact tissue for less than 1 hour and therefore are exempt from 510(k) requirements and are described only generally.
- VII. Labels and Labeling: Draft labels of the ENA CEM and instructions for use are provided.
- VIII. Substantial Equivalence: The ENA CEM system is substantially equivalent to some other dual composite cements currently on the market used by dentist for luting prosthesis made with several types of dental materials as well as for post and core. A list of these is provided.

The composition of these dental composite cements is very similar: resin matrix made of polymerizable monomers, (glass) fillers, and all of them are dual cure materials, The packaging is similar because all products are packaged in self mixing syringes. Regarding indications, all products can be used for luting of almost all dental appliances. Physical characteristics are similar although one version of Ena Cem, Ena Cem HF, has better physical properties that allow also the core build up and a better performance regarding wear resistance.

File: ENACEM 510K new.doc

V 2.0 - 11/2006

ag. 11/ 13

# XI.

{1}------------------------------------------------

## VIII.1 Risks to Health

Potential adverse affects and complications common to composite materials include:

- Allergy to one of the ingredients -
- Incorrect / insufficient filling of the root canal ।
- Incorrect / incomplete curing of erroneous polymerisation -

Cytotoxicity tests appears in appendix V.1

- Indications for Use. EnaCem is a dual curing radiopaque fluorescent luting IX. composite, available in several dentine colours, for cementation of posts, ceramic and composite inlay, onlay veneers and crowns. It can be used also for core build-up and as a liner.
(End of Summary)

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like symbol with three curved lines forming its body and wings. The logo is surrounded by a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in capital letters.

#### Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

DEC 2 7 2006

Mr. Eugenio Miceli Quality Assurance Manager Micerium S.p.A. Via Marconi 83 16030 Avengo ITALY

Re: K062355

Trade/Device Name: ENA CEM Regulation Number: 21 CFR 872.3275 Regulation Name: Dental Cement Regulatory Class: II Product Code: EMA Dated: December 19, 2006 Received: December 22, 2006

Dear Mr. Miceli:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

#### Page 2 - Mr. Miceli

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Shule A. Murphy, M.D. for Chiu Lin, M.D.

Chiu Lin, Ph.D.

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

#### Enclosure

{4}------------------------------------------------

# Indications for Use

510(k) Number (if known): K062355

Device Name:

ENA CEM

Indications For Use:

EnaCem is a dual curing radiopaque fluorescent luting composite, available in several dentine colours, for cementation of posts, ceramic and composite inlay, onlay veneers and crowns. It can be used also for core build-up and as a liner.

Prescription Use __ メ (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Angela Blackwell for MSR

Page 1 of 1

nesthesiology, General Hospital. ontrol, Dental Devices

KD62355

---

**Source:** [https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/EMA/K062355](https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/EMA/K062355)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com