← Product Code [EMA](/submissions/DE/subpart-d%E2%80%94prosthetic-devices/EMA) · K013008

# NEXUS 2 MODIFIED (K013008)

_Sybron Dental Specialties, Inc. · EMA · Oct 2, 2001 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/EMA/K013008

## Device Facts

- **Applicant:** Sybron Dental Specialties, Inc.
- **Product Code:** [EMA](/submissions/DE/subpart-d%E2%80%94prosthetic-devices/EMA.md)
- **Decision Date:** Oct 2, 2001
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3275
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

The intended use of Nexus 2 Modified is for luting inlays, onlays, crowns, bridges, porcelain veneers and encdodontic posts.

## Device Story

Nexus 2 Modified is a dual-cured (chemical and light-cured), resin-based dental cement. It is supplied as a two-part, base/catalyst paste system. The device is used by dental professionals in a clinical setting for the permanent cementation (luting) of indirect dental restorations, including inlays, onlays, crowns, bridges, porcelain veneers, and endodontic posts. The clinician mixes the two-part paste and applies it to the restoration or tooth surface; the material then hardens via dual-curing mechanisms to secure the restoration in place. This provides a stable, durable bond between the dental structure and the prosthetic device, benefiting the patient by restoring function and aesthetics to damaged or missing teeth.

## Clinical Evidence

No clinical data provided; substantial equivalence is based on technological characteristics and intended use.

## Technological Characteristics

Dual-cured (chemical and light-cured) resin-based dental cement. Two-part, base/catalyst paste/paste system. Class II device (21 CFR 872.3275).

## Regulatory Identification

Zinc oxide-eugenol is a device composed of zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp. Dental cement other than zinc oxide-eugenol is a device composed of various materials other than zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp.

## Predicate Devices

- Nexus 2 ([K013008](/device/K013008.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# OCT - 2 2001

## Section III - 510(k) Summary of Safety and Effectiveness

#### Submitter:

Sybron Dental Specialties, Inc. 1717 W. Collins Avenue Orange, California 92867 (714) 516-7484 - Phone (714) 516-7488 - Facsimile Colleen Boswell - Contact Person

Date Summary Prepared: August 2001

#### Device Name:

- Trade Name Nexus 2 Modified .
- Common Name Luting Cement .
- Classification Name Dental Cement, per 21 CFR § 872.3275 .

### Devices for Which Substantial Equivalence is Claimed:

- Kerr Corporation, Nexus 2 o
### Device Description:

Nexus 2 Modified is a dual cured, chemical and light cured, resin-based, dental cement designed to be used for luting inlays, onlays, crowns, bridges, porcelain veneers and encdodontic posts. Nexus 2 Modified is a two-part, base/catalyst - paste/paste, system.

#### Intended Use of the Device:

The intended use of Nexus 2 Modified is for luting inlays, onlays, crowns, bridges, porcelain veneers and encdodontic posts.

#### Substantial Equivalence:

Nexus 2 Modified is substantially equivalent to other legally marketed devices in the United States. Nexus 2 Modified functions in a manner similar to and is intended for the same use as the original Nexus 2 formulation that is currently manufactured by Kerr Corporation.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure with three overlapping wings, creating a sense of movement or flight.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OCT - 2 2001

Ms. Colleen Boswell Director, Corporation Compliance Sybron Dental Specialties, Incorporated 1717 West Collins Avenue Orange, California 92867

Re: K013008

Trade/Device Name: Nexus 2 Modified Regulation Number: 872.3274 Regulation Name: Luting Cement Regulatory Class: II Product Code: EMA Dated: August 31, 2001 Received: September 6, 2001

Dear Ms. Boswell:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{2}------------------------------------------------

Page 2 - Ms. Boswell

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Timothy A. Ulatowski Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health

{3}------------------------------------------------

# KO13008

Section I

# Indications for Use Statement

Ver/ 3 - 4/24/96

Applicant: Kerr Dental Materials Center

| <strong>Labels</strong>   | <strong>Values</strong> |
|---------------------------|-------------------------|
| 510(k) Number (if known): | K0/3008                 |
| Device Name:              | Nexus 2 Modified        |

Indications For Use:

*Nexus 2 Modified* is a dual cured, resin-based, dental cement designed to be used for luting
inlays, onlays, crowns, bridges, porcelain veneers and endodontic posts.

Susan Runne
(Division Sign-Off)
Division of Dental, Infection Control,
and General Hospital Devices

| <strong>Labels</strong> | <strong>Values</strong> |
|-------------------------|-------------------------|
| 510(k) Number           | K0/3008                 |

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF
NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Per 21 CFR 801.109)
(Optional Format 1-2-96)

---

**Source:** [https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/EMA/K013008](https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/EMA/K013008)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
