← Product Code [ELW](/submissions/DE/subpart-d%E2%80%94prosthetic-devices/ELW) · K963766

# 3M PROPLUS SYSTEM (K963766)

_3M Company · ELW · Oct 31, 1996 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/ELW/K963766

## Device Facts

- **Applicant:** 3M Company
- **Product Code:** [ELW](/submissions/DE/subpart-d%E2%80%94prosthetic-devices/ELW.md)
- **Decision Date:** Oct 31, 1996
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3660
- **Device Class:** Class 2
- **Review Panel:** Dental

## Indications for Use

3M™ ProPlus System is designed for making precise crown and bridge and denture impressions.

## Device Story

Hydrophilic vinyl polysiloxane dental impression material system; consists of tray material and two wash materials of varying viscosities. Base and catalyst components extruded from two-part cartridges via mixing tip; material inserted into patient mouth; cures into rigid dental impression. Used by dental professionals for heavy body/light body impression techniques; tray material compatible with 3M Imprint monophase materials. Output is a physical dental impression used for crown, bridge, and denture fabrication.

## Clinical Evidence

Bench testing only. Comparative tests performed on strain in compression, compression set, mixed consistency, and linear dimensional change.

## Technological Characteristics

Hydrophilic vinyl polysiloxane addition cure chemistry. Two-part cartridge delivery system with mixing tip. Material cures to rigid state. No electronic or software components.

## Regulatory Identification

Impression material is a device composed of materials such as alginate or polysulfide intended to be placed on a preformed impression tray and used to reproduce the structure of a patient's teeth and gums. The device is intended to provide models for study and for production of restorative prosthetic devices, such as gold inlays and dentures.

## Special Controls

*Classification.* Class II (Special Controls).

## Predicate Devices

- 3M™ Imprint™ Vinyl Polysiloxane Impression Material Single Phase System ([K882690](/device/K882690.md))
- Caulk®/Dentsply® Reprosil® Cartridge System ([K863287](/device/K863287.md))
- Kerr® Extrude™ Polyvinylsiloxane Impression Material System ([K940379](/device/K940379.md))
- Examix™ Hydrophilic Vinyl Polysiloxane Impression Material ([K880849](/device/K880849.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

3M Dental Products Laboratory
3M Center, Building 260-2B-09
St. Paul, MN 55144-1000
612 733 3149
K963766
OCT 31 1996

# 510(K) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

Name: Amy E. Fowler
Address: 3M Dental Products Laboratory
3M Center, Building 260-2B-09
St. Paul, MN 55144-1000
Phone: (612) 736-1590
Fax: (612) 736-0990
Date: September 6, 1996
Trade Name: 3M™ ProPlus System
Common Names: Dental impression material
Classification Name: Impression material
(21 CFR 872.3660)
Predicate devices: 3M™ Imprint™ Vinyl Polysiloxane Impression Material Single Phase System (K882690)
Caulk®/Dentsply® Reprosil® Cartridge System (K863287)
Kerr® Extrude™ Polyvinylsiloxane Impression Material System (K940379)
Examix™ Hydrophilic Vinyl Polysiloxane Impression Material (K880849)

3M™ ProPlus System is composed of three hydrophilic vinyl polysiloxane dental impression materials. The system includes a tray material and two wash materials with different viscosities. The base and catalyst are extruded from two-part cartridges through a mixing tip. The pastes are inserted into the mouth and they cure into a rigid dental impression. 3M™ ProPlus System tray and wash materials are specially formulated to meet the demands of the heavy body/light body impression technique. The tray material may also be used in combination with 3M Imprint™ monophase materials.

3M™ ProPlus System, like the predicate devices, is designed for making precise crown and bridge and denture impressions.

3M™ ProPlus System and the predicate devices have similar technological characteristics as indicated by their vinyl polysiloxane addition cure chemistry. This is further validated by the comparative results of the bench tests conducted. These tests include strain in compression, compression set, mixed consistency, and linear dimensional change.

Based on the safety analysis and the bench tests, 3M™ ProPlus System is safe, effective, and performs as well or better than the predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/ELW/K963766](https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/ELW/K963766)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
