← Product Code [ELW](/submissions/DE/subpart-d%E2%80%94prosthetic-devices/ELW) · K254112

# Dia-X Sil Bite (K254112)

_DiaDent Group International · ELW · Mar 19, 2026 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/ELW/K254112

## Device Facts

- **Applicant:** DiaDent Group International
- **Product Code:** [ELW](/submissions/DE/subpart-d%E2%80%94prosthetic-devices/ELW.md)
- **Decision Date:** Mar 19, 2026
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3660
- **Device Class:** Class 2
- **Review Panel:** Dental

## Indications for Use

Indicated for bite registration in prosthodontics, orthodontics, and implantology. Used for recording the occlusal relationship to aid in the fabrication of crowns, bridges, dentures, and orthodontic appliances. Suitable of occlusal analysis and adjustment in clinical dental procedures.

## Device Story

Dia-X Sil Bite is a vinyl polysiloxane (VPS)-based impression material; used by dental professionals in clinical settings to capture precise occlusal relationships. The material is supplied in cartridge form; applied to the patient's teeth/gums; hardens upon contact with oral moisture. The resulting physical impression serves as a template for the fabrication of dental prosthetics (crowns, bridges, dentures) and orthodontic appliances. It facilitates clinical occlusal analysis and adjustment. The device is non-sterile, permanent, and prescription-only.

## Clinical Evidence

No clinical data was collected or provided. Substantial equivalence is supported by bench testing, including physical/mechanical property evaluation and biocompatibility testing (cytotoxicity, skin sensitization, and oral mucosa irritation) per ISO 10993 and ISO 4823 standards.

## Technological Characteristics

Vinyl polysiloxane (VPS) silicone-based impression material; Type B per ISO 4823:2021. Supplied in 50ml cartridges. Non-sterile. Biocompatibility tested per ISO 10993-1, 5, 10, and 23. Mechanical properties include defined setting and working times.

## Regulatory Identification

Impression material is a device composed of materials such as alginate or polysulfide intended to be placed on a preformed impression tray and used to reproduce the structure of a patient's teeth and gums. The device is intended to provide models for study and for production of restorative prosthetic devices, such as gold inlays and dentures.

## Special Controls

*Classification.* Class II (Special Controls).

## Predicate Devices

- Imprint Bite ([K882690](/device/K882690.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

March 19, 2026

DiaDent Group International
Jin An
RA team
16, Osongsaengmyeong 4-Ro, Osong-Eup, Heungdeok-Gu
Cheongju-Si, 28186
REPUBLIC OF KOREA

Re: K254112
Trade/Device Name: Dia-X Sil Bite
Regulation Number: 21 CFR 872.3660
Regulation Name: Impression Material
Regulatory Class: Class II
Product Code: ELW
Dated: December 16, 2025
Received: December 19, 2025

Dear Jin An:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K254112 - Jin An
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

{2}

K254112 - Jin An
Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

MICHAEL E. ADJODHA -S

Michael E. Adjodha, MChE, RAC, CQIA
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

Submission Number (if known)

K254112

Device Name

Dia-X Sil Bite

Indications for Use (Describe)

- Indicated for bite registration in prosthodontics, orthodontics, and implantology.
- Used for recording the occlusal relationship to aid in the fabrication of crowns, bridges, dentures, and orthodontic appliances.
- Suitable of occlusal analysis and adjustment in clinical dental procedures.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}

DiaDent Group International
Product Name: Dia-X Sil Bite

# 510(k) Summary

This summary of 510(k) substantial equivalence information is submitted in accordance with the requirements of 21 CFR 807.92.

## 1. Application Information

|  510(k) Number | K254112  |
| --- | --- |
|  Date Prepared | Mar 19, 2026  |
|  Company Name and Address | DiaDent Group International
16, Osongsaengmyeong 4-ro, Osong-eup, Heungdeok-gu,
Cheongju-si, Chungcheongbuk-do, 28161, Republic of Korea  |
|  Contact Person | Byung Chul Kim
Quality Assurance Manager
Phone: +82-43-266-2315
FAX: +82-43-235-2315
Email: diadent68@diadent.co.kr  |

## 2. Device Information

|  Device Type | Material, Impression  |
| --- | --- |
|  Regulation Description | Impression material  |
|  Review Panel | Dental  |
|  Regulation Number | 21 CFR 872.3660  |
|  Product Code | ELW  |
|  Device Class | II  |
|  Device Name | Dia-X Sil Bite  |

## 3. Primary Predicate Device

|  510(k) Number | K882690  |
| --- | --- |
|  Applicant | 3M ESPE  |
|  Device Name | Imprint Bite  |
|  Regulation Number | 21 CFR 872.3660  |
|  Product Code | ELW  |
|  Device Class | II  |

## 4. Device Description

Dia-X Sil Bite is vinyl polysiloxane (VPS) – based bite registration material designed for occlusal capture. Classified as Type B under ISO 4823 standards.

|  No. | Model Name | Composition  |
| --- | --- | --- |
|  1 | Dia-X Sil Bite A Type | Dia-X Sil Bite 50ml Cartridge 1ea  |
|  2 | Dia-X Sil Bite C Type | Dia-X Sil Bite 50ml Cartridge 2ea  |
|  3 | Dia-X Sil Bite E Type | Dia-X Sil Bite 50ml Cartridge 4ea + Mixing Tip 10ea  |
|  4 | Dia-X Sil Bite G type | Dia-X Sil Bite 50ml Cartridge 10ea  |

Page 1

{5}

DiaDent Group International

Product Name: Dia-X Sil Bite

# 5. Indications for Use

- Indicated for bite registration in prosthodontics, orthodontics, and implantology.
- Used for recording the occlusal relationship to aid in the fabrication of crowns, bridges, dentures, and orthodontic appliances.
- Suitable of occlusal analysis and adjustment in clinical dental procedures.

# 6. Technological Characteristics

The subject device, Dia-X Sil Bite has similar characteristics to the primary predicate device, Imprint Bite.

[Comparison table]

|   | Subject Device | Predicate Device | Discuss  |
| --- | --- | --- | --- |
|  510(k) Number | - | K882690 | -  |
|  Product Code | ELW | ELW | Equivalent  |
|  Device Class | II | II | Equivalent  |
|  Manufacturer | DiaDent Group International | 3M ESPE | -  |
|  Device Name | Dia-X Sil Bite | Imprint Bite | -  |
|  Intended Use | to obtain the structure of a patient's teeth and gums | to obtain the structure of a patient's teeth and gums | Equivalent  |
|  Raw Materials | - Silicone Base | -Silicone Base | * See the below table.  |
|  Principle of operation | It is an elastic occlusal recording material in a convenient cartridge form designed to harden when in contact with oral moisture. It is used to capture precise occlusal relationships | It is an elastic occlusal recording material in a convenient cartridge form designed to harden when in contact with oral moisture. It is used to capture precise occlusal relationships. | Equivalent  |
|  Performance Standard Conformance | Conformed ISO 4823 | Conformed ISO 4823 | Equivalent  |
|  Physical and Mechanical properties | - Setting time - Working time - Hardness | - Setting time - Working time - Hardness | Equivalent  |
|  Biocompatibility | Biocompatible | Biocompatible | Equivalent  |
|  Use | Prescription / Hospital | Prescription / Hospital | Equivalent  |
|  Period of use | Permanent | Permanent | Equivalent  |
|  Sterility | Non-sterile | Non-sterile | Equivalent  |
|  Shelf-life | 2 years | 2 years | Equivalent  |

{6}

DiaDent Group International
Product Name: Dia-X Sil Bite

* Difference table

|  Subject Device | Predicate Device | Discussion  |
| --- | --- | --- |
|  - Silicone Base | -Silicone Base | Both this device and the existing device contain the main ingredient, silicone base. Through the biocompatibility test results, the difference does not raise any issue of safety and effectiveness.  |

7. Non-Clinical Performance Data

The performance and biological tests were conducted on the subject device; Dia-X Sil Bite according to the following standards.

ISO 4823:2021, Dentistry — Elastomeric impression and bite registration materials
ISO 7405:2018, Dentistry — Evaluation of biocompatibility of medical devices used in dentistry
ISO 10993-1:2018, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
ISO 10993-5:2009, Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
ISO 10993-10:2021, Biological evaluation of medical devices - Part 10: Tests for skin sensitization
ISO 10993-23:2021, Biological evaluation of medical devices - Part 23: Tests for irritation

To demonstrate equivalence in the biological safety of the subject device, the following specific biocompatibility tests were performed:

In vitro Cytotoxicity Test (per ISO 10993-5:2009)
Skin Sensitization Test (per ISO 10993-10:2021)
Oral Mucosa Irritation Test (per ISO 10993-23:2021)

The test results corresponded to the requirements of standards. Therefore, the subject device is substantially equivalent to the predicate device.

8. Clinical Performance Data

No clinical data was collected or provided to support substantial equivalence between the subject and predicate device.

9. Conclusions

Based on the above information and all data provided in this submission, the subject device is substantially equivalent to the legally marketed device identified in this submission.

Page 3

---

**Source:** [https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/ELW/K254112](https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/ELW/K254112)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
