AURECAST GF

K972319 · Aurex Precious Metal Industries (Pty)Ltd · EJT · Aug 26, 1997 · Dental

Device Facts

Record IDK972319
Device NameAURECAST GF
ApplicantAurex Precious Metal Industries (Pty)Ltd
Product CodeEJT · Dental
Decision DateAug 26, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3060
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Dental casting alloy for making restoration. (NOT FOR USE WITH DENTAL PORCELAIN )

Device Story

Aurecast GF is a silver-palladium based dental casting alloy used for Type III dental restorations. The device is intended for use by dental professionals in a clinical or laboratory setting to create cast restorations. The alloy is melted and cast to form the final dental prosthetic. It serves as a restorative material to replace missing tooth structure or restore function. The device benefits patients by providing a durable, biocompatible material for dental crowns, bridges, or other cast restorations.

Clinical Evidence

Bench testing only. Physical and mechanical properties (melting point, hardness, yield strength, elongation, density) were tested and compared to the predicate device using ANSI/ADA 5 standards.

Technological Characteristics

Silver-palladium based casting alloy. Composition: 72.1% Ag, 24% Pd, 0.2% Cu, 1% Zn, 2.5% In, 0.2% Ru, 0.3% Ir. Physical properties: Melting range 1100-1115°C, Vickers hardness 130-155, density 10.6 g/cm3. Tested per ANSI/ADA 5 standards.

Indications for Use

Indicated for use as a dental casting alloy for the fabrication of dental restorations. Not for use with dental porcelain.

Regulatory Classification

Identification

A noble metal alloy is a device composed primarily of noble metals, such as gold, palladium, platinum, or silver, that is intended for use in the fabrication of cast or porcelain-fused-to-metal crown and bridge restorations.

Special Controls

*Classification.* Class II (special controls). The special control for these devices is FDA's “Class II Special Controls Guidance Document: Dental Noble Metal Alloys.” The devices are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9. See § 872.1(e) for availability of guidance information.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ AUG 26 1997 K972319 Image /page/0/Picture/2 description: The image shows a logo with two overlapping triangles and the word "AUREX" underneath. The triangles are black and appear to be slightly rounded at the corners. The word "AUREX" is in a simple, sans-serif font and is also black. P.O. Box 509 Edenvale 1610 Republic of South Africa Telephone:+27 (11) 609-8640 Telefax : +27 (11) 452-3918 ## 510 (k) Summary Trade name: AURECAST GF Dental casting alloy Common name: Gold based alloys and precious metal alloys for clinical use Classification name: EJT Class ification number: Legally marketed device: Jelenko Albacast Description of the device: Silver-palladium based casting alloy Intended use of the device: Type III restorations ## Summary of the technological characteristics Test methods applied: as in ANSI/ADA 5 Comparison of composition: - 1 - | ALLOY | | COMPOSITION (weight%) | | | | | | | |-------|------------|-----------------------|----|-----|----|-----|-----|-----| | | Name | Ag | Pd | Cu | Zn | In | Ru | Ir | | Legal | Albacast | 70 | 25 | 0 | 2 | 3 | 0 | x | | New | AurecastGF | 72.1 | 24 | 0.2 | 1 | 2.5 | 0.2 | 0.3 | x is less than 1 % Comparison of physical and mechanical properties | ALLOY | | Melting point range (°C) | | Hardness<br>(Vickers 5/30) | | Yield strength<br>(MPa) | | Elongation<br>(%) | | Density<br>(g/cm3) | |-------|------------|--------------------------|--------|----------------------------|------|-------------------------|------|-------------------|------|--------------------| | | Name | solid | liquid | soft | hard | soft | hard | soft | hard | | | Legal | Albacast | 1020 | 1100 | 158 | 162 | 300 | 350 | 14 | 10 | 10.6 | | New | AurecastGF | 1100 | 1115 | 130 | 155 | 330 | 350 | 18 | 13 | 10.6 | ## Discussion The constituents above 1% are the same. The concentration difference in the major component (silver) is less than 3 %. The base metal content is slightly less in the new alloy. ## Conclusion The mechanical properties of the two alloys are very similar, and considering the constitutions it may be assumed that also the chemical properties and biological effect should be very similar. Procious Motal Industries (Ptv) Limited (Reg.No. 88/04891/07) Image /page/0/Figure/18 description: The image shows a logo with the letters "TUV" in bold, sans-serif font. The letters are stacked on top of each other, with the "T" on top and the "UV" below. The logo is enclosed in a rounded rectangle with multiple outlines. The logo appears to be a certification mark, possibly indicating that a product or service has been tested and certified by TUV. {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized design featuring three abstract human profiles facing to the right, stacked on top of each other. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. Jerome D. Davis Managing Director Aurex Precious Metal Industries (PTY) LTD P.O. Box 509 Edenvale 1610 Republic of South Africa Re : K972319 Aurecast GF Trade Name: II Regulatory Class: Product Code: EJT Dated: June 13, 1997 Received: June 20, 1997 AUG 26 1997 Dear Mr. Davis: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, qood manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 {2}------------------------------------------------ Page 2 - Mr. Davis through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours Timothy A. Ulatowski Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radioloqical Health Enclosure {3}------------------------------------------------ 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ AURECAST GF evice Name:___________________________________________________________________________________________________________________________________________________________________ Indications For Use: Dental casting alloy for making restoration. (NOT FOR USE WITH DENTAL PORCELAIN ) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUED ON OTHER PAGE IF NEEDED Concurrence of CDRH, Office of Device Evaluation (ODE) Suoon Rumpr (Division Sign-Off and General Ho 5, 1 O(k) Numbe Prescription Use " (Per 21 CFR 801.109) OR Over-The-Counter Use_ (Optional Format 1-2-96)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...