BTI STERILE DENTAL DRILLS KIT, BTI ABUTMENTS AND CAPS

K061383 · B.T.I. Biotechnology Institute, Sl. · EJL · Aug 3, 2006 · Dental

Device Facts

Record IDK061383
Device NameBTI STERILE DENTAL DRILLS KIT, BTI ABUTMENTS AND CAPS
ApplicantB.T.I. Biotechnology Institute, Sl.
Product CodeEJL · Dental
Decision DateAug 3, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3240
Device ClassClass 1
AttributesTherapeutic

Indications for Use

BTI Dental Implant System comprising Endosseous titanium implants and prosthetic elements to be attached to the implants, as well as auxiliary elements for surgical and prosthetic procedures. The intended use of the system is the restoration of missing teeth in partially or fully edentulous patients and/or the fixation of overdentures to restore or enhance the chewing capacity of patients. Intended use for every modified product: Dental burs are intended to drill bone tissue and provisionally soft tissues during, before and after a surgical intervention of dental implants. Endosseous dental implant abutments are intended to model the gingival tissue during the process of healing after the first or second surgery and to be attached to the implant to hold single or multiple teeth restorations.

Device Story

BTI Dental Implant System consists of endosseous titanium implants, prosthetic elements, and auxiliary surgical/prosthetic tools. Dental burs act as rotary cutting devices for bone and soft tissue during implant surgery. Endosseous dental implant abutments serve as premanufactured prosthetic components connecting to implants to facilitate gingival healing and support single or multiple tooth restorations. Used in clinical dental settings by dental professionals. System restores missing teeth and enhances chewing capacity in edentulous patients.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

System components include endosseous titanium implants, dental burs, and abutments. Burs are rotary cutting devices. Abutments are premanufactured prosthetic components. Materials are consistent with previously cleared BTI Dental Implant System predicates.

Indications for Use

Indicated for partially or fully edentulous patients requiring dental implant restoration or overdenture fixation. Dental burs are used for drilling bone and soft tissue during implant surgery. Abutments are used for gingival tissue modeling during healing and supporting dental restorations.

Regulatory Classification

Identification

A dental bur is a rotary cutting device made from carbon steel or tungsten carbide intended to cut hard structures in the mouth, such as teeth or bone. It is also intended to cut hard metals, plastics, porcelains, and similar materials intended for use in the fabrication of dental devices.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K061383 TRADITIONAL 510 (K) SUBMISSION BTI Dental Implant System Modification #### 510 (K) SUMMARY 510 (K) SUMMARY. SAFETY AND EFFECTIVENESS INFORMATION BTI DENTAL IMPLANT SYSTEM MODIFICATION | SUBMITTER'S NAME. ADDRESS AND TELEPHONE NUMBER: | B.T.I. Biotechnology Institute, S.L.<br>Parque Tecnológico de Álava<br>Leonardo da Vinci, 14 B<br>Miñano (Álava)<br>01510 Spain<br>PH: 34 945 297030<br>FAX: 34 945 297031 | |-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | CONTACT PERSON | Leyre Zúñiga Hernando<br>Quality and Regulatory Affairs | SUMARY PREPARATION DATE: ESTABLISHMENT REGISTRATION No: PROPRIETARY NAME: CLASSIFICATION NAME: S A C Sterile dental drills kit/ Abutments and caps Dental bur/ Endosseous dental COMMON NAME: implant abutment Pharmacist March 2006 3004417597 Dental bur (Sec. 872.3240)/ Endosseous dental implant abutment (Sec. 872.3630) PRODUCT CODE: EJL/ NHA DEVICE CLASSIFICATION: Class I/ II #### PREDICATE DEVICE The modified BTI Dental Implant System is claimed to be substantially equivalent in material, design, and function to BTI Dental Implant System cleared by FDA under 510 (k) K022258 on Sep 11, 2003 and 510 (k) 053355 on Mar 14, 2006. {1}------------------------------------------------ K061383 p. 2 of 2 TRADITIONAL 510 (K) SUBMISSION BTI Dental Implant System Modification #### DEVICE DESCRIPTION Dental bur are rotary cutting devices intended to cut hard structures in the mouth, such as teeth or bone. They are also intended to cut hard metals, plastics, porcelains, and similar materials intended for use in the fabrication of dental devices. Endosseous dental implant abutments are premanufactured prosthetic component directly connected to the Endosseous dental implant and are intended for use as aids in prosthetic rehabilitation. #### INTENDED USE BTI Dental Implant System comprising Endosseous titanium implants and prosthetic elements to be attached to the implants, as well as auxiliary elements for surgical and prosthetic procedures. The intended use of the system is the restoration of missing teeth in partially or fully edentulous patients and/or the fixation of overdentures to restore or enhance the chewing capacity of patients. Intended use for every modified product: Dental burs are intended to drill bone tissue and provisionally soft tissues during, before and after a surgical intervention of dental implants. Endosseous dental implant abutments are intended to model the gingival tissue during the process of healing after the first or second surgery and to be attached to the implant to hold single or multiple teeth restorations. #### SUBSTANTIAL EQUIVALENCE The modified Sterile dental drills kit, Abutments and caps are considered to be substantially equivalent to the BTI Dental Implant System. #### CONCLUSION The modified BTT Dental Implant System is considered to be substantially equivalent in design, material and function to BTI Dental Implant System. {2}------------------------------------------------ Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 AUG - 3 2006 Mr. Leyre Zuniga Hernando Quality and Regulatory Affairs B.T.I. Biotechnology Institute, S.L. Parque Tecnologico de Alava Leonardo Da Vinci, 14 B Minano (Alava), Spain 01510 Re: K061383 Trade/Device Name: BTI Sterile Dental Drills Kit/ BTI Abutments and Caps Regulation Number: 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA, EJL Dated: April 4, 2006 Received: May 18, 2006 Dear Mr. Hernando: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the referenced to reastated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device merslate ectime to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general approvisions of the Act. The general controls provisions of the Act include controls provident of w registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (sec above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting (1 Wri). It may of caejoes to annot in the Code of Federal Regulations, Title 21, Parts 800 to 898. In your device ear or router announcements concerning your device in the Federal Register. {3}------------------------------------------------ ### Page 2 - Mr. Hernando Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the clectronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Charles Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ TRADITIONAL 510 (K) SUBMISSION BTI Dental Implant System Modification ## Indications for Use 510(k) Number (if known): __K061383 Device Name: BTI Sterile dental drills kit/ BTI Abutments and caps Indications for Use: Dental burs are intended to drill bone tissue and provisionally soft tissues Dental Dors' are Theendou-crosical intervention of dental implants. Endosseous dental implant abutments are intended to model the gingival Endosseous dental Implant abutinents are internet second surgery and to be tissue uuring the process of neaking and multiple teeth restorations. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) # Concurrence of CDRH, Office of Device Evaluation (ODE) Susan Penner on Sign-Off) ്ഥാ of Anesthesiology, General Hospital, - Control, Dental Devices ) Number P061353
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...