RELINE-IT MODEL RELINE-IT

K101771 · Majestic Drug Co., Inc. · EBP · Sep 20, 2010 · Dental

Device Facts

Record IDK101771
Device NameRELINE-IT MODEL RELINE-IT
ApplicantMajestic Drug Co., Inc.
Product CodeEBP · Dental
Decision DateSep 20, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3560
Device ClassClass 2
AttributesTherapeutic

Indications for Use

An over the counter temporary Denture Reliner, intended to replace a worn denture lining.

Device Story

Reline-It is a two-part OTC denture reliner kit consisting of a polymer powder and a monomer liquid. The user mixes these components to create a plastic resin paste, which is applied to the denture plate and cured in the mouth. Once set, the user trims excess material with an included spatula. The relining is reversible; the material is removed by soaking the denture in hydrogen peroxide and manual removal. The device is intended for home use by patients to temporarily replace worn denture linings. It contains no methyl methacrylate monomer, phthalate plasticizer, bisphenol A, leachable alcohol, or cadmium colors.

Clinical Evidence

No clinical data. Substantial equivalence is based on conformance to FDA guidance for OTC denture reliners and comparison of chemical composition and intended use to existing marketed devices.

Technological Characteristics

Two-part system: polymer powder and monomer liquid. Forms a plastic resin paste upon mixing. Reversible via hydrogen peroxide soak. Free of methyl methacrylate, phthalates, bisphenol A, leachable alcohol, and cadmium. Manual application and trimming.

Indications for Use

Indicated for over-the-counter temporary replacement of soft denture lining for individuals requiring temporary denture adjustment until a dentist can be seen.

Regulatory Classification

Identification

An OTC denture reliner is a device consisting of a material such as plastic resin that is intended to be applied as a permanent coating or lining on the base or tissue-contacting surface of a denture. The device is intended to replace a worn denture lining and may be available for purchase over the counter.

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ” and (2) “OTC Denture Reliners, Repair Kits, and Partially Fabricated Denture Kits.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## Majestic Drug Inc. Reline-Itm 510(k) Summary K101771 ## Date Prepared [21CFR 807.92(a)(1)] June 18, 2010 ## Submitter's Information [21CFR 807.92(a)(1)] #### Sponsor: SEP 2 0 2010 Mr. Larry Fishman President & CEO Majestic Drug Inc. 4996 Main St., Route 42 PO Box 490 S. Fallsburg, NY 12779 Telephone: 845-436-0011 The establishment registration number for Majestic Drug Inc. is 2411564 ## Contact Information: Roger S. Mastrony Quality Systems/Regulatory Consulting Services 848-1/2 Derby Avenue Orange, CT. 06477 Telephone: 203-640-5047 Fax: 203-298-0438 ## Trade Name, Common Name, Classification [21 CFR 807.92(2)[2)] #### Trade Name: Reline-It™ #### Device common, usual or Classification Names Device: Reliner, denture, over the counter Regulation Description: OTC Denture Reliner #### Class Classification: Class II Product Code: EBP, 21 CFR 872.3560 {1}------------------------------------------------ ## Majestic Drug Inc. Reline-It™ 510(k) Summary The subject device is packaged with the following items: - Container of clear polymer powder ● - Container of liquid monomer . - Mixing/Application Spatula � - . Mixing Tray - Instructions for Use . #### Intended Use [21 CFR 807.92(a)(5)] The device has the following "Indications for Use": - For over the counter temporary replacement of soft denture lining until a Dentist can be seen. . #### Technological Characteristics [21 CFR 807.92(a)(6)[ Majestic Drug Inc. maintains that the subject device is substantially equivalent to the predicate devices in that they all are similar in technology, function and intended use. #### Labels, Labeling Per section "Labels and Instruction For Use" #### Performance Data [21 CFR 807.92(b)(1)] Articles and supporting documents are included in sections "Conformance to Voluntary Standards" and "Clinical and Scientific Studies" #### Predicate Devices [21 CFR 807.92(a)(3)(1)] - DEN MAT CORP "DENTURE-TIGHT" K790589 . - LEE PHARMACEUTICALS "ACRYLINE 2"( 510(k) unknown) . - PERMA LABORATORIES PERMA SOFT II. ( 510(k) unknown) ● The subject device is substantially equivalent (materials, technology) to the above listed devices. {2}------------------------------------------------ #### Majestic Drug Inc. Reline-It™ 510(k) Summary ## Device Description [21 CFR 807.92(a)(4)] The subject device is a temporary OTC soft, clear denture reliner consisting of 2 parts (1) a monomer liquid and (2) a polymer powder which are mixed together prior to application. This results in a plastic resin paste which is then applied to the denture plate and set in the mouth to cure for a short period of time. Once setting is complete, the excess soft material is able to be trimmed by an enclosed spatula. The application is completely reversible by soaking the denture plate overnight in an over the counter solution of hydrogen peroxide and then removing it by hand. The material contains no Methyl Methacrylate Monomer, no Phthalate Plasticizer, no Bisphenol A, no leachable alcohol and no Cadmium colors. The subject device is similar to the device that was cleared under premarket submission to DEN - MAT CORP K790589 DENTURE-TIGHT, LEE PHARMACEUTICALS "ACRYLINE 2" a currently available OTC denture reliner and PERMA LABORATORIES PERMA SOFT II which is also marketed on the internet as an OTC product. #### Background: The purpose of this submission is to follow the Food and Drug Administrations' Guidance document issued Aug 18, 1998 "OTC Denture Cushions, Pads, Reliners, Repair Kits, and Partially Fabricated Denture Kits, " along with verbal technical guidance from FDA technical contacts to present a successful representation of the subject device to obtain an OTC marketing clearance by the FDA. There is some confusion on the present availability of technical information on 3 existing marketed devices; (1) Den-Mat's "Denture-Tight" K790589 obtained clearance, however technical data on formulation and a marketed sample was not available from present research efforts, and (2) Lee Pharmaceuticals "Acryline 2" is presently available on the internet, however a 5100k) could not be found for the device. (3) In addition, the FDA database did have one other device listed under product code EBP. The device name is "PERMA SOFT II, manufactured by PERMA LABORATORIES. This device is currently marketed on the internet, however a 510(k) could not be found. Given these circumstances, a recent telephone conversation with a technical contact at the FDA indicated that a successful clearance could be obtained by demonstrating conformance to the above referenced document "OTC Denture Cushions, Pads, Reliners, Repair Kits, and Partially Fabricated Denture Kits", and by clearly stating all warnings and precautions in the instructions for use and labeling. Each of these devices along with the subject device, "Reline-It™" are soft, removable over the counter denture reliners. The chemistry, methods of application and intended use are essentially identical. The information in this submission will clearly detail the formulation, safety and effectiveness data of the subject device without the benefit of having technical information of the devices listed above and all similar pre amendment devices currently on the market. {3}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized symbol featuring three abstract shapes resembling human figures or waves. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Majestic Drug Company, Inc. C/O Mr. Roger S. Mastrony Quality Systems/Regulary Consulting Services 848-1/2 Derby Avenue Orange, Connecticut 06477 SEP 2 0 2010 Re: K101771 Trade/Device Name: Reline-It Regulation Number: 21 CFR 872.3560 Regulation Name: OTC Denture Reliner Regulatory Class: II Product Code: EBP Dated: June 24, 2010 Received: June 24, 2010 Dear Mr. Mastrony: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ #### Page 2 - Mr. Mastrony Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely vours. Antions D. Mntr Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## "INDICATIONS FOR USE" Statement 510(k) Number (if known):_ # K101771 SEP 2 0 2010 Device Name: Reline-It ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ The Device has the following "Indications for Use": ## An over the counter temporary Denture Reliner, intended to replace a worn denture lining. Prescription Use (per 21 CFR 801.109) · OR Over-The Counter Use _ X (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Sina Rusone (Optional Format 1-2-96) (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices K101271 ్10(k) Number: Majestic Reline-It/RSM
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...