K220042 · Dentca, Inc. · EBI · Nov 17, 2022 · Dental
Device Facts
Record ID
K220042
Device Name
Dentca Base Premium, Dentca Base Hi-Impact
Applicant
Dentca, Inc.
Product Code
EBI · Dental
Decision Date
Nov 17, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3760
Device Class
Class 2
Indications for Use
DENTCA Base Resin is a light-curable resin indicated for the fabrication and repair of removable denture bases in dental laboratories, including full and partial dentures as well as immediate dentures, and baseplates. DENTCA Base Resin can also be used for the fabrication of try-in dentures for the evaluation before fabricating the final dentures. Fabrication of these prostheses with DENTCA Base Resin requires a digital denture file, additive printer, and curing light equipment. DENTCA Base Resin can be utilized as an aid in bonding the denture teeth onto denture base.
Device Story
DENTCA Base Resin is a light-curable, acrylate-based resin used by dental professionals in laboratories to fabricate removable denture bases, partials, and try-in appliances. The process utilizes CAD/CAM systems; digital denture files are processed via additive 3D printing to build the prosthesis in layers, followed by post-curing in a light-curing unit. The resin also serves as an aid for bonding denture teeth to the base and for repairs. By replacing traditional heat-cured or auto-polymerizing resins with an automated additive manufacturing workflow, the device enables precise, digital-based production of dental prosthetics, potentially improving fit and efficiency for the patient.
Clinical Evidence
Bench testing only. Performance evaluated against ISO 20795-1:2013 standards for Type 4 resins, including flexural modulus, flexural strength, water sorption, water solubility, maximum intensity factor, total fracture work, and residual methylmethacrylate. Biocompatibility testing conducted per ISO 7405:2018 and ISO 10993-1.
Technological Characteristics
Acrylate-based resin with photoinitiator and pigments. Type 4 light-cure resin per ISO 20795-1:2013. Form factor: liquid resin for additive 3D printing. Requires digital denture file, additive printer, and light-curing unit. Chemical bonding for teeth assembly.
Indications for Use
Indicated for dental laboratory fabrication and repair of removable denture bases, including full, partial, and immediate dentures, baseplates, and try-in dentures. Requires digital denture file, additive printer, and curing light.
Regulatory Classification
Identification
A denture relining, repairing, or rebasing resin is a device composed of materials such as methylmethacrylate, intended to reline a denture surface that contacts tissue, to repair a fractured denture, or to form a new denture base. This device is not available for over-the-counter (OTC) use.
K143033 — DENTCA DENTURE BASE · Dentca, Inc. · Jul 21, 2015
K221678 — SprintRay High Impact Denture Base · Sprintray, Inc. · Nov 10, 2022
K250946 — Denture Base Resin DT20 · Hangzhou SHINING3D Dental Technology Co., Ltd. · Mar 31, 2025
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 17, 2022
Dentca, Inc. Jason Lee Principal Scientist 357 Van Ness Way, Ste 250 Torrance, California 90501
Re: K220042
Trade/Device Name: Dentca Base Premium, Dentca Base Hi-Impact Regulation Number: 21 CFR 872.3760 Regulation Name: Denture Relining, Repairing, Or Rebasing Resin Regulatory Class: Class II Product Code: EBI Dated: October 18, 2022 Received: October 19, 2022
Dear Jason Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Michael E. Adjodha -S
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K220042
Device Name DENTCA Base Resin
Indications for Use (Describe)
DENTCA Base Resin is a light-curable resin indicated for the fabrication and repair of removable denture bases in dental laboratories, including full and partial dentures as well as immediate dentures, and baseplates. DENTCA Base Resin can also be used for the fabrication of try-in dentures for the evaluation before fabricating the final dentures.
Fabrication of these prostheses with DENTCA Base Resin requires a digital denture file, additive printer, and curing light equipment.
DENTCA Base Resin can be utilized as an aid in bonding the denture teeth onto denture base.
Type of Use (Select one or both, as applicable)
| <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|-----------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary - K220042
# Submitter
DENTCA, INC. 357 Van Ness Way, Suite 250 Torrance, CA 90501
Phone: 424-558-8726 Fax: 424-558-8738 Contact Person: Jason Lee Date Prepared: December 3, 2021 Date Revised: October 14, 2022
# Device Classification
| Proprietary Name: | DENTCA Base Resin |
|----------------------|------------------------------------------------|
| Trade Name: | DENTCA Base Premium, DENTCA Base Hi-Impac |
| Common Name: | Dental Acrylic Resin |
| Classification Name: | Denture relining, repairing, or rebasing resin |
| Regulation Number: | 21 CFR 872.3760 |
| Product Code: | EBI |
| Regulatory Class: | II |
| Review Panel: | Dental |
# Predicate Device
The following predicate is a legally marketed, post-amendment device:
| 510(k) Number: | K190043 |
|--------------------|------------------------|
| Clearance Date: | April 24, 2019 |
| Trade Name: | Halley Resin |
| Actual Trade Name: | Lucitone Digital Print |
| Manufacturer: | Dentsply Sirona |
| Regulation Number: | 21 CFR 872.3760 |
| Product Code: | EBI |
{4}------------------------------------------------
# Reference Device
The following reference is a legally marketed, post-amendment device:
| 510(k) Number: | K160244 |
|--------------------|------------------------|
| Clearance Date: | March 3, 2017 |
| Actual Trade Name: | DENTCA Denture Base II |
| Manufacturer: | DENTCA, Inc. |
| Regulation Number: | 21 CFR 872.3760 |
| Product Code: | EBI |
# Device Description
DENTCA Base Resin is a light-curable resin intended to fabricate removable dentures in a CAD/CAM system using an additive printing process. This material is used as an alternative to traditional heat cured and auto polymerizing denture base resins and is available in multiple shades. DENTCA Base Resin can also be utilized to repair the printed denture and to bond printed teeth onto denture base.
# Indications For Use
DENTCA Base Resin is a light-curable resin indicated for the fabrication and repair of removable denture bases in dental laboratories, including full and partial dentures as well as immediate dentures, and baseplates. DENTCA Base Resin can also be used for the fabrication of try-in dentures for the evaluation before fabricating the final dentures. Fabrication of these prostheses with DENTCA Base Resin requires a digital denture file, additive printer, and curing light equipment.
DENTCA Base Resin can be utilized as an aid in bonding the denture teeth onto denture base.
Both the subject and predicate are light-curable resins indicated for fabrication and repair of dental prostheses.
# Comparison of Technological Characteristics With Predicate
The following table compares technological and other characteristics of the subject, predicate, and reference devices.
{5}------------------------------------------------
## Table 5.1 Comparison of Technical Features
| | Subject Device | Predicate Device | Reference Device | Similarities and<br>Differences |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|
| Device Names | DENTCA Base Resin | Halley Resin | DENTCA Denture<br>Base II | NA |
| Regulation &<br>Product Code | 872.3760; EBI | 872.3760; EBI | 872.3760; EBI | Same Classification |
| Intended Use | Fabrication and repair of<br>removable denture bases<br>and dental appliances | Fabrication and repair of<br>removable denture bases<br>and dental appliances. | Fabrication and repair of<br>removable denture bases<br>and dental prosthesis | Same Intended Use |
| Indications For<br>Use | DENTCA Base Resin is a<br>light-curable resin indicated<br>for the fabrication and<br>repair of removable denture<br>bases in dental laboratories,<br>including full and partial<br>dentures as well as<br>immediate dentures, and<br>baseplates. DENTCA Base<br>Resin can also be used for<br>the fabrication of try-in<br>dentures for the evaluation<br>before fabricating the final<br>dentures. | Halley resin is a light-<br>cured resin indicated for<br>the fabrication of denture<br>bases in dental<br>laboratories, including full<br>and partial dentures as well<br>as implant overdentures,<br>and other dental<br>appliances. Halley resin<br>can be used as a try-in<br>material for evaluation<br>prior to fabrication of the<br>final restoration.<br>Fabrication of these<br>prostheses require a | DENTCA Denture Base<br>II is a light-curable resin<br>indicated for fabrication<br>and repair of full and<br>partial removable<br>dentures and baseplates.<br>The material is an<br>alternative to traditional<br>heat-curable and auto<br>polymerizing resins.<br>Fabrication of dental<br>prosthetics with<br>DENTCA Denture Base<br>II requires a computer-<br>aided design and<br>manufacturing | All three devices<br>are light-cure resins<br>indicated for<br>fabrication and<br>repair of dental<br>prostheses. |
| Fabrication of<br>these prostheses with<br>DENTCA Base Resin<br>requires a digital denture<br>file, additive printer, and<br>curing light equipment.<br>DENTCA Base Resin can<br>be utilized as an aid in<br>bonding the denture teeth<br>onto denture base. | computer-aided design and<br>manufacturing<br>(CAD/CAM) system using<br>an additive printer.<br>Halley resin can be utilized<br>as an aid in bonding<br>denture base to denture<br>teeth as well as repair using<br>traditional techniques. | (CAD/CAM) system<br>that includes the<br>following components:<br>digital denture base files<br>based on a digital<br>impression,<br>stereolithographic<br>additive printer, and<br>curing light equipment. | | |
{6}------------------------------------------------
## Device Description
| Resin Type | Type 4, Light-cure resin | Type 4, Light-cure resin | Type 4, Light-cure resin | Same Technology |
|------------------------------------------|--------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|-----------------|
| Chemical<br>Composition | Acrylate-based resins with<br>photoinitiator, and pigments | Acrylate-based resins with<br>photoinitiator, and<br>pigments | Acrylate-based resins<br>with photoinitiator, and<br>pigments | Same Technology |
| Fabrication<br>Method of<br>Denture Base | Automated 3D printing of<br>resin in multiple layers, and<br>post-curing in light curing<br>unit | Automated 3D printing of<br>resin in multiple layers,<br>and post-curing in light<br>curing unit | Automated 3D printing<br>of resin in multiple<br>layers, and post-curing<br>in light curing unit | Same Technology |
| Teeth Assemble | Chemical Bonding | Chemical Bonding | Chemical Bonding | Same Technology |
| Requirement | Meets the specification of<br>ISO 20795-1 for Type 4<br>material | Meets the specification of<br>ISO 20795-1 for Type 4<br>material | Meets the specification<br>of ISO 20795-1 for<br>Type 4 material | Same Standard |
| Intended User | Fabricated by Dental<br>professionals in a dental lab | Fabricated by Dental<br>professionals in a dental<br>lab | Fabricated by Dental<br>professionals in a dental<br>lab | Same User |
{7}------------------------------------------------
The above comparison shows the similarities on indications for use and technological aspects among the devices. A comparison proved that all devices fabricate the denture bases and dental prosthesis in the same manner, via automated application of a light-cure resins in an additive 3D printer.
# Performance Data
The following performance data were provided in support of the substantial equivalence determination.
## Biocompatibility Testing
The biocompatibility evaluation for DENTCA Base Resin was conducted in accordance with the standard ISO 7405:2018 Dentistrv – Evaluation of biocompatibility of medical devices used in dentistry and the FDA Guidance "Use of International Standard ISO 10993-1, Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process."
The denture base is considered tissue contacting for a period longer than 30 days (a removable prosthesis).
#### Software Verification and Validation Testing
The operation software for additive printing (3D printer) was verified and validated in accordance with the FDA guidance "FDA Reviewers and Compliance on Off-The-Shelf Software Use in Medical Devices" and "Technical Considerations for Additive Manufactured Medical Devices."
#### Performance Bench Test
Bench test for DENTCA Base Resin was conducted in accordance with ISO 20795-1:2013, as recognized by FDA. The data support a conclusion that the subject device meets the requirements per ISO 207495-1 standard as presented in Table 5.2 below.
{8}------------------------------------------------
| Representative<br>Physical Properties | ISO 20795-1 | Subject Device | | Predicate<br>Device | Reference<br>Device |
|---------------------------------------|-------------------------------------|------------------------------|------------------------------|------------------------------|------------------------------|
| | Specification<br>of Type 4<br>resin | DENTCA Base<br>Hi-Impact | Premium | Halley<br>Resin* | DENTCA<br>Denture<br>Base II |
| Flexural Modulus | ≥ 2000 MPa | Meet the<br>consensus<br>STD | Meet the<br>consensus<br>STD | Meet the<br>consensus<br>STD | Meet the<br>consensus<br>STD |
| Flexural Strength | ≥ 65 MPa | Meet the<br>consensus<br>STD | Meet the<br>consensus<br>STD | Meet the<br>consensus<br>STD | Meet the<br>consensus<br>STD |
| Water Sorption | < 32 µg/mm³ | Meet the<br>consensus<br>STD | Meet the<br>consensus<br>STD | Meet the<br>consensus<br>STD | Meet the<br>consensus<br>STD |
| Water Solubility | < 1.6 µg/mm³ | Meet the<br>consensus<br>STD | Meet the<br>consensus<br>STD | Meet the<br>consensus<br>STD | Meet the<br>consensus<br>STD |
| Maximum Intensity<br>Factor | ≥ 1.9 MPa<br>m1/2 | Meet the<br>consensus<br>STD | Not<br>claimed | Meet the<br>consensus<br>STD | Not<br>claimed |
| Total Fracture Work | ≥ 900 J/m² | Meet the<br>consensus<br>STD | Not<br>claimed | Meet the<br>consensus<br>STD | Not<br>claimed |
| Residual<br>Methylmethacrylate | < 2.2% | Not<br>Detectable | Not<br>Detectable | Meet the<br>consensus<br>STD | Not<br>Detectable |
#### Table 5.2 Comparison of Performance Bench Test
* Data cited from the predicate device K190043 510(k) Summary, Table 5.2
# Conclusions
The above comparison of the subject and predicate devices shows that both devices fabricate the denture in the same manner, via automated application of a light-curable resin in an additive 3D printer.
{9}------------------------------------------------
The subject and predicate devices have the same intended use, similar indications for use, and comprise the same technological characteristics.
Furthermore, the non-clinical data support the safety and effectiveness of the device and demonstrates that DENTCA Base Resin should perform as intended in the specified use conditions.
In conclusion, DENTCA Base Resin is substantially equivalent to the legally marketed Halley Resin, and thus warrants clearance from FDA for premarketing activities in the United States.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.