3Delta Denture Base - 3Delta Denture Bond - 3Delta Denture Fix

K212541 · Deltamed GmbH · EBI · Jan 10, 2022 · Dental

Device Facts

Record IDK212541
Device Name3Delta Denture Base - 3Delta Denture Bond - 3Delta Denture Fix
ApplicantDeltamed GmbH
Product CodeEBI · Dental
Decision DateJan 10, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3760
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Production of dentures. 3Delta Denture Base: Light curing resin for additive manufacturing of denture bases 3Delta Denture Bond: Light curing bonding agent for denture teeth and denture bases · 3Delta Denture Fix: Dual curing cementation composite for denture teeth and denture bases

Device Story

3Delta system comprises three components for denture fabrication: 3Delta Denture Base (light-curing resin for additive manufacturing), 3Delta Denture Bond (light-curing bonding agent), and 3Delta Denture Fix (dual-curing cementation composite). Used by dental professionals in clinical or laboratory settings to manufacture and assemble dentures. The base resin is processed via additive manufacturing; bonding agent and cement are applied to secure denture teeth to the base. These materials facilitate the creation of custom dental prosthetics, improving patient oral function and aesthetics.

Clinical Evidence

Bench testing only.

Technological Characteristics

Light-curing resin (Denture Base/Bond) and dual-curing cementation composite (Denture Fix). Designed for additive manufacturing and dental prosthetic assembly. No specific ASTM standards or software algorithms described.

Indications for Use

Indicated for the production of dentures, including denture bases and the bonding/cementation of denture teeth to denture bases. Intended for prescription use only.

Regulatory Classification

Identification

A denture relining, repairing, or rebasing resin is a device composed of materials such as methylmethacrylate, intended to reline a denture surface that contacts tissue, to repair a fractured denture, or to form a new denture base. This device is not available for over-the-counter (OTC) use.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services-USA. To the right, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION". January 10, 2022 DeltaMed GmbH % Gary Chuven Director of Regulatory Affairs Cosmedent Inc. 401 North Michigan Ave. Suite 2500 Chicago, Illinois 60611 Re: K212541 Trade/Device Name: 3Delta Denture Base - 3Delta Denture Bond - 3Delta Denture Fix Regulation Number: 21 CFR 872.3760 Regulation Name: Denture Relining, Repairing, Or Rebasing Resin Regulatory Class: Class II Product Code: EBI, KLE Dated: October 4, 2021 Received: October 12, 2021 Dear Gary Chuven: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) ## K212541 Device Name 3Delta Denture Base - 3Delta Denture Bond - 3Delta Denture Fix Indications for Use (Describe) - Production of dentures. - 3Delta Denture Base: Light curing resin for additive manufacturing of denture bases - 3Delta Denture Bond: Light curing bonding agent for denture teeth and denture bases - · 3Delta Denture Fix: Dual curing cementation composite for denture teeth and denture bases Type of Use (Select one or both, as applicable) | <span style="font-size: 10pt;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> | |---------------------------------------------------------------------------------------------------------------------------------| | <span style="font-size: 10pt;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> | ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...