Halley resin is a light-cured resin indicated for the fabrication of dentures bases in dental laboratories, including full and partial dentures as well as implant overdentures, and other dental appliances. Halley resin can be used as a try-in material for evaluation prior to fabrication of the final restoration. Fabrication of these prostheses require a computer-aided design and manufacturing (CAD/CAM) system using an additive printer. Halley resin can be utilized as an aid in bonding denture base to denture teeth as well as repair using traditional techniques. Halley Denture-Tooth Bonding Agent is indicated for use in enhancing the bond of denture base to denture base. Halley Sealer is a light-cured sealant that produces a smooth, glossy surface finish on the denture.
Device Story
Halley resin system consists of light-cured methacrylate resins; includes Halley resin (base material), Halley Denture-Tooth Bonding Agent, and Halley Sealer. Used by dental professionals in laboratories to fabricate/repair dentures and implant overdentures. Workflow utilizes CAD/CAM additive printing for base fabrication; bonding agent and sealer applied via traditional techniques. Sealer provides smooth, glossy surface finish. System components support modification of existing Halley resin (K190043) workflow. Benefits include improved bonding and surface finishing of 3D-printed dental prostheses.
Clinical Evidence
No clinical data included. Substantial equivalence supported by bench testing and comparison to predicate/reference devices.
Technological Characteristics
Light-cured methacrylate (mono and dimethacrylate) resins; photoinitiator-based polymerization. No pigments in bonding/sealing components. System supports 3D printing workflow. Tested per ISO 20795-1:2013 and ISO 10477:2018. Shelf life: 3 years.
Indications for Use
Indicated for fabrication of full/partial dentures, implant overdentures, and other dental appliances in dental laboratories; includes try-in material for final restoration evaluation; requires CAD/CAM additive printer; includes bonding agent for denture base-to-teeth or base-to-base bonding and sealer for surface finish.
Regulatory Classification
Identification
A denture relining, repairing, or rebasing resin is a device composed of materials such as methylmethacrylate, intended to reline a denture surface that contacts tissue, to repair a fractured denture, or to form a new denture base. This device is not available for over-the-counter (OTC) use.
K231578 — Lucitone Digital Print Denture System · Dentsply Sirona · Oct 20, 2023
K243336 — Lucitone Digital Print Denture System · Dentsply Sirona, Inc. · Dec 17, 2024
K220042 — Dentca Base Premium, Dentca Base Hi-Impact · Dentca, Inc. · Nov 17, 2022
K210298 — GR-14 Resin System · Pro3Dure Medical GmbH · May 5, 2021
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 26, 2019
Dentsply Sirona Karl Nittinger Director, Corporate Regulatory Affairs 221 West Philadelphia Street Suite 60W York, Pennsylvania 17401
Re: K191427
Trade/Device Name: Halley Resin System Regulation Number: 21 CFR 872.3760 Regulation Name: Denture relining, repairing, or rebasing resin Regulatory Class: Class II Product Code: EBI, KLE, EBD Dated: May 23, 2019 Received: May 29, 2019
Dear Karl Nittinger:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Srinivas Nandkumar, Ph.D. Acting Assistant Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K191427
Device Name Halley resin system
### Indications for Use (Describe)
Halley resin is a light-cured resin indicated for the fabrication of dental laboratories, including full and partial dentures as well as implant overdentures. Halley resin can be used as a try-in material for evaluation prior to fabrication of the final restoration of these prostheses require a computer-aided design and manufacturing (CAD/CAM) system using an additive printer.
Halley resin can be utilized as an aid in bonding denture teeth as well as repair using traditional techniques.
Halley Denture-Tooth Bonding Agent is indicated for use in enhancing the bond of denture base.
Halley Sealer is a light-cured sealant that produces a smooth, glossy surface finish on the denture.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2020
See PRA Statement below.
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Dentsply Sirona 221 West Philadelphia Street Suite 60W York, PA 17401
Image /page/3/Picture/1 description: The image shows the logo for Dentsply Sirona. The logo consists of a stylized leaf-like shape on the left, with a white curved line running through it. To the right of the shape are the words "Dentsply" on top and "Sirona" on the bottom, both in a simple, sans-serif font. The logo is presented in a muted gray color.
K191427
# SECTION 5. 510(k) SUMMARY for
Halley resin system
- 1. Submitter Information:
Dentsply Sirona 221 West Philadelphia Street Suite 60W York, PA 17401
Contact Person: Karl Nittinger Telephone Number: 717-849-4424 Fax Number: 717-849-4343
Date Prepared: May 23, 2019
- 2. Device Name:
- Proprietary Name: .
- Classification Name: .
- CFR Number:
- 21 C.F.R. 872.3760 . Device Class: 11 ●
- Product Code: EBI .
- 3. Predicate Device:
| Predicate Device Name | 510(k) | Company Name |
|-----------------------|---------|----------------------|
| Halley resin | K190043 | Dentsply Prosthetics |
Halley resin system
Resin, Denture, Relining, Repairing, Rebasing
# Reference Device:
| Reference Device Name | 510(k) | Company Name |
|---------------------------------------|---------|----------------------|
| Eclipse® Bonding Agent | K051707 | Dentsply Prosthetics |
| TEMPFX Esthetic<br>Provisional System | K061264 | Dentsply Prosthetics |
{4}------------------------------------------------
#### 4. Description of Device:
The proposed device, Halley resin system introduces two additional components (Halley Denture-Tooth Bonding Agent and Halley Sealer) to the Halley resin (K190043) workflow. The proposed Halley resin system consists of light-cured resins that support the modification of the Halley resin (K190043) workflow to include bonding and sealing steps using conventional techniques. The two additional components of the Halley resin system have previously been cleared under Eclipse® Bonding Agent. K051707 and TEMPEX Esthetic Provisional System. K061264. For comparison purposes, the reference device VLC Sealer (part of the TEMPFX Esthetic Provisional System, K061264) is being included. These devices are being included as reference devices (Eclipse® Bonding Agent, K051707 and TEMPFX Esthetic Provisional System, K061264) to support chemical composition and characteristics to support the Halley resin (K190043) workflow.
# 5. Indications for Use:
Halley resin (K190043) indications will be modified as follows (language to be added to the modified indications for use is identified in bold font):
Halley resin is a light-cured resin indicated for the fabrication of dentures bases in dental laboratories, including full and partial dentures as well as implant overdentures, and other dental appliances. Halley resin can be used as a try-in material for evaluation prior to fabrication of the final restoration. Fabrication of these require a computeraided design and manufacturing (CAD/CAM) system using an additive printer.
Halley resin can be utilized as an aid in bonding denture base to denture teeth as well as repair using traditional techniques.
Halley Denture-Tooth Bonding Agent is indicated for use in enhancing the bond of denture base to denture base.
Halley Sealer is a light-cured sealant that produces a smooth, glossy surface finish on the denture.
- 6. Substantial Equivalence:
For the purpose of substantial equivalence, the proposed device, Halley resin system is compared to the legally marketed predicate device Halley resin (K190043) manufactured by Dentsply Prosthetics. The predicate device, Halley resin received U.S. premarket clearance in April 2019 under premarket notification K190043.
The proposed Halley resin system introduces two additional components to the Halley resin (K190043) workflow. In addition to the primary predicate device. Eclipse® Bonding Agent, K051707 and TEMPFX Esthetic Provisional System, K061264 are included to support substantial equivalence based on similarities in composition and characteristics to the proposed device.
The predicate device Halley resin (K190043) has been tested in accordance with ISO 20795-1:2013 - Dentistry - Base polymers - Part 1: Denture base polymers. The proposed Halley resin system and reference devices (Eclipse® Bonding Agent, K051707 and TEMPFX Esthetic Provisional System, K061264) have been tested to support the performance characteristics contributing to the Halley resin (K190043) workflow.
Table 5.1. details similarities and differences between the proposed, predicate and reference devices
{5}------------------------------------------------
| Table 5.1 | Proposed Device<br>Halley resin system | Predicate Device<br>Halley resin | Reference Device<br>Eclipse® Bonding<br>Agent | Reference Device<br>TEMPFX Esthetic<br>Provisional System | Similarities and<br>Differences |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Element | | | | | |
| 510(k) | To be assigned | K190043 | K051707 | K061264 | N/A |
| Indications for use | Halley resin is a light-cured<br>resin indicated for the<br>fabrication of dentures bases<br>in dental laboratories,<br>including full and partial<br>dentures as well as implant<br>overdentures, and other<br>dental appliances. Halley<br>resin can be used as a try-in<br>material for evaluation prior to<br>fabrication of the final<br>restoration. Fabrication of<br>these prostheses require a<br>computer-aided design and<br>manufacturing (CAD/CAM)<br>system using an additive<br>printer.<br>Halley resin can be utilized as<br>an aid in bonding denture<br>base to denture teeth as well<br>as repair using traditional<br>techniques.<br>Halley Denture-Tooth Bonding<br>Agent is indicated for use in<br>enhancing the bond of<br>denture base to denture base.<br>Halley Sealer is a light-cured<br>sealant that produces a<br>smooth, glossy surface finish<br>on the denture. | Halley resin is a light-<br>cured resin indicated for<br>the fabrication of<br>dentures bases in dental<br>laboratories, including<br>full and partial dentures<br>as well as implant<br>overdentures, and other<br>dental appliances.<br>Halley resin can be<br>used as a try-in material<br>for evaluation prior to<br>fabrication of the final<br>restoration. Fabrication<br>of these prostheses<br>require a computer-<br>aided design and<br>manufacturing<br>(CAD/CAM) system<br>using an additive printer.<br>Halley resin can be<br>utilized as an aid in<br>bonding denture base to<br>denture teeth as well as<br>repair using traditional<br>techniques. | Eclipse® Bonding Agent<br>is indicated for use in<br>enhancing the bond of<br>acrylic teeth to acrylic<br>removable denture<br>bases. | TEMPFX Esthetic<br>Provisional System is<br>indicated for fabrication<br>of provisional dental<br>restorations. | -The proposed Halley<br>resin system is a<br>modification of the<br>workflow cleared under<br>the predicate device,<br>Halley resin (K190043).<br>Two additional<br>components are<br>introduced the Halley<br>resin (K190043)<br>workflow.<br>-Additional components<br>proposed for the Halley<br>resin system have been<br>previously cleared,<br>Eclipse® Bonding Agent<br>(K051707) and as part<br>of TEMPFX Esthetic<br>Provisional System<br>(K061264) and are<br>listed as reference<br>devices.<br>-Two additional<br>components test reports<br>support modifications to<br>the Halley resin<br>(K190043) workflow to<br>include a bonding agent<br>and sealer.<br>-Both reference devices<br>are currently utilized for<br>similar uses in different<br>workflows under<br>K051707 and K061264.<br>VLC Sealer is included<br>within the system<br>components as a light-<br>curable resin material<br>which forms a glossy<br>surface. The identical<br>VLC Sealer has been<br>re-branded as Halley<br>Sealer to support similar<br>use for Halley resin<br>system workflow. |
| Table 5.1 | Proposed Device | Predicate Device | Reference Device | Reference Device | |
| Element | Halley resin system | Halley resin | Eclipse® Bonding<br>Agent | TEMPFX Esthetic<br>Provisional System | Similarities and<br>Differences |
| Composition | Halley resin:<br>Methacrylate (mono and<br>dimethacrylate) resins<br>polymerized via<br>photoinitiators<br>Pigments are added for color<br>of the denture.<br>Halley Denture-Tooth<br>Bonding Agent:<br>Solvent vehicle containing<br>Methacrylate (mono and<br>dimethacrylate) resins<br>polymerized via<br>photoinitiators<br>Halley Sealer:<br>Methacrylate (mono and<br>dimethacrylate) resins<br>polymerized via<br>photoinitiators | Methacrylate (mono<br>and dimethacrylate)<br>resins polymerized via<br>photoinitiators in a 3D<br>printer settings.<br>Pigments are added<br>for color of the denture. | Solvent vehicle<br>containing<br>Methacrylate (mono<br>and dimethacrylate)<br>resins polymerized via<br>photoinitiators | Methacrylate (mono<br>and dimethacrylate)<br>resins polymerized via<br>photoinitiators | -Composition of the<br>proposed, predicate<br>and reference device<br>VLC Sealer (K061264)<br>are similar. All devices<br>are resins polymerized<br>via photoinitiators.<br>-No pigments are<br>added to the proposed<br>and reference devices.<br>Additional color<br>variations are not<br>required for these<br>devices.<br>-The reference device,<br>Eclipse Bonding Agent<br>(K051707) uses a<br>different solvent<br>vehicle, however, after<br>polymerization, the<br>device composition is<br>similar to the<br>proposed, predicate<br>devices.<br>-The predicate device,<br>Halley resin utilizes 3D<br>printing workflow for<br>device fabrication. The<br>proposed and<br>reference devices<br>support the device<br>fabrication workflow. |
| Application | Light-cured resin | Light-cured resin | Light-cured resin | Light-cured resin | -All devices are light-<br>cured resins. |
| Use | Fabricated by Dental<br>professionals in a dental<br>laboratory | Fabricated by Dental<br>professionals in a<br>dental laboratory | Fabricated by Dental<br>professionals in a<br>dental laboratory | Fabricated by Dental<br>professionals in a<br>dental laboratory | -All devices are<br>marketed and<br>fabricated or printed by<br>dental laboratories. |
| Table 5.1 | Proposed Device | Predicate Device | Reference Device | Reference Device | |
| Element | Halley resin system | Halley resin | Eclipse® Bonding | TEMPFX Esthetic | Similarities and |
| | | | Agent | Provisional System | Differences |
| Physical<br>properties | No ISO standard available<br>for these components; a<br>combination of ISO 20795-<br>1:2013, ISO 10477:2018 and<br>internal requirements were<br>tested to confirm<br>performance. | ISO 20795-1:2013<br>(Dentistry - Denture<br>base polymers)<br>Type 4 | No ISO standard<br>available for these<br>components: a<br>combination of ISO<br>20795-1:2013, ISO<br>10477:2018 and<br>internal requirements<br>were tested to confirm<br>performance. | No ISO standard<br>available for these<br>components; a<br>combination of ISO<br>20795-1:2013, ISO<br>10477:2018 and<br>internal requirements<br>were tested to confirm<br>performance. | -The predicate device<br>Halley resin (K190043)<br>complies with ISO<br>20795-1 for Type 4<br>materials.<br>-The proposed Halley<br>resin system and<br>reference devices are<br>minor parts of the<br>restorations and do not<br>have a unique<br>standard requirement.<br>-A combination of ISO<br>standards and internal<br>requirements were<br>considered to support<br>use as bonding agent<br>and sealer.<br>-Performance is based<br>on testing. |
| Shelf life | 3 years | 3 years | Not submitted in the<br>reference device<br>510(k) | Not submitted in the<br>reference device<br>510(k) | Shelf life of the<br>proposed device<br>Halley resin system is<br>based on performance<br>testing. |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
# 7. Non-Clinical Performance Data
Physical Properties:
There are no specific ISO standards for supporting materials such as a bonding agent or a sealer. Performance of these proposed devices are linked to the Hallev resin (K190043) workflow. These components are minor part of the overall device and the performance testing is limited to support its characteristic as a bonding agent or a sealer. These components were added and tested to align with predicate device, Halley resin (K190043).
Data derived from testing supports conclusion that the proposed device Halley resin system support the predicate device Halley resin (K190043) workflow.
# Biocompatibility:
The predicate device. Hallev resin (K190043) biocompatibility and test rationale remains unchanged from the current clearance. The patient contacting level and duration of the proposed Halley Denture-Tooth Bonding Agent and Halley Sealer are identical in material formulation and produced by the identical manufacturing methods as the reference devices. Therefore, no new biocompatibility testing was conducted for the proposed Halley resin system.
# 8. Clinical Performance Data
No data from human clinical studies has been included to support the substantial equivalence of the proposed device, Halley resin system.
# 9. Conclusion Regarding Substantial Equivalence
The proposed device. Halley resin system testing supports the cleared predicate device, Halley resin (K190043) workflow for the following reasons:
- . Both components were tested to support the modification of the cleared predicate device, Halley resin (K190043) and its workflow.
- Both components have been previously cleared for similar use and the ● material composition is identical to the cleared reference devices, Eclipse® Bonding Agent, K051707 and TEMPFX Esthetic Provisional System, K061264.
- The test data to verify the performance and biological safety of the proposed ● device, Halley resin system, has been provided. The results of this testing combined with the characteristics supporting the modified workflow support substantial equivalence.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.