← Product Code [EBI](/submissions/DE/subpart-d%E2%80%94prosthetic-devices/EBI) · K190940

# Tokuyama Rebase III (K190940)

_Tokuyama Dental Corporation · EBI · Oct 4, 2019 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/EBI/K190940

## Device Facts

- **Applicant:** Tokuyama Dental Corporation
- **Product Code:** [EBI](/submissions/DE/subpart-d%E2%80%94prosthetic-devices/EBI.md)
- **Decision Date:** Oct 4, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3760
- **Device Class:** Class 2
- **Review Panel:** Dental

## Indications for Use

The TOKUYAMA REBASE III device is for relining dentures and extension of denture borders.

## Device Story

Acrylic resin kit for chairside hard denture relining and border extension; consists of powder (poly(ethyl methacrylate), benzoyl peroxide) and liquid (methacrylate monomer) components; includes REBASE adhesive for bonding to acrylic surfaces. Principle of operation: chemical polymerization initiated by mixing powder and liquid. Used in dental clinics by dentists. Output: cured hard resin layer on denture base. Benefits: restores denture fit and comfort for patients. Does not bond to nylon, metal, or silicone.

## Clinical Evidence

No clinical tests performed. Evidence is based on bench testing of physical properties (consistency, peak temperature, porosity, surface gloss, water sorption, water solubility, hardness) per ISO 20795-1:2013 and JIS T 6521:2005, and biocompatibility assessment via comparison of ingredients with previously cleared devices.

## Technological Characteristics

Self-curing acrylic resin; powder (poly(ethyl methacrylate), benzoyl peroxide) and liquid (methacrylate monomer). Adhesive contains ethyl acetate, acetone, and methacrylate polymer. No energy source. Non-sterile. Complies with ISO 20795-1 and JIS T 6521 standards.

## Regulatory Identification

A denture relining, repairing, or rebasing resin is a device composed of materials such as methylmethacrylate, intended to reline a denture surface that contacts tissue, to repair a fractured denture, or to form a new denture base. This device is not available for over-the-counter (OTC) use.

## Predicate Devices

- GC RELINE ([K982695](/device/K982695.md))

## Reference Devices

- TOKUSO REBASE ([K896981](/device/K896981.md))
- TOKUYAMA REBASE II ([K022641](/device/K022641.md))
- TOKUYAMA CURECGRACE ([K170549](/device/K170549.md))
- PALFIQUE ESTELITE PASTE ([K980051](/device/K980051.md))
- Bonding agent ([K963144](/device/K963144.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue. Underneath the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 4, 2019

Tokuyama Dental Corporation % Keith Barritt Principal Fish & Richardson P.C. 1001 Maine Ave., SW, Suite 1000 Washington, District of Columbia 20024

Re: K190940

Trade/Device Name: Tokuyama Rebase III Regulation Number: 21 CFR 872.3760 Regulation Name: Denture Relining, Repairing, Or Rebasing Resin Regulatory Class: Class II Product Code: EBI Dated: September 5, 2019 Received: September 6, 2019

Dear Keith Barritt:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Srinivas Nandkumar, Ph.D. Acting Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

#### Indications for Use

510(k) Number (if known) K190940

Device Name

TOKUYAMA REBASE III

Indications for Use (Describe)

The TOKUY AMA REBASE III device is for relining dentures and extension of denture borders.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (7/17)

Page 1 of 1

EF PSC Publishing Services (301) 443-4710

{3}------------------------------------------------

## K190940

# 510(k) Summary Tokuyama Dental Corporation TOKUYAMA REBASE III Denture relining, repairing, and rebasing resin

### Submitter

球

商

న్న

в

元

ﺗﻘﺪﻳ

e

. .

### (i) 510(k) Submitter

Tokuyama Dental Corporation 38-9 Taitou 1-chome, Taitou-ku Tokyo 110-0016 Japan Phone: 011-81-3-3835-2261

#### (ii) 510(k) Submitter Contact

Keith A. Barritt Fish & Richardson P.C. 1000 Maine Ave., S.W, Suite 1000 Washington, DC 20024 Phone: (202) 783-5070 Facsimile: (202) 783-2331 Email: barritt@fr.com

#### (ii) Preparation Date

September 5, 2019

#### Device

Trade or Proprietary Name: Common Name: Classification Name: Product Code: Class:

#### TOKUYAMA REBASE III

denture relining, repairing, or rebasing resin resin, denture, relining, repairing, rebasing EBI, 21 CFR 872.3760 2

{4}------------------------------------------------

#### Predicate Device

Primary Predicate: GC America, Inc.'s GC RELINE (K#982695).

Reference Devices for purposes of establishing biocompatibility (i.e. all ingredients of the TOKUYAMA REBASE III are already cleared for marketing in similar Tokuyama devices): TOKUSO REBASE (K#896981), TOKUYAMA REBASE II (K#022641), TOKUYAMA CURECGRACE (K#170549), and PALFIQUE ESTELITE PASTE (K#980051).

#### Device Description

The TOKUYAMA REBASE III device is an acrylic resin used for chairside hard denture linings and extension of denture borders. The powder contains poly(ethyl methacrylate) and benzoyl peroxide. The liquid contains methacrylate monomer. The REBASE adhesive contains ethyl acetate, acetone and methacrylate polymer. The REBASE adhesive is used solely for bonding to acrylic surfaces, and cannot be used to bond nylon, metal, and silicone surfaces, etc.

The device is composed of both a powder and liquid that is cured by chemical polymerization. The powder is packaged in a plastic bottle, and the liquid is packaged in a glass bottle. A group of polymerization agents of the device is contained in the powder and liguid separately to avoid the reaction during storage, and the polymerization reaction starts by mixing the powder and liquid. The device is cured by polymerization of methacrylic monomers contained in the liquid.

The TOKUYAMA REBASE III kit contains powder, liquid, REBASE adhesive, TOKUSO RESIN HARDENER II, and other accessories, namely, a measuring cup, dropper, rubber cup, spatula, drip nozzle, drip cup, brush, and spoon.

The TOKUYAMA REBASE III device, like the primary predicate K#982695, is regulated under FDA product code EBI, 21 CFR 872.3760.

The device does not come sterile and is not intended to be sterilized prior to use.

The powder comes in "fast" and "normal" depending on the desired setting time, and in three shades designated as Light Pink, Pink, and Live Pink.

#### Indications for Use

The TOKUYAMA REBASE III device is for relining dentures and extension of denture borders

{5}------------------------------------------------

## Comparison of Technological Characteristics

The TOKUYAMA REBASE III device has the same basic technological characteristics in terms of design, material, and chemical composition as the predicate device identified above. The TOKUYAMA REBASE III device does not have its own energy source.

For purposes of performance characteristics for obtaining FDA marketing authorization, the TOKUYAMA REBASE III is substantially equivalent to GC America Inc.'s GC RELINE (K#982695), as shown below:

|                           | Subject device                                                                                                                                              | Primary predicate                                                                                                                                           | Difference                                                                                                            |                                                                                                                                                                                                                                                                                                                         |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device name               | TOKUYAMA REBASE III                                                                                                                                         | GC RELINE                                                                                                                                                   | -                                                                                                                     |                                                                                                                                                                                                                                                                                                                         |
| Manufacturer              | Tokuyama Dental                                                                                                                                             | GC America Inc.                                                                                                                                             | -                                                                                                                     |                                                                                                                                                                                                                                                                                                                         |
| 510(k) No.                | (Pending)                                                                                                                                                   | K982695<br>(Bonding agent is separately<br>authorized under K963144)                                                                                        | -                                                                                                                     |                                                                                                                                                                                                                                                                                                                         |
| Classification<br>name    | Resin, Denture, Relining,<br>Repairing, Rebasing                                                                                                            | Resin, Denture, Relining,<br>Repairing, Rebasing                                                                                                            | -                                                                                                                     |                                                                                                                                                                                                                                                                                                                         |
| Product Code              | EBI                                                                                                                                                         | EBI                                                                                                                                                         | -                                                                                                                     |                                                                                                                                                                                                                                                                                                                         |
| Device Class              | 2                                                                                                                                                           | 2                                                                                                                                                           | -                                                                                                                     |                                                                                                                                                                                                                                                                                                                         |
| Indications for<br>Use    | Relining dentures and extension<br>of denture<br>borders                                                                                                    | GC Reline is an acrylic resin<br>for direct denture linings and<br>extension of denture borders 2)                                                          | Similar<br><br>The Indications<br>for Use of the<br>subject device<br>are within those<br>of the predicate<br>device. |                                                                                                                                                                                                                                                                                                                         |
| Component                 | Powder, Liquid, and Adhesive                                                                                                                                | Powder, Liquid, and Bonding<br>agent                                                                                                                        | Similar<br><br>It only differs in<br>expression of<br>adhesive, there<br>are no<br>differences in<br>components.      |                                                                                                                                                                                                                                                                                                                         |
| Principle of<br>operation | Acrylic resin that is cured by<br>chemical polymerization<br>reaction starting with mixing<br>powder and liquid component<br>(Self - curing acrylic resin). | Acrylic resin that is cured by<br>chemical polymerization<br>reaction starting with mixing<br>powder and liquid component<br>(Self - curing acrylic resin). | Identical                                                                                                             |                                                                                                                                                                                                                                                                                                                         |
| Sterilization             | Not provided sterile and not<br>intended to be sterilized                                                                                                   | Not provided sterile and not<br>intended to be sterilized                                                                                                   | Identical                                                                                                             |                                                                                                                                                                                                                                                                                                                         |
| Material                  | P<br>o<br>w<br>d<br>e<br>r                                                                                                                                  | - Poly(ethyl methacrylate)<br>- Benzoyl peroxide                                                                                                            | - Poly(ethyl methacrylate)<br>- Benzoyl peroxide                                                                      | Identical                                                                                                                                                                                                                                                                                                               |
|                           | L<br>i<br>q<br>u<br>i<br>d                                                                                                                                  | - Methacrylate monomer                                                                                                                                      | - Benzyl methacrylate<br>- Dimethacrylate<br>- 1,6-hexanediyl<br>bismethacrylate<br>- Catalyst<br>- UV-light absorber | Similar<br>The subject<br>device consists<br>mainly of<br>methacrylates as<br>with the predicate<br>device.<br>Since all of the<br>ingredients to be<br>used in the subject<br>device have<br>already been used<br>in such existing<br>devices, the safety<br>of subject device<br>is assured.                          |
|                           | A<br>d<br>h<br>e<br>s<br>i<br>v<br>e                                                                                                                        | REBASE Adhesive<br>- Ethyl acetate<br>- Acetone<br>- Methacrylate polymer                                                                                   | Bonding agent<br>- Methyl methacrylate<br>(MMA)<br>- Acetone<br>- 2-Hydroxyethyl<br>methacrylate (HEMA)               | Similar<br>The subject<br>device contains<br>organic solvents<br>as a part of its<br>ingredients as<br>with the predicate<br>device.<br>Since all of the<br>ingredients to be<br>used in the subject<br>device have<br>already been used<br>in such existing<br>devices, the safety<br>of subject device<br>is assured. |

{6}------------------------------------------------

్లో గ్రామం

a

:

"和

广

ू में

្រុក 1

ﺮﺓ

المواليد الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الم

ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘ

e

网 18

,对

{7}------------------------------------------------

#### Shelf Life Testing

Shelf life test data of TOKUYAMA REBASE III based on ISO 20795-1 are not available. However, Tokuyama conducted real-time shelf-life testing to establish a shelf life of three years at a temperature between 0-25°C (32-77F°) based on JIS T 6521, and no deterioration is expected in real time shelf life data based on the ISO standard. In particular, since deterioration of curability is easily observed as deterioration of surface hardness, water absorption, and water solubility, it is expected that there will be no deterioration in flexural properties, which is one of the indicators of curability.

#### Material And Chemical Composition

For purposes of performance, as noted above the primary predicate is GC America, Inc.'s GC RELINE (K#982695).

For purposes of material and chemical composition, the TOKUYAMA REBASE III device has the same basic characteristics as Tokuyama's own TOKUSO REBASE (K#896981), TOKUYAMA REBASE II (K#022641), TOKUYAMA CURECGRACE (K#170549), and PALFIQUE ESTELITE PASTE (K#980051). Thus, no further biocompatibility testing was conducted on the device.

Tokuyama examined the biocompatibility of TOKUYAMA REBASE III according to ISO 10993-1: 2009 and the FDA's related guidance "Use of International Standard ISO 10993-1. Biological evaluation of medical devices -Part 1: Evaluation and testing within a risk management process." As a result of comparing of each ingredient and its purpose of use, operating principle, manufacturing process, nature of body contact, contact duration, and other factors with existing devices, it was concluded that TOKUYAMA REBASE III has no new biocompatibility risk and required no further biocompatibility tests.

#### Performance Data Summary

Tokuyama examined the performance based on ISO 20795-1: 2013, including testing of the cured material for surface characteristics, polishability, shape capability, color stability, translucency, porosity, water sorption, and water solubility, and TOKUYAMA REBASE III conforms to the standard for each characteristic, as does the primary predicate GC America, Inc.'s GC RELINE (K#982695).

"Residual methyl methacrylate monomer" was not applicable as the material does not contain methyl methacrylate monomer. "Colour match" test could not be conducted because the material has no shade guide. "Bonding to artificial teeth" data were not available, but the need for the test was low as relining material

{8}------------------------------------------------

tends to be applied to the mucosal surface of the denture base in a thin layer, and not applied to the occlusal surface, so it does not contact with artificial teeth. Because the TOKUYAMA REBASE III device is not impact resistant, related tests were not conducted.

Additional non-clinical testing of the physical properties of the TOKUYAMA REBASE III device was conducted in accordance with JIS T 6521:2005 "Denture Base Hard Relining Materials," including testing of the cured material for consistency, peak temperature, porosity and defects, surface gloss, water sorption, water solubility, and hardness, and TOKUYAMA REBASE III conforms to the standard for each characteristic, as does the primary predicate GC America, Inc.'s GC RELINE (K#982695).

There were no clinical tests performed for the TOKLUYAMA REBASE III device.

#### Conclusion

Based on the non-clinical testing conducted of the physical properties of the TOKUYAMA REBASE III device in comparison to the predicate device identified above, and on the biocompatibility of FDA-authorized devices for the same use with the same ingredients, it is concluded that the TOKUYAMA REBASE III device is substantially equivalent to the predicate device.

41383636.doc

---

**Source:** [https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/EBI/K190940](https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/EBI/K190940)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
