DEFLEX ACRILATO

K113788 · Nuxen S.R.L. · EBI · Mar 16, 2012 · Dental

Device Facts

Record IDK113788
Device NameDEFLEX ACRILATO
ApplicantNuxen S.R.L.
Product CodeEBI · Dental
Decision DateMar 16, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3760
Device ClassClass 2

Indications for Use

This product is indicated for the manufacturing of bases of partial or full removable dentures, r nis products indication for the mating and pneumatic injection of the material, following the Indications of Use that come with the product.

Device Story

Deflex Acrilato is a pigmented polymethyl-methacrylate (PMMA) thermoplastic resin supplied as pellets in vacuum-sealed packaging. It is used by dental professionals for the fabrication of partial, complete, and combined removable dentures. The material is processed via heating and pneumatic injection. The device provides a flexible, resilient denture base material with characterized water absorption and solubility properties. It is intended to serve as a stable, inert material for dental prosthetics.

Clinical Evidence

No clinical data. Bench testing only. Testing included material characterization, mechanical properties, residual MMA monomer content, water absorption, and water solubility, performed according to EN 20795-1:2009 (equivalent to ANSI/ADA 12-2002/R2008).

Technological Characteristics

Thermoplastic PMMA resin; supplied as pellets. Processing involves heating and pneumatic injection. Tested per EN 20795-1:2009 and ANSI/ADA 12-2002 (R2008) for mechanical and chemical properties including water absorption, solubility, and residual monomer content.

Indications for Use

Indicated for the manufacturing of bases for partial or full removable dentures.

Regulatory Classification

Identification

A denture relining, repairing, or rebasing resin is a device composed of materials such as methylmethacrylate, intended to reline a denture surface that contacts tissue, to repair a fractured denture, or to form a new denture base. This device is not available for over-the-counter (OTC) use.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ KII3788 ### 510(k) Summary (Rev 01) ### (1) Submitter: NUXEN S.R.L.. 1053 Ayacucho, 3rd A (C1111AAE) Ciudad Autónoma de Buenos Aires, Argentina Tel +54 11 4241 7762 Contact person: Pablo Lagostena Preparation Date: August 1, 2011 ### (2) Device Proprietary Information | Trade Name: | DEFLEX ACRILATO | |----------------------|--------------------------------------------------------------------------------------| | Common Name: | Polymer for Injection of Denture Bases | | Classification Name: | Denture relining, repairing, or rebasing resin (21 CFR<br>872.3760 Product Code EBI) | #### (3) Predicate Devices | 510(k)<br>Number | Trade Name | Manufacturer | |------------------|-----------------|-------------------------------------------| | K053060 | TCS unbreakable | Thermoplastic<br>Comfort<br>Systems, Inc. | | K102874 | Veracryl | New Stetic | #### Device Description (4) Deflex Acrilato is a pigmented polimethyl-methacrilate (PMMA) resin that belongs to the group of thermoplastic polymers which is used for the manufacturing of partial, complete and combined dentures. The product is presented in pellets, packed in an aluminum canister (as a primary packaging), and then in a vacuum sealed aluminized plastic bag (as a secondary packaging). the PMMA granules are pigmented so as to achieve aesthetic effects on the manufactured denture. ### (5) Intended Use l This product is indicated for the manufacturing of bases of partial or full removable dentures, r mis products indication for the mating and pneumatic injection of the material, following the Indications of Use that come with the product. ### Comparison to predicate devices (6) {1}------------------------------------------------ Deflex Acrilato is a thermoplastic injectable polymer that belong to the group of the PMMA. As no thermoplastic injectable PMMA could be identified as predicate, the properties of the material were devided into two groups and compared to different predicate devices (all of them having the same intended use), the first one sharing similar indications of use and mechanical properties and the second one sharing similar composition and chemical properties Intended use, mechanical properties and indications of use have been compared to the following currently marketed thermoplastic injectable material: | 510(k)<br>Number | Trade Name | Manufacturer | |------------------|-----------------|-------------------------------------------| | K053060 | TCS unbreakable | Thermoplastic<br>Comfort<br>Systems, Inc. | Intended use, chemical composition and chemical properties have been compared to the following currently marketed thermoplastic injectable material: | 510(k)<br>Number | Trade Name | Manufacturer | |------------------|------------|--------------| | K102874 | Veracryl | New Stetic | Finding that: ### Comparison of Intended Use - -The two predicates and Deflex Acrilato have the same intended use ### Comparison to TCS - Deflex Acrilato and TCS have similar indications of use • - Both products show flexible and resilient behavior being DEFLEX ACRILATO more resistant than its predicate ### Comparison to Veracryl - Both materials have been characterized as poly methyl methacrylate - - Both materials show similar residual MMA monomer content - - Both materials show similar water absorption and solubility levels - - Deflex acrilato shows much lower solubility levels than its predicate, which is an indication of long term stability and inert behavior Considering that both TCS and VERACRYL have essentially the same intended use of DEFLEX ACRILATO, that mechanical properties and indications of use are comparable to those of TCS and AONEATO, that moonanied propenties are comparable to those of VERACRYL, It is our that onemised componion arilato is substantially equivalent to predicate devices in safety and effectiveness. l : {2}------------------------------------------------ ### (7) Non Clinical Testing The following tests have been carried out in order to demonstrate product's safety and effectiveness | TEST | Standard or Test<br>Method applied | |---------------------------------|------------------------------------| | Material<br>characterization | Internal laboratory<br>method | | Mechanical properties | EN 20795-1:2009 * | | Residual MMA<br>monomer content | EN 20795-1:2009 * | | Water absorption | EN 20795-1:2009 * | | Water solubility | EN 20795-1:2009 * | * equivalent to test method described in ANSI+ADA+12-2002+(R2008) ### (8) Clinical Investigation Clinical investigation is not required for this device in order to demonstrate its safety and effectiveness. #### Conclusion (9) NUXEN S.R.L. concludes that DEFLEX ACRILATO Resin for dental prostheses is substantially equivalent to its predicate devices. Based upon test data submitted, the product included in this submission can be safely and effectively used for manufacturing bases of partial removable or complete dentures through heating processes and pneumatic injection of the material, following the "indications of use" that come with the product. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its wing and body. The eagle is positioned within a circular border that contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a sans-serif font. Food and Drug Administration 10903 New Hampshire-Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 MAR 1 6 2012 Nuxen S.R.L. C/O Mr. Hernan Ilari Consultant IC INGENIERIA J B Justo 9741, PB Buenos Aires, C1408ALB ARGENTINA Re: K113788 Trade/Device Name: DEFLEX ACRILATO Regulation Number: 21 CFR 872.3760 Regulation Name: Denture Relining, Repairing, or Rebasing Resin Regulatory Class: II Product Codes: EBI and MQC Dated: March 12, 2012 Received: March 12, 2012 Dear Mr. Ilari: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ ## Page 2 - Mr. Hernan Ilari: Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Anthony V. Watson Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indications for Use Statement 510(k) Number (if known): K//3788 Device Name: DEFLEX ACRILATO Indications for Use: This product is indicated for the manufacturing of bases of partial or full removable dentures, r nis products thatoated for the marketing and pneumatic injection of the material, following the Indications of Use that come with the product. Prescription Use X (Part 21 CFR 801 Subpart D) Over-The-Counter Use AND/OR (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Page of Susan Reager (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: K113781
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...