P pro Crown & Bridge is a flowable, light-curing, acrylate-based composite material for the additive production of temporary anterior and posterior restorations.
Device Story
P pro Crown & Bridge is a flowable, light-curing, acrylate-based composite material used for additive manufacturing of temporary dental restorations. Input consists of digital design data processed by 3D printers operating at 385 nm wavelength. The device transforms these inputs into physical temporary crowns, bridges, veneers, inlays, and onlays. Used in dental clinics or laboratories by dental professionals; the output is a finished temporary restoration placed in the patient's mouth to provide esthetic and functional support during the interim period before permanent restoration. Benefits include efficient fabrication of custom temporary dental prosthetics.
Clinical Evidence
Bench testing only. Biocompatibility evaluation conducted per ISO 10993-1, including cytotoxicity (ISO 10993-5), sensitization (ISO 10993-10), and irritation (ISO 10993-10). Mechanical testing performed per ISO 4049 and ISO 10477, showing flexural strength of 146 MPa, solubility of 0.5 µg/mm3, water absorption of 16 µg/mm3, and compressive strength of 358 MPa. No clinical data provided.
Technological Characteristics
Flowable, light-curing, acrylate-based composite material containing multifunctional acrylic resins and inorganic fillers (0.04 - 0.2 µm). Additive manufacturing process using 385 nm wavelength printers. Complies with ISO 4049 and ISO 10477 standards. Mechanical properties: 146 MPa flexural strength, 358 MPa compressive strength, 5988 MPa modulus of elasticity.
Indications for Use
Indicated for additive production of temporary anterior and posterior restorations, including single-tooth crowns, bridges (max. six units, max. two consecutive pontics), veneers, inlays, and onlays.
Regulatory Classification
Identification
A temporary crown and bridge resin is a device composed of a material, such as polymethylmethacrylate, intended to make a temporary prosthesis, such as a crown or bridge, for use until a permanent restoration is fabricated.
K192806 — DENTCA Crown and Bridge · Dentca, Inc. · Feb 5, 2020
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 31, 2020
DeltaMed GmbH % Gary Chuven Director of Regulatory Affairs Cosmedent, Inc 401 N Michigan Ave, Suite 2500 Chicago, Illinois 60611
Re: K200039
Trade/Device Name: P pro Crown & Bridge Regulation Number: 21 CFR 872.3770 Regulation Name: Temporary Crown and Bridge Resin Regulatory Class: Class II Product Code: EBG Dated: June 22, 2020 Received: July 6, 2020
Dear Gary Chuven:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm_identifies_combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Srinivas "Nandu" Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K200039
Device Name P pro Crown & Bridge
Indications for Use (Describe)
Additive production of temporary anterior and posterior restorations, e.g.
- Single tooth crowns.
- Bridges (max. six units, max. two consecutive pontics).
- Veneers, Inlays, Onlays.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
{3}------------------------------------------------
# Section 5: 510(k) Summary
This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92(c)
| 510(k) No.: | K200039 |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared: | 2020-07-29 |
| 501(k) Submitter: | DeltaMed GmbH<br>Raiffeisenstr. 8a<br>61169 Friedberg<br>Deutschland / Germany<br>Tel. +49 (0) 6031 / 7283- 0<br>Fax. +49 (0) 6031 / 7283- 29<br>E-mail: info@deltamed.de |
| Contact Person: | Michael Zimmermann<br>Regulatory Affairs<br>DeltaMed GmbH<br>Tel: +49 (0) 6031 / 7283- 17<br>Email: michael.zimmermann@deltamed.de |
| Trade Names: | P pro Crown & Bridge |
| Common Name: | Crown and bridge, temporary, resin |
| Regulatory Class: | Class II |
| Device Panel: | Dental Panel |
| Regulation Number: | 21 CFR 872.3770 |
| Classification Name: | Temporary crown and bridge resin |
| Product Code: | EBG |
## Predicate Devices:
(1) Primary Predicate e-Dent Temporary; DeltaMed GmbH, Friedberg, DE
- (2) Predicate Device Vita CAD-Temp for in Lab: VITA Zahnfabrik GmbH & Co KG, Säckingen, GE
- (3) Predicate Device Luxatemp Ultra/Luxatemp Star: DMG Inc., Ayer Ma. USA
## Description of Device:
P pro Crown & Bridge is a flowable, light-curing, acrylate-based composite material for the additive production of temporary anterior and posterior restorations.
It consists of multifunctional acrylic resins and fillers made of inorganic fillers ranging in size from 0.04 - 0.2 um. The raw materials used in the preparation have been used for many years in dental products.
Exposition: The temporary crown and bridge resin has an intended gingival, enamel/dentine contact for more than 30 days.
## Intended Use:
{4}------------------------------------------------
P pro Crown & Bridge is a flowable, light-curing, acrylate-based composite material for the additive production of temporary anterior and posterior restorations.
## Indications for Use:
- Additive production of temporary anterior and posterior restorations, e.g.
- Single-tooth crowns.
- Bridges (max. six units, max. two consecutive pontics).
- Veneers, inlays, onlays.
## Biocompatibility testing
The biocompatibility evaluation for P pro Crown & Bridge was conducted in accordance with the Guidance Document: Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" and the International Standard ISO 10993-1:2018(R) 2019 "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" as recognized by FDA. The battery of testing included the following tests:
- Cytotoxicity (ISO 10993-5) i
- Sensitization (ISO 10993-10)
- Irritation (ISO 10993-10)
According to ISO 10993-2 "Biological evaluation of medical devices ----------------------------------------------------------------------------------------------------------no additional animal testing were necessary to ensure human safety. For the evaluation of the biological risk of endpoints: systemic- and CMR toxicity, sufficient literature data were available.
### Substantial Equivalence Discussion:
| Subject Device | : P pro Crown & Bridge |
|------------------------------|---------------------------------|
| (1) Primary Predicate Device | : e-Dent Temporary |
| (2) Predicate Device | : Vita CAD-Temp, for in Lab |
| (3) Predicate Device | : Luxatemp Ultra / Luxatemp Sta |
| Submission: | Class. Name: | Product Code | Regulation No. |
|-------------|------------------------------------|--------------|----------------|
| (1) K102776 | Crown and Bridge, Temporary, Resin | EBG | 872.3770 |
| (2) K070991 | Crown and Bridge, Temporary, Resin | EBG | 872.3770 |
| (3) K101710 | Crown and Bridge, Temporary, Resin | EBG | 872.3770 |
The table on the following page compares the subject device with its predicates:
{5}------------------------------------------------
| Property | Subject Device | (1) Primary Predicate | (2) Predicate Device | (3) Predicate Device |
|---------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|
| | P pro Crown &<br>Bridge | e-Dent Temporary<br>(K102776) | Vita CAD-Temp for<br>inLab (K070991) | Luxatemp Ultra /<br>Star<br>(K101710) |
| Intended use,<br>Indications | Additive production<br>of temporary anterior<br>and posterior<br>restorations, e.g.<br>Single-tooth crowns.<br>Bridges (max. six<br>units, max. two<br>consecutive pontics).<br>Veneers, inlays,<br>onlays. | e-dent Temp is used for<br>the fabrication of multi-<br>unit, fully or partially<br>anatomical long-term<br>temporary crown and<br>bridge restorations,<br>anterior crowns.<br>posterior crowns,<br>anterior/posterior<br>bridges 3 or 4 units, 1<br>pontic | CAD-Temp is used<br>for the fabrication of<br>multi-unit, full or<br>partial anatomical<br>long term temporary<br>bridge restorations<br>with up to 2 pontics<br>using the inLab or<br>CEREC systems of<br>Sirona. The bridge<br>block can also be<br>used for temporary<br>crowns. | Self-curing composite<br>for the fabrication of<br>temporary crowns and<br>bridges, inlays, onlays<br>and veneers. |
| Processing | Additive<br>manufacturing, using<br>printers with 385 nm<br>wavelength | Additive manufacturing<br>using Perfactory DDP,<br>350-420nm; (Dental<br>Digital Printer),<br>Envisiontec GmbH,<br>Germany | Substractive<br>(mechanical)<br>processing | Self-curing composite |
| Requirements<br>1. ISO 4049<br>2. ISO 10477 | | Mechanical Properties according to ISO 4049 and ISO 10477 | | |
| Flexural strength | 146 MPa | 123 MPa | > 80 MPa | 125 MPa |
| 1. ≥ 100 MPa | pass | pass | ./. | pass |
| 2. ≥ 50 MPa | pass | pass | pass | pass |
| Solubility | 0.5 µg/mm3 | 0 µg/mm3 | No data | No data |
| 1. & 2.<br>≤ 7,5 µg/mm3 | pass | pass | Complies with<br>ISO 10477 | --- |
| Water absorption | 16 µg/mm3 | 18 µg/mm3 | No data | 10 µg/mm3 |
| 1. & 2.<br>≤ 40 µg/mm3 | pass | pass | Complies with<br>ISO 10477 | pass |
| Color stability<br>1. & 2. | pass | pass | Complies | No data |
| Requirements | Mechanical Properties according to DeltaMed internal standard | | | |
| Compressive<br>Strength (MPa) | 358 MPa | 354 MPa | n.e. | 376 MPa |
| > 300 MPa | pass | pass | --- | pass |
{6}------------------------------------------------
| Property | Subject Device | (1) Primary Predicate | (2) Predicate Device | (3) Predicate Device |
|--------------------------------------------------------------------------------------------------|-------------------------|------------------------------------------------------------|---------------------------------------------------------------------------------------------|---------------------------------------|
| | P pro Crown &<br>Bridge | e-Dent Temporary<br>(K102776) | Vita CAD-Temp for<br>inLab (K070991) | Luxatemp Ultra /<br>Star<br>(K101710) |
| Modulus of<br>elasticity | 5988 MPa | 5300 MPa | >2800 MPa | No data |
| > 5000 MPa | pass | pass | ./. | --- |
| Composition/Chemical properties | | | | |
| Acrylate monomers with initiators,<br>stabilizers and pigments, approx. 50%<br>inorganic fillers | | Modified<br>polymethacrylates with<br>14% inorganic filler | Matrix of multi-<br>functional<br>methacrylates;<br>catalysts, stabilizers<br>and additives | |
## Discussion:
Primary Predicate Device: All four products have substantially the same intended use as temporary esthetic (dental) restorative materials but differ in the indication for use. The predicate device (1) as the primary predicate device is used to demonstrate substantially equivalence in manufacturing method, indication (fully or partially anatomical long-term temporary crown and bridge -but limited by 4 units and 1 ponic), chemical- and biological properties.
(2) Reference Device: Is referenced based on its physical properties to demonstrate equivalence of the expanded indication of 2 pontics.
(3) Predicate Device: Concerning the additional claims (indications) "veneers, inlays, onlays", was selected to show the equivalence with the subject device.
The physical, mechanical and chemical properties of the subject device, the primary predicate device (K102776) and the two predicate devices (K070991 and K101710) are very similar. Although the indication refers to temporary restoration, the subject device complies with the standards ISO 4049 (Polymer-based restorative materials) and ISO 10477 (Polymer-based crown and veneering materials) in terms of flexural strength, water sorption, solubility and color stability.
Biocompatibility studies were performed on the subjected device. The data were analyzed and the results of the biocompatibility test (10993-xx) substantiate that P pro Crown & Bridge with its modification in composition is as safe (biocompatible) as the primary predicate.
## Results/summary of the Substantial Equivalence
The subject device P pro Crown & Bridge has been compared with its primary predicate device e-Dent Temporary (K102776) and the two predicate devices- Vita CAD-Temp for inLab (K070991) and Luxatemp Ultra / Star (K101710), with regard to indication (intended use), performance data, chemical composition and biocompatibility.
It can be concluded, that the subject device P pro Crown & Bridge is substantially equivalent to the predicates.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.