Hytac® Aplitip® is a compomer filling material indicated for the following uses: (1) fillings in deciduous teeth; (2) fillings in Class V cavities; (3) eroded and abraded cervical areas; (4) fillings in Class III cavities; and (5) temporary fillings in Class I and II cavities. Hytac® OSB is a light cured bonding agent for use with compomer filling materials.
Device Story
Hytac® Aplitip® (compomer filling material) and Hytac® OSB (light-cured bonding agent) used by dentists for restorative dental procedures. Aplitip® applied to tooth cavities; OSB used as adhesive interface. Materials provide structural filling and bonding for specific cavity classes and deciduous teeth. Benefit: restorative dental care. Clinical decision-making: dentist selects material based on cavity classification and tooth type.
Clinical Evidence
Bench testing only. Biocompatibility studies performed on final products including cytotoxicity, Ames test, skin/eye irritation, acute oral toxicity, and sensitization. Results showed no adverse toxicity or sensitization.
Technological Characteristics
Compomer filling material and light-cured bonding agent. Materials meet relevant DIN and ISO standards for dental composite materials. Contains water-soluble fluoride.
Indications for Use
Indicated for dental restorations in deciduous teeth, Class V cavities, eroded/abraded cervical areas, Class III cavities, and temporary fillings in Class I and II cavities.
Regulatory Classification
Identification
Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.
K962440 — PERTAC II/PERTAC II APLITIP · Espe GmbH & Co. Kg. · Sep 11, 1996
K080480 — NANOCOMPOSITE RESTORATIVE KIT · Dmg USA, Inc. · May 30, 2008
K231696 — Fusion Bond 5, Fusion Bond 7, Fusion Bond DC, Renew MDP, Renew Universal · Prevest Denpro Limited · Aug 11, 2023
Submission Summary (Full Text)
{0}
K962442
SEP 11 1996
## II. 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
ESPE is submitting a 510(k) premarket notification for modifications to its 510(k) submissions for tooth shade resin material, tradenamed Pertac®-Hybrid® (K900510), and the corresponding bonding agent, tradenamed Pertac® Universal Bond (K904289). The modified materials, referred to by the tradenames Hytac® Aplitip® and Hytac® OSB, respectively, are intended as compomer filling material and corresponding bonding agent. Hytac® Aplitip® is a compomer filling material indicated for the following uses: (1) fillings in deciduous teeth; (2) fillings in Class V cavities; (3) eroded and abraded cervical areas; (4) fillings in Class III cavities; and (5) temporary fillings in Class I and II cavities. Hytac® OSB is a light cured bonding agent for use with compomer filling materials.
With respect to Hytac® Aplitip®, ESPE is claiming substantial equivalence to its previously cleared Pertac®-Hybrid® product and Ivoclar's Compoglass® (K951836). These products have similar intended uses and, in the case of Compoglass®, the same principal ingredient composition. To support substantial equivalence to predicate products, the physical and technical
WAHA/A36360.1
{1}
characteristics, as well as the water soluble fluoride content, of Hytac® Aplitip® have been compared to those of Pertac®-Hybrid® and Ivoclar's Compoglass®. In addition, certain tests have been undertaken on the final Hytac® product, including cytotoxicity, Ames, skin irritation, eye irritation, acute oral toxicity and sensitization. Results of these studies indicate that the final product shows no adverse toxicity results.
With respect to Hytac® OSB, ESPE is claiming substantial equivalence to its previously cleared Pertac® Universal Bond and to Ivoclar's Compoglass® SCA (K951836). All products have the same intended use and all but one of the same components. To support substantial equivalence to predicate products, the physical and technical characteristics of Hytac® OSB have been compared to those of Pertac® Universal Bond and Ivoclar's Compoglass® SCA. In addition, the sensitization test has been performed on the final Hytac® OSB product to confirm safe use of the product, and test results show no sensitization.
Hytac® Aplitip® and Hytac® OSB meet the requirements of relevant DIN and ISO standards for dental composite material.
WADIA/A38368.1
{2}
ESPE's 510(k) has been submitted on June 17, 1996, by Dr. Barbara Wagner at Am Griesberg 2, D-82229 Seefeld, Germany (011-49-8152-700395).
WADIA/A38360.1
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.