← Product Code [EBF](/submissions/DE/subpart-d%E2%80%94prosthetic-devices/EBF) · K180463

# Apoller Flow Light Cure Flowable Composite (K180463)

_Greenway Yw, LLC · EBF · Nov 14, 2018 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/EBF/K180463

## Device Facts

- **Applicant:** Greenway Yw, LLC
- **Product Code:** [EBF](/submissions/DE/subpart-d%E2%80%94prosthetic-devices/EBF.md)
- **Decision Date:** Nov 14, 2018
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 872.3690
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

Class III and V restorations Restoration of minimally invasive cavity preparations (including small, non stress-bearing occlusal restorations) Base/liner under direct restorations Repair of small defects in esthetic indirect restorations Pit and fissure sealant Undercut blockout Repair of resin and acrylic temporary materials

## Device Story

Apoller Flow is a low-viscosity, light-activated flowable nanocomposite for dental restorative procedures. Device consists of syringe with dispensing tips containing methmethacrylate monomers (BisGMA, UDMA, TEGDMA), photo initiator, stabilizer, and inorganic fillers (barium glass, strontium glass, nanosilica). Filler loading is ~56% by weight; particle sizes range 0.03 to 1 micron. Used by dentists in clinical settings for cavity restoration, lining, and repair. Material is applied via syringe and needle tip, then polymerized using a light-curing source. Output is a hardened, tooth-colored restorative material. Benefits include ease of application in minimally invasive preparations and aesthetic repair of dental defects.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing per ISO 4049 (sensitivity to ambient light, depth of cure, flexural strength, water sorption/solubility, shade stability, radio-opacity) and biocompatibility testing per ISO 10993-1, -3, -5, -11 and ISO 7405.

## Technological Characteristics

Low-viscosity light-cure nanocomposite. Composition: BisGMA, UDMA, TEGDMA monomers, photo initiator, stabilizer, and inorganic fillers (barium/strontium glass, nanosilica). Filler loading: 56% wt. Particle size: 0.03-1 micron. Form factor: plastic syringe with bendable metal injection needle. Polymerization: light-cured. Standards: ISO 4049 (physical properties), ISO 10993/7405 (biocompatibility).

## Regulatory Identification

Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.

## Predicate Devices

- Filtek™ Supreme Ultra Flowable Restorative ([K100235](/device/K100235.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 14, 2018

Greenway YW LLC % Maritza Moncayo VP Regulatory Affairs Iris Global LLC 40931 W. Hopper Dr. Maricopa, Arizona 85138

Re: K180463

Trade/Device Name: Apoller Flow Light Cure Flowable Composite Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth Shade Resin Material Regulatory Class: Class II Product Code: EBF Dated: August 23, 2018 Received: August 29, 2018

Dear Maritza Moncayo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Digitally signed by Mary S. Mary S. Runner -S3 Runner - S3 12:23:42 -05'00' 12:23:42 -05'00'

> for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

K180463

Device Name

APOLLER FLOW LIGHT CURE FLOWABLE COMPOSITE

Indications for Use (Describe)

Indications for use:

*Class III and V restorations

- * Restoration of minimally invasive cavity preparations (including small, non stress-bearing occlusal restorations)
*Base/liner under direct restorations

*Repair of small defects in esthetic indirect restorations

*Pit and fissure sealant

*Undercut blockout

*Repair of resin and acrylic temporary materials

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows a green logo with the letters "GW" in a stylized font. There are three horizontal lines running through the letters, giving them a striped appearance. A small "R" in a circle is located in the upper right corner of the logo, indicating a registered trademark.

Greenway YW LLC

### 5. 510 (k) Summary

# K180463 - Apoller Flow Light Cure Flowable Composite

Date prepared: November 09, 2018

#### Applicant, Contact and Manufacturer Information: 5.1.

| Applicant Information: | GREENWAY YW LLC             |
|------------------------|-----------------------------|
|                        | 21 N. Skokie Hwy, Suite 204 |
|                        | Lake Bluff, II. 60044,      |
|                        | United States of America    |
|                        | Tel: 1(847) 525-5258        |
|                        | Fax: 1(847) 478-5258        |
|                        | Contact: Scott Yin          |
|                        | Email: yinradtech@gmail.com |

Maritza Moncayo Contact Person: Regulatory Affairs IRIS GLOBAL, LLC 40931 W. Hopper Dr. Maricopa, AZ 85138 Phone/Fax: 1 (847)-226-9176 Email: mmoncayo@irisgra.com

- SINO-DENTEX CO., LTD. Manufacturer: 12357E Jinhu Road, High-Tech Development District, Changchun, China 130103 Contact: Pengtao Yan Email: dentex-ypt@163.com Tel: (086) 431-87855156 Fax: (086) 431-87855157

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows a green logo with the letters "GW" in a stylized font. There are three horizontal lines below the letters, giving the impression of a solid base. A small registered trademark symbol is located in the upper right corner of the logo. The logo is simple and modern.

### 5.2. General Device Information:

Greenway is providing general product information and classification per FDA guidelines, in Table 1.

# Table 1.- Product Classification

| Device Trade Name<br>(Commercial): | Apoller Flow Light Cure Flowable Composite |
|------------------------------------|--------------------------------------------|
| Common Name:                       | Tooth Shade Resin Material                 |
| Classification:                    | 21 CFR 872.3690                            |
| Regulation Name:                   | Tooth Shade Resin Material                 |
| Regulatory Class:                  | Class II                                   |
| Product Code:                      | EBF                                        |

### 5.3. Primary Predicate Device Information to which Equivalence is Claimed:

The primary predicate device is Filtek™ Supreme Ultra Flowable Restorative, summary of the predicate device in Table 2:

| Device Trade Name | Filtek™ Supreme Ultra Flowable Restorative |
|-------------------|--------------------------------------------|
| Common Name       | Tooth shade resin material                 |
| 510 (k) Number    | K100235                                    |
| Decision Date     | May 12, 2010                               |
| Manufacturer      | 3M Company                                 |
| Classification    | 21 CFR 872.3690                            |
| Product Code      | EBF                                        |
| Class             | II                                         |

## Table 2- Summary Predicate Device

### 5.4. Device Description:

Apoller Flow-Light Cure Flowable Composite is a low viscosity light activated flowable nanocomposite device. The device is comprised of a syringe with dispensing tips. It consists of methmethacrylate monomers, inorganic fillers and nanoparticles. Apoller Flow Light Cure Flowable Composite contains BisGMA, UDMA, TEGDMA, photo initiator, stabilizer and inorganic fillers including barium glass powder, strontium glass powder and nanosilica. The filler loading is about 56% by weight. The particle size varies from 0.03 to 1 micron. Apoller Flow Light Cure Flowable Composite series is packed in a plastic syringe with a bend metal injection needle heads for ease of usage, its model is Flowable and comes with Refill Syringe of 3.0g /syringe and 1.5 g /syringe. It is available in a variety of 15 tooth-colored shades.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows a green logo with the letters "GW" in a stylized font. The letters are bold and have horizontal lines running through them, creating a striped effect. There is a registered trademark symbol in the upper right corner of the logo. The logo is simple and modern.

Apoller Flow - Light Cure Flowable Composite Abbreviated 510(k) Notification

Greenway YW LLC

### 5.5. Indications for Use:

- Class III and V restorations
- . Restoration of minimally invasive cavity preparations (including small, non stress-bearing occlusal restorations)
- Base/liner under direct restorations
- Repair of small defects in esthetic indirect restorations .
- Pit and fissure sealant
- Undercut blockout
- . Repair of resin and acrylic temporary materials

### 5.6. Technological Characteristics:

The assessment of the technological characteristic of the Apoller Flow Light Cure Flowable Composite and of the predicate device "Filtek™ Supreme Ultra Flowable Restorative" are equivalent as the currently marketed predicate device manufactured by 3M Company-3M ESPE Dental Products, as described in Table 3.

Both Apoller Flow Light Cure Flowable Composite and primary predicate device are very similar in all comparison areas including chemical composition, intended use, method of polymerization, mechanical properties and application process and use. A slight difference is the filler loading which impact the flowability of the composite, doesn't influence the application and all physical properties.

| Technological<br>Characteristics/ Properties | Apoller Flow Light<br>Cure Flowable<br>Composite                                                                                                                                                                                                                                         | Filtek™ Supreme<br>Ultra Flowable<br>Restorative                                                                                                                                                                                                                                      | Discussion                                   |
|----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------|
| Intended use                                 | *Class III and V<br>restorations<br>*Restoration of<br>minimally invasive<br>cavity preparations<br>(including small, non<br>stress-bearing<br>occlusal<br>restorations)<br>*Base/liner under<br>direct restorations<br>*Repair of small<br>defects in esthetic<br>indirect restorations | *Class III and V<br>restorations<br>*Restoration of<br>minimally invasive<br>cavity preparations<br>(including small, non<br>stress-bearing<br>occlusal restorations)<br>*Base/liner under<br>direct restorations<br>*Repair of small<br>defects in esthetic<br>indirect restorations | Equivalent to<br>primary predicate<br>device |

Table 3.- Technological Characteristics

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows a logo with the letters 'GW' in a stylized, green font. The letters are bold and slightly overlapping, with horizontal lines running through them, creating a sense of depth or texture. A registered trademark symbol (®) is positioned in the upper right corner of the logo.

Greenway YW LLC

## Apoller Flow - Light Cure Flowable Composite Abbreviated 510(k) Notification

| Technological<br>Characteristics/ Properties    | Apoller Flow Light<br>Cure Flowable<br>Composite                                                         | Filtek™ Supreme<br>Ultra Flowable<br>Restorative                                                         | Discussion                                   |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|----------------------------------------------|
|                                                 | *Pit and fissure<br>sealant *Undercut<br>blockout *Repair of<br>resin and acrylic<br>temporary materials | *Pit and fissure<br>sealant *Undercut<br>blockout *Repair of<br>resin and acrylic<br>temporary materials |                                              |
| Method of Polymerization                        | Light Cured                                                                                              | Light Cured                                                                                              | Equivalent to<br>primary predicate<br>device |
| Chemical composition                            | Glass filled<br>Methmethacrylate<br>monomer resins<br>Nanoparticle<br>Filler loading-56 %<br>by weight   | Glass filled<br>Methmethacrylate<br>monomer resins<br>Nanoparticle<br>Filler loading-65 %<br>by weight   | Equivalent to<br>primary predicate<br>device |
| Mechanical/physical<br>properties               | Low viscosity<br>Flowable                                                                                | Low viscosity<br>Flowable                                                                                | Equivalent to<br>primary predicate<br>device |
| Application process and use                     | Similar                                                                                                  | Similar                                                                                                  | Equivalent to<br>primary predicate<br>device |
| Camphorquinone / amine<br>photoinitiator system | Equivalent                                                                                               | Equivalent                                                                                               |                                              |
| Methacrylate-based resin<br>matrix              | Equivalent                                                                                               | Equivalent                                                                                               |                                              |
| Silane treated fillers                          | Equivalent                                                                                               | Equivalent                                                                                               |                                              |
| Bonded with permanent<br>dental adhesive        | Equivalent                                                                                               | Equivalent                                                                                               |                                              |

### 5.7. Summary of Physical Properties:

The Physical properties of this 510(k) submission includes data of the comparison testing per ISO4049, performed to the Apoller Flow Light Cure Flowable Composite and the primary predicate device "Filtek™ Supreme Ultra Flowable Restorative". The properties evaluated included Sensitivity to ambient light, Depth of cure (mm), Flexural Strength (MPa), Water sorption (µg/mm³), Water solubility(µg/mm³), Shade Color stability after irradiation and water sorption, Radio-opacity (mm). Based on the tests performed the results demonstrate substantial equivalence to the primary predicate device.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows a green logo with the letters "GW" in a stylized font. The letters are connected and have horizontal lines running through them. There is a small registered trademark symbol in the upper right corner of the logo.

Apoller Flow - Light Cure Flowable Composite Abbreviated 510(k) Notification

Greenwav YW LLC

### 5.8. Biocompatibility

Evaluation of biocompatibility was conducted using ISO7405, ISO 10993-1; ISO10993-3; ISO10993-5:2009 and ISO10993-11 to determine the biocompatibility of Apoller Flow Light Cure Flowable Composite. Based on the comparative test report it was concluded that Apoller Flow Light Cure Flowable Composite is substantial equivalent to the primary predicate device.

### 5.9. Non-Clinical Performance Testing:

Based on the technological characteristics, physical property and biocompatibility comparisons between Apoller Flow Light Cure Flowable Composite and the primary predicate device, it is demonstrated that Apoller Flow Light Cure Flowable Composite is equivalent to 3M Filtek™ Supreme Ultra Flowable Restorative.

### Conclusion - Substantial Equivalence: 5.10.

The information provided in the 510(k) submission shows that Apoller Flow Light Cure Flowable Composite is substantially equivalent to the primary predicate device Filtek™ Supreme Ultra Flowable Restorative, the two products have the same indications for use, are composed of similar materials, have the same intended use, physical properties and technological characteristics as well as a similar design.

---

**Source:** [https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/EBF/K180463](https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/EBF/K180463)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
